K Number
K960556

Validate with FDA (Live)

Date Cleared
1996-05-08

(89 days)

Product Code
Regulation Number
872.6640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BACSTOP is intended for use in dental unit waterlines as a means of preventing the retraction of orally contaminated finids into the coolant and irrigant water hoses. The unit is intended to be used for one day only, and is then to be discarded, and replaced daily.

Device Description

BACSTOP is a disposable, normally closed Anti-retraction check valve, consisting of a flat silastic rubber disc valve membrane in a housing made up of top and bottom parts, united by a sonically sealed joint. The two parts of the valve housing are made of clear Lexan 124R Polycarbonate, with the top (entry) port being in the form of a female Luer fitting, and the bottom (exit) port being in the form of a male Luer fitting, with locking collar. Total length of the device is 0.93 inches, width 0.5 inches. Nominal Cracking (Opening) pressure for the valve is 1.5 PSI.

AI/ML Overview

The provided text is a 510(k) summary for the BACSTOP Valve and describes non-clinical performance testing. It does not contain information about clinical studies with human participants. Therefore, I will respond to your request based on the non-clinical performance testing described.

Here's an analysis of the provided text in relation to your questions regarding acceptance criteria and the study proving the device meets them:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Implied)Reported Device Performance (BACSTOP Valve)
Prevent backflow of orally contaminated fluids into dental waterlines.Effective in preventing backflow: "Both devices [BACSTOP and Aseptico] were effective in preventing backflow of an aqueous suspension of a marker bacterium, Chromobacterium (ATCC # 553), into dental PVC hoses connected to either high speed handpieces or air/water syringes."
No detectable bacteria in growth broth after challenge.No bacterial growth: "No bacteria grew from water samples collected with either the Aseptico device or the BACSTOP valve in place."
Maintain performance after repeated on/off cycling (simulating one day's use).Maintains performance: "Again, both devices were equivalent in completely preventing backflow of the challenge suspension of bacteria into the waterlines," after "250 five-second pulses of water flow at 35 PSI" (air/water syringe) or "50 twenty-second pulses" (handpiece).
Permit no backflow (zero microliters) during its normal working life (one day).Zero microliters backflow: "BACSTOP, by comparison, permits no backflow (zero microliters) during its normal working life (one day)." This implicitly relies on the non-failure of the valve over this period and the daily disposal preventing biofilm formation. The experiments confirmed this by showing no bacterial growth after a simulated single day's use.
Biocompatibility as per USP Class VI and other tests.Satisfactory biocompatibility: "Safety testing of the silastic rubber and polycarbonate polymer component parts of BACSTOP has included USP Class VI in vivo and in vitro biological reactivity testing, 24 Hour MEM elution test in vitro, hemolysis extract test, in vivo irritation, sensitisation, mutagenicity, and subchronic toxicity tests; all gave satisfactory results (see enclosed report)."

Study Proving Device Meets Acceptance Criteria:

The study described is a non-clinical laboratory test focused on the ability of the BACSTOP valve to prevent backflow of a bacterial suspension, simulating its intended use in dental waterlines.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: The text does not specify the exact number of BACSTOP valves or Aseptico valves tested in each experiment. It refers to "Both devices" and "BACSTOP valves" in plural, implying more than one, but a specific number is not provided for the backflow tests or the cycling tests.
  • Data Provenance: The study is reported as "Non-clinical Performance Testing" and "Non-clinical Laboratory Tests." This implies prospective testing specifically designed to evaluate the device's performance. The country of origin of the data is not explicitly stated, but the submission is from "SciTech Dental, Inc., Seattle, WA 98104," suggesting it's likely US-based.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. This was a non-clinical laboratory study involving bacterial growth and flow dynamics, not human interpretation or assessment. Therefore, no "experts" in the sense of medical professionals were used to establish ground truth. The "ground truth" was determined by the presence or absence of bacterial growth in the broth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. Adjudication methods are relevant for studies involving human interpretation or subjective assessments. This was a laboratory test with objective endpoints (bacterial growth).

