(112 days)
The Healgen® Accurate Oral Fluid Drug Test is a competitive binding lateral flow immunochromatographic assay for the qualitative and simultaneous detection of Amphetamine, Cocaine, Marijuana, Methadone, Methamphetamine, Opiates, Oxycodone, and Phencyclidine in human oral fluid at the cutoff concentrations listed below and their metabolites:
| Test | Calibrator | Cutoff (ng/mL) |
|---|---|---|
| Amphetamine (AMP) | d-Amphetamine | 50 |
| Cocaine (COC) | Benzoylecgonine | 20 |
| Marijuana (THC) | Delta-9-Tetrahydrocannabinol | 40 |
| Methadone (MTD) | Methadone | 30 |
| Methamphetamine (MET) | d-Methamphetamine | 50 |
| Opiates (OPI) | Morphine | 40 |
| Oxycodone (OXY) | Oxycodone | 20 |
| Phencyclidine (PCP) | Phencyclidine | 10 |
The test provides only preliminary test results. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Liquid Chromatography/Mass Spectrometry/ Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. It is not intended to distinguish between prescription use or abuse of the drug. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
The Healgen® Accurate Oral Fluid Drug Test is a lateral flow chromatographic immunoassay for the qualitative detection of Amphetamine, Cocaine, marijuana, Methadone, Methamphetamine, Opiates, Oxycodone and Phencyclidine in oral fluids. The tests are the first step in a two-step process. The second step is to send the sample for laboratory testing if preliminary positive results are obtained.
N/A
U.S. Food & Drug Administration 510(k) Clearance Letter
Page 1
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
Doc ID # 04017.08.04
March 16, 2026
Healgen Scientific, LLC
℅ Jenny Xia
Director
LSI International, Inc.
504e Diamond Ave., Suite H
Gaithersburg, Maryland 20877
Re: K253705
Trade/Device Name: Healgen® Accurate Oral Fluid Drug Test
Regulation Number: 21 CFR 862.3610
Regulation Name: Methamphetamine test system
Regulatory Class: Class II
Product Code: DJC, DKZ, DIO, LDJ, LCM, DJG, DJR
Dated: November 16, 2025
Received: November 24, 2025
Dear Jenny Xia:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Page 2
K253705 - Xia Jenny Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
Page 3
K253705 - Xia Jenny Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JOSEPH A. KOTAREK -S
Digitally signed by JOSEPH A. KOTAREK -S
Date: 2026.03.16 10:45:20 -04'00'
Joseph Kotarek
Branch Chief
Division of Chemistry and Toxicology Devices
OHT7: Office of In Vitro Diagnostics
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
Page 4
FORM FDA 3881 (8/23) Page 1 of 1
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known): K253705
Device Name: Healgen® Accurate Oral Fluid Drug Test
Indications for Use (Describe)
The Healgen® Accurate Oral Fluid Drug Test is a competitive binding lateral flow immunochromatographic assay for the qualitative and simultaneous detection of Amphetamine, Cocaine, Marijuana, Methadone, Methamphetamine, Opiates, Oxycodone, and Phencyclidine in human oral fluid at the cutoff concentrations listed below and their metabolites:
| Test | Calibrator | Cutoff (ng/mL) |
|---|---|---|
| Amphetamine (AMP) | d-Amphetamine | 50 |
| Cocaine (COC) | Benzoylecgonine | 20 |
| Marijuana (THC) | Delta-9-Tetrahydrocannabinol | 40 |
| Methadone (MTD) | Methadone | 30 |
| Methamphetamine (MET) | d-Methamphetamine | 50 |
| Opiates (OPI) | Morphine | 40 |
| Oxycodone (OXY) | Oxycodone | 20 |
| Phencyclidine (PCP) | Phencyclidine | 10 |
The test provides only preliminary test results. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Liquid Chromatography/Mass Spectrometry/ Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. It is not intended to distinguish between prescription use or abuse of the drug. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 5
510(k) SUMMARY
-
Date: February 27, 2026
-
Submitter: HEALGEN SCIENTIFIC LLC
5213 Maple Street
Bellaire, TX 77401 -
Contact person: Jenny Xia
LSI International Inc.
504 East Diamond Ave., Suite H
Gaithersburg, MD 20877
Telephone: 301-525-6856
Email: jxia@lsi-consulting.org -
Device Names: Healgen® Accurate Oral Fluid Drug Test
Classification: Class II
| Product Code | CFR # | Panel |
|---|---|---|
| DKZ | 21CFR 862.3100, Amphetamine Test System | Toxicology |
| DIO | 21 CFR 862.3250, Cocaine and metabolites Test System | Toxicology |
| LDJ | 21 CFR, 862.3870 Cannabinoids Test System | Toxicology |
| DJC | 21 CFR 862.3610, Methamphetamine Test System | Toxicology |
| LCM | Unclassified, Enzyme immunoassay Phencyclidine | Toxicology |
| DJG | 21 CFR 862.3650, Opiate Test System | Toxicology |
| DJR | 21 CFR, 862.3610 Methadone Test System | Toxicology |
-
Predicate Devices:
Premier Biotech OralTox® Oral Fluid Drug Test, K181305 -
Intended Use
The Healgen® Accurate Oral Fluid Drug Test is a competitive binding lateral flow immunochromatographic assay for the qualitative and simultaneous detection of Amphetamine, Cocaine, Marijuana, Methadone, Methamphetamine, Opiates, Oxycodone, and Phencyclidine in human oral fluid at the cutoff concentrations listed below and their metabolites:
| Test | Calibrator | Cutoff (ng/mL) |
|---|---|---|
| Amphetamine (AMP) | d-Amphetamine | 50 |
| Cocaine (COC) | Benzoylecgonine | 20 |
| Marijuana (THC) | Delta-9-Tetrahydrocannabinol | 40 |
| Methadone (MTD) | Methadone | 30 |
| Methamphetamine (MET) | d-Methamphetamine | 50 |
| Opiates (OPI) | Morphine | 40 |
| Oxycodone (OXY) | Oxycodone | 20 |
Page 6
| Phencyclidine (PCP) | Phencyclidine | 10 |
The test provides only preliminary test results. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Liquid Chromatography/Mass Spectrometry/ Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. It is not intended to distinguish between prescription use or abuse of the drug. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.
- Device Description
The Healgen® Accurate Oral Fluid Drug Test is a lateral flow chromatographic immunoassay for the qualitative detection of Amphetamine, Cocaine, marijuana, Methadone, Methamphetamine, Opiates, Oxycodone and Phencyclidine in oral fluids. The tests are the first step in a two-step process. The second step is to send the sample for laboratory testing if preliminary positive results are obtained.
- Substantial Equivalence Information
A summary comparison of features of the Healgen® Accurate Oral Fluid Drug Test and the predicate devices are provided in following tables.
Table 1: Features Comparison of Healgen® Accurate Oral Fluid Drug Test and the Predicate Device
| Item | Device | Predicate – K181305 |
|---|---|---|
| Indication(s) for Use | For the qualitative determination of drug analytes in human oral fluid. | Same |
| Calibrators | d-AmphetamineBenzoylecgonineDelta-9-TetrahydrocannabinolMethadoned-MethamphetamineMorphineOxycodonePhencyclidine | Same |
| Methodology | Competitive binding, lateral flow immunochromatographic assays | Same |
| Type of Test | Qualitative | Same |
| Specimen Type | Human Oral fluid | Same |
| Cut-Off Values | AMP 50 ng/mLCOC 20 ng/mLTHC 40 ng/mLMET 50 ng/mLOPI 40 ng/mLPCP 10 ng/mLOXY 20 ng/mLMTD 30 ng/mL | Same |
| Intended Use | For prescription use | Same |
Page 7
Differences
| Device Format | Rectangular prism shaped cup | Half cylinder shaped cup |
- Test Principle
The Healgen® oral fluid drug test device is an immunoassay based on the principle of competitive binding. Drugs that may be present in the oral fluid specimen compete against their respective drug conjugates for binding sites on their specific antibody.
During testing, a portion of the oral fluid specimen migrates upward by capillary action. A drug, if present in the oral fluid specimen below its cut-off concentration, will not saturate the binding sites of its specific antibody. The antibody will then react with the drug-protein conjugate and a visible colored line will show up in the test line region of the specific drug strip. The presence of drug above the cut-off concentration in the oral fluid specimen will saturate all the binding sites of the antibody. Therefore, the colored line will not form in the test line region.
A drug-positive oral fluid specimen will not generate a colored line in the specific test line region of the strip because of drug competition, while a drug-negative oral fluid specimen will generate a line in the test line region because of the absence of drug competition.
To serve as a procedural control, a colored line will always appear at the control line region, indicating that proper volume of specimen has been added and membrane wicking has occurred.
- Performance Characteristics
1. Analytical Performance
a. Precision-Reproducibility-Cut-Off
Precision-Reproducibility-Cut-Off studies were carried out for samples with concentrations of -100% cut off, -75% cut off, -50% cut off, -25% cut off, cut off, +25% cut off, +50% cut off, +75% cut off and +100% cut off. These samples were prepared by spiking target drug in negative oral fluid samples. Each drug concentration was confirmed by LC/MS/MS. All sample aliquots were blindly labeled by the person who prepared the samples and didn't take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days per device lot in a randomized order. The following are summaries.
