K Number
K960734

Validate with FDA (Live)

Manufacturer
Date Cleared
1996-12-10

(292 days)

Product Code
Regulation Number
868.5830
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Haemocell System 350 is an autologous transfusion system designed to collect blood during or after an elective or emergency surgical procedure, filter the blood, process as required by the physician, and reinfuse the autologous blood.

The Haemocell System 350 is designed to be used in accordance with the American Association of Blood Banks, Guidelines for Blood Salvage and Reinfusion in Surgery and Trauma (1993). No change to the intended use of the System 350 is proposed as compared to the predicate devices.

Device Description

The Haemocell System 350 is a device designed for collecting and reinfusing a patient's own blood. It comprises an electromechanical system called a Filter Pump and a range of sterile, single use disposable products which when used in accordance with their operating instructions are safe and effective in various surgical procedures.

The single use disposables comprise:

Collection Sets: These are plastic, biocompatible reservoirs of 600 and 1 200ml capacity which are used to contain blood suctioned from the patient during or after surgery. In both cases a regulated vacuum is applied to the collection reservoir by means of the hospital or a portable vacuum system. During surgery, a suction catheter provided with the collection reservoir is used to collect blood, whereas collection after surgery is effected by connecting a tube provided in the set to drains located at the surgical site by the surgeon.

Filter Sets: These are plastic, biocompatible, preassembled membrane filters connected to tubing and flexible containers. The sets are designed to locate on the electromechanical processing system in such a way that there is no contact between the processing system and the fluid content of the filter set, thus preserving the sterility of the fluid.

Reinfusion blood bags are located within either the collection set or filter set depending on the function of the set. The reinfusion bag is connected to the output of the filter set when processing is required and is used to reinfuse the processed blood when processing has been completed. Reinfusion is carried out under gravity, no pump operated or pressurised reinfusion is provided nor recommended.

The electromechanical hardware comprises:

  • . A mobile stand on which the Filter Pump is located for ease of movement and which provides the location for fluid reservoirs.
  • . A mains powered Filter Pump which when fitted with the correct disposables and set up according to the instructions for use will automatically process collected blood if required and selected by the operator. The Filter Pump is not required to be connected to the patient at any time in the procedure and can be operated by a circulating nurse or operating department assistant remote from the patient.

Various reusable hardware components are used with the system, such as a . vacuum canister for use with the 1200ml sterile collection reservoir to facilitate suction.

AI/ML Overview

The provided text describes modifications to an existing device, the Haemocell System 350, and asserts its substantial equivalence to predicate devices. However, the text does not contain a specific study designed to prove the device meets pre-defined acceptance criteria in terms of performance metrics like accuracy, sensitivity, specificity, or improvement in human reader performance with AI assistance.

Instead, the document details a series of non-clinical tests to confirm that the modified device maintains the safety and effectiveness of the predicate device, rather than establishing new performance benchmarks for a novel AI or diagnostic system.

Here's an analysis of the provided information based on your requested headings, highlighting what is present and what is missing:


1. A table of acceptance criteria and the reported device performance

The document does not explicitly state acceptance criteria in terms of specific performance metrics (e.g., accuracy, sensitivity, specificity) for device performance. Instead, it relies on demonstrating substantial equivalence to a predicate device through various non-clinical tests.

The reported performance is qualitative: "in in vitro tests to perform in a substantially equivalent way to the predicate devices with clinically insignificant differences."

Acceptance Criteria (Not explicitly defined in quantitative terms for novel performance)Reported Device Performance (Qualitative)
Maintain safety and effectiveness of predicate devicePerforms as well as the predicate devices and the safety and effectiveness of the predicate devices has been maintained.
Meet requirements of electromechanical testingHas been shown to meet the requirements of the electromechanical testing as well as the predicate device.
No indication of bio-incompatibility in material testsIn material tests to give no indication of bio-incompatibility.
Perform in substantially equivalent way to predicate devices in in vitro testsIn in vitro tests to perform in a substantially equivalent way to the predicate devices with clinically insignificant differences.

2. Sample size used for the test set and the data provenance

The document refers to "in vitro biological function testing" but does not specify the sample size used for these tests. It also does not mention "data provenance" in terms of country of origin or whether it was retrospective or prospective, as the testing described is primarily for mechanical and biological compatibility/function rather than diagnostic performance on patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided. The document describes technical and biological tests, not studies requiring expert consensus for ground truth on disease diagnosis or similar clinical outcomes.

4. Adjudication method for the test set

This information is not provided. As no ground truth establishment by experts is described, no adjudication method would apply.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not done. This device is an autotransfusion apparatus, not a diagnostic AI system, so the concept of human readers improving with AI assistance is not applicable here.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The device is an electromechanical system with disposables, not a software algorithm for diagnostic purposes. Therefore, the concept of "standalone performance" in the context of an algorithm is not applicable. The device's function is automated processing of blood, operated by a circulating nurse or operating department assistant.

7. The type of ground truth used

The concept of "ground truth" in the context of diagnostic accuracy (e.g., pathology, outcomes data) is not applicable to the described device. The "ground truth" in the reported non-clinical tests would have been established by engineering specifications, validated biological standards, and comparisons to the performance of the predicate devices. For example, electromechanical tests would compare against standards like BS EN 60601-1-1, BS EN 60601-1-2, UL 544. Biocompatibility testing would be against Tripartite Biocompatibility Guidance for Medical Devices.

8. The sample size for the training set

This information is not applicable as the device is not an AI/machine learning system that requires a "training set."

9. How the ground truth for the training set was established

This information is not applicable as the device is not an AI/machine learning system.


In summary: The provided document is a 510(k) summary for device modification, focusing on demonstrating substantial equivalence to a predicate device through non-clinical engineering and biological tests. It does not describe a clinical study of diagnostic performance or AI effectiveness. Therefore, many of your specific questions related to AI performance, reader studies, and ground truth establishment in a diagnostic context are not addressed by this document.

{0}------------------------------------------------

HAEMOCELL plc

Dec. 10, 1996

12 Napier Court, Abingdon Science Park, Barton Lane, Abingdon, Oxon. OX14 3YT Tel: +44 (0)1235 527242 Fax: +44 (0)1235 535

510(K) SUMMARY

  • Mr R C Hall Submitter's Name .
  • 12 Napier Court Address Abingdon Science Park Abingdon OX14 3YT Oxon.
  • +44 1235 527242 Telephone Number .
  • +44 1235 535466 Fax number .
  • Mr R C Hall Contact Person .
  • 31 January 1996 Date .
  • Device Identity .
Trade nameHaemocell System 350
Proprietary nameSystem 350 or S350
Common or usual nameAutotransfusion Apparatus (per 21 CFR 868.5830)

{1}------------------------------------------------

Legally Marketed Device 1.

Haemocell plc is proposing that the modified device which is the subject of this submission is equivalent to the Haemocell System 350 granted marketing clearance following submission of a 510(k) document (control number K914053B) and to the Baxter Medi-Vac ATS granted marketing clearance following the submission of a 510(k) document (control number K861677).

Description of the Device 2.

Device Function 2.1

The Haemocell System 350 is a device designed for collecting and reinfusing a patient's own blood. It comprises an electromechanical system called a Filter Pump and a range of sterile, single use disposable products which when used in accordance with their operating instructions are safe and effective in various surgical procedures.

The single use disposables comprise:

Collection Sets: These are plastic, biocompatible reservoirs of 600 and 1 200ml capacity which are used to contain blood suctioned from the patient during or after surgery. In both cases a regulated vacuum is applied to the collection reservoir by means of the hospital or a portable vacuum system. During surgery, a suction catheter provided with the collection reservoir is used to collect blood, whereas collection after surgery is effected by connecting a tube provided in the set to drains located at the surgical site by the surgeon.

Filter Sets: These are plastic, biocompatible, preassembled membrane filters connected to tubing and flexible containers. The sets are designed to locate on the electromechanical processing system in such a way that there is no contact between the processing system and the fluid content of the filter set, thus preserving the sterility of the fluid.

Reinfusion blood bags are located within either the collection set or filter set depending on the function of the set. The reinfusion bag is connected to the output of the filter set when processing is required and is used to reinfuse the processed blood when processing has been completed. Reinfusion is carried out under gravity, no pump operated or pressurised reinfusion is provided nor recommended.

The electromechanical hardware comprises:

  • . A mobile stand on which the Filter Pump is located for ease of movement and which provides the location for fluid reservoirs.
  • . A mains powered Filter Pump which when fitted with the correct disposables and set up according to the instructions for use will automatically process collected blood if required and selected by the operator. The Filter Pump is not required to be connected to the patient at any time in the procedure and can be operated by a circulating nurse or operating department assistant remote from the patient.

{2}------------------------------------------------

  • Various reusable hardware components are used with the system, such as a . vacuum canister for use with the 1200ml sterile collection reservoir to facilitate suction.

2.2 Principle of Operation

Following blood collection, and when processing is required, the Filter Pump controls the flow of fluid around the system by means of a peristaltic pump. pinch clamps, sensors, operator selection via a key-pad and an integral control system.