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This was a non-clinical laboratory study of a valve, not a study involving human readers or AI.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Yes, in essence. The device (BACSTOP valve) was tested in a standalone manner, meaning its performance was evaluated independently without human intervention during the "backflow prevention" process itself. Humans set up the experiment and observed the results, but the valve's function was autonomous.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The ground truth for the backflow prevention tests was determined by laboratory culture for bacterial growth. The absence of bacterial growth in the broth after exposure to the challenge suspension indicated effective backflow prevention.

8. The sample size for the training set

  • Not applicable. This device (a physical valve) does not involve a "training set" in the context of machine learning or AI algorithms. The "training" for its design and manufacturing would be engineering and material science principles, not data-driven learning.

9. How the ground truth for the training set was established

  • Not applicable. As above, there is no "training set" for a physical device in this context.

{0}------------------------------------------------

19605

510 (k) SUMMARY

Submitted by:

Simon Johnston SciTech Dental, Inc., 562 First Avenue South, Suite 700 Seattle, WA 98104 FDA Establishment Registration No. 3027751 Owner/Operator I. D. 9011186 TEL # 800 524 6984 or 206 382 0880 FAX # 206 382 0823 e-mail 102477.2560 @ compuserve. com

Date of preparation:

January 29, 1996

Trade Name of Device:

BACSTOP VALVE

Common Name of Device:

Daily disposable in-line Anti-retraction check valve

Classification name of Device:

Unknown

Legally Marketed Device to which BACSTOP is equivalent

Aseptico Anti-retraction Valve for Dental Lines, from Aseptico, Inc., Kirkland, WA. This check valve was marketed prior to 1976.

Description of the device:

BACSTOP is a disposable, normally closed Anti-retraction check valve, consisting of a flat silastic rubber disc valve membrane in a housing made up of top and bottom parts, united by a sonically sealed joint. The two parts of the valve housing are made of clear Lexan 124R Polycarbonate, with the top (entry) port being in the form of a female Luer fitting, and the bottom (exit) port being in the form of a male Luer fitting, with locking collar. Total length of the device is 0.93 inches, width 0.5 inches. Nominal Cracking (Opening) pressure for the valve is 1.5 PSI.

{1}------------------------------------------------

Intended use of the Device:

BACSTOP is intended for use in dental unit waterlines as a means of preventing the retraction of orally contaminated finids into the coolant and irrigant water hoses. The unit is intended to be used for one day only, and is then to be discarded, and replaced daily.

Technical Comparison to Predicate Device:

BACSTOP is a polycarbonate device manufactured for SciTech Dental Inc. by B. Braun Medical Inc. It is commonly used as a backflow prevention component in I/V Infusion sets, and is cleared for marketing for medical use under K790062. It is "normally closed". When placed in-line in the dental water hose, it opens only under sufficient water pressure, equivalent to 1.5 PSI. The valve snaps closed and seals against backflow when the pressure drops below this point. The device is inserted by connection to luer mounts placed permanently in-line.

By comparison, the in-line Anti-retraction Valve sold by Aseptico Inc., has a plated brass housing surrounding a spring-controlled plunger with an O-ring seal at one end. As designed, the device is intended to fit "permanently" in-line by means of barbed hose mounts. The internal plunger and O-ring normally close an orifice through which water enters the central chamber of the housing, and then flows out through the exit barb. Water is permitted to pass only when sufficient pressure is applied against the spring the "cracking pressure" -- , to overcome its resistance. This pressure in the Aseptico device is 2.50 PSI.

Both valves remain open at working pressures up to 50 PSI or more, although normal working water pressures are in the 35-40 PSI range.