AMP
| Results | -100% cut off | -75% cut off | -50% cut off | -25% cutoff | cut off | +25% cut off | +50% cut off | +75% cut off | +100% cut off |
|---|---|---|---|---|---|---|---|---|---|
| Device Lot | 0 | 12.2 | 26.0 | 37.1 | 51.0 | 64.6 | 77.1 | 87.8 | 101.2 |
| LC-MS (ng/mL) | |||||||||
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 27-/23+ | 48+/2- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 26-/24+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 49-/1+ | 25-/25+ | 48+/2- | 50+/0- | 50+/0- | 50+/0- |
COC
| Results | -100% cut off | -75% cut off | -50% cut off | -25% cutoff | cut off | +25% cut off | +50% cut off | +75% cut off | +100% cut off |
|---|---|---|---|---|---|---|---|---|---|
| Device Lot | 0 | 5.3 | 10.2 | 16.2 | 21.3 | 26.8 | 31.9 | 35.2 | 40.7 |
| LC-MS (ng/mL) | |||||||||
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 49-/1+ | 26-/24+ | 49+/1- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 49-/1+ | 26-/24+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 27-/23+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
MET
| Results | -100% cut off | -75% cut off | -50% cut off | -25% cutoff | cut off | +25% cut off | +50% cut off | +75% cut off | +100% cut off |
Page 8
| LC-MS (ng/mL) | 0 | 12.2 | 25.3 | 36.7 | 50.2 | 62.3 | 78.0 | 87.0 | 99.1 |
|---|---|---|---|---|---|---|---|---|---|
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 29-/21+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 49-/1+ | 28-/22+ | 49+/1- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 28-/22+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
MTD
| Results | -100% cut off | -75% cut off | -50% cut off | -25% cutoff | cut off | +25% cut off | +50% cut off | +75% cut off | +100% cut off |
|---|---|---|---|---|---|---|---|---|---|
| Device Lot | 0 | 7.8 | 15.8 | 23.4 | 31.6 | 38.2 | 44.5 | 52.9 | 59.7 |
| LC-MS (ng/mL) | |||||||||
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 48-/2+ | 27-/23+ | 49+/1- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 47-/3+ | 26-/24+ | 49+/1- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 48-/2+ | 27-/23+ | 48+/2- | 50+/0- | 50+/0- | 50+/0- |
OPI
| Results | -100% cut off | -75% cut off | -50% cut off | -25% cutoff | cut off | +25% cut off | +50% cut off | +75% cut off | +100% cut off |
|---|---|---|---|---|---|---|---|---|---|
| Device Lot | 0 | 10.4 | 20.7 | 30.4 | 40.1 | 51.4 | 61.2 | 70.8 | 81.7 |
| LC-MS (ng/mL) | |||||||||
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 49-/1+ | 26-/24+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 49-/1+ | 26-/24+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 26-/24+ | 49+/1- | 50+/0- | 50+/0- | 50+/0- |
OXY
| Results | -100% cut off | -75% cut off | -50% cut off | -25% cutoff | cut off | +25% cut off | +50% cut off | +75% cut off | +100% cut off |
|---|---|---|---|---|---|---|---|---|---|
| Device Lot | 0 | 5.4 | 10.6 | 15.7 | 20.4 | 25.2 | 32.8 | 37.8 | 43.6 |
| LC-MS (ng/mL) | |||||||||
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 26-/24+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 49-/1+ | 26-/24+ | 49+/1- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 25-/25+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
PCP
| Results | -100% cut off | -75% cut off | -50% cut off | -25% cutoff | cut off | +25% cut off | +50% cut off | +75% cut off | +100% cut off |
|---|---|---|---|---|---|---|---|---|---|
| Device Lot | 0 | 2.7 | 5.3 | 8.0 | 10.7 | 13.2 | 16.0 | 18.8 | 21.5 |
| LC-MS (ng/mL) | |||||||||
| Lot 1 | 50-/0+ | 50-/0+ | 50-/0+ | 49-/1+ | 27-/23+ | 48+/2- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 27-/23+ | 50+/0- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3 | 50-/0+ | 50-/0+ | 50-/0+ | 50-/0+ | 25-/25+ | 49+/1- | 50+/0- | 50+/0- | 50+/0- |
THC
| Result | -100% cut off | -75% cut off | -50% cut off | -25% cutoff | cut off | +25% cut off | +50% cut off | +75% cut off | +100% cut off |
|---|---|---|---|---|---|---|---|---|---|
| Device Lot | 0 | 10.3 | 20.7 | 30.8 | 41.3 | 51.1 | 60.7 | 72.8 | 80.2 |
| LC-MS (ng/mL) | |||||||||
| Lot 1: | 50-/0+ | 50-/0+ | 50-/0+ | 47-/3+ | 26-/24+ | 49+/1- | 50+/0- | 50+/0- | 50+/0- |
| Lot 2: | 50-/0+ | 50-/0+ | 50-/0+ | 48-/2+ | 28-/22+ | 48+/2- | 50+/0- | 50+/0- | 50+/0- |
| Lot 3: | 50-/0+ | 50-/0+ | 50-/0+ | 49-/1+ | 26-/24+ | 47+/3- | 50+/0- | 50+/0- | 50+/0- |
The following cut-off values are verified.
