FLEXSURE HP,CONTROLS FOR FLEXSUREHP,FLEXPACK HP(ABBOTT DIAGNOSTICS)ANDCONTROLS FOR FLEXPACK HP (ABBOTT DIAGNOSTICS)
K960183 · Smithkline Diagnostics, Inc. · LYR · Jun 7, 1996 · Microbiology
Device Facts
| Record ID | K960183 |
| Device Name | FLEXSURE HP,CONTROLS FOR FLEXSUREHP,FLEXPACK HP(ABBOTT DIAGNOSTICS)ANDCONTROLS FOR FLEXPACK HP (ABBOTT DIAGNOSTICS) |
| Applicant | Smithkline Diagnostics, Inc. |
| Product Code | LYR · Microbiology |
| Decision Date | Jun 7, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.3110 |
| Device Class | Class 1 |
Intended Use
The FlexSure® HP test for IgG antibodies to H. pylori in whole blood is a rapid, visually read, qualitative immunochromatographic method. The test is for use by health professionals as an aid in the diagnosis of H. pylori infection in patients with clinical signs and symptoms of gastrointestinal disease.
Device Story
Rapid, visually read, qualitative immunochromatographic test; detects IgG antibodies to H. pylori in whole blood. Used by health professionals in clinical settings. Input: whole blood sample; output: visual qualitative result on test card. Assists clinicians in diagnosing H. pylori infection in symptomatic patients. Benefits patient by providing rapid diagnostic information to guide management of gastrointestinal disease.
Clinical Evidence
Multi-center trial (n=173 symptomatic patients) compared whole blood test to serum test and reference methods (histology/urease). Whole blood test showed 95% sensitivity, 94% specificity, and 95% agreement after resolving discordants. Volunteer study (n=233) showed 99% specificity vs. serum test. Reproducibility (within-site 96%, between-site 89%) and interference studies (lipemic/hemolytic samples) performed. No cross-reactivity with C. jejuni, C. fetus, C. coli, or H. mustalae.
Technological Characteristics
Immunochromatographic test card; qualitative visual readout. Analyte: IgG antibodies to H. pylori. Sample type: whole blood. No electronic components, software, or energy source. Sterilization not specified.
Indications for Use
Indicated for health professionals to aid in the diagnosis of H. pylori infection in patients presenting with clinical signs and symptoms of gastrointestinal disease.
Regulatory Classification
Identification
Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identify Campylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases. Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.
Predicate Devices
- FlexSure® HP Test for Serum IgG Antibodies to H. pylori (K934863)
Related Devices
- K982362 — ABBOTT TESTPACK PLUS H. PYLORI · Abbott Laboratories · Nov 18, 1998
- K151935 — Alere Signify H. pylori Whole Blood, Serum, Plasma; Alere Signify H. pylori Whole Blood Only;Alere Clearview H. pylori Whole Blood, Serum, Plasma; Alere Clearview H. pylori Whole Blood Only · Alere San Diego, Incorporated · Aug 12, 2015
- K024350 — ACON H. PYLORI RAPID TEST STRIP AND ACON H. PYLORI RAPID TEST DEVICE · ACON Laboratories, Inc. · Apr 2, 2003
- K980798 — BIOSIGN H. PYLORI WB-NEW ONE-STEP ANTI-H. PYLORI ANTIBODY TEST · Princeton BioMeditech Corp. · Sep 22, 1998
- K980782 — STAT-SIMPLE PYLORI · Saliva Diagnostic Systems, Inc. · Aug 5, 1998
Submission Summary (Full Text)
{0}
225 Baypointe Parkway, San Jose, CA 95134-1622 • (408) 435-2660
K960183
# SmithKline Diagnostics, Inc.
JUN - 7 1996
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
*FlexSure® HP* Test for IgG Antibodies to *H. pylori* in Whole Blood
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
**Manufacturer:** SmithKline Diagnostics, Inc.
225 Baypointe Parkway
San Jose, CA 95134-1622
**Attention:** Marshall C. McCarty
**Proprietary Name:** *FlexSure® HP*
**Classification Name:** Test for IgG antibodies to *H. pylori* in whole blood
**Intended Use:** The *FlexSure® HP* test for IgG antibodies to *H. pylori* in whole blood is a rapid, visually read, qualitative immunochromatographic method. The test is for use by health professionals as an aid in the diagnosis of *H. pylori* infection in patients with clinical signs and symptoms of gastrointestinal disease.
**Predicate Product:** *FlexSure® HP* Test for Serum IgG Antibodies to *H. pylori*; Manufactured by SmithKline Diagnostics, Inc.
**Performance Summary:** The SKD *FlexSure® HP* Test for IgG Antibodies to *H. pylori* in Whole Blood is substantially equivalent to the predicate device, *FlexSure® HP* Test for Serum IgG Antibodies to *H. pylori* (K934863). The performance of the *FlexSure® HP* whole blood product was verified by sensitivity, specificity and reproducibility and interference studies in symptomatic patients and volunteer blood donors, mainly asymptomatic individuals, not being treated for gastrointestinal disease. Refer to attached PERFORMANCE CHARACTERISTICS.
Marshall C. McCarty
Manager, Regulatory Affairs
SKD
Facsimile: (408) 435-1953
1-11-96
Date
000011
{1}
000012
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
## FlexSure® HP Test for IgG Antibodies to *H. pylori* in Whole Blood
## PERFORMANCE CHARACTERISTICS
The basic methodology and format for the FlexSure® HP Whole Blood test were established previously with the FlexSure® HP Serum test. The tests differ principally in the method for sample collection and application of the sample to the Test Card. Both tests are equally effective in determining the presence or absence of IgG antibodies against *H. pylori*.
## FlexSure® HP Serum Test
The performance characteristics of the FlexSure® HP Serum test were previously reported (Premarket Notification K934863) for a group of 551 individuals (196 symptomatic and 355 asymptomatic) by comparison with a commercial microwell ELISA serological test. After eliminating 15 samples from the calculations with indeterminate ELISA Values, in accordance with the manufacturer's instructions, the remaining 536 samples yielded a relative sensitivity of 95% (285/299), a relative specificity of 94% (222/237) and overall agreement of 95% (507/536). The FlexSure® HP Serum test was also compared with two other reference methods in this group of individuals: histology and $^{13}\mathrm{C}$-urea breath test.
## FlexSure® HP Whole Blood Test
The FlexSure® HP Whole Blood test was evaluated in a multi-center trial at six different gastroenterology clinics located in the United States, Canada and the United Kingdom. A group of 173 patients who presented with gastrointestinal symptoms were evaluated with the FlexSure® HP Whole Blood test, the FlexSure® HP Serum test and upper endoscopy where multiple biopsy specimens were studied by histology and/or a urease test.
Direct comparison of the FlexSure® HP Whole Blood test with the FlexSure® HP Serum test yielded a relative sensitivity of 92%, a relative specificity of 91% and overall agreement of 92% (Table 1). After resolution of discordant serological results by histology and/or a urease test, the FlexSure® HP Whole Blood test had a relative sensitivity of 95%, a relative specificity of 94% and overall agreement of 95%.
Page 2 of 5
{2}
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
FlexSure® HP Test for IgG Antibodies to *H. pylori* in Whole Blood
## Table 1
FlexSure® HP Serum Test vs. FlexSure® HP Whole Blood Test Symptomatic Patients*
| | | FlexSure® HP Serum | | |
| --- | --- | --- | --- | --- |
| | | + | - | Total |
| FlexSure® HP | + | 98 | 6 | 104 |
| Whole Blood | - | 8 | 61 | 69 |
| | | 106 | 67 | 173 |
Relative Sensitivity: 92%
Relative Specificity: 91%
Overall Agreement: 92%
* without resolution of discordants by histology and/or urease test
In this same group of patients, the FlexSure® HP Whole Blood test was compared directly with histology and/or a urease test, yielding a relative sensitivity of 88%, a relative specificity of 74% and overall agreement of 82% (Table 2).
