Automatic Core Biopsy Instrument/ Semi-Automatic Core Biopsy Instrument/ Disposable Coaxial Biopsy Needle
K241793 · Zhejiang Curaway Medical Technology Co., Ltd. · KNW · Jan 24, 2025 · Gastroenterology, Urology
Device Facts
| Record ID | K241793 |
| Device Name | Automatic Core Biopsy Instrument/ Semi-Automatic Core Biopsy Instrument/ Disposable Coaxial Biopsy Needle |
| Applicant | Zhejiang Curaway Medical Technology Co., Ltd. |
| Product Code | KNW · Gastroenterology, Urology |
| Decision Date | Jan 24, 2025 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1075 |
| Device Class | Class 2 |
Intended Use
Automatic Core Biopsy Instrument is intended for use in obtaining biopsies from soft tissues such as breast, liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors. It is not intended for use in bone. Semi-Automatic Core Biopsy Instrument is intended for use in obtaining biopsies from soft tissues such as lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid, and various soft tissue tumors. It is not intended for use in bone. Disposable Coaxial Biopsy Needle is intended for use as a guiding needle in obtaining core biopsy samples from soft tissue such as lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid and various soft tissue tumors. It is designed for use with CuraWay Automatic Core Biopsy Instrument and CuraWay Semi-Automatic Core Biopsy Instrument. It is not intended for use in bone.
Device Story
Device suite includes Automatic Core Biopsy Instruments, Semi-Automatic Core Biopsy Instruments, and Disposable Coaxial Biopsy Needles. Instruments utilize spring-loaded mechanisms (automatic or semi-automatic) to rapidly advance an inner stylet and cutting cannula to capture soft tissue samples. Coaxial needles serve as guides for biopsy instruments. Used in clinical settings under imaging guidance (ultrasound, X-ray, CT; excluding MRI) by physicians. Features include centimeter depth markings, ultrasound-enhanced tips, and color-coded gauge indicators. Output is a core tissue sample for histological examination. Benefits include precise tissue acquisition for diagnostic evaluation of lesions.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including appearance, dimensions, sampling performance, connection firmness, stiffness, toughness, corrosion resistance, puncture force, and chemical properties. Biocompatibility testing performed per ISO 10993 series.
Technological Characteristics
Materials: Stainless steel (SUS304), TPE. Energy: Mechanical (spring-actuated). Form factor: Handheld, single-use, sterile (EO gas). Features: Centimeter depth markings, ultrasound-enhanced tips, color-coded gauge indicators. Connectivity: None. Software: None.
Indications for Use
Indicated for obtaining core biopsy samples from soft tissues (breast, liver, kidney, prostate, spleen, lymph nodes, lung, thyroid, soft tissue tumors) in patients requiring diagnostic tissue sampling. Contraindicated for use in bone.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Predicate Devices
- BARD® MAX-CORE® Disposable Core Biopsy Instrument (K133948)
- Achieve Programmable Automatic Biopsy System (K141552)
- Bard® Mission® Disposable Core Biopsy Instrument (K171953)
- Bard® TruGuide® Disposable Coaxial Biopsy Needle (K171953)
Related Devices
- K162588 — Möller Medical Biopsy Needles and Systems · Moller Medical GmbH · Jun 19, 2017
- K171953 — Bard Mission Disposable Core Biopsy Instrument · Bard Peripheral Vascular, Inc. · Sep 14, 2017
- K233031 — M·Biopsy /SureCore Automatic Disposable Biopsy Needle, M·Biopsy /SureCore Semi-automatic Disposable Biopsy Needle, M·Biopsy /SureAim Coaxial Biopsy Needle · Canyon Medical, Inc. · Jan 17, 2024
- K210946 — Promisemed Automatic Biopsy Needles, Promisemed Co-Axial Biopsy Devices, Promisemed Semi-automatic Biopsy Needles · Promisemed Hangzhou Meditech Co., Ltd. · Aug 12, 2021
- K222865 — M Biopsy /SureCore Automatic Disposable Biopsy Needle, M Biopsy /SureCore Semi-Automatic Disposable Biopsy Needle, M Biopsy /SureAim Coaxial Biopsy Needle · Canyon Medical, Inc. · Apr 3, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
January 24, 2025
Image /page/0/Picture/1 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Zhejiang CuraWay Medical Technology Co., Ltd. Li Yin Regulatory affairs specialist Room 106, Building 1, No. 600, 21st Avenue, Baiyang Sub-district, Qiantang New District Hangzhou, Zhejiang 310018 China
Re: K241793
Trade/Device Name: Automatic Core Biopsy Instrument/ Semi-Automatic Core Biopsy Instrument/ Disposable Coaxial Biopsy Needle Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: December 27, 2024 Received: December 27, 2024
Dear Li Yin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jessica Carr -S
Jessica Carr Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K241793
Device Name
Automatic Core Biopsy Instrument/ Semi-Automatic Core Biopsy Instrument/ Disposable Coaxial Biopsy Needle
Indications for Use (Describe)
Automatic Core Biopsy Instrument is intended for use in obtaining biopsies from soft tissues such as breast, liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors. It is not intended for use in bone.