| Calibrator | Cutoff (ng/mL) |
|---|---|
| d-Amphetamine | 50 |
| Benzoylecgonine | 20 |
| Delta-9-Tetrahydrocannabinol | 40 |
| Methadone | 30 |
| d-Methamphetamine | 50 |
| Morphine | 40 |
| Oxycodone | 20 |
| Phencyclidine | 10 |
Page 9
b. Linearity
Not applicable.
c. Stability
The devices are stable at 2-30 °C for 24 months based on the real time stability study at 2°C and 30°C.
d. Interference
Potential interfering substances were added to drug-free oral fluid and target drugs oral fluid with concentrations at 50% below and 50% above Cut-Off levels. These oral fluid samples were tested using three batches of the Healgen device. Compounds that showed no interference for drugs at a concentration of 100µg/mL are summarized in the following tables.
| Acetaminophen | Deoxycorticosterone | Naproxen |
|---|---|---|
| Acetylcodeine | Dextromethorphan | Nicotinamide |
| Allobarbital | Digoxin | Nicotine |
| Alprazolam | Diltiazem HCl | Noscapine |
| Amobarbital | Diphenhydramine HCl | Omeprazole |
| Apomorphine | DL-Propranolol | Papaverine |
| Atenolol | Estradiol | Pentazocine |
| Atropine | Estrone | Phenytoin |
| Baclofen | Fluconazole | Pioglitazone HCl |
| Benzocaine | Furosemide | Prednisolone |
| Butabarbital | Hexobarbital | Prednisone |
| Caffeine | Hydrochlorothiazide | Procainamide HCl |
| Carbamazepine | Ibuprofen | Promethazine |
| Chlordiazepoxide | Imipramine | Quinine HCl |
| Chlorpromazine | Lamotrigine | R,R(-)-Pseudoephedrine |
| Cimetidine | Levetiracetam | Salicylic Acid |
| Citalopram HBr | Lidocaine | Sertraline HCL |
| Clobazam | Lormetazepam | Simvastin |
| Clomipramine | L-Thyroxine | Theophylline |
| Clonazepam | Metformin HCl | Thiamine |
| Clonidine | Methylphenidate HCl | Topiramate |
| Clopidogrel bisulfate | Metoprolol | Valproic Acid |
| Cortisol | Metronidazole | Verapamil |
| Cotinine | Montelukast sodium salt | Zonisamide |
| d,l-Salbutamol | Naltrexone HCl (except for OXY) | |
| Dihydrocodeine (except for OPI & OXY) | Doxylamine | Ecgonine methylester (except for COC) |
| Phentermine (except for AMP) | Naloxone HCl (except for OXY) |
Food items such as methanol cough drops, cough syrup, cola, mouthwash, coffee, tea, milk, sugar, chewing gum, alcohol, baking soda, salt, cranberry juice, orange juice, food coloring (red, blue, green), toothpaste, tomatoes and MSG were added in either drug-free oral fluid or oral fluid containing the target drugs with concentrations of 50% below and 50% above cutoff levels to a concentration of 5%. None of the substances showed interference.
Page 10
Hemoglobin showed no interference at 10mg/dL.
e. Specificity
To test specificity, drug metabolites and other components that are likely to interfere in oral fluid samples were tested using three batches of the Healgen device. The following are summaries.
