K Number
K250384

Validate with FDA (Live)

Device Name
IPL400
Date Cleared
2025-05-23

(101 days)

Product Code
Regulation Number
878.4810
Age Range
2 - 12
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Ice Cooling IPL Hair Removal Device with sapphire treatment window is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

Device Description

Ice Cooling IPL Hair Removal Device is an over-the-counter, home-use and personal device for hair reduction by using Intense Pulsed Light (IPL). It is designed with a lamp that can emit single pulse, double pulse or triple pulse per shot. It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with nearly painless pain. The device is only powered by the external power adapter and its IPL emission activation is by finger switch. This product adopts sapphire treatment window that is suitable for multiple hair removal areas. It contains a skin sensor to detect appropriate skin contact, if the device is not in full contact with the skin, the device cannot emit the treatment light pulses. Besides, the device has the ice cooling function that will be activated throughout the whole hair removal process to provide users with a more comfortable experience.

AI/ML Overview

Based on the provided FDA 510(k) clearance letter for the Ice Cooling IPL Hair Removal Device, here's a breakdown of the acceptance criteria and the study (performance data) that proves the device meets them:

Important Note: The provided document is an FDA Clearance Letter and a 510(k) Summary. These documents primarily focus on demonstrating substantial equivalence to predicate devices and detailing the safety and performance testing that supports this claim. They do not contain detailed clinical study results such as those typically found in a clinical trial report (e.g., specific hair reduction percentages, statistical significance, or details on human reader studies for AI). The performance data listed focuses on engineering and safety standards, not clinical efficacy metrics with human subjects.

Therefore, for aspects related to clinical efficacy (like "how much human readers improve with AI vs without AI assistance" or detailed ground truth establishment for a clinical outcome), the information below is derived from what can be inferred or is explicitly stated regarding the device's technical specifications and safety profile, rather than a direct clinical performance study with defined acceptance criteria for efficacy. The "acceptance criteria" primarily relate to safety, electrical performance, and biocompatibility, as evidenced by compliance with international standards.


1. Table of Acceptance Criteria and Reported Device Performance

Given that this is an IPL hair removal device, the "acceptance criteria" in this context are primarily related to safety, electrical performance, and biocompatibility standards, rather than direct efficacy metrics like those for an AI diagnostic device. Efficacy is typically established by comparing the device's operational characteristics and intended use to legally marketed predicate devices that have prior clearance for similar indications, and by showing that it achieves the "permanent reduction in hair regrowth" as stated in the Indications for Use. The "reported device performance" is then the successful demonstration of compliance with these standards and comparable technological characteristics.

Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance
BiocompatibilitySkin Sensitization (ISO 10993-10:2021)Passed
Irritation (ISO 10993-23:2021)Passed
In Vitro Cytotoxicity (ISO 10993-5:2009)Passed
Electrical Safety & EMCGeneral Requirements for Basic Safety & Essential Performance (IEC 60601-1:2005+A1:2012+A2:2020)Passed
Electromagnetic Disturbances (IEC 60601-1-2:2014+A1:2020)Passed
Home Healthcare Environment Requirements (IEC 60601-1-11:2015+A1:2020)Passed
Non-Laser Source Equipment for Therapeutic/Cosmetic Use (IEC 60601-2-57:2011)Passed
Home Light Therapy Equipment (IEC 60601-2-83:2022)Passed
Light SafetyPhotobiological Safety of Lamps and Lamp Systems (IEC 62471:2006)Passed
Software V&VAll software requirement specifications met, all software hazards mitigated to acceptable risk levels (Consistent with Basic Documentation Level)Demonstrated
UsabilityGeneral Requirements for Basic Safety & Essential Performance - Collateral Standard: Usability (IEC 60601-1-6:2005+2012+2020)Evaluated and Validated
Compliance with FDA Human Factors and Usability Engineering Guidance (FEBRUARY 2016)Evaluated and Validated
Technological Equivalence to Predicate DevicesSimilar intended use, design, specifications, and performance (e.g., wavelength range, energy density within comparable ranges, pulse characteristics, safety features like skin sensor)Deemed "Substantially Equivalent" despite minor differences, with supporting engineering explanations (e.g., Notes 1-5 regarding dimensions, energy density, output energy, pulse duration, and output intensity levels). Key safety features like sapphire treatment window, skin sensor, and ice cooling function are highlighted.
Clinical Efficacy (stated aim)Permanent reduction in hair regrowth when measured at 6, 9 and 12 months after the completion of a treatment regime.The summary does not detail the clinical study that specifically proves this acceptance criterion with quantitative results. It relies on the substantial equivalence to predicate devices that have previously proven this. The Usability evaluation also verified that users can use the device "safely and effectively."

2. Sample Size Used for the Test Set and Data Provenance

The provided document does not specify a sample size for a clinical test set for hair reduction efficacy or any "test set" in the context of an AI/algorithm-driven diagnostic device. The performance data is primarily focused on engineering and safety testing conducted to established international standards.

  • Sample Size for Testing: Not applicable in the context of a clinical test set for efficacy, as no such study is detailed. For electrical and biocompatibility testing, "samples" would refer to a statistically appropriate number of units or materials for the specified tests, but these numbers are not disclosed.
  • Data Provenance: The manufacturer is Shenzhen Ulike Smart Electronics Co., Ltd., located in Shenzhen, China. The testing standards are international (IEC, ISO). The nature of the studies discussed (biocompatibility, electrical safety, light safety, software V&V, Usability) are laboratory-based and engineering assessments, not multi-center clinical trials. There is no indication of retrospective or prospective clinical data from human subjects mentioned.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not applicable or provided in the context of the 510(k) summary for an IPL hair removal device. "Ground truth" established by experts (e.g., radiologists) is typically relevant for diagnostic devices that analyze medical images or data. For this device, "ground truth" for safety specifications is established by compliance with published international standards (IEC, ISO), and human factors/usability experts would likely assess the usability aspects. No details are given about such experts or their roles in establishing a "ground truth" for clinical efficacy.


4. Adjudication Method for the Test Set

Not applicable/provided. Adjudication methods (e.g., 2+1, 3+1) are typically used in clinical studies for diagnostic devices, especially when establishing ground truth from multiple expert readers. This type of clinical study is not detailed in the provided 510(k) summary for this device.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done according to the provided 510(k) summary. MRMC studies are specific to evaluating the diagnostic performance of algorithms (often AI) in conjunction with human interpretation, and assessing the improvement in human reader performance with AI assistance. This device is an IPL hair removal device, not a diagnostic AI system.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

While the device does contain software ("Software Verification and Validation" is listed as a performance data category), this is embedded control software for the device's functionality (e.g., controlling pulse emission, energy levels, safety features like the skin sensor). There is no indication of a "standalone" algorithm performance study in the sense of an AI model independently performing a diagnostic task without human interaction. The software functions to ensure the device operates safely and effectively as designed, not to perform an independent diagnostic or analysis function typically evaluated in a standalone study.


7. The Type of Ground Truth Used

For the technical and safety performance, the "ground truth" is adherence to established international standards (IEC, ISO). For example:

  • Biocompatibility: In vitro and in vivo test results against material safety standards.
  • Electrical Safety & EMC: Measurements of electrical parameters and electromagnetic emissions against IEC limits.
  • Light Safety: Measurements of light output parameters against IEC limits for photobiological safety.
  • Software V&V: Verification that software functions as specified and validation that it meets user needs/hazard mitigation requirements.
  • Usability: Observational and empirical data from user testing against usability standards and FDA guidance.

For clinical efficacy (hair reduction), the ground truth would inherently be a measured reduction in hair regrowth over time, typically observed by clinicians or participants. However, the document does not describe the detailed clinical study methodology (e.g., type of outcome data, how it was collected, or if it involved pathology/histology to confirm hair follicle damage) that would have been used to establish this clinical "ground truth" for the device, beyond stating the "permanent reduction in hair regrowth" as an indication for use, implying it was demonstrated previously by predicate devices.


8. The Sample Size for the Training Set

Not applicable/provided. The concept of a "training set" is relevant for machine learning or AI models. This 510(k) summary describes a physical medical device (an IPL hair removal device), not an adaptive AI system requiring a training data set for its primary function. While the device contains software, it's for control and safety, not for learning or diagnostic predictions from historical data.


9. How the Ground Truth for the Training Set Was Established

Not applicable/provided. As there is no "training set" in the context of an AI/ML algorithm discussed, the establishment of ground truth for such a set is not relevant to this 510(k) submission.

FDA 510(k) Clearance Letter - Ice Cooling IPL Hair Removal Device

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue Doc ID# 04017.07.05
Silver Spring, MD 20993
www.fda.gov

May 23, 2025

Shenzhen Ulike Smart Electronics Co., Ltd.
Blue Yang
Registration Director
810,Bldg 1,Xunmei Science and Tech Plaza, No.8 Keyuan Rd.
Science Park Community, Yuehai Sub-District, Nanshan Dist
Shenzhen,
China

Re: K250938
Trade/Device Name: Ice Cooling IPL Hair Removal Device (MI01 LP, MI01 CB, MI01 GR, MI01 PP, MI01 RM, MI01 WG, MI01 WH, MI01 BK;UI04S PP, UI04S BU, UI04S WH, UI04S PN)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHT
Dated: March 28, 2025
Received: March 28, 2025

Dear Blue Yang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

Page 2

K250938 - Blue Yang Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).

Page 3

K250938 - Blue Yang Page 3

Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

TANISHA L. HITHE -S Digitally signed by TANISHA L. HITHE -S Date: 2025.05.23 11:26:03 -04'00'

Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.

K250938

Please provide the device trade name(s).

Ice Cooling IPL Hair Removal Device (MI01 LP, MI01 CB, MI01 GR, MI01 PP, MI01 RM, MI01 WG, MI01 WH, MI01 BK ; UI04S PP, UI04S BU, UI04S WH, UI04S PN)

Please provide your Indications for Use below.

Ice Cooling IPL Hair Removal Device with sapphire treatment window is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

Please select the types of uses (select one or both, as applicable).

  • ☐ Prescription Use (Part 21 CFR 801 Subpart D)
  • ☑ Over-The-Counter Use (21 CFR 801 Subpart C)

Ice Cooling IPL Hair Removal Device Page 9 of 38
K250938

Page 5

510(k) Summary of K250938

I. Submitter

Shenzhen Ulike Smart Electronics Co.,Ltd.
Address:810, Building 1, Xunmei Science and Technology Plaza, No. 8 Keyuan Road, Science Park Community, Yuehai Sub-District, Nanshan District, Shenzhen
518000, Guangdong, P.R. China
Contact person: Blue Yang
Email: blue@ulike.com
The date the summary was prepared:05/15/2024

II. Device

Name of Device: Ice Cooling IPL Hair Removal Device
Model(s): MI01 LP, MI01 CB, MI01 GR, MI01 PP, MI01 RM, MI01 WG, MI01 WH, MI01 BK;UI04S PP, UI04S BU, UI04S WH, UI04S PN
Common or Usual Name: Light Based Over-The-Counter Hair Removal
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: II
Product Code: OHT
Regulation Number: 21 CFR 878.4810

III. Predicate Devices

Predicate DevicePredicate Device 1Predicate Device 2Predicate Device 3
510(k) numberK242039K130315K240016

Page 6

Trade NameIce Cooling IPL Hair Removal DeviceiPulse Hair Removal SystemIPL Hair Removal Device
ManufacturerShenzhen Ulike Smart Electronics Co., Ltd.CyDen LtdShenzhen Jianrong Biomedical Electronics Co., Ltd
Regulation number21 CFR 878.481021 CFR 878.481021 CFR 878.4810
Product codeOHTOHTOHT
Device classificationClass IIClass IIClass II

IV. Device Description

Ice Cooling IPL Hair Removal Device is an over-the-counter, home-use and personal device for hair reduction by using Intense Pulsed Light (IPL). It is designed with a lamp that can emit single pulse, double pulse or triple pulse per shot. It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with nearly painless pain. The device is only powered by the external power adapter and its IPL emission activation is by finger switch. This product adopts sapphire treatment window that is suitable for multiple hair removal areas. It contains a skin sensor to detect appropriate skin contact, if the device is not in full contact with the skin, the device cannot emit the treatment light pulses. Besides, the device has the ice cooling function that will be activated throughout the whole hair removal process to provide users with a more comfortable experience.

V. Indications for Use

Ice Cooling IPL Hair Removal Device with sapphire treatment window is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.

Page 7

VI. Comparison of Technological Characteristics With the Predicate Device

The Ice Cooling IPL Hair Removal Devices have the same intended use and similar operational characteristics as the predicate devices. Any minor differences between the subject devices and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the devices are as safe and as effective as the predicate devices for its intended use. Therefore, the IPL Hair Removal Devices may be found substantially equivalent to its predicate devices. Ice Cooling IPL Hair Removal Devices are compared with the following legally marketed devices in terms of intended use, design, specifications and performance:

Comparison ItemsSubject DevicePredicate Device 1Predicate Device 2Predicate Device 3Remark
510(k) numberK250938K242039K130315K240016/
Trade NameIce Cooling IPL Hair Removal DeviceIce Cooling IPL Hair Removal DeviceiPulse Hair Removal SystemIPL Hair Removal Device/
ManufacturerShenzhen Ulike Smart Electronics Co., Ltd.Shenzhen Ulike Smart Electronics Co., Ltd.CyDen LtdShenzhen Jianrong Biomedical Electronics Co., Ltd/
Regulation number21 CFR 878.481021 CFR 878.481021 CFR 878.481021 CFR 878.4810Same
Product codeOHTOHTOHTOHTSame
Device classificationClass IIClass IIClass IIClass IISame
Indication for use/ Intended useIce Cooling IPL Hair Removal Device with sapphire treatment window is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hairIce Cooling IPL Hair Removal Device with sapphire treatment window is indicated for theThe iPulse Hair Removal System is an Over-the-counter device intended for the removal of unwanted hair.IPL Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanentSame

Page 8

regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Prescription or OTCOTCOTCOTC
Applicable skinFitzpatrick Skin Types I-VFitzpatrick Skin Types I-Vunknown
Treatment areaLarge areas (e.g. arms, legs, chest) and small areas (e.g. lip)Large areas (e.g. arms, legs, chest) and small areas (e.g. lip)leg, arm, bikini and underarm
Device design
Source energySupplied by external adapterSupplied by externalSupplied by external adapter

Page 9

Power supply100-240V~, 50/60Hz100-240V~, 50/60Hz100-240V~, 50/60Hz100-240V~, 50/60HzSame
DimensionMI01 series: 58.6(W)35(H)180.3(L)mm UI04S series: 179.458.736.4mm206.73mm68.68mm54.29mmunknownunknownSE Note 1
SterilizationNot requiredNot requiredNot requiredNot requiredSame
Output specification
Light sourceIntense Pulsed LightIntense Pulsed LightIntense Pulsed LightIntense Pulsed LightSame
Energy mediumXenon Arc FlashlampXenon Arc FlashlampXenon Arc FlashlampXenon Arc FlashlampSame
Wavelength range550-1200mm550-1200mm530-1100nm550-1200nmSame
Energy densityMI01 series: 1.8J/cm26.67J/cm2 Gentle mode:1.8J/cm² Face mode:2.8J/cm² Body mode:4.3J/cm² Bikini mode:6.67J/cm² UI04S series: 2.126.67J/cm² Soft Mode:2.12J/cm² Body Mode:3.52J/cm² Power Mode:4.55J/cm²2.79J/cm2~6.41J/cm27-10J/cm²1.2-4.3J/cm²SE Note 2

Page 10

USHR Mode:6.67J/cm²
Output energyMI01 series:5.4~20J Gentle mode:5.4J Face mode:8.4J Body mode:12.9J Bikini mode:20J UI04S series:7-22J Soft Mode:7J Body Mode:11.6J Power Mode:15J USHR Mode:22J10.9J~25J21~30J5.4~19.35J
Spot sizeMI01 series:3cm² UI04S series:3.3cm²3.9cm²3cm²4.5cm²
Pulse durationMI01 series: 1.2~7ms Single pulse Triple pulse UI04S series: 1.47-7.52ms Double pulse Triple pulse0.93ms~3.50ms Single pulse Double pulse Triple pulseVariable - Single pulse 25milliseconds. to Double Pulse 20ms on, 60 ms off.unknown
Pulsing controlFinger switchFinger switchFinger switchFinger switch
Delivery deviceDirect illumination to tissueDirect illumination to tissueDirect illumination to tissueDirect illumination to tissue

Page 11

Output intensity level4 Levels1-10 LevelsunknownunknownSE Note 5
Software/Firmware/Microprocess or Control?YesYesYesYesSame
Additional features
Electrical safetyIEC 60601-1 IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-57 IEC 60601-2-83IEC 60601-1 IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-57 IEC 60601-2-83unknownIEC 60601-2-83Same
Eye safetyIEC 62471IEC 62471unknownIEC 62471Same
BiocompatibilityISO 10993-5 ISO 10993-10 ISO 10993-23ISO 10993-5 ISO 10993-10 ISO 10993-23unknownunknownSame

Note 1:
The dimension belongs to basic characteristics. Although there is the dimension difference between subject devices and the predicate devices, but it will not affect the function and intended use of the device, and they all comply with IEC 60601-1 requirements, so the difference will not raise safety and effectiveness issue.

Note 2:
Although there is a minor difference of the energy density between the subject devices and predicate devices, but the energy density of subject devices is within the range of the value of the predicate devices(the Maximum energy density of the subject devices is higher than the predicate device 1 but lower than the predicate device 2, the Minimum energy density of the subject devices is higher than the predicate device 3), and they all comply with IEC 60601-2-57/IEC 60601-2-83 and IEC 62471 requirements, so such minor difference would not raise safety or effectiveness issue.

Page 12

Note 3:
Although there is a minor difference of the output energy between the subject devices and predicate devices, but the output energy of subject devices is within the range of the value of the predicate devices(the Maximum output energy of the subject devices is lower than the predicate device 1 but higher than the predicate device 3, the Minimum output energy of the subject devices is higher than the predicate device 3), and they all comply with IEC 60601-2-57/IEC 60601-2-83 and IEC 62471 requirements, so such minor difference would not raise safety or effectiveness issue.

Note 4:
Although there is a minor difference of the pulse width between the subject devices and predicate devices, it is within the pulse width value of the predicate devices(the Maximum pulse width of the subject devices is lower than the predicate device 2 but higher than the predicate device 1, the Minimum pulse width of the subject devices is higher than the predicate device 1). In addition, the subject devices comply with IEC 60601-2-83 and IEC 62471 requirement, so this difference will not raise any safety or effectiveness issue.

Regarding the double pulse or triple pulse, the subject devices operated on multipulse while the predicate device 2 also operated on multipulse, and Temperature Test Report and Usability evaluation has been conducted to verify the users can use the device with multiple pulses safely and effectively, so such difference would not raise safety or effectiveness issue.

Note 5:
Though the subject devices has 4 energy levels, which is different from the predicate device, this difference is insignificant and do not raise any safety or effectiveness problems.

VII. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

Summary of performance testing

The following performance data were provided in support of the substantial equivalence determination.

Page 13

1) Biocompatibility Testing

The biocompatibility evaluation for the body-contacting components of the IPL Hair Removal Device was conducted in accordance with the "Use of International Standard ISO 10993- 1, 'Biological Evaluation of Medical Devices–Part 1: Evaluation and Testing Within a Risk Management Process, Document", as recognized by FDA. The following testing was performed to, and passed, including:

  • ISO 10993-10:2021, Biological evaluation of medical devices Part 10: Tests for skin sensitization
  • ISO 10993-23:2021, Biological evaluation of medical devices Part 23: Tests for irritation
  • ISO 10993-5: 2009, Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

2) Electrical Safety and EMC

Electrical safety and EMC testing was performed to, and passed, as per the following standards:

  • IEC 60601- 1:2005+A1:2012+A2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
  • IEC 60601- 1-2:2014+A1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances - Requirements and tests
  • IEC 60601- 1- 11:2015+A1:2020 Medical Electrical Equipment– Part 1- 11: General Requirements for Basic Safety and Essential Performance –Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
  • IEC 60601-2-57:2011 Medical electrical equipment–Part 2-57: Particular requirements for the basic safety and essential performance of non-laser source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
  • IEC 60601-2-83:2022 Medical electrical equipment - Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment

3) Light Safety

  • IEC 62471:2006 Photobiological safety of lamps and lamp systems

4) Software Verification and Validation

Software documentation consistent with Basic Documentation Level was submitted in this 510(k). System testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.

5) Usability

Page 14

The product usability has been evaluated and validated according to the following standard and FDA guidance.

  • IEC 60601-1-6:2005+2012+2020 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
  • Applying Human Factors and Usability Engineering to Medical Devices, issued on FEBRUARY 2016

Conclusion: Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device IPL Hair Removal Device is as safe, as effective, and performs as well as the legally marketed predicate device.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.