When newly inserted, the performance of the two devices is equivalent in preventing backflow of orally contaminated fluids into the dental irrigant and coolant hoses. However, no replacement schedule has been defined for the Aseptico valve, nor are there any limits established for its working life-span under normal conditions. It is considered to be appropriate for the Aseptico valve to allow the backflow of up to 45 microliters of fluid from the tip of the dental handpiece waterline, according to ANSI specification # 47, promulgated by the American Dental Association. Routine testing to ensure compliance with this standard is recommended, and deviations from it constitute grounds for Aseptico valve replacement. Failure to meet this standard may come about because of mechanical failure of moving parts, or because bacterial slime accumulation on the apposing surfaces of the valve prevents effective seal formation.

BACSTOP, by comparison, permits no backflow (zero microliters) during its normal working life ( one day ). Over this brief period, mechanical failure due to declining resilience of the silastic rubber valve disc does not occur, and there is insufficient time for

{2}------------------------------------------------

bacterial adherence and slime formation to interfere with the working of the valve surfaces.

Safety testing of the silastic rubber and polycarbonate polymer component parts of BACSTOP has included USP Class VI in vivo and in vitro biological reactivity testing, 24 Hour MEM elution test in vitro, hemolysis extract test, in vivo irritation, sensitisation, mutagenicity, and subchronic toxicity tests; all gave satisfactory results (see enclosed report).

Non-clinical Performance Testing for Substantial Equivalence:

Both devices were effective in preventing backflow of an aqueous suspension of a marker bacterium, Chromobacterium (ATCC # 553), into dental PVC hoses connected to either high speed handpieces or air/water syringes. Suction presure of 3-5 PSI was applied unstream of the valves, which were placed in-line approximately 6 inches from the connection to the respective instrument, while the tip of each instrument was inserted into the bacterial suspension. After cessation of the suction pressure, the PVC tubing between the valve device and the instrument was removed and the entire contents flushed into a bacterial growth broth and incubated at 22 C for 3 days. If any bacteria had been drawn back up the waterline they should be detectable by their growth in this broth medium. No bacteria grew from water samples collected with either the Aseptico device or the BACSTOP valve in place. Under these conditions the devices were equivalent in performance.

Dental handpieces and air/water syringes are normally operated intermittently in a "pulsatile" pattern throughout the working day. Repeated on/off cvcling of a backflow valve could cause exhaustion of the closure mechanism, and result in valve failure. In order to mimic these working conditions the experiment was repeated, but the efficacy challenge to each device was done only after a previous sequence of 250 five-second pulses of water flow at 35 PSI had been completed (intended to mimic air/water syringe use patterns), or after a sequence of 50 twenty-second pulses (intended to mimic handpiece use). Again, both devices were equivalent in completely preventing backflow of the challenge suspension of bacteria into the waterlines.

However, when the Aseptico valve was left in place in a dental waterline for 3 weeks with daily water flow, a layer of biofilm slime formed within the valve. These valves were challenged as before, and they were no longer effective in completely preventing retraction of bacteria into the waterline. BACSTOP valves were not challenged in these experiments because they are disposable and are to be discarded after each day's use; thus biofilm formation on the apposing surfaces does not take place. In this regard the two devices differ; BACSTOP has a defined use life-span. The Aseptico Anti-retraction valve has no defined life-span, and may fail whenever biofilm formation leads to interference with effective valve closure.

{3}------------------------------------------------

Conclusions Drawn from the Non-clinical Laboratory Tests:

Freshly inserted, new valves of the BACSTOP and Aseptico types are substantially equivalent in preventing backflow of bacterially contaminated suspensions into dental waterlines connected to handpieces or air/water syringes, when negative pressures of 3-5 PSI are applied upstream of the valves. Both valves perform equally well after being subjected to repeated on/off cycling, intended to mimic one day's use in a dental operatory. Aseptico Anti-retraction valves fail when build up of biofilm slime occurs inside the valves, after continuous use in-line. BACSTOP valves are discarded daily, obviating this possibility.

§ 872.6640 Dental operative unit and accessories.

(a)
Identification. A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.(b)
Classification. Class I (general controls). Except for dental operative unit, accessories are exempt from premarket notification procedures in subpart E of part 807 of this chapter subject to § 872.9.