| Drug | Concentration (ng/mL) | % Cross-Reactivity |
|---|---|---|
| AMPHETAMINE (AMP) | ||
| D-Amphetamine | 50 | 100% |
| l-Amphetamine | 2,500 | 2% |
| d/l-Amphetamine | 100 | 50% |
| Methylenedioxyamphetamine (MDA) | 62.5 | 80% |
| 3-Hydroxy Tyramine | 2,500 | 2% |
| d,l-Phenylpropanolamine | 10,000 | 0.5% |
| Phenethylamine | 1,000 | 5% |
| Phentermine | 12,500 | 0.4% |
| d-Methamphetamine | 50,000 | 0.1% |
| l-Methamphetamine | 50,000 | 0.1% |
| Hydroxyamphetamine | 12,500 | 0.4% |
| Dimethylamylamine (DMAA) | 50,000 | 0.1% |
| Methoxyamphetamine | 200 | 25% |
| Benzodioxolylbutanamine(BDB) | 10,000 | 0.5% |
| d,l-p-Chloramphetamine | 300 | 16.7% |
| Methylenedioxyethylamphetamine | >100,000 | <0.05% |
| Methylenedioxymethamphetamine | >100,000 | <0.05% |
| Methylbenzodioxolylbutanamine | >100,000 | <0.05% |
| para-Methoxymethamphetamine | >100,000 | <0.05% |
| Phendimetrazine | >100,000 | <0.05% |
| Phenmetrazine | >100,000 | <0.05% |
| Ephedrine (d-,or l-,or d-l form) | >100,000 | <0.05% |
| d-Pseudoephedrine | >100,000 | <0.05% |
| Isoxsuprine | >100,000 | <0.05% |
| l-Pseudoephedrine | >100,000 | <0.05% |
| Fenfluramine | >100,000 | <0.05% |
| Mephentermine | >100,000 | <0.05% |
| COCAINE (COC) | ||
| Benzoylecgonine | 20 | 100% |
| Cocaine | 20 | 100% |
| Cocaethylene | 25 | 80% |
| Ecgonine | 15,00 | 1.3% |
| Ecgonine methylester | 12,500 | 0.16% |
| Norcocaine | 500 | 4% |
| Procaine | >100,000 | <0.02% |
Page 11
| Drug | Concentration (ng/mL) | % Cross-Reactivity |
|---|---|---|
| MARIJUANA (THC) | ||
| Δ9-Tetrahydrocannabinol | 40 | 100% |
| 11-nor-Δ9-THC-9 COOH | 4 | 1000% |
| Δ8-Tetrahydrocannabinol | 80 | 50% |
| 11-hydroxy-Δ9-THC | 45 | 89% |
| Cannabinol | 200 | 20% |
| Cannabidiol (CBD) | 2,200 | 1.8% |
| 11-Nor-Δ9-THC-carboxy-glucuronide | 60 | 66.7% |
| (+)-11-nor-9-carboxy-Δ9-THC | 50 | 80% |
| 11-nor-Δ8-THC-9-COOH | 20 | 200% |
| 8-beta-11-dihydroxy-Δ9-THC | 200 | 20% |
| 8-beta-hydroxy-Δ9-THC | 200 | 20% |
| Exo-THC | 75 | 53.3% |
| l-11-Nor-Δ9-THC-9-Carboxylic Acyl-Glucuronide | 15 | 266.7% |
| Δ8-THC Carboxylic Acid | 20 | 200% |
| Δ9-THC Carboxylic Acid | 4 | 1000% |
| (-)-Δ9-THC-D3 | 3,000 | 1.3% |
| 11-Nor-9betahydroxyhexahydrocannabinol (11-Nor-9beta-HHC) solution | 125 | 32% |
| METHADONE (MTD) | ||
| Methadone | 30 | 100% |
| Phencyclidine | 5,000 | 0.6% |
| LAAM | 10,000 | 0.3% |
| Alpha-Methadol | 150 | 20% |
| Doxylamine | >100,000 | <0.03% |
| 2-Ethylidene-1,5-dimethyl-3,3-diphenyl(EDDP) pyrrolidine(EDDP) | >100,000 | <0.03% |
| 2-Ethyl-5-methyl-3,3-diphenyl pyrrolidine(EMDP) | >100,000 | <0.03% |
| METHAMPHETAMINE (MET) | ||
| d-Methamphetamine | 50 | 100% |
| l-Methamphetamine | 1,250 | 4% |
| Methoxymethamphetamine | 25 | 200% |
| Ephedrine | 800 | 6.25% |
| Phenylephrine | 4,000 | 1.25% |
| Procaine | 2,000 | 2.5% |
| Methylephedrine | 5,000 | 1% |
| Methylenedioxyethylamphetamine | 250 | 20% |
| 3,4-methylenedioxymethamphetamine(MDMA) | 50 | 100% |
Page 12
| Drug | Concentration (ng/mL) | % Cross-Reactivity |
|---|---|---|
| d-Amphetamine | 25,000 | 0.2% |
| 3,4-methylenedioxyamphetamine | 25,000 | 0.2% |
