Intended Use

The models 436-07 and 437-07 pacing leads are intended for use with implantable cardiac pulse generators for long-term pacing of the heart. The indications for ventricular pacing include, but are not limited to: Sick sinus syndrome, sinus bradycardia, complete heart block, symptomatic second-degree heart block, and certain conditions of asymptomatic second-degree block.

In the presence of normal A-V conduction the indications for atrial pacing include, but are not limited to: Sinus arrest, sick sinus syndrome, sinus bradycardia and conditions requiring increased cardiac efficiency, enhanced cardiac output or the overdrive of certain cardiac arrhythmias.

In the absence of normal A-V conduction, an atrial lead may be used with a ventricular lead in a dual chamber pacing system to restore A-V synchrony.

Device Description

The Intermedics Models 436-07 (bipolar) and 437-07 (unipolar) endocardial pacing leads are designed for use with implantable cardiac pulse generators configured for ventricular applications for long term cardiac pacing. The nominal lead length for the model 436-07 lead is 58 cm and is available in lengths from 40 to 110 cm. The nominal lead length for the model 437-07 lead is also 58 cm and is available in lengths from 40 to 110 cm. The lead connectors are designed to meet VS-1 specifications (an intra-industry agreement standardizing lead-to-pulse generator connection dimensions) for lead connectors.

The models 436-07 and 437-07 pacing leads are packaged with the model 367-01 vein lifter and the following stylets: model 367-11 (straight stylet, limber), model 365-12 (straight stylet, firm), model 365-81(tapered stylet, limber), and model 365-82 (tapered stylet, firm).

The model 437-07 is also packaged with the following adapters: model 366-29 (Step up adapter, VS-1 unipolar lead to 5 mm unipolar pacer) and model 366-30 (Step up adapter, VS-1 unipolar lead to 6 mm unipolar pacer).

The minimum introducer size recommended for both lead models is 10 French.

The models 436-07 and 437-07 pacing leads are identical to the Intermedics models 436-02 (K883602, K954719) and 437-02 (K883602, K954719) respectively, except for the use of iridium-oxide (IROX) coated titanium for the electrodes. The IROX-coated electrodes are identical to those used in models 430-07 (K902672, K954719) and 431-07 (K902672, K954719).

AI/ML Overview

The provided text is a 510(k) summary for pacing leads, not an AI/ML device. Therefore, it does not contain information about acceptance criteria, device performance metrics, sample sizes, ground truth establishment, expert adjudication, or comparative effectiveness studies typically associated with AI/ML device submissions.

The document focuses on demonstrating substantial equivalence to predicate devices for traditional medical devices (pacing leads) through:

  • Device Description: Highlighting similarities in design and materials.
  • Biocompatibility Studies: Testing the materials to ensure safety.
  • Qualification Testing: Relying on previous testing of predicate leads.
  • Clinical Summary: Referencing the low incidence of complications and effective performance of existing, similar devices.

Therefore, I cannot fulfill the request as the necessary information for Acceptance Criteria and Study details for an AI/ML device is not present in the provided text.

{0}------------------------------------------------

K555122

NOV 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is:

(To be completed by FDA)

●●●●●●

TABLE OF CONTENTS

A. GENERAL INFORMATION2
B. DESCRIPTION OF THE DISEASES AND CONDITIONS FOR WHICH THEMODELS 436-07 AND 437-07 PACING LEADS ARE INDICATED2
C. DEVICE DESCRIPTION3
1. Labeling4
2. Design and Materials4
3. Performance6
D ALTERNATIVES7
E. POTENTIAL ADVERSE EFFECTS7
F. SUMMARY OF STUDIES8
1. Biocompatibility Studies8
2. Qualification Testing9
G. CLINICAL SUMMARY9
H. MANUFACTURING/STERILIZATION9
I. CONCLUSION10

l

{1}------------------------------------------------

510(K) NOTIFICATION - MODELS 436-07 AND 437-07 ENDOCARDIAL PACING LEADS SECTION 2 - 510(K) SUMMARY , SURGARY OF ADVERSE SAFETY AND EFFECTIVENESS DATA AND CERTIFICATION

팀 홈 홈 홈 바 레

GENERAL INPORMATION A.

