(108 days)
The MiniMed Flex pump is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.
The MiniMed Flex pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices.
The MiniMed Flex pump is indicated for use in persons 7 years of age and older.
The MiniMed Flex pump is intended for single patient use and requires a prescription.
The MiniMed Flex pump is an alternate controller enabled (ACE) pump intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. It can reliably and securely communicate with compatible digitally connected devices, including an integrated continuous glucose monitor (iCGM), interoperable Medtronic continuous glucose monitor (CGM), and interoperable automated glycemic controller (iAGC). The MiniMed Flex pump is intended to be used in conjunction with compatible, digitally connected medical devices (i.e., iAGC and CGM) for the purpose of drug delivery. The MiniMed Flex pump is provided with compatible iAGCs installed, refer to the compatible iAGCs for indications related to iAGC use with MiniMed Flex pump.
The MiniMed Flex pump is an ambulatory, battery-operated, rate-programmable micro-infusion pump. It is a screenless, tubed pump that houses electronics, a pumping mechanism, and a medication reservoir within the same physical device. It is intended for use with compatible insulin reservoir and insulin infusion set for delivery of basal and bolus insulin according to settings selected or inputs entered by the user or compatible iAGC software based on healthcare provider recommendations.
The MiniMed Flex pump is an interoperable device that communicates via a Bluetooth Low Energy (BLE) wireless electronic interface with digitally connected devices. The MiniMed Flex pump is able to integrate the compatible MiniMed iAGC algorithms into the pump firmware to receive, execute, and confirm commands from an iAGC to adjust delivery of insulin. The pump receives sensor glucose (SG) data from a compatible iCGM or a compatible interoperable Medtronic CGM via BLE and transmits the SG and other CGM data to the iAGCs.
The MiniMed Flex pump is a screenless pump with minimal user interface and manual control capabilities available from the pump itself. The primary user interface and primary display for the MiniMed Flex pump is the MiniMed app, also the user interface for the compatible MiniMed iAGCs, which runs on a compatible iOS or Android device.
N/A
FDA 510(k) Clearance Letter - MiniMed Flex Pump
Page 1
March 12, 2026
Medtronic Minimed, Inc.
Omar Becerra
Principal Regulatory Affairs Specialist
18000 Devonshire St.
Northridge, California 91325
Re: K253743
Trade/Device Name: MiniMed Flex pump
Regulation Number: 21 CFR 880.5730
Regulation Name: Alternate Controller Enabled Infusion Pump
Regulatory Class: Class II
Product Code: QFG
Dated: November 21, 2025
Received: November 24, 2025
Dear Omar Becerra:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not
Page 2
required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Page 3
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JOSHUA BALSAM -S
Joshua M. Balsam, Ph.D.
Branch Chief
Division of Chemistry and
Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
March 12, 2026
Medtronic Minimed, Inc.
Omar Becerra
Principal Regulatory Affairs Specialist
18000 Devonshire St.
Northridge, California 91325
Re: K253743
Trade/Device Name: MiniMed Flex pump
Regulation Number: 21 CFR 880.5730
Regulation Name: Alternate Controller Enabled Infusion Pump
Regulatory Class: Class II
Product Code: QFG
Dated: November 21, 2025
Received: November 24, 2025
Dear Omar Becerra:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not
Page 5
required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Page 6
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JOSHUA BALSAM -S
Joshua M. Balsam, Ph.D.
