MIDAS REX MOTORS, (MIDAS I, MIDAS II & CONVERTIBLE MODELS)
K953434 · Midas Rex Pneumatic Tools, Inc. · HBB · May 16, 1996 · Neurology
Device Facts
Record ID
K953434
Device Name
MIDAS REX MOTORS, (MIDAS I, MIDAS II & CONVERTIBLE MODELS)
Applicant
Midas Rex Pneumatic Tools, Inc.
Product Code
HBB · Neurology
Decision Date
May 16, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4370
Device Class
Class 2
Attributes
Therapeutic
Intended Use
To provide power to operate removable rotating surgical cutting tools or drill bits on a patient's skull.
Device Story
Midas Rex Motors (Midas I, Midas II, Convertible) are air-powered surgical motors; designed to drive rotating cutting tools or drill bits for cranial surgery. System components include stainless steel/aluminum/brass motor housing, air supply hose, and joint connectors. Operates at 20-120 psi or 200 psi. Features include optional rotating cuffs at joints to allow independent swivel of motor and hose, preventing kinking. Used in clinical surgical settings by physicians. Device provides mechanical power to surgical attachments; enables bone/tissue cutting. Supplied non-sterile; requires hospital-performed sterilization (steam or ethylene oxide) prior to use.
Pneumatic motor; stainless steel, aluminum, and brass construction. Operating pressure: 20-120 psi or 200 psi. RPM: 30,000-75,000. Stall torque: 1-5.3 in-oz. Form factor: round or hexagonal casing (0.375, 0.5, or 0.75 inch diameter). Connectivity: air hose connection with optional swivel joints. Sterilization: compatible with ethylene oxide, high vacuum steam (270° F), and gravity displacement steam (250° F or 270° F).
Indications for Use
Indicated for use in surgical procedures requiring the operation of rotating cutting tools or drill bits on a patient's skull.
Regulatory Classification
Identification
A pneumatic cranial drill motor is a pneumatically operated power source used with removable rotating surgical cutting tools or drill bits on a patient's skull.
K962483 — MIDAS REX III MOTOR · Midas Rex Pneumatic Tools, Inc. · Nov 19, 1996
K980686 — AESCULAP HILAN MOTOR SYSTEM · Aesculap, Inc. · Sep 2, 1998
K972289 — MIDAS REX MOTOR · Midas Rex Pneumatic Tools, Inc. · Sep 12, 1997
K083112 — PRIMADO · Nakanishi, Inc. · Apr 29, 2009
K150801 — Stryker Maestro Air Motors · Stryker Corporation · May 22, 2015
Submission Summary (Full Text)
{0}
K953434
# 510 (k) SUMMARY
MAY 16 1995
## I. ADMINISTRATIVE
Submitter: Midas Rex Pneumatic Tools, Inc.
3001 Race Street
Fort Worth, Texas 76111
(817) 831-2604
Contact Person: Forest Barber, Jr.
Date of Preparation: July 20, 1995
## II. DEVICE NAME
Proprietary Name: Midas Rex Motors (Midas I, Midas II and Convertible Models)
Common Name: Surgical Instrument Motor
Classification Name: Pneumatic Cranial Drill Motor (84HBB)
## III. PREDICATE DEVICE
Midas Rex Whirlwind™ Motor, legally marketed prior to May 28, 1976.
## IV. DEVICE DESCRIPTION
The Midas Rex Motors (Midas I, Midas II and Convertible models) are air-powered, rotating motors designed to operate at 20-120 psi or at up to 200 psi. The motor is connected to an air-supply source by means of a stainless steel joint and a plastic air hose with inner rubber pressure hose.
The motors are constructed of stainless steel, aluminum and brass and are designed to operate at 20-120 psi or at up to 200 psi with the same performance characteristics. Depending on user preference, the 200 psi motor is available in either a round or hexagonal casing, approximately 0.75 inch in diameter. Similarly, the 120 psi motor can be supplied in either a round or hexagonal casing in diameters of approximately 0.375, 0.5 or 0.75 inch.