The purpose of the flow of fluid (principally blood) around the system is to create a flow across the surface of a membrane filter. The filter is designed to allow the liquid component of the fluid to pass through the membrane. whereas the particulate component (principally red blood cells) is maintained in the flow around the system. The membrane which effects the separation is a microporous material with a pore size smaller than the cells that are being conserved. Separation of the cells from the liquid creates a concentration of the cellular component and a reduction of the liquid component.

The separation is assisted by a proprietory process called vortex mixing which is produced by an oscillatory flow component being applied to the fluid flow by means of mechanical pistons acting on flexible diaphragms at the invut and output ends of the filter.

This process results in the reduction of a major proportion of the contaminants in collected blood which is not required for reinfusion to the patient. The operator may select two different processes depending on their requirements and the Filter Pump will process the collected blood in accordance with the selected process.

The principle of this process is identical in the Haemocell System 350 as compared to the predicate device.

3. Intended Use

The Haemocell System 350 is an autologous transfusion system designed to collect blood during or after an elective or emergency surgical procedure, filter the blood, process as required by the physician, and reinfuse the autologous blood.

The Haemocell System 350 is designed to be used in accordance with the American Association of Blood Banks, Guidelines for Blood Salvage and Reinfusion in Surgery and Trauma (1993). No change to the intended use of the System 350 is proposed as compared to the predicate devices.

{3}------------------------------------------------

Technological Characteristics 4.

Summary of Differences in Commarison to the Predicate Devices 4.1

4.1.1 Filter Pump

  • . Changes to the design of the Filter Pump to alter the peristaltic pump flowrate and ancillary features to ensure the effective operation of the system with the new pump flowrate.
  • Introduction of an in-house manufactured power supply to reduce cost and . accommodate changes to international regulatory requirements.
  • Modifications to mechanical design to improve ease of use, reduce . manufacturing cost and accommodate changes to international regulatory requirements.

4.1.2 Disposables

  • Changes to the location of manufacture of the sterile disposables, . including adoption of updated national and international standards for sterilisation by irradiation.
  • Modifications to the design of the clamp used to meter anticoagulant . during blood collection.
  • Incorporation of a 1200ml collection system to offer a larger capacity . collection reservoir.
  • Introduction of a new filter set to accommodate the 1200ml collection set. .
  • . Certain defined components have been added to the predicate System 350.

4.3 Summary of Similarities to the Predicate Devices

4.3.1 Filter Pump

  • The system is identical in appearance and operator interface to the . predicate device System 350.
  • . The principle of operation of the control system and function of the System 350 is identical to the predicate System 350.
  • The principle of mechanical and electrical engineering construction of the . System 350 and its accessories are identical to the predicate devices.

4.3.2 Disposables

  • . With the exception of certain defined components the materials used in the modified device are identical to those used in the predicate Haemocell 350. No change in biocompatibility has been detected.

{4}------------------------------------------------

  • The components and materials used in the new Haemocell 1200ml . collection set are identical to those used in the predicate 1200ml collection set.
  • The principles of operation and design of collection, processing and . reinfusion disposables are identical to those in the predicate devices.

Non-Clinical Tests ર .

Throughout the period since the predicate device Haemocell System 350 has been marketed Haemocell has monitored the requirements of customers, distributors, suppliers and regulators and determined that a range of modifications have been applicable.

Depending on the change, appropriate testing has been planned, implemented and the results and conclusions analysed to ensure the continuity of Haemocell's assertion that the device is safe and effective. Three principal methods of test have been adopted to ensure adequate validation of proposed modifications; electromechanical type testing (BS EN 60601-1-1, BS EN 60601-1-2, UL 544), biocompatibility testing (Tripartite Biocompatibility Guidance for Medical Devices) and in vitro biological function testing.

The Haemocell System 350 has been shown to meet the requirements of the electromechanical testing as well as the predicate device: in material tests to give no indication of bio-incompatibility and in in vitro tests to perform in a substantially equivalent way to the predicate devices with clinically insignificant differences.

Conclusion 6.

The modified Haemocell System 350 is substantially equivalent to the predicate devices in all areas of assessment: principle of operation, intended use, materials used, design and construction, compliance with standards, safety and effectiveness.

It is Haemocell's assertion that testing designed to establish the safety and effectiveness of the modified device, where changes have occurred, has shown that it performs as well as the predicate devices and the safety and effectiveness of the predicate devices has been maintained.

§ 868.5830 Autotransfusion apparatus.

(a)
Identification. An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.(b)
Classification. Class II (performance standards).