PERCEPTOR COMPENSATOR MORSE MANIFOLD
Device Facts
| Record ID | K951722 |
|---|---|
| Device Name | PERCEPTOR COMPENSATOR MORSE MANIFOLD |
| Applicant | Namic |
| Product Code | DRS · Cardiovascular |
| Decision Date | Mar 25, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.2850 |
| Device Class | Class 2 |
Indications for Use
The PERCEPTOR® Compensator MORSE® Manifold, like its predicates, the MORSE® DT Manifold and the Stand Alone Transducer, is a physiological blood pressure transducer. Both devices are intended for invasive pressure monitoring, catheterization procedures, fluid delivery and/or blood access maintenance.
Device Story
Physiological blood pressure transducer integrated into manifold side port lumen; catheter fluid column transmits heart pressure to transducer; gel membrane passes signal to sensor; signal transmitted via cable to monitor. Used in clinical settings for invasive monitoring/catheterization. Compensator design redirects lumen above/below sensor to enable accurate zero balancing and height-independent monitoring. Discontinued over-pressure protection band. Output used by clinicians for real-time hemodynamic monitoring.
Clinical Evidence
Bench testing only; biocompatibility testing performed on predicate devices; no changes in device materials.
Technological Characteristics
Physiological pressure transducer; integrated into manifold side port; gel membrane sensor; ANSI/AAMI BP22-1994 compliant; materials identical to predicate devices.
Indications for Use
Indicated for patients requiring invasive blood pressure monitoring, catheterization, fluid delivery, or blood access maintenance.
Regulatory Classification
Identification
An extravascular blood pressure transducer is a device used to measure blood pressure by changes in the mechanical or electrical properties of the device. The proximal end of the transducer is connected to a pressure monitor that produces an analog or digital electrical signal related to the electrical or mechanical changes produced in the transducer.
Predicate Devices
- MORSE® DT Manifold
- Stand Alone Transducer
Related Devices
- K152472 — Disposable Pressure Transducer · Zhejiang Haisheng Medical Device Co., Ltd. · Oct 31, 2016
- K102033 — COMBITRANS MONITORING SETS AND ACCESSORIES · B.Braun Melsungen AG · Apr 7, 2011
- K091408 — DISPOSABLE PRESSURE TRANSDUCER · Shenzhen Ant Hi-Tech Industrial Co., Ltd. · Aug 18, 2009
- K242909 — FloTrac sensors; FloTrac Jr sensors; Acumen IQ sensors; VolumeView sensors · Edwards Lifesciences, LLC · Mar 21, 2025
- K120624 — ELCAM DISPOSABLE INTEGRATED PRESSURE TRANSDUCER (DIPT) · Elcam Medical Acal · Jun 8, 2012