| (±)-Amphetamine | 10,000 | 0.5% |
| l-Amphetamine | >100,000 | <0.05% |
| OPIATES (OPI) | ||
| Morphine | 40 | 100% |
| Acetylmorphine | 50 | 80% |
| Codeine | 10 | 400% |
| Ethylmorphine | 24 | 166.7% |
| Heroin(diacetylmorphine) | 50 | 80% |
| Hydromorphone | 100 | 40% |
| Thebaine | 1,500 | 2.67% |
| Norcodeine | 1,500 | 2.67% |
| Oxycodone | 25,000 | 0.16% |
| Oxymorphone | 25,000 | 0.16% |
| Nalorphine | 10,000 | 0.4% |
| Hydrocodone | 100 | 40% |
| 6-monoacetylmorphine(6-AM) | 25 | 160% |
| Morphine 3- β-d-glucuronide | 50 | 80% |
| Bilirubin | 3,500 | 1.14% |
| Dihydrocodeine | 50 | 80% |
| Normorphine | 12,500 | 0.32% |
| Morphine-6-β-d-glucuronide | 100 | 40% |
| OXYCODONE (OXY) | ||
| Oxycodone | 20 | 100% |
| Hydrocodone | 1,562.5 | 1.28% |
| Hydromorphone | 750 | 2.67% |
| Naloxone HCl | 5,000 | 0.4% |
| Oxymorphone | 48.8 | 40.98% |
| Dihydrocodeine | 3,125 | 0.64% |
| Naltrexone HCl | 2,000 | 1% |
| Heroin(diacetymorphine) | 12,500 | 0.16% |
| Buprenorphine | >100,000 | <0.02% |
| Codeine | >100,000 | <0.02% |
| Morphine | >100,000 | <0.02% |
| Morphine 3- β-d-glucuronide | >100,000 | <0.02% |
| Ethylmorphine | >100,000 | <0.02% |
| 6-monoacetylmorphine(6-AM) | >100,000 | <0.02% |
| PHENCYCLIDINE (PCP) | ||
| Phencyclidine | 10 | 100% |
| Cyclizine | 5,000 | 0.2% |
Page 13
| Drug | Concentration (ng/mL) | % Cross-Reactivity |
|---|---|---|
| Venlafaxine | 50,000 | 0.02% |
| Tenocyclidine (TCP) | 2,000 | 0.5% |
| l-(l-phenylcyclohexyl) morpholine | 15 | 66.7% |
| 4-hydroxyphencylidine | 10 | 100% |
| Hydrocodone | >100,000 | <0.01% |
| Hydromorphone | >100,000 | <0.01% |
| Nalorphine | >100,000 | <0.01% |
| EDDP | >100,000 | <0.01% |
| Ketamine | >100,000 | <0.01% |
| Prazepam | >100,000 | <0.01% |
| Amitriptyline | >100,000 | <0.01% |
| (+) Brompheniramine | >100,000 | <0.01% |
| (+) Chlorphenamine | >100,000 | <0.01% |
| Desmethylvenlafaxine | >100,000 | <0.01% |
| Chlorpromazine | >100,000 | <0.01% |
| Clomipramine | >100,000 | <0.01% |
| Cyclobenzaprine | >100,000 | <0.01% |
| Dextromethorphan | >100,000 | <0.01% |
| Doxepin | >100,000 | <0.01% |
| Doxylamine | >100,000 | <0.01% |
| Imipramine | >100,000 | <0.01% |
| Thioridazine | >100,000 | <0.01% |
| Dexbrompheniramine | >100,000 | <0.01% |
f. Effect of Oral fluid pH
To investigate the effect of oral fluid pH, oral fluid samples with pH 4 to 9 were spiked with target drugs at 50% below and 50% above Cut-Off levels. These samples were tested using three lots of the device. Results were all positive for samples at and above +50% Cut-Off and all negative for samples at and below -50% Cut-Off.
g. Drug Recovery Study
Negative oral fluid samples in glass bottles were spiked with target drugs to concentrations of -50% and +50% of the cutoffs. The samples were transferred to Healgen devices and store at room temperature, at 2-8℃, at -20oC and at 40oC. Over 90% recoveries were observed for the drugs in the Healgen devices. Oral fluid samples can be stored in the device at -20oC for at least 100 days. Oral fluid samples can be shipped overnight in the device for LC-MS confirmation.
Room Temperature (20 to 25℃) (3-day storage)
| -50% cutoff | AMP 25 | COC 10 | MET 25 | MTD 15 | OPI 20 | OXY 10 | PCP 5 | THC 20 |
|---|---|---|---|---|---|---|---|---|
| MAX | 107.05% | 98.13% | 109.77% | 102.80% | 105.55% | 109.55% | 97.51% | 106.12% |
| MIN | 93.32% | 91.16% | 93.58% | 98.67% | 91.19% | 97.06% | 91.54% | 93.81% |
| +50% cutoff | AMP 75 | COC 30 | MET 75 | MTD 45 | OPI 60 | OXY 30 | PCP 15 | THC 60 |
|---|---|---|---|---|---|---|---|---|
| MAX | 109.38% | 109.47% | 109.66% | 107.71% | 105.76% | 108.91% | 98.00% | 108.44% |
| MIN | 102.87% | 95.44% | 104.06% | 100.46% | 92.56% | 101.85% | 92.67% | 94.44% |
Page 14
2-8℃ (10-day storage)
| -50% cutoff | AMP 25 | COC 10 | MET 25 | MTD 15 | OPI 20 | OXY 10 | PCP 5 | THC 20 |
|---|---|---|---|---|---|---|---|---|
| MAX | 109.91% | 103.56% | 108.91% | 105.96% | 108.01% | 107.54% | 105.58% | 109.38% |
| MIN | 97.71% | 91.16% | 90.75% | 91.39% | 93.12% | 90.20% | 91.54% | 96.91% |
| +50% cutoff | AMP 75 | COC 30 | MET 75 | MTD 45 | OPI 60 | OXY 30 | PCP 15 | THC 60 |
|---|---|---|---|---|---|---|---|---|
| MAX | 109.79% | 109.03% | 109.81% | 105.18% | 106.81% | 109.18% | 106.45% | 108.72% |
| MIN | 91.17% | 91.15% | 97.92% | 92.66% | 90.50% | 93.19% | 92.00% | 93.25% |
-20℃ (100-day storage)
| -50% cutoff | AMP 25 | COC 10 | MET 25 | MTD 15 | OPI 20 | OXY 10 | PCP 5 | THC 20 |
|---|---|---|---|---|---|---|---|---|
| MAX | 109.27% | 98.81% | 108.91% | 100.00% | 108.28% | 98.04% | 105.58% | 106.12% |
| MIN | 98.31% | 90.26% | 91.13% | 92.72% | 96.18% | 91.18% | 91.54% | 100.00% |
| +50% cutoff | AMP 75 | COC 30 | MET 75 | MTD 45 | OPI 60 | OXY 30 | PCP 15 | THC 60 |
|---|---|---|---|---|---|---|---|---|
| MAX | 95.63% | 103.07% | 108.95% | 101.96% | 103.82% | 100.68% | 103.65% | 108.90% |
| MIN | 92.62% | 95.00% | 97.92% | 96.30% | 91.74% | 93.19% | 92.67% | 92.57% |
40℃ (1-day storage)
| -50% cutoff | AMP 25 | COC 10 | MET 25 | MTD 15 | OPI 20 | OXY 10 | PCP 5 | THC 20 |
|---|---|---|---|---|---|---|---|---|
| MAX | 108.30% | 105.83% | 109.58% | 105.12% | 105.01% | 104.90% | 106.67% | 109.37% |
| MIN | 95.44% | 94.29% | 92.83% | 98.05% | 93.61% | 96.45% | 98.46% | 96.29% |
| +50% cutoff | AMP 75 | COC 30 | MET 75 | MTD 45 | OPI 60 | OXY 30 | PCP 15 | THC 60 |
|---|---|---|---|---|---|---|---|---|
| MAX | 103.68% | 107.00% | 109.76% | 104.55% | 102.43% | 102.42% | 109.03% | 107.12% |
| MIN | 99.12% | 99.75% | 106.56% | 99.67% | 95.84% | 93.73% | 103.60% | 92.06% |
2. Comparison Studies
Method comparison studies for the Healgen Device were performed at six testing sites. Operators tested total 1430 samples for target drugs. The obtained test results are compared to LC/MS/MS results. The results are presented in the tables below
AMP
| Drug Concentration | Number of Negative | Number of Positive | Total | The percentage of correct result |
|---|---|---|---|---|
| Drug free | 833 | 0 | 833 | 100% |
| ˂ -50% Cut off | 261 | 0 | 261 | 100% |
| -50% Cut off ~ Cut off | 30 | 1 | 31 | 97% |
| Cut off ~ +50% Cut off | 2 | 12 | 14 | 86% |
| >+50% Cut off | 0 | 95 | 95 | 100% |
Discordant Results
| Sample # | LC/MS Concentration (ng/mL) | Healgen Test Results |
|---|---|---|
| HA02-043 | 41.377 | Pos |
| HA03-096 | 52.62 | Neg |
| HA01-188 | 56.82 | Neg |
Page 15
COC
| Drug Concentration | Number of Negative | Number of Positive | Total | The percentage of correct result |
|---|---|---|---|---|
| Drug free | 833 | 0 | 833 | 100% |
| ˂ -50% Cut off | 176 | 0 | 176 | 100% |
| -50% Cut off ~ Cut off | 17 | 4 | 21 | 81% |
| Cut off ~ +50% Cut off | 2 | 21 | 23 | 91% |
| >+50% Cut off | 0 | 181 | 181 | 100% |
Discordant Results
| Sample # | LC/MS Concentration (ng/mL) | Healgen Test Results |
|---|---|---|
| HA02B-020 | 16.55 | Pos |
