K Number
K960865

Validate with FDA (Live)

Date Cleared
1996-09-20

(200 days)

Product Code
Regulation Number
892.1200
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is intended to be used for diagnostic imaging of organs and lesions. There is no change of intended use from that of the predicate device.

Device Description

The STEP Option to the Prism Dual Head system is a modification of a gamma camera system. This device includes adding hardware and software to a gamma camera system.

AI/ML Overview

This 510(k) notification describes a modification to an existing gamma camera system (STEP Option to the Prism Dual Head system), rather than a novel device requiring extensive performance studies as seen in AI/ML or new therapeutic devices. The primary claim is substantial equivalence to a predicate device. Therefore, many of the typical "acceptance criteria" and "study types" for assessing diagnostic performance are not explicitly detailed in the provided text in the way one might expect for a new diagnostic algorithm or complex software.

Based on the provided text, the focus is on functional equivalence and safety.

Here's an analysis of the provided information in the requested format, with caveats where the information is not present:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria (from text)Reported Device Performance (from text)
Intended UseDiagnostic imaging of organs and lesions."There is no change of intended use from that of the predicate device."
Functional EquivalenceSubstantially equivalent to legally marketed devices.Claimed: "The STEP option to the Prism Dual Head system is substantially equivalent to legally marketed devices." (supported by a comparative matrix, not provided)
Attenuation CorrectionMinimize degrading effects of body attenuation."Laboratory tests have shown that the STEP option has minimized the degrading effects of body attenuation when compared to a standard gamma camera without STEP correction."
PerformancePerform in accordance with development specifications."The product will perform in accordance with the development specifications."
SafetyCertified to electrical safety standards (IEC-601 or UL-544)."The STEP option will be certified to electrical safety standards... by a third party organization prior to use on human patients."

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not explicitly stated. The text refers to "Laboratory tests" but does not quantify the number of tests, phantoms, or subjects used to demonstrate minimized attenuation effects.
  • Data Provenance: Not explicitly stated. The "laboratory tests" context suggests internal validation, likely in a controlled environment using phantoms or simulated data, rather than human patient data, given the phrase "prior to use on human patients."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: Not applicable/not stated. The validation appears to be based on physical measurements of attenuation effects in laboratory settings, not on expert interpretation of diagnostic images.
  • Qualifications of Experts: Not applicable/not stated.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Adjudication Method: Not applicable. The validation described is technical performance (attenuation correction, adherence to specifications) not diagnostic performance requiring expert adjudication of images.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study Done: No, an MRMC study was not described. This device is a hardware/software modification to an imaging system, not an AI/ML diagnostic aid intended to improve human reader performance. Its benefit is in image quality, which could indirectly aid interpretation, but this is not measured by an MRMC study in the provided text.
  • Effect Size: Not applicable.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Standalone Performance Study Done: Yes, in a limited sense. The "laboratory tests" demonstrating minimized attenuation effects appear to be a standalone assessment of the system's technical function (the "algorithm" here being the hardware/software combination for attenuation correction). This study evaluated the device's physical performance without human diagnostic interpretation as its primary outcome.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Type of Ground Truth: For the attenuation correction claim, the ground truth would likely be physical measurements or established theoretical models of attenuation in controlled laboratory settings (e.g., using phantoms whose composition and density are precisely known). For functional specifications, the ground truth would be the engineering specifications themselves.

8. The sample size for the training set

  • Sample Size for Training Set: Not applicable/not stated. This is a conventional gamma camera system modification, not an AI/ML algorithm that typically requires a large training set of labeled data in the way modern AI is developed. The "training" here would refer to engineering design and calibration based on known physics, not machine learning.

9. How the ground truth for the training set was established

  • How Ground Truth for Training Set Was Established: Not applicable in the context of AI/ML. For a conventional medical device, the "ground truth" for its development and calibration would be established through physics principles, engineering specifications, and established medical imaging standards, rather than labeled data sets for machine learning. This would involve designing the attenuation correction method based on known physical properties of radiation interaction with matter.

Summary of the Study Proving Acceptance Criteria:

The provided text indicates that the STEP option meets acceptance criteria primarily through laboratory tests and a comparison matrix to a predicate device.

  • Attenuation Correction: "Laboratory tests have shown that the STEP option has minimized the degrading effects of body attenuation when compared to a standard gamma camera without STEP correction." This is the core performance study mentioned.
  • Functional Specifications: The device "will perform in accordance with the development specifications." This implies internal verification and validation against those specifications, likely through engineering tests.
  • Substantial Equivalence: A "matrix was enclosed comparing the STEP option to a predicate device." This comparative analysis, though not detailed, is foundational to the 510(k) claim of substantial equivalence.
  • Safety: The device "will be certified to electrical safety standards (IEC-601or UL-544) by a third party organization." This is a regulatory compliance pathway.

The 510(k) process for this type of device modification relies heavily on demonstrating that the changes do not raise new questions of safety or effectiveness and that the modified device performs comparably to or better than a legally marketed predicate device on relevant technical performance parameters.

{0}------------------------------------------------

EP 2 0 1996

960865

PICKER INTERNATIONAL 510(k) NOTICE

STEP OPTION TO THE PRISM DUAL HEAD SYSTEM

E: SUMMARY OF SAFETY AND EFFECTIVENESS

This is a summary of the information submitted by Picker International, Inc. To the Office of Device Evaluation (DRAERD) of the FDA as required by the Federal Food, Drug, and Cosmetic Act as amended on November 18, 1990 in section 513(f)(3) for the STEP for Prism Dual Head (system).

The STEP Option to the Prism Dual Head system is a modification of a gamma camera system. This device is intended to be used for diagnostic imaging of organs and lesions. There is no change of intended use from that of the predicate device. This device includes adding hardware and software to a gamma camera system.

Functional specifications and operator's instructions (preliminary) are included in the attachments. Final documentation will be provided with productions units.

The STEP option to the Prism Dual Head system is substantially equivalent to legally marketed devices. The STEP option will be operated by trained health care professionals who are responsible for Nuclear Medicine diagnostic examinations. The STEP option will be certified to electrical safety standards (IEC-601or UL-544) by a third party organization prior to use on human patients. Labeling (Product Bulletin and Operator's Manual) will be provided to the user of the equipment.

Laboratory tests have shown that the STEP option has minimized the degrading effects of body attenuation when compared to a standard gamma camera without STEP correction. The product will perform in accordance with the development specifications. A matrix was enclosed comparing the STEP option to a predicate device and therefore we concluded that it is substantially equivalent to that legally marketed predicate device.

Picker has reviewed all known information and performed an investigation as to the causes of safety and effectiveness concerning the STEP option. In addition, all information contained in this 510(k) Notice is accurate and complete.

§ 892.1200 Emission computed tomography system.

(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.