Comparison of the FlexSure® HP Serum test with histology and/or a urease test in these patients yielded similar results: a relative sensitivity of 92%, a relative specificity of 76% and overall agreement of 84%.
The relative specificity of any serological test, when compared directly with histology or urease tests, may be lower if the bacterium was previously eradicated or suppressed as a result of taking antimicrobial drugs in connection with other medical treatments. In addition, sampling errors may occur due to the patchy distribution of the bacteria in the gastric mucosa. It is known that atrophy of the gastric mucosa often develops in persons with chronic active gastritis due to long-term *H. pylori* infection. This may lead to reduced bacterial loads making it difficult to detect the bacterium by histology or urease tests.
Page 3 of 5
000013
{3}
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
*FlexSure® HP* Test for IgG Antibodies to *H. pylori* in Whole Blood
## Table 2
Histology and/or Urease Test vs. *FlexSure® HP* Whole Blood Test Symptomatic Patients
| | | Histology and/or Urease | | |
| --- | --- | --- | --- | --- |
| | | + | - | Total |
| FlexSure® HP | + | 84 | 20 | 104 |
| Whole Blood | - | 11 | 58 | 69 |
| | | 95 | 78 | 173 |
Relative Sensitivity: 88%
Relative Specificity: 74%
Overall Agreement: 82%
The *FlexSure® HP* Whole Blood test was further evaluated in a group of 233 volunteer blood donors and yielded a relative specificity of 99% and overall agreement of 97% when compared with the *FlexSure® HP* Serum test. Since diagnostic confirmation of *H. pylori* infection by a second method was not done in this volunteer group, a meaningful determination of relative sensitivity could not be made (Table 3).
## Table 3
*FlexSure® HP* Serum Test vs. *FlexSure® HP* Whole Blood Test Volunteer Blood Donors
| | | FlexSure® HP Serum | | |
| --- | --- | --- | --- | --- |
| | | + | - | Total |
| FlexSure® HP | + | 41 | 1 | 42 |
| Whole Blood | - | 7 | 184 | 191 |
| | | 48 | 185 | 233 |
Relative Sensitivity: Not Determined
Relative Specificity: 99%
Overall Agreement: 97%
Page 4 of 5
000014
{4}
000015
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
*FlexSure® HP* Test for IgG Antibodies to *H. pylori* in Whole Blood
## Reproducibility
*Within-* and *between-site* reproducibility of the *FlexSure® HP* Whole Blood test was done with fresh fingerstick blood samples from three volunteers with known levels of antibody to *H. pylori* (negative, low/borderline positive, positive). Each volunteer gave three fingerstick samples on each day of testing. The *within-site* reproducibility was 26/27 (96%) and the *between-site* reproducibility was 24/27 (89%); the reproducibility of the between-site sample sets was 9/9 for the negative, 9/9 for the positive and 6/9 for the low/borderline positive.
## Interference
The *FlexSure® HP* Whole Blood test was evaluated for possible interference from visibly lipemic or hemolytic samples. Whole blood was obtained by venipuncture and stored in Vacutainer tubes containing EDTA. Each sample was spiked with cholesterol, triglycerides or hemoglobin to obtain concentrations above physiological levels. Blood samples that were positive or negative for IgG antibodies to *H. pylori* were run before and after spiking on the *FlexSure® HP* Whole Blood test. None of the biological substances tested interfered with the procedure or yielded inaccurate test results.
## Cross Reactivity
The cross reactivity of the test was determined previously for the *FlexSure® HP* Serum test. Sera containing known levels of antibody against *H. pylori* were evaluated according to the method of Perez-Perez, et al., (*Ann. Int. Med.* 109:11-17, 1988) with the following bacteria:
| Campylobacter jejuni | Escherichia coli |
| --- | --- |
| Campylobacter fetus | Helicobacter mustalae |
| Campylobacter coli | Helicobacter pylori |
All species tested showed no cross reactivity, indicating that the test has high specificity for human antibodies against *H. pylori*. *Helicobacter pylori* was tested as a control and found to be reactive.
Page 5 of 5