| HA02-075 | 16.788 | Pos |
| HA03-026 | 17.037 | Pos |
| HA02-038 | 19.574 | Pos |
| HA01-121 | 22.596 | Neg |
| HA02B-104 | 22.60 | Neg |
MTD
| Drug Concentration | Number of Negative | Number of Positive | Total | The percentage of correct result |
|---|---|---|---|---|
| Drug free | 1040 | 0 | 1040 | 100% |
| ˂ -50% Cut off | 223 | 0 | 223 | 100% |
| -50% Cut off ~ Cut off | 18 | 1 | 19 | 95% |
| Cut off ~ +50% Cut off | 1 | 10 | 11 | 91% |
| >+50% Cut off | 0 | 32 | 32 | 100% |
Discordant Results
| Sample # | LC/MS Concentration (ng/mL) | Healgen Test Results |
|---|---|---|
| WA011 | 28.8 | Pos |
| HA01-297 | 33.04 | Neg |
MET
| Drug Concentration | Number of Negative | Number of Positive | Total | The percentage of correct result |
|---|---|---|---|---|
| Drug free | 838 | 0 | 838 | 100% |
| ˂ -50% Cut off | 296 | 0 | 296 | 100% |
| -50% Cut off ~ Cut off | 29 | 1 | 30 | 97% |
| Cut off ~ +50% Cut off | 0 | 17 | 17 | 100% |
| >+50% Cut off | 0 | 55 | 55 | 100% |
Discordant Results
| Sample # | LC/MS Concentration (ng/mL) | Healgen Test Results |
|---|---|---|
| HA01-149 | 41.36 | Pos |
OPI
| Drug Concentration | Number of Negative | Number of Positive | Total | The percentage of correct result |
|---|---|---|---|---|
| Drug free | 799 | 0 | 799 | 100% |
| ˂ -50% Cut off | 233 | 0 | 233 | 100% |
| -50% Cut off ~ Cut off | 27 | 1 | 28 | 96% |
| Cut off ~ +50% Cut off | 1 | 11 | 12 | 92% |
| >+50% Cut off | 0 | 162 | 162 | 100% |
Discordant Results
| Sample # | LC/MS Concentration (ng/mL) | Healgen Test Results |
Page 16
| HA02-119 | 33.62 | Pos |
| HA01-201 | 42.81 | Neg |
OXY
| Drug Concentration | Number of Negative | Number of Positive | Total | The percentage of correct result |
|---|---|---|---|---|
| Drug free | 854 | 0 | 854 | 100% |
| ˂ -50% Cut off | 269 | 0 | 269 | 100% |
| -50% Cut off ~ Cut off | 16 | 0 | 16 | 100% |
| Cut off ~ +50% Cut off | 1 | 38 | 39 | 97% |
| >+50% Cut off | 0 | 56 | 56 | 100% |
Discordant Results
| Sample # | LC/MS Concentration (ng/mL) | Healgen Test Results |
|---|---|---|
| HA03-344 | 22.53 | Neg |
PCP
| Drug Concentration | Number of Negative | Number of Positive | Total | The percentage of correct result |
|---|---|---|---|---|
| Drug free | 1023 | 0 | 1023 | 100% |
| ˂ -50% Cut off | 233 | 0 | 233 | 100% |
| -50% Cut off ~ Cut off | 23 | 0 | 23 | 100% |
| Cut off ~ +50% Cut off | 0 | 15 | 15 | 100% |
| >+50% Cut off | 0 | 47 | 47 | 100% |
THC
| Drug Concentration | Number of Negative | Number of Positive | Total | The percentage of correct result |
|---|---|---|---|---|
| Drug free | 725 | 0 | 725 | 100% |
| ˂ -50% Cut off | 252 | 0 | 252 | 100% |
| -50% Cut off ~ Cut off | 87 | 4 | 91 | 96% |
| Cut off ~ +50% Cut off | 11 | 14 | 25 | 56% |
| >+50% Cut off | 0 | 141 | 141 | 100% |
Discordant Results
| Sample # | LC/MS Concentration (ng/mL) | Healgen Test Results |
|---|---|---|
| HA01-288 | 35.42 | Pos |
| HA03-097 | 37.094 | Pos |
| HA03-103 | 38.145 | Pos |
| HA02-071 | 38.982 | Pos |
| HA01-141 | 40.67 | Neg |
| HA03-030 | 40.680 | Neg |
| HA02-100 | 40.95 | Neg |
| HA03-145 | 43.409 | Neg |
| HA02-166 | 43.91 | Neg |
| HA01-081 | 44.868 | Neg |
| HA01-035 | 45.231 | Neg |
| HA02B-244 | 45.27 | Neg |
| HA03-163 | 47.22 | Neg |
| HA01-221 | 49.26 | Neg |
| HA01-144 | 49.75 | Neg |
3. Clinical Studies
Not applicable.
Page 17
11. Conclusion
Based on the test principle and performance characteristics of the device, it's concluded that the Healgen® Accurate Oral Fluid Drug Test is substantially equivalent to the predicate.
§ 862.3610 Methamphetamine test system.
(a)
Identification. A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.(b)
Classification. Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).