Intended Use

This device is a disposable device to be used to inflate, deflate and monitor pressure for gastrointestinal dilation balloons.

Device Description

The Wilson-Cook Balloon Inflation Device is a hand-held mechanical device consisting of a 20cc capacity syringe, a 300 psi/20atm pressure gauge, a plunger mechanism, lever, extension line, and stopcock. This device is to be used to inflate, deflate and monitor pressure for gastrointestinal dilation balloons.

AI/ML Overview

The provided text describes a medical device, the Wilson-Cook Balloon Inflation Device, and its premarket notification. However, it does not include detailed information regarding specific acceptance criteria for device performance or a comprehensive study report proving the device meets those criteria in the way typically expected for AI/ML-based medical devices (e.g., sensitivity, specificity, AUC).

Instead, the document focuses on regulatory compliance, description of the device, its intended use, and comparison to predicate devices, along with basic functional testing.

Therefore, for the given input, most of the requested information cannot be extracted. I will fill in what can be inferred and explicitly state what is not available.


Acceptance Criteria and Device Performance Study for the Wilson-Cook Balloon Inflation Device

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicAcceptance CriteriaReported Device Performance
Functional TestingAll samples tested to meet testing specifications for device integrity."All samples tested were deemed acceptable in that testing specifications were met."
SterilityValidated EO cycle following AAMI Overkill Method to SAL 10⁻⁶."Validated EO cycle following AAMI Overkill Method to SAL 10⁻⁶."
Intended UseCapable of inflating, deflating, and monitoring pressure for gastrointestinal dilation balloons.Device description and intended use confirm this functionality.
Pressure Gauge AccuracyNot specified in detail. (Implied to be accurate enough for intended use.)Device includes a 300 psi/20 atm pressure gauge. Functional testing "met testing specifications," which would presumably include pressure accuracy.
Syringe Capacity20cc capacity.Device includes a 20cc capacity syringe.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document states "All samples tested" for in-house functional testing. A specific number of samples is not provided.
  • Data Provenance: The testing was "in-house functional testing." This indicates the testing was conducted by Wilson-Cook Medical Inc. The data is retrospective in the sense that it was performed on manufactured samples. Country of origin of data is not specified but assumed to be the USA where Wilson-Cook Medical Inc. is located.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This information is not applicable and not provided. The device is a mechanical balloon inflation device, not an AI/ML diagnostic or prognostic tool requiring expert-established ground truth for its performance assessment. Ground truth here would be based on engineering specifications and physical measurements.

4. Adjudication Method for the Test Set

  • This information is not applicable and not provided. As it's a mechanical device, performance is typically assessed against engineering specifications, not through expert adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs. without AI assistance

  • This information is not applicable and not provided. The device is a mechanical instrument, not an AI system that assists human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • This information is not applicable and not provided. The device is a mechanical instrument, not an algorithm. Its standalone performance is inherently what was tested through "in-house functional testing."

7. The type of ground truth used

  • The ground truth for the "functional testing" would be based on engineering specifications and metrological standards for pressure accuracy, syringe capacity, inflation/deflation capabilities, and sterility. It is not expert consensus, pathology, or outcomes data.

8. The Sample Size for the Training Set

  • This information is not applicable and not provided. The device is a mechanical device, not an AI/ML model that requires a "training set."

9. How the Ground Truth for the Training Set was Established

  • This information is not applicable and not provided for the same reason as point 8.

{0}------------------------------------------------

Page 7 of 9

APR 2 6 1996

RE: PREMARKET NOTIFICATION FOR THE WILSON-COOK BALLOON INFLATION DEVICE

I. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

Submitted Bv:

Wilson-Cook Medical Inc. 4900 Bethania Station Road Winston-Salem, NC 27105

Device Description:

The Wilson-Cook Balloon Inflation Device is a hand-held mechanical device consisting of a 20cc capacity syringe, a 300 psi/20atm pressure gauge, a plunger mechanism, lever, extension line, and stopcock. This device is to be used to inflate, deflate and monitor pressure for gastrointestinal dilation balloons.

Classification Name: Syringe, Balloon Inflation, 74 MAV

Classification: FDA has classified similar devices as Class II. This device falls within the purview of the Gastroenterology and Urology Devices Panel.

Trade/Proprietary Name: Wilson-Cook Dilation Balloon Inflation Device.

Common/Usual Name: Balloon Inflation Device

Performance Standards To the best of our knowledge, performance standard for this device do not exist.

Sterility: Validated EO cycle following AAMI Overkill Method to SAL 10°.

Intended Use: This device is a disposable device to be used to inflate, deflate and monitor pressure for gastrointestinal dilation balloons.

Predicate Devices: This device is similar with respect to technological characteristics and intended use to predicate balloon inflation devices, including the Microvasive Digiflator™ (K860589) and the Wilson-Cook Balloon Inflation Device (K936168).

{1}------------------------------------------------

PREMARKET NOTIFICATION FOR THE WILSON-COOK BALLOON RE: INFLATION DEVICE

I. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

CHARACTERISTICWilson-Cook BalloonInflation Device[Subject of 510(K)]MicrovasiveDigiflator™(K860589)Wilson-CookQuantum InflationDevice (K936168)
Intended UseUsed to inflate, deflateand monitor pressureof the Wilson-Cookline of balloon dilatorsfor GI use.Used to inflate, deflateand monitor pressureof balloon dilators forGI use.Used to inflate, deflateand monitor pressureof the Wilson-Cookline of balloon dilatorsfor GI use.
ConstructionPlunger mechanism tocontrol pressure.20 cc syringeHand held design.Extension line withstopcock forconnection to balloon.Plunger mechanism tocontrol pressure.60 cc syringeHand held design.Extension line forconnection to balloon.Plunger mechanism tocontrol pressure.60 cc syringeHand held design.Extension line forconnection to balloon.
SterilityAnalog pressure gaugeSterile and non-sterile.Digital pressure gaugeAvailable sterile andnon-sterile.Analog pressure gaugeNon-sterile.
DisposableDisposableDisposable

Testing: The Wilson-Cook Balloon Inflation device has been subjected to in-house functional testing to demonstrate device integrity. All samples tested were deemed acceptable in that testing specifications were met.

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.