Semi-Automatic Core Biopsy Instrument is intended for use in obtaining biopsies from soft tissues such as lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid, and various soft tissue tumors. It is not intended for use in bone.
Disposable Coaxial Biopsy Needle is intended for use as a guiding needle in obtaining core biopsy samples from soft tissue such as lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid and various soft tissue tumors. It is designed for use with CuraWay Automatic Core Biopsy Instrument and CuraWay Semi-Automatic Core Biopsy Instrument. It is not intended for use in bone.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(K) Summary K241793
| Company Name/Owner | Zhejiang CuraWay Medical Technology Co., Ltd. |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person/Author | Yin Li |
| Date prepared | January 24, 2025 |
| Contact details Address | Room 106, Building 1, No. 600, 21st Avenue, Baiyang<br>Sub-district, Qiantang New District, 310018 Hangzhou<br>City, Zhejiang Province, China |
| Contact phone number | 86-571-87016876 |
| Trade name | Automatic Core Biopsy Instrument<br>Semi-Automatic Core Biopsy Instrument<br>Disposable Coaxial Biopsy Needle |
| Common name | Core Biopsy Instrument, Biopsy Needle |
| Classification name | Instrument, biopsy |
| Review panel | Gastroenterology/Urology |
| Regulation number | 21 CFR 876.1075 |
| Product code | KNW |
| Predicate device | BARD® MAX-CORE®Disposable Core Biopsy<br>Instrument (K133948)<br>Achieve Programmable Automatic Biopsy System<br>(K141552)<br>Bard® Mission® Disposable Core Biopsy Instrument<br>(K171953)<br>Bard® TruGuide® Disposable Coaxial Biopsy Needle<br>(K171953) |
{5}------------------------------------------------
### 1. Device Description
#### ● Automatic Core Biopsy Instrument
Automatic Core Biopsy Instrument has seven models: BN-OCR-1, BN-OCR-2, BN-OCR-3, BN-OCR-6, BN-OCR-7, BN-OCR-8 and BN-OCR-9.
BN-OCR-1, BN-OCR-2 and BN-OCR-3 of Automatic Core Biopsy Instrument is automatic firing, and the structure is consisted of Inner Stylet, Cutting Cannula, Operating Handle, Actuation button, Depth stopper, Slide block and Protective sheath.
BN-OCR-6, BN-OCR-7, BN-OCR-8 and BN-OCR-9 of Automatic Core Biopsy Instrument has both automatic firing and delay firing options, the automatic firing mode releases the cannula and stylet in rapid sequence and captures a tissue sample with the push of a button. The delay firing method is used to verify the position of the sample notch in the target area prior to depressing the actuation button "A", which activates the cutting cannula, capturing the tissue sample. The structure is consisted of Inner Stylet, Cutting Cannula, Adjustable Knob (only for BN-OCR-7 and BN-OCR-9), Charging handle, Operating Handle, Security Lock (only for BN-OCR-8 and BN-OCR-9) and Protective Sheath.
Each model has various specifications depending on different sizes such as needle lengths, needle diameters, etc.
The cutting cannula has centimeter scales on it to provide reference for depth placement. In the distal area of the cutting cannula, an ultrasound enhancement is available to promote accurate placement under ultrasound or X-ray guidance, excluding MRI.
The Automatic Core Biopsy Instrument is single-use device, supplied in a sterile state sterilized by EO gas. A re-sterilization by users is not allowed. The shelf life is defined for 3 years, which is verified.