Again depending on user preference, the joint connecting the motor to the air hose can be equipped with a rotating cuff at one or both ends, allowing the motor and/or air hose to swivel independently with respect to the joint, thus providing greater freedom of movement during surgical procedures and avoiding possible "kinking" of the air hose. Thus, a variety of device configurations are possible based on the surgeon's personal preference, as shown below:
Premarket Notification: Midas Rex Motors
{1}
| | Motor | Joint | Hose |
| --- | --- | --- | --- |
| Midas I | Fixed | Straight (180°) | Fixed |
| | Swivels | Straight (180°) | Swivels |
| Midas II | Fixed | Angled (45°) | Fixed |
| | Fixed | Angled (45°) | Swivels |
| | Swivels | Angled (45°) | Fixed |
| | Swivels | Angled (45°) | Swivels |
| Convertible | Fixed | 180° or 45° | Fixed |
| | Fixed | 180° or 45° | Swivels |
| | Swivels | 180° or 45° | Fixed |
| | Swivels | 180° or 45° | Swivels |
Midas Rex Motors are supplied non-sterile for sterilization by hospital staff immediately prior to each use. The materials of construction of the motor and air hose are compatible with both ethylene oxide and steam sterilization (flash or standard autoclave procedures). A 4-minute cycle is recommended for high vacuum steam sterilization at 270° F. Gravity displacement steam sterilization may be performed at 250° F for 35 minutes or at 270° F for 15 minutes. A minimum of 18 hours mechanical aeration is recommended following ethylene oxide sterilization.
## V. INTENDED USE
To provide power to operate removable rotating surgical cutting tools or drill bits on a patient's skull.
## VI. COMPARISON TO PREDICATE DEVICES
See Table.
Premarket Notification: Midas Rex Motors
{2}
Table of Comparison to Legally Marketed Device
| | Midas I | Midas II | Convertible | Midas I, II and Convertible (K952300) | Whirlwind™ (pre-1976) |
| --- | --- | --- | --- | --- | --- |
| Construction | Stainless steel, brass and aluminum | Stainless steel, brass and aluminum | Stainless steel, brass and aluminum | Stainless steel, brass and aluminum | Stainless steel, brass and aluminum |
| Case Shape/Diameter | Round or hexagonal: 0.375, 0.5 or 0.75 inch | Round or hexagonal: 0.375, 0.5 or 0.75 inch | Round or hexagonal: 0.375, 0.5 or 0.75 inch | Round or hexagonal: 0.375, 0.5 or 0.75 inch | Round/0.75 inch |
| Motor/Air hose Joint | Straight | Angled (45°) | Straight or Angled (swivel joint) | Straight, Angled and Swivel Joint | Straight |
| Motor Alignment (relative to joint) | Fixed or Swivel | Fixed or Swivel | Fixed or Swivel | Fixed or Swivel | Fixed |
| Hose Alignment (relative to joint) | Fixed or Swivel | Fixed or Swivel | Fixed or Swivel | Fixed or Swivel | Fixed |
| Air Operating Pressure | 20-120 psi or 200 psi | 20-120 psi or 200 psi | 20-120 psi or 200 psi | 20-120 psi or 200 psi | 200 psi |
| RPM (000's) | 40-75 (20-120 psi)
30-75 (200 psi) | 40-75 (20-120 psi)
30-75 (200 psi) | 40-75 (20-120 psi)
30-75 (200 psi) | 40-75 (20-120 psi)
30-75 (200 psi) | 30-75 |
| Stall Torque | 1-5.3 (20-120 psi)
1.3-5.3 (200 psi) | 1-5.3 (20-120 psi)
1.3-5.3 (200 psi) | 1-5.3 (20-120 psi)
1.3-5.3 (200 psi) | 1-5.3 (20-120 psi)
1.3-5.3 (200 psi) | 1.3-5.3 |
| Intended Use | To provide power to operate removable rotating surgical cutting tools or drill bits on a patient's skull. | To provide power to operate removable rotating surgical cutting tools or drill bits on a patient's skull. | To provide power to operate removable rotating surgical cutting tools or drill bits on a patient's skull. | To provide power to operate various accessories or attachments to cut hard tissue or bone and soft tissue. | To provide power to operate various accessories or attachments to cut hard tissue or bone and soft tissue. |
Premarket Notification: Midas Rex Motors
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.