Branch Chief
Division of Chemistry and
Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 7
510(K) Summary - MiniMed Flex Pump
510(k) Submitter and Device Information
| Submitter's Name and Address | Medtronic MiniMed, Inc.18000 Devonshire StNorthridge, CA 91325 USA |
|---|---|
| Primary Contact Person | Omar BecerraPrincipal Regulatory Affairs SpecialistMedtronic MiniMed Inc.Omar.Becerra@medtronic.com |
| Alternate Contact Person | Siddhi RasamSr Manager, Regulatory AffairsMedtronic MiniMed Inc.Siddhi.Rasam@medtronic.com |
| Device Trade Name | MiniMed Flex Pump |
| Device Common Name | Alternate Controller Enabled Insulin Infusion Pump (ACE Pump) |
| Device Classification Name | Alternate Controller Enabled (ACE) Insulin Infusion Pump |
| Regulation Number | 21 CFR 880.5730 |
| Product Codes | QFG |
| Device Panel | Clinical Chemistry |
| Device Class | Class II |
| Predicate Device | K251032 - MiniMed 780G Flex Pump with Interoperable Technology |
Device Description
The MiniMed Flex pump is an alternate controller enabled (ACE) pump intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. It can reliably and securely communicate with compatible digitally connected devices, including an integrated continuous glucose monitor (iCGM), interoperable Medtronic continuous glucose monitor (CGM), and interoperable automated glycemic controller (iAGC). The MiniMed Flex pump is intended to be used in conjunction with compatible, digitally connected medical devices (i.e., iAGC and CGM) for the purpose of drug delivery. The MiniMed Flex pump is provided with compatible iAGCs installed, refer to the compatible iAGCs for indications related to iAGC use with MiniMed Flex pump.
The MiniMed Flex pump is an ambulatory, battery-operated, rate-programmable micro-infusion pump. It is a screenless, tubed pump that houses electronics, a pumping mechanism, and a
Page 8
medication reservoir within the same physical device. It is intended for use with compatible insulin reservoir and insulin infusion set for delivery of basal and bolus insulin according to settings selected or inputs entered by the user or compatible iAGC software based on healthcare provider recommendations.
The MiniMed Flex pump is an interoperable device that communicates via a Bluetooth Low Energy (BLE) wireless electronic interface with digitally connected devices. The MiniMed Flex pump is able to integrate the compatible MiniMed iAGC algorithms into the pump firmware to receive, execute, and confirm commands from an iAGC to adjust delivery of insulin. The pump receives sensor glucose (SG) data from a compatible iCGM or a compatible interoperable Medtronic CGM via BLE and transmits the SG and other CGM data to the iAGCs.
The MiniMed Flex pump is a screenless pump with minimal user interface and manual control capabilities available from the pump itself. The primary user interface and primary display for the MiniMed Flex pump is the MiniMed app, also the user interface for the compatible MiniMed iAGCs, which runs on a compatible iOS or Android device.
Indications for Use / Intended Use
The MiniMed Flex pump is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.
The MiniMed Flex pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices.
The MiniMed Flex pump is indicated for use in persons 7 years of age and older.
The MiniMed Flex pump is intended for single patient use and requires a prescription.
Summary of Technological Characteristics of Subject Device Compared to Predicate Device
The table below provides a side-by-side comparison of the subject device, MiniMed Flex pump compared to its predicate device, MiniMed 780G insulin pump.
Page 9
Table 1: Technological Characteristics Comparison between MiniMed Flex pump (Subject Device) and MiniMed 780G insulin pump (Predicate Device, cleared in K251032)
| Predicate Device (K251032)MiniMed 780G insulin pump | Subject Device (K253743)MiniMed Flex pump | |
|---|---|---|
| Device Classification | Alternate Controller Enabled (ACE) Insulin Infusion Pump | SAME |
| Product Code | QFG | SAME |
| Device Type/Regulation | 21 CFR 880.5730 | SAME |
| Indications for Use/Intended Use | The MiniMed 780G insulin pump is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.The MiniMed 780G insulin pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices.The MiniMed 780G insulin pump contains a bolus calculator that calculates an insulin dose based on user-entered data.The MiniMed 780G insulin pump is indicated for use in individuals 7 years of age and older.The MiniMed 780G insulin pump is intended for single patient use and requires a prescription. | The MiniMed Flex pump is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.The MiniMed Flex pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices.The MiniMed Flex pump is indicated for use in individuals 7 years of age and older.The MiniMed Flex pump is intended for single patient use and requires a prescription. |
