K955529 · Mark H. Friedman · ILO · Jun 10, 1996 · Physical Medicine
Device Facts
Record ID
K955529
Device Name
CRYTRON SYSTEM
Applicant
Mark H. Friedman
Product Code
ILO · Physical Medicine
Decision Date
Jun 10, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5720
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
To cool a small portion of the body to provide temporary relief of minor aches and pains and muscle spasms or bruises.
Device Story
Cryotron Cold Probe is a water-circulating cold therapy device. It consists of a hollow metal tip (chrome-plated brass) shaped for application to specific body areas, such as the maxillary nerve section. A small pump circulates ice water through the probe via plastic tubing; the probe is covered by a disposable plastic sheath. Used in clinical settings by healthcare providers to provide localized cooling. The device provides temporary relief of minor aches, pains, muscle spasms, or bruises. The system is sold non-sterile, with pre-sterilized disposable sleeves provided separately.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Materials: chrome-plated brass probe, plastic pump, cooler, and tubing. Energy source: electric pump for water circulation. Form factor: hollow metal tip with disposable plastic sheath. Sold non-sterile. No software or electronic algorithms.
Indications for Use
Indicated for patients requiring temporary relief of minor aches, pains, muscle spasms, or bruises via localized cold therapy.
Regulatory Classification
Identification
A water circulating hot or cold pack is a device intended for medical purposes that operates by pumping heated or chilled water through a plastic bag and that provides hot or cold therapy for body surfaces.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 890.9.
K962537 — INCARE HOT/ICE LITE MACHINE · Hollister, Inc. · Nov 27, 1996
K964354 — ARTOCOOL-S WATER CIRCULATING COLD PACK · Ormed GmbH · Aug 29, 1997
K070402 — MODIFICATION TO POLAR CARE 500 · Breg, Inc. · Mar 29, 2007
K955057 — MODEL 1100 COLD THERAPY DEVICE · Smith and Nephew Donjoy, Inc. · Mar 8, 1996
K040317 — IVC SYSTEM · Mark H. Friedman · May 7, 2004
Submission Summary (Full Text)
{0}
PAGE 04
# XL SAFE MEDICAL DEVICES ACT OF 1990 SUMMARY OF SAFETY AND EFFECTIVENESS.
March 20, 1996
A. Submitter: Dr. Mark Friedman, 660 Gramatan Ave., Mt. Vernon, NY.
I. Classification Names and numbers:
Hot or cold pack, water-circulating, 89ILO*
Cold pack, reusable, 89IME
K955529
JUN 10 1996
II. Common/Usual Names: Cold Pack
III. Proprietary Names: Cryotron™ Cold Probe
IV. Establishment Registration Number: In process
V. Classification: Circulating water cold packs have been classified in the Code of Federal Regulations, in class II, CFR 890.5720 by the Physical Medicine Panel, and placed in Tier II by the Office of Device Evaluation
VI. Performance Standard: None established under section 514.
VII. Description of the Device: Cryotron is fabricated by joining two hollow metal or plastic tubes together into a hollow metal tip which is shaped to fit over the maxillary nerve section or other area selected for cold therapy. Ice water driven by a small pump enters the smaller tube and exits the large tube, chilling the metal which is covered by a disposable plastic sheath. The product will be sold non-sterile; prepackaged, presterilized plastic sleeves will be provided with the Cryotron.
VIII. Labels of the product and competitive devices are provided.
IX. Substantial Equivalence Statement. The "510(k) "Substantial Equivalence" Decision-Making Process (Detailed) from ODE Guidance Memorandum #86-3 was followed as described below:
1. These products have the same intended use, to cool a small portion of the body to provide temporary relief of minor aches and pains and muscle spasms or bruises. These are the same as those of the predicate devices. These products also have the same intended uses as similar products currently cleared for marketing by the 510(k) process, and several that are exempt (CFR 890.5700) that are exempt from 510(k) requirements.
2. The technological characteristics for this product are the same as those for the predicate devices and those currently on the market. The Cryotron probe is a common reusable type, made of chrome-plated brass and is sold non-sterile. The pump, cooler, and tubing are of plastic similar to that used in equivalent products. The plastic cover is disposable and is sold packaged separately.
3. Descriptive information provided shows that the materials from which the Cryotron device are made is substantially equivalent to (nearly identical with) those of similar products, used for identical purposes, currently on the market.
000000
{1}
3. Predicate devices
This device is substantially equivalent to preamendment devices classified as shown above. Specifically, the Cryotron device for which we are seeking marketing clearance is described in item 1. However, variations of this device, used in this office for the same purpose include an ice-extrusion pack which could be considered as an item 2 product, exempt from 510(k) requirements.
The device also is substantially equivalent to devices currently on the market such those sold by Kinetic Concepts, Inc. K-891152 and K-891151; Seabrook Medical Systems, K-902269; Breg, Inc., K-920581, K-914434; Toronto Medical Corp., K-935841; Electro-Biology, Inc., K-945067 and others. Labels and labeling for the Seabrook Medical Systems, Electri-Cool as well as for simpler systems such as EBIce by Electron Biology, Inc. and DeltaTherm by DeltaTherm, Inc. (to which the Cryotron is most similar) are shown for comparison in Attachment III. The device is also very similar to Hollister's System 3 (K-915196).
4. Descriptive information provided shows that the materials from which the Cryotron device are made is substantially equivalent to (nearly identical with) those of similar products, used for identical purposes, currently on the market.
End of Summary
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches