K Number
K252101

Validate with FDA (Live)

Device Name
JWG-1
Date Cleared
2026-03-19

(259 days)

Product Code
Regulation Number
890.3860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device JWG-1 is a Powered Wheelchair Conversion Kit and suitable for a disabled person who can not walk or have restrictions on walking. It is intended for medical purposes to provide a means for a disabled person to take over the propulsion of the wheelchair and increase mobility and flexibility.

Device Description

The JWG-1 is an add-on drive for wheelchairs. The JWG-1 turns a manually propelled wheelchair into a powered wheelchair. It is intended for medical purposes to provide a means for a disabled person to take over the propulsion of the wheelchair and increase mobility and flexibility.

The device consists of the left/ right drive units, controller, battery and the battery charger. The drive units replace the original wheels of the manual wheelchair.

Yamaha Li-ion battery JWB3 can be used for the device. They are recharged with the charger JWC4. Battery can be charged in detached 'desktop' condition and on the wheelchair condition. JWB3 does not supply the driving current during the charging to prevent the wheelchair to move.

AI/ML Overview

N/A

FDA 510(k) Clearance Letter - Yamaha JWG-1 Powered Wheelchair

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.04

March 19, 2026

Yamaha Motor Co., Ltd
Yasutoshi Morikawa
General Manager
2500, Shingai
Iwata city, Shizuoka 4388501
Japan

Re: K252101
Trade/Device Name: JWG-1
Regulation Number: 21 CFR 890.3860
Regulation Name: Powered Wheelchair
Regulatory Class: Class II
Product Code: ITI
Dated: February 17, 2026
Received: February 17, 2026

Dear Yasutoshi Morikawa:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K252101 - Yasutoshi Morikawa
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K252101 - Yasutoshi Morikawa
Page 3

Sincerely,

Digitally signed by MARY S. KESZLER -S
Date: 2026.03.19 09:03:38 -04'00'

for Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and Physical Medicine Devices
OHT5: Office of Neurological and Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Page 11 of 66

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.

K252101

Please provide the device trade name(s).

JWG-1

Please provide your Indications for Use below.

The device JWG-1 is a Powered Wheelchair Conversion Kit and suitable for a disabled person who can not walk or have restrictions on walking. It is intended for medical purposes to provide a means for a disabled person to take over the propulsion of the wheelchair and increase mobility and flexibility.

Please select the types of uses (select one or both, as applicable).

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

JWG-1
Page 11 of 66

Page 5

510(k) Summary

510(k) #: K252101
Prepared on: 2026-03-18

Contact Details (21 CFR 807.92(a)(1))

FieldInformation
Applicant NameYamaha Motor Co., Ltd
Applicant Address2500, Shingai Iwata city Shizuoka 4388501 Japan
Applicant Contact Telephone+81-538-32-2106
Applicant ContactMr. Yasutoshi Morikawa
Applicant Contact Emailmorikawaya@yamaha-motor.co.jp

Device Name (21 CFR 807.92(a)(2))

FieldInformation
Device Trade NameJWG-1
Common NamePowered wheelchair
Classification NameWheelchair, Powered
Regulation Number890.3860
Product Code(s)ITI

Legally Marketed Predicate Devices (21 CFR 807.92(a)(3))

Predicate #Predicate Trade Name (Primary Predicate is listed first)Product Code
K203806JWX-1 PLUSITI
K161241e-fix E35/E36ITI

Device Description Summary (21 CFR 807.92(a)(4))

The JWG-1 is an add-on drive for wheelchairs. The JWG-1 turns a manually propelled wheelchair into a powered wheelchair. It is intended for medical purposes to provide a means for a disabled person to take over the propulsion of the wheelchair and increase mobility and flexibility.

The device consists of the left/ right drive units, controller, battery and the battery charger. The drive units replace the original wheels of the manual wheelchair.

Yamaha Li-ion battery JWB3 can be used for the device. They are recharged with the charger JWC4. Battery can be charged in detached 'desktop' condition and on the wheelchair condition. JWB3 does not supply the driving current during the charging to prevent the wheelchair to move.

Intended Use/Indications for Use (21 CFR 807.92(a)(5))

The device JWG-1 is a Powered Wheelchair Conversion Kit and suitable for a disabled person who can not walk or have restrictions on walking. It is intended for medical purposes to provide a means for a disabled person to take over the propulsion of the wheelchair and increase mobility and flexibility.

Indications for Use Comparison (21 CFR 807.92(a)(5))

The indications for use are the same.

Page 6

JWG-1
Page 2 of 2

Technological Comparison (21 CFR 807.92(a)(6))

The indications for use are the same.

Non-Clinical and/or Clinical Tests Summary & Conclusions (21 CFR 807.92(b))

The subject device was tested with the following standards:

  • ISO 7176-1:1999, Wheelchairs - Part 1: Determination of static stability
  • ISO 7176-2:2001, Wheelchairs - Part 2: Determination of dynamic stability of electrically powered wheelchairs
  • ISO 7176-3 Third edition 2012-12-15, Wheelchairs - Part 3: Determination of effectiveness of brakes
  • ISO 7176-4 Third edition 2008-10-01, Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range
  • ISO 7176-6 Second edition 2001-10-01, Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs
  • ISO 7176-8:1998, Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths
  • ISO 7176-9 Third edition 2009-11-15, Wheelchairs - Part 9: Climatic tests for electric wheelchairs
  • ISO 7176-11 Second edition 2012-12-01, Wheelchairs - Part 11: Test dummies
  • ISO 7176-14 Second edition 2008-02-15, Wheelchairs - Part 14: Power and control systems for electrically powered wheelchairs and scooters - Requirements and test methods
  • ISO 7176-15 First edition 1996-11-15, Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling
  • ISO 7176-21 Second edition 2009-04-01, Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers
  • ISO 10993-1 Fifth edition 2018-10, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
  • ISO 10993-5 Third edition 2009-06-01, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity
  • ISO 10993-10 Third edition 2010-08-01, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization

Clinical Test results of JWG-1 were not used to establish substantial equivalence.

The device JWG-1 has the same intended use and similar technological characteristics as the predicate device. The device does not raise any new questions of safety or effectiveness. The device is substantially equivalent to the predicate device.

JWG-1
Page 2 of 2

§ 890.3860 Powered wheelchair.

(a)
Identification. A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.(b)
Classification. Class II (performance standards).