| Prescription Use | Prescription is required | SAME |
| Environment of Use | Professional healthcare facilities and home environments | SAME |
| Patient Environment | On-body wearable ambulatory pump | SAME |
| Intended Population | Persons with diabetes mellitus ages 7 and up | SAME |
| Technological Characteristics | The MiniMed 780G insulin pump is an ambulatory, battery-operated, rate-programmable micro-infusion pump designed for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The device houses electronics, a pumping mechanism, a user interface, and a medication reservoir to deliver patient-programmed basal | The MiniMed Flex pump is a screenless ambulatory, battery-operated, rate-programmable micro-infusion pump designed for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The device houses electronics, a pumping mechanism, and a medication reservoir to deliver patient-programmed basal rates and |
Page 10
| Predicate Device (K251032)MiniMed 780G insulin pump | Subject Device (K253743)MiniMed Flex pump | |
|---|---|---|
| rates and boluses through an infusion set into the subcutaneous tissue. | boluses through an infusion set into the subcutaneous tissue. | |
| Specific Drug/Biological Use | U-100 insulin¹:Novolog®Humalog®Admelog® | U-100 insulin:Novolog®Humalog®Admelog®Fiasp®Lyumjev® |
| Principles of Operation | Delivery of insulin (Bolus and Basal) programmed by the patient or compatible iAGC software based on healthcare provider recommendations. | SAME |
| Pump Operating Modes | Manual ModeSmartGuard Mode | SAME |
| Pump Device Accessories | MiniMed Mobile AppCompatible FDA cleared infusion setsCompatible FDA cleared reservoirsBelt ClipPump Covers/Cases | MiniMed Flex ChargerCompatible FDA cleared infusion setsCompatible FDA cleared reservoirsBelt ClipPump Covers/Sleeves |
| Compatible Interoperable Devices | Integrated Continuous Glucose Monitors (iCGMs)Interoperable Automated Glycemic Controllers (iAGCs)Compatible Interoperable Medtronic Continuous Glucose Monitors (CGMs) | SAME |
| Compatible Connected Devices | N/A | MiniMed App (installed on compatible smartphone or App Manager) |
| Communication with Compatible Devices | Bluetooth Low Energy (BLE) | SAME |
| Battery Type/ Power Requirements | The MiniMed 780G pump requires an internal lithium-ion rechargeable battery plus one AA (1.5V) battery | The MiniMed Flex pump includes an internal lithium-ion rechargeable battery. |
| Pump Screen/Controls | Liquid Crystal Display (LCD) Screen + Keypad | The Flex pump is a screenless device. Users receive safety critical alerts from the pump's lights and speaker. The user has limited interaction with the pump through two buttons (the action button and the acknowledge button) and the status light. Alerts are communicated to the MiniMed app where notifications are presented for the user to review and acknowledge. |
¹ MiniMed 780G pump use with Fiasp and Lyumjev was cleared in K253470.
Page 11
| Predicate Device (K251032)MiniMed 780G insulin pump | Subject Device (K253743)MiniMed Flex pump | |
|---|---|---|
| Wireless control of bolus insulin therapy | The MiniMed 780G insulin pump does not have the capability to allow users to wirelessly interact with insulin therapy via a mobile app. | The MiniMed Flex pump does have the capability to allow users to wirelessly interact with insulin therapy via the MiniMed app. |
| Predetermined changed control plan (PCCP) | The MiniMed 780G insulin pump was cleared with an authorized PCCP that included modifications for integrating with potential additional commercialized interoperable devices in the future. | SAME |
Summary of Non-Clinical Performance Data
Medtronic MiniMed conducted extensive performance bench testing for MiniMed Flex pump to demonstrate substantial equivalence to the predicate device and to ensure that the subject device meets all applicable ACE Special Controls requirements defined in 21 CFR 880.5730. These are summarized below:
Special Controls:
Evaluation and adherence to the Special Controls of the Predicate Device (K251032) demonstrates continued assurance of the safety and effectiveness of the Subject Device
Delivery Volume Accuracy
Delivery volume accuracy (DVA) testing for the MiniMed Flex pump was conducted to assess the delivery volume accuracy performance of the pump for foreseeable use conditions as required by the ACE special controls requirements stated in 21 CFR 880.5730(b). The test results support that the device is safe to use under foreseeable use conditions and expected environments.
Catheter Occlusion Detection
Catheter Occlusion Detection testing was conducted with U100 insulins (Humalog, NovoLog, Admelog, Fiasp, and Lyumjev) to demonstrate the incidence of catheter blockage due to insulin crystallization. The test results showed there were no drug-fluid-path occlusions and were confirmed to be acceptable.
Drug Stability and Compatibility
Drug stability and compatibility testing was performed with U100 insulins (Humalog, NovoLog, Admelog, Fiasp, and Lyumjev) used with the reservoir and infusion sets compatible to MiniMed Flex pump. The test results demonstrated that the compatible
Page 12
reservoir and infusion sets do not adversely affect the insulins being delivered, and that the insulin types do not adversely affect the pump.
Software Verification and Validation
Software verification activities were performed in accordance with ISO 14971:2019 "Medical Devices - Application of Risk Management to Medical Devices," IEC 62304:2006/A1:2016, "Medical Device Software - Software Life Cycle Processes," and FDA guidance "Content of Premarket Submissions for Device Software Functions" (June 2023).
Additionally, cybersecurity activities were all completed per cybersecurity plan and cybersecurity risks were assessed for impact to confidentiality, integrity, and availability. A robust cybersecurity risk assessment was conducted, all cybersecurity risks with potential to impact safety were mitigated.
Interoperability
Interoperability documentation was provided in accordance with FDA Guidance "Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices" (September 2017) and the requirements defined by the ACE special controls 21 CFR 880.5730(b)(3)-(5).
Human Factors Validation
A human factors and usability engineering process was performed on MiniMed Flex pump with compatible Medtronic CGMs and compatible MiniMed iAGCs in accordance with IEC 62366-1:2015, HE75:2009 and FDA's guidance document, "Applying Human Factors and Usability Engineering to Medical Devices" (February 2016). Results of the human factors validation testing demonstrated that the device is safe and effective for the intended users, intended uses and expected tasks, and intended use environments.
Labeling
The MiniMed Flex pump labeling for users and healthcare practitioners is sufficient and satisfies applicable requirements of 21 CFR 801.
Page 13
Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical Safety and EMC testing were conducted on MiniMed Flex pump in accordance with IEC 60601-1, IEC 60601-1-2, collateral standards of IEC 60601-1, and additional standards as applicable.
Pre-determined Change Control Plan (PCCP)
A Pre-determined change control plan (PCCP) for planned modifications to the MiniMed Flex Pump, was provided in accordance with the FDA Draft Guidance, "Predetermined Change Control Plans for Medical Devices" (August 2024). It included modifications for integrating with potential additional commercialized interoperable devices in the future. The PCCP included a description of modifications, a modification protocol, traceability from modifications to the modification protocols and an impact assessment.
Conclusion
The MiniMed Flex pump has the same intended use and indications as the Predicate Device, MiniMed 780G insulin pump, cleared in K251032. The required technical documentation provided in this 510(k) demonstrates the MiniMed Flex pump is as safe and effective as the Predicate Device. Furthermore, the subject device meets all the Special Controls requirements for Alternate controller enabled infusion pump defined in 21 CFR 880.5730. Therefore, the MiniMed Flex pump has been evaluated to be substantially equivalent to the Predicate Device and does not raise new or different questions of safety or effectiveness.
§ 880.5730 Alternate controller enabled infusion pump.
(a)
Identification. An alternate controller enabled infusion pump (ACE pump) is a device intended for the infusion of drugs into a patient. The ACE pump may include basal and bolus drug delivery at set or variable rates. ACE pumps are designed to reliably and securely communicate with external devices, such as automated drug dosing systems, to allow drug delivery commands to be received, executed, and confirmed. ACE pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of drug delivery.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Evidence demonstrating that device infusion delivery accuracy conforms to defined user needs and intended uses and is validated to support safe use under actual use conditions.
(A) Design input requirements must include delivery accuracy specifications under reasonably foreseeable use conditions, including ambient temperature changes, pressure changes (
e.g., head-height, backpressure, atmospheric), and, as appropriate, different drug fluidic properties.(B) Test results must demonstrate that the device meets the design input requirements for delivery accuracy under use conditions for the programmable range of delivery rates and volumes. Testing shall be conducted with a statistically valid number of devices to account for variation between devices.
(ii) Validation testing results demonstrating the ability of the pump to detect relevant hazards associated with drug delivery and the route of administration (
e.g., occlusions, air in line, etc.) within a clinically relevant timeframe across the range of programmable drug delivery rates and volumes. Hazard detection must be appropriate for the intended use of the device and testing must validate appropriate performance under the conditions of use for the device.(iii) Validation testing results demonstrating compatibility with drugs that may be used with the pump based on its labeling. Testing must include assessment of drug stability under reasonably foreseeable use conditions that may affect drug stability (
e.g., temperature, light exposure, or other factors as needed).(iv) The device parts that directly or indirectly contact the patient must be demonstrated to be biocompatible. This shall include chemical and particulate characterization on the final, finished, fluid contacting device components demonstrating that risk of harm from device-related residues is reasonably low.
(v) Evidence verifying and validating that the device is reliable over the ACE pump use life, as specified in the design file, in terms of all device functions and in terms of pump performance.
(vi) The device must be designed and tested for electrical safety, electromagnetic compatibility, and radio frequency wireless safety and availability consistent with patient safety requirements in the intended use environment.
(vii) For any device that is capable of delivering more than one drug, the risk of cross-channeling drugs must be adequately mitigated.
(viii) For any devices intended for multiple patient use, testing must demonstrate validation of reprocessing procedures and include verification that the device meets all functional and performance requirements after reprocessing.
(2) Design verification and validation activities must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following:
(i) Risk control measures shall be implemented to address device system hazards and the design decisions related to how the risk control measures impact essential performance shall be documented.
(ii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design.
(3) The device shall include validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following:
(i) Secure authentication (pairing) to external devices.
(ii) Secure, accurate, and reliable means of data transmission between the pump and connected devices.
(iii) Sharing of necessary state information between the pump and any digitally connected alternate controllers (
e.g., battery level, reservoir level, pump status, error conditions).(iv) Ensuring that the pump continues to operate safely when data is received in a manner outside the bounds of the parameters specified.
(v) A detailed process and procedure for sharing the pump interface specification with digitally connected devices and for validating the correct implementation of that protocol.
(4) The device must include appropriate measures to ensure that safe therapy is maintained when communications with digitally connected alternate controller devices is interrupted, lost, or re-established after an interruption (
e.g., reverting to a pre-programmed, safe drug delivery rate). Validation testing results must demonstrate that critical events that occur during a loss of communications (e.g., commands, device malfunctions, occlusions, etc.) are handled appropriately during and after the interruption.(5) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the system must, at a minimum, include:
(i) A record of all drug delivery
(ii) Commands issued to the pump and pump confirmations
(iii) Device malfunctions
(iv) Alarms and alerts and associated acknowledgements
(v) Connectivity events (
e.g., establishment or loss of communications)(6) Design verification and validation must include results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use.
(7) Device labeling must include the following:
(i) A prominent statement identifying the drugs that are compatible with the device, including the identity and concentration of those drugs as appropriate.
(ii) A description of the minimum and maximum basal rates, minimum and maximum bolus volumes, and the increment size for basal and bolus delivery, or other similarly applicable information about drug delivery parameters.
(iii) A description of the pump accuracy at minimum, intermediate, and maximum bolus delivery volumes and the method(s) used to establish bolus delivery accuracy. For each bolus volume, pump accuracy shall be described in terms of the number of bolus doses measured to be within a given range as compared to the commanded volume. An acceptable accuracy description (depending on the drug delivered and bolus volume) may be provided as follows for each bolus volume tested, as applicable: Number of bolus doses with volume that is 250 percent of the commanded amount.
(iv) A description of the pump accuracy at minimum, intermediate, and maximum basal delivery rates and the method(s) used to establish basal delivery accuracy. For each basal rate, pump accuracy shall be described in terms of the amount of drug delivered after the basal delivery was first commanded, without a warmup period, up to various time points. The information provided must include typical pump performance, as well as worst-case pump performance observed during testing in terms of both over-delivery and under-delivery. An acceptable accuracy description (depending on the drug delivered) may be provided as follows, as applicable: The total volume delivered 1 hour, 6 hours, and 12 hours after starting delivery for a typical pump tested, as well as for the pump that delivered the least and the pump that delivered the most at each time point.
(v) A description of delivery hazard alarm performance, as applicable. For occlusion alarms, performance shall be reported at minimum, intermediate, and maximum delivery rates and volumes. This description must include the specification for the longest time period that may elapse before an occlusion alarm is triggered under each delivery condition, as well as the typical results observed during performance testing of the pumps.
(vi) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device.
(vii) For any infusion pumps intended for multiple patient reuse, instructions for safely reprocessing the device between uses.