Device Classification Name:

Electrode, Pacemaker, Permanent and Temporary

Device Trade Name:

  • Intermedics Model 436-07 Bipolar Implantable Endocardial Pacing Lead i
  • Intermedics Model 437-07 Unipolar Implantable Endocardial Pacing Lead .

Applicant's Name and Address:

Intermedics Inc. (Establishment Registration Number: 1640319) 4000 Technology Drive Angleton, TX 77515

Primary Company Representative:

Alaine Medio, RAC Regulatory Affairs Specialist (409) 848-4528 FAX: (409) 848-4533

Alternate Company Representative:

Kathleen M. Chester Senior Regulatory Affairs Specialist (409) 848-4527 FAX: (409) 848-4533

Date of Notification:

November 7, 1995

B. DESCRIPTION OF THE DISEASES AND CONDITIONS FOR WHICH THE MODELS 436-07 AND 437-07 PACING LEADS ARE INDICATED

The models 436-07 and 437-07 pacing leads are intended for use with implantable cardiac pulse generators for long-term pacing of the heart. The indications for ventricular pacing include, but are not limited to: Sick sinus syndrome, sinus bradycardia, complete heart block, symptomatic second-degree heart block, and certain conditions of asymptomatic second-degree block.

In the presence of normal A-V conduction the indications for atrial pacing include, but are not limited to: Sinus arrest, sick sinus syndrome, sinus

목 및 홈 홈 트 트

{2}------------------------------------------------

510(K) NOTIPICATION - MODELS 436-07 AND 437-07 ENDOCARDIAL PACING LEADS SECTION 2 - 510(K) SUMMARY, SUMMARY OF ADVERSE SAFETY AND EFFECTIVERESS DATA AND CERTIFICATION

일본부분들

bradycardia and conditions requiring increased cardiac efficiency, enhanced cardiac output or the overdrive of certain cardiac arrhythmias.

In the absence of normal A-V conduction, an atrial lead may be used with a ventricular lead in a dual chamber pacing system to restore A-V synchrony.

Contraindications

The use of endocardial leads may be contraindicated in the presence of tricuspid atresia. Ebstein's malformation, and various forms of atrial or ventricular transposition, and in patients with mechanical tricuspid heart valves.

C. Device Description

The Intermedics Models 436-07 (bipolar) and 437-07 (unipolar) endocardial pacing leads are designed for use with implantable cardiac pulse generators configured for ventricular applications for long term cardiac pacing. The nominal lead length for the model 436-07 lead is 58 cm and is available in lengths from 40 to 110 cm. The nominal lead length for the model 437-07 lead is also 58 cm and is available in lengths from 40 to 110 cm. The lead connectors are designed to meet VS-1 specifications (an intra-industry agreement standardizing lead-to-pulse generator connection dimensions) for lead connectors.

The models 436-07 and 437-07 pacing leads are packaged with the model 367-01 vein lifter and the following stylets: model 367-11 (straight stylet, limber), model 365-12 (straight stylet, firm), model 365-81(tapered stylet, limber), and model 365-82 (tapered stylet, firm).

The model 437-07 is also packaged with the following adapters: model 366-29 (Step up adapter, VS-1 unipolar lead to 5 mm unipolar pacer) and model 366-30 (Step up adapter, VS-1 unipolar lead to 6 mm unipolar pacer).

The minimum introducer size recommended for both lead models is 10 French.

The models 436-07 and 437-07 pacing leads are identical to the Intermedics models 436-02 (K883602, K954719) and 437-02 (K883602, K954719) respectively, except for the use of iridium-oxide (IROX) coated titanium for the

행 및 목록 및 원

{3}------------------------------------------------

5 10(K) NOTESCATION - MODELS 436-07 AND 437-07 ENDOCARDIAL PACING LIBADS SECTION 2 - 510(K) SUMMARY, SURMARY OF ADVERSE SAFBTY AND Expectiveness DATA AND CERTIFICATION

를 둘 때 데 있

electrodes. The IROX-coated electrodes are identical to those used in models 430-07 (K902672, K954719) and 431-07 (K902672, K954719).

1. Labeling

The generic Intermedics Endocardial Pacing Leads Package insert accompanies each lead. The sterile package labeling for the leads contains the nominal product specification for the particular lead model enclosed. Other than the change in lead model numbers and electrode material; there will be no changes to the labels and/or labeling for these leads.

2. Design and Materials

  • a. Electrodes
    The cathode tip electrodes transfer the clectrical charge from the pulse generator to the endocardial surface via the pacing lead. This electrical charge stimulates the myocardium, causing cardiac depolarization.

The cathode tip of the model 436-07 and 437-07 leads is constructed of IROX-coated titanium and has a blunt shape with three radial cross slots and contact surface area of approximately 8 mm. This electrode is identical to the cathode tip of the commercially available Intermedics Models 430-07 and 431-07.

The anode sleeve electrode of the model 436-07 (bipolar), also constructed of IROX-coated titanium, has an approximate surface area of 50 mm2 and is 1 inch from the cathode tip. It is identical to the anode sleeve of the commercially available Intermedics Model 430-07.

IROX is a metal oxide of iridium and oxygen. IROX belongs to a family of oxides that has an inherently high surface area because of its sub-micron particle size. Other oxides have sub-micron particle size and a high surface area, but the IROX family has additional beneficial properties that yield a more conductive oxide: they are highly corrosion resistant and undergo reversible oxidation-reduction reactions. IROX-coated titanium is currently used in several Intermedics commercially available bradycardia pacing leads.

속 배 목 회 로

{4}------------------------------------------------

510(K) NOTTFICATION - MODELS 436-07 AND 437-07 ENDOCARDIAL PACING LEADS SECTION 2 - 510(K) SUMMARY, SUBINARY OF ADVERSE SAFETY AND EFFECTIVENESS DATA AND CERTIFICATION

님 님 이 글 홀

b. Lead Body

The conductor coils, constructed of three nickel-cobalt alloy wires wound uniaxially (trifilar), transmit electrical activity to and from the heart. Electrical isolation between the connectors and the body environment is provided by a silicone rubber insulation sheath surrounding the conductor coil. The silicone insulation also contributes to the structural strength of the leads.

The bipolar lead, model 436-07, consists of an inner coll (ioining the connector to the cathode) surrounded by silicone rubber insulation. The second coil (joining the connector to the anode) is wound around the insulation of the first coil. Silicone rubber insulation rubing surrounds this coil and prevents an electrical interface berween the body fluids and the coil. The lead body of the model 436-07 is identical in design and material to the commercially available Intermedics Model 436-02.

The model 437-07 unipolar lead, consisting of one nickel-cobalt alloy coil surrounded by silicone rubber insulation tubing, is identical in design and material to the commercially available Intermedics Model 437-02.

c. Tip Fixation

The passive fixation mechanism of the lead tip lodges among the trabeculae in order to prevent dislodgement or movement of the tip electrode from the endocardium. The tip fixation mechanism, made of silicone rubber, is a trailing tines design with radial projections 45 to the axis of the lead body. This tined lead tip fixation mechanism is identical in design and material to that of the commercially available Intermedics Models 436-02 and 437-02.

d. Connector Assembly

The connector pin, made of 316L stainless steel, provides the electrical and mechanical connection between the pulse generator and pacing lead. The connector sleeve with O-ring seals is made of silicone

{5}------------------------------------------------

510(K) NOT181CATION - MODELS 436-07 AND 437-07 ENDOCARDIAL PACING LEADS SECTION 2 - 510(K) SUMMARY, SURGARY OF ADVERSE SAFETY AND EXPRECTIVERESS DATA AND CERTIFICATION

월 에 올라 내 올라

rubber and provides an insulating seal between the lead and the header of the pulse generator. The connector dimensions are designed to meet the requirements of the VS-1 standard. The connector assembly of the pacing leads is identical in design and material to that of the commercially available Intermedics Models 436-02 and 437-02.

3. Performance

The Intermedics Models 436-07 and 437-07 pacing leads with IROXcoated titanium electrodes are intended for use with implantable pulse generators. Their performance is expected to demonstrate equivalent clinical performance to commercially available unipolar and bipolar ventricular endocardial pacing leads.

D. ALTERNATIVES

The alternatives to the use of these devices are similar to those described for pulse generators. Surgery or drug therapy have been stated as alternates to cardiac pacing in certain instances. However, when a cardiac pacing system is employed, the side effects of drugs and/or the risks of surgery make these alternatives less desirable. Commercially available pacing leads provide another alternative to the use of the Intermedics Models 436-07 and 437-07 endocardial pacing leads.

13. POTENTIAL ADVERSE EFFECTS

The potential adverse effects associated with the use of the Intermedics Models 438-07 and 435-07 endocardial pacing leads, as well as other pacing leads, may include intermittent or continuous loss of pacing or sensing produced by factors suck as displacement of the electrode, unsatisfactory electrode position, breakage of the conductor or its insulation, an increase in thresholds, or poor electrical connection to the pulse generator.

When subclavian venipuncture is used for endocardial lead introduction, "extremely medial" insertion of a lead and/or "anatomic abnormalities" may

{6}------------------------------------------------

5 10(K) NOTIFICATION - MODELS 436-07 AND 437-07 ENDOCARDIAL PACING LEADS SECTION 2 - 510(K) SUMMARY, SOMEARY OF ADVERSE SAFETY AND EFFECTIVENESS DATA AND CERTISCATION

의 발 등 등 용

contribute to conductor fracture'2,344, Perforation of the ventricular wall may cause phrenic nerve stimulation. Perforation of the ventricle may also cause diaphragmatic muscle stimulation. Cardiac tamponade has been reported from instances of lead perforation.

If the connector is cut off when removing an implanted endocardial lead, the lead's insulation tubing (under sufficient traction) may separate from the lead conductor and slide off, leaving an exposed conductor coil in the heart and vein.

As with the introduction of any foreign object into the body, infection can result from the use of endocardial leads.

These complications can occur during implantation, explanation, or at any time postoperatively and may require noninvasive or invasive management techniques.

F. SUMMARY OF STUDIES

1. Biocompatibility Studies

The models 436-07 and 437-07 pacing leads have been evaluated for biocompatibility in both in vitro and in vivo test systems, and were subjected to the following tests:

  • Hemolysis Test
  • MEM Elution Cytology

2 Suzuki Y, Pujimori S, Makoto S, et al. A case of pacemaker lead fracture associated with thoracic outlet syndrome. PACE 1988; 11:326.

Furman S. Venous cutdown for pacemaker implantation. Ann Thorac Surg 1986; 41:438-439.

Furman S. Subclavian puncture for pacemaker lead placement. PACE 1986; 9:467.

*Fyke FB, III. Simultaneous insulation deterioration associated with side-by-side subclavian placement of two polyurethane leads. PACE 1988; 11:1571.

뒤 득 박 옥 로

Stokes K, et al. A possible new complication of subclavian stick: Conductor fracture. PACE 1987; 10:748 (Abstract).

{7}------------------------------------------------

510(K) NOTIFICATION - MODELS 436-07 AND 437-07 ENDOCARDIAL PACING LEADS SECTION 2 - 510(K) SUMMARY , SUNPHARY OF ADVERSE SAFETY AND ERRECTVENESS DATA AND CERTIFICATION

월 목록 및 밀

  • Ames Mutagenicity 트
  • USP Class V
  • Intramuscular Implantation Test
  • Maximization Sensitization Test I
  • USP Pyrogen Test I

Based upon the test data, the materials above were not found to present any toxic liability under physiological conditions. Therefore, the tissue/fluid contacting materials of this pacing system are considered biocompatible.

{8}------------------------------------------------

510(K) NOTERCATION - MODELS 436-07 AND 437-07 ENDOCARDIAL PACING LEADS SECTION 2 - 510(K) SUMMARY, SUMMARY OF ADVERSE SAFETY AND EFFECTIVENESS DATA AND CERTIFICATION

비 역 일 동 동

2. Qualification Testing

Qualification testing was performed on the predicate leads. Since the design of the new leads only incorporates features of the predicate leads (that have already been subjected to qualification testing), no further qualification testing is believed to be necessary.

G. CLINICAL SUMMARY

The clinical experience of the commercially available Intermedics Models 436-02, 437-02, 430-07, and 431-07 endocardial pacing leads has resulted in an extremely low incidence of clinical complications, demonstrating its effective performance in human implantation. Because of the similarity in design and materials, the models 436-07 and 437-07 pacing leads can be expected to perform with comparable efficacy.

H. MANUFACTURING/STERILIZATION

Intermedics utilizes environmental controls in the manufacturing facilities which are designed, maintained and closely monitored to achieve an efficacious environment for manufacturing all products. The manufacturing environment is routinely monitored for particle counts, humidity, temperature, and static electricity controls. Additionally, bioburden testing is performed on all products.

Vendors of purchased material must be "approved" based upon a quality survey in which Intermedics' field engineers determine whether the vendor has the capability of consistently supplying material that will meet Intermedics' standards. Each quantity of purchased material is assigned a lot number and acceptance or rejection is determined based upon an inspection (conducted on a statistical sampling basis) performed to Intermedics' Engineering specifications.

The manufacturing process for the models 436-07 and 437-07 consists of the following assembly processes: connector assembly, molded O-ring assembly, sandblast/coat/mask anode sleeve assembly (model 436-07 only), tip assembly, and final assembly.

{9}------------------------------------------------

S10(K) NOTITICATION - MODELS 436-07 AND 437-07 ENDOCARDIAL PACING LEADS SECTION 2 - SIO(K) SUMMARY, SUMMARY OF ADVERSE SAFETY AND EFFECTIVENESS DATA AND CERTIFICATION

동 대 발 응원 및

The finished pacing lead and accessories are packaged in the formed pockets of an inner blister tray which is closed by heat sealing a peelable cover around the periphery. This inner blister is then placed into an outer blister which is closed by heat sealing a second peelable cover to it. Cover material is microbial penetration resistant, water resistant, and puncture resistant as suitable for this application.

Intermedics sterilization procedures for pacing leads utilize ethylene oxide (EO) sterilizers set for specific parameters derived using methods described in the 1988 AAMI guideline, Guideline for Industrial Ethylene Oxide Sterilization of Medical Devices.

Upon release from sterilization, pacing lead and accessories are packaged into a carton and the boxed product is inspected prior to being routed to the Finished Goods Inventory area.

I. CONCLUSION

The information presented in this submission for the Intermedics Models 436-07 (bipolar) and 437-07 (unipolar) pacing leads demonstrates that the intended use of these devices does not differ from that of the commercially available models 436-02, 437-02, 430-07, and 431-07.

The clinical experience of the Intermedics endocardial pacing leads has resulted in an extremely low incidence of clinical complications, demonstrating their effective performance in human implantation. Because of the similarity in design and materials, the models 436-07 and 437-07 pacing leads can be expected to perform with comparable efficacy.

§ 870.3680 Cardiovascular permanent or temporary pacemaker electrode.

(a)
Temporary pacemaker electrode —(1)Identification. A temporary pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to anexternal pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class II (performance standards).(b)
Permanent pacemaker electrode —(1)Identification. A permanent pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class III (premarket approval).(c)
Date PMA or notice of completion of PDP is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before October 4, 2012, for any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976, or that has, on or before October 4, 2012, been found to be substantially equivalent to any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976. Any other pacemaker repair or replacement material device shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.