#### ● Semi-Automatic Core Biopsy Instrument
The Semi-Automatic Core Biopsy Instrument has two models (BN-OCR-4, BN-OCR-5) according to whether there is a Safety limit slot.
The Semi-Automatic Core Biopsy Instrument consists of Inner stylet, Cutting cannula, Fixed handle, Windows, Safety switch, Plunger and Protective sheath. The needle is protected in a protective sheath.
{6}------------------------------------------------
Each model has various specifications depending on different sizes such as needle lengths, needle diameters, etc. The plunger is color-coded, which indicates gauge size of the needle.
The Cutting cannula has centimeter scales on it to provide reference for depth placement. In the distal area of the Cutting cannula, an ultrasound enhancement is available to promote accurate placement under ultrasound or X-ray guidance, excluding MRI.
The Semi-Automatic Core Biopsy Instrument is single-use device, supplied in a sterile state sterilized by EO gas. A re-sterilization by users is not allowed. The shelf life is defined for 3 years, which is verified.
#### ● Disposable Coaxial Biopsy Needle
Disposable Coaxial Biopsy Needle has two models: BN-MAR-5, BN-MAR-6.
The Disposable Coaxial Biopsy Needle consists of Inner Stylet, Cutting cannula, Operating handle of Cutting cannula. Operating handle of Inner stylet, Depth stopper, Protective Sheath, Adaptor (optional) and Blunt needle (optional).
An adjustable depth stopper allows the user to restrict forward movement, localizing the needle tip to the biopsy site. The Depth Stopper is color-coded, which indicates gauge size of the needle. It is available in several needle gauge sizes and lengths.
Disposable Coaxial Biopsy Needle is intended for soft-tissue biopsy and designed as guiding needle for use with CuraWay disposable Biopsy Instruments/Needles such as CuraWay's Automatic Core Biopsy Instrument and CuraWay's Semi-Automatic Core Biopsy Instrument. The outer cannula is one gauge-size less than the compatible Biopsy Instrument/Needle, e.g., 19-gauge CuraWay Coaxial Biopsy Needle for a 20 gauge CuraWay Disposable Biopsy Needle or CuraWay Disposable biopsy instrument.
The Cutting cannula has numerical scales on it to provide reference for depth placement. In the distal area of the cutting cannula, an ultrasound enhancement is available to promote accurate placement under ultrasound or X-ray guidance excluding MRI.
{7}------------------------------------------------
The Disposable Coaxial Biopsy Needle is single-use device, supplied in a sterile state sterilized by EO gas. A re-sterilization by users is not allowed. The shelf life is defined for 3 years, which is verified.
### 2. Indications for use
Automatic Core Biopsy Instrument is intended for use in obtaining biopsies from soft tissues such as breast, liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors. It is not intended for use in bone.
Semi-Automatic Core Biopsy Instrument is intended for use in obtaining biopsies from soft tissues such as lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid, and various soft tissue tumors. It is not intended for use in bone.
Disposable Coaxial Biopsy Needle is intended for use as a guiding needle in obtaining core biopsy samples from soft tissue such as lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid and various soft tissue tumors. It is designed for use with CuraWay Automatic Core Biopsy Instrument and CuraWay Semi-Automatic Core Biopsy Instrument. It is not intended for use in bone.
# 3. Substantial equivalence comparison with predicate device
Detailed substantial comparison was made between subject device and predicate
| Subject device | Predicate device 1 | Predicate device 2 | Comparison | |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Automatic Core<br>Biopsy<br>Instrument | BARD®<br>MAX-CORE®Disposable<br>Core Biopsy Instrument<br>(K133948) | Achieve<br>Programmable<br>Automatic Biopsy<br>System (K141552) | Refer to Table<br>1 below | |
| Semi-Automatic<br>Core Biopsy<br>Instrument | Bard® Mission®<br>Disposable Core Biopsy<br>Instrument (K171953) | / | Refer to Table<br>2 below | |
| Disposable<br>Coaxial Biopsy<br>Needle | BARD® TRUGUIDE®<br>Disposable Coaxial<br>Biopsy Needle<br>(K171953) | / | Refer to Table<br>3 below | |
| | Subject Device | Predicate Device 1 | Predicate Device 2 | |
| Comparison<br>Elements | Automatic Core Biopsy<br>Instrument | BARD® MAX-CORE®<br>Disposable Core Biopsy<br>Instrument | Achieve Programmable<br>Automatic Biopsy System | Comment |
| Classification<br>Regulation | 876.1075 | 876.1075 | 876.1075 | Same |
| Product Code | KNW | KNW | KNW | Same |
| 510(k) Number | - | K133948 | K141552 | / |
| Indications for<br>use | The Automatic Core Biopsy<br>Instrument is intended for use<br>in obtaining biopsies from soft<br>tissues such as breast, liver,<br>kidney, prostate, spleen, lymph<br>nodes and various soft tissue<br>tumors. It is not intended for<br>use in bone. | The core needle biopsy device is<br>intended for use in obtaining<br>biopsies from soft tissues such as<br>liver, kidney, prostate, spleen,<br>lymph nodes and various soft<br>tissue tumors. It is not intended<br>for use in bone. | The Achieve Programmable<br>Automatic Biopsy System is<br>intended for use in obtaining<br>core biopsy samples from soft<br>tissue such as kidney, liver,<br>prostate, spleen, lymph nodes,<br>and various soft tissue masses.<br>Not intended for use in bone.<br><br>The Achieve Programmable<br>Automatic Biopsy System is also<br>indicated to provide breast tissue<br>samples for diagnostic sampling<br>of breast abnormalities. It is<br>designed to provide breast tissue<br>for histologic examination with | Same |
| | | | | partial or complete removal of<br>the imaged abnormality. |
| Operation<br>mechanics | Automatic activation and<br>Automatic mode and delay<br>mode | Automatic activation | Automatic mode and delay mode | Same |
| Material of<br>patient contact<br>component | Stainless steel | Stainless steel | Stainless steel | Same |
| Body contact<br>site | Breast, liver, kidney, prostate,<br>spleen, lymph nodes and<br>various soft tissue tumors | Liver, kidney, prostate, spleen,<br>lymph nodes and various soft<br>tissue tumors | Kidney, liver, prostate, spleen,<br>lymph nodes, various soft tissue<br>masses and breast | Same |
| Structure | The Automatic Core Biopsy<br>Instrument contains automatic<br>activation type (BN-OCR-1,<br>BN-OCR-2 and BN-OCR-3)<br>and automatic and delay type<br>(BN-OCR-6, BN-OCR-7,<br>BN-OCR-8 and BN-OCR-9).<br>Automatic activation type is<br>consisted of inner stylet, cutting<br>cannula, operating handle, slide<br>block, actuation button, depth<br>stopper and protective sheath.<br>Automatic and delay type is | The device is consisted of handle<br>and biopsy instrument. The<br>biopsy instrument is composed of<br>inner stylet and cutting cannula.<br>Each needle has numerically<br>ordered centimeter markings on<br>the outer cannula to provide<br>reference for depth placement.<br>Needles feature an adjustable<br>needle stop which allows the user<br>to restrict forward movement,<br>localizing the needle tip to the<br>biopsy site. | The device is consisted of<br>notched stylet, cutting cannula,<br>firing buttons and charging<br>handle. | Difference<br>1 |
| | consisted of inner stylet, cutting<br>cannula, adjustable knob,<br>charging handle, operating<br>handle, security lock and<br>protective sheath. | Color coded stylet hubs indicate<br>gauge size of needles available in<br>a variety of gauge sizes and<br>centimeter lengths. | | |
| | The cutting cannula has<br>numerically ordered centimeter<br>markings on it to provide<br>reference for depth placement. | | | |
| | An adjustable depth stopper<br>allows the user to restrict<br>forward movement, localizing<br>the needle tip to the biopsy site.<br>The operating handle is<br>color-coded, which indicates<br>gauge size of the needle. | | | |
| Visualization<br>technique | The introduction of the needle<br>into the body should be carried<br>out under imaging guidance<br>(ultrasound, X-Ray, CT, etc.) | The introduction of the needle<br>into the body should be carried<br>out under imaging control<br>(ultrasound, X-Ray, CT, etc.). | Echogenic markings on both<br>stylet tip and cannula help<br>confirm the placement of the<br>sample notch for precise<br>ultrasound positioning. | Same |
| Nominal length<br>of Cutting<br>Cannula (mm) | 80, 90, 100, 130, 150, 160,180,<br>200, 250 | 100, 160, 200, 250 | 60, 90, 110, 150, 200, 250 | Difference<br>2 |
device.
{8}------------------------------------------------
# Table 1: Substantial equivalence comparison of Automatic Core Biopsy Instrument
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
| Needle size | 14G, 16G, 18G, 20G | 14G, 16G, 18G, 20G | 14G, 16G, 18G, 20G | Same |
|-----------------------------------|----------------------------|----------------------------------------------|----------------------------------------------|-----------------|
| Length of<br>sample notch<br>(mm) | 18.5 mm | 18mm, 19mm | 20mm | Difference<br>3 |
| Biocompatibility | ISO 10993 series standards | ISO 10993-1, ISO 10993-5 and<br>ISO 10993-10 | ISO 10993-1, ISO 10993-5 and<br>ISO 10993-10 | Same |
| Sterilization<br>method | EO sterilization | EO sterilization | EO sterilization | Same |
| Single use | Yes | Yes | Yes | Same |
# ● Difference 1: Structure
The structure and working principle of subject devices are basically the same. The components of subject device and predicate devices both include inner stylet, cutting cannula, operating handle, actuation button, and depth stopper, but they are named differently. The function of components of subject devices are the same. The differences in design do not affect the safety and effectiveness of product. In sum, subject device and predicate devices have similar structure.
- Difference 2: Nominal length of Cutting Cannula
It can be seen from the table above that subject device provides more selections on the nominal length of Cutting Cannula, namely 80mm, 130mm and 180mm. Considering the different shape of patients and the distance between needle insertion point and lesion, 80mm. 130mm and 180mm are provided by subject device. Besides, 80mm and 180mm are within the length range (60mm to 250mm) of predicate devices. So this difference does not affect the safety and effectiveness of device.
- Difference 3: Length of sample notch
{12}------------------------------------------------
The length of subject device is 18.5 mm, and the length of sample notch of predicate devices are 18mm, 19mm and 20mm. Although the lengths are not the same, the length of within the length range of predicate devices. So this difference does not affect the safety and effectiveness of device.
### Conclusion
Comparison of the device in question with the predicate devices K141552 is performed, and comparison items include indications for use, operation mechanics, material of paient, component, body contact site, structure, visualization technique, nominal length of Cutting Cannula, needle size, length of sample notch, biocompatibility, sterilization method and single we. It can be concluded that subject device and predicate device perform the same clinical conditions. The differences between the subject device and predicate devices do not affect the safety and effectiveness. Therefore, they can be considered as equivalent devices.
{13}------------------------------------------------
| Comparison<br>Elements | Subject Device | Predicate Device | Comment |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Proprietary Name | Semi-Automatic Core Biopsy Instrument | Bard® Mission® Disposable Core Biopsy<br>Instrument | - |
| Classification<br>Regulation | 876.1075 | 876.1075 | Same |
| Product Code | KNW | KNW | Same |
| 510(k) Number | - | K171953 | - |
| Indications for use | Semi-Automatic Core Biopsy Instrument is<br>intended for use in obtaining biopsies from soft<br>tissues such as lung, liver, spleen, kidney, prostate,<br>lymph nodes, breast, thyroid, and various soft<br>tissue tumors. It is not intended for use in bone. | The BARD® MISSION® Disposable Core Biopsy<br>Instrument is intended for use in obtaining biopsy<br>samples from soft tissues such as from the lung,<br>liver, spleen, kidney, prostate, lymph nodes, breast,<br>thyroid, and various soft tissue tumors. It is not<br>intended for use in bone. | Same |
| Operation<br>mechanics | Semi-Automatic | Semi-Automatic | Same |
| Comparison<br>Elements | Subject Device | Predicate Device | Comment |
| Material of patient<br>content component | Stainless Steel | Stainless Steel | Same |
| Body contact site | Lung, liver, spleen, kidney, prostate, lymph nodes,<br>breast, thyroid, and various soft tissue tumors. | Lung, liver, spleen, kidney, prostate, lymph<br>nodes, breast, thyroid, and various soft tissue<br>tumors. | Same |
| Structure | Component/ Design features: Inner stylet, Cutting<br>cannula, Fixed handle, Windows, Safety switch,<br>Plunger and Protective sheath, Centimeter marking,<br>fixed handle, windows, ultrasound enhancement<br>(offers enhanced ultrasound visibility of the Inner<br>stylet and Cutting cannula). The plunger is<br>color-coded, which indicates gauge size of the<br>needle. | Component/Design features: Trocar stylet, Coaxial<br>cannula, Depth stop, centimeter marks, Ergonomic<br>grip design, Fire Ready Indicator, Penetration<br>depth indicator, Echogenic technology (offers<br>enhanced ultrasound visibility of the stylet needle<br>and cutting cannula needle), Protective needle<br>sheath. The plunger (is color coded according to<br>the various gauge sizes. | Difference 1 |
| Visualization<br>technique | The introduction of the needle into the body should<br>be carried out under imaging guidance (ultrasound, | The introduction of the needle into the body<br>should be carried out under imaging guidance | Same |
| Comparison<br>Elements | Subject Device | Predicate Device | Comment |
| | X-Ray, CT, etc.). excluding MRI. | (ultrasound, X-Ray, CT, etc.). excluding MRI. | |
| Nominal length of<br>Cutting Cannula | 8cm, 9cm, 10cm, 13cm, 15cm, 16cm, 18cm,<br>20cm, 25 cm | 10cm, 16cm, 20cm, 25cm | Difference 2 |
| Needle size | 14G, 16G, 18G, 20G | 14G, 16G, 18G, 20G | Same |
| Penetration depth | 10mm and 20mm | 10mm and 20mm | Same |
| Biocompatibility | ISO 10993 series standards | ISO 10993 series standards | Same |
| Sterilization method | EO sterilization | EO sterilization | Same |
| Single use | YES | YES | Same |
Table 2: Substantial equivalence comparison of Semi-Automatic Core Biopsy Instrument
{14}------------------------------------------------
{15}------------------------------------------------
#### Difference 1: Structure ●
The structure and working principle of Subject device are basically the same. The components are both including the Inner stylet, Cutting cannula, fixed handle, Plunger, and Protective sheath with identical appearances, but named differently. Both products are designed with two firing positions and support single-handed operation.
Major differences in structure Compared to Predicate device:
- Fixed Handle: Fixed handle of Subject device adopts a structure of a rectangular cover shell and a base shell. Two complete circular a. rings are designed on the base shell to facilitate effective gripping during operation. In contrast, Predicate device features a symmetrical cover and base shell design, forming open semi-circular rings on both sides when combined, allowing effective gripping.
{16}------------------------------------------------
- Safety Switch: Subject device includes a safety switch at the end of the fixed handle to prevent accidental firing during the puncture b. process, ensuring safe usage.
- Window Design: The window of Subject device is marked with "10" and "20" indicator lines (with numbers) on both sides. When the C. plunger is pulled, a indicator line appears to show the firing length. Predicate device displays numbers ("(0", "10") directly within the window without using indicator lines.
Above all, The subject device and predicate device have fundamental Scientific Technology, including Mechanism/Mode of Action, and Energy Used/Delivered. These differences changes of structure are ergonomically designed and do not affect the safety and effectiveness of the product.
- Difference 2: Nominal length of Cutting Cannula ●
Due to the different body shapes of patients and the different distances between different lesions and the needle entry point, clinical and market demand, subject product has increased 5 kinds of needle length of 8cm, 15cm and 18cm compared with the predicate product, this will not raise any safety issue.
### The above mentioned differences do not raise new questions of safety and effectiveness.
### Conclusion
Comparison of the device in question with the predicate devices K171953 is performed, and comparison for use, operation mechanics, material of patient component, body contact site, structure, visualization technique, nominal length of Cutting Cannula, needle size, penetration depths, biocompatibility, sterlization method and single that subject device and predicate device perform the same finical conditions. The differences between the subject device and predicate devices do not affect the safety and effectiveness. Therefore, subject device can be considered as equivalent devices.
{17}------------------------------------------------
| Comparison Elements | Subject Device | Predicate Device | Comment |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Proprietary Name | Disposable Coaxial Biopsy Needle | BARD® TRUGUIDE® Disposable<br>Coaxial Biopsy Needle | - |
| Classification Regulation | 876.1075 | 876.1075 | Same |
| Product Code | KNW | KNW | Same |
| 510(k) Number | - | K171953 | - |
| Indications for use | Intended for use as a guiding needle in<br>obtaining core biopsy samples from soft<br>tissue such as lung, liver, spleen, kidney,<br>prostate, lymph nodes, breast, thyroid and<br>various soft tissue tumors. It is designed for<br>use with CuraWay Automatic Core Biopsy<br>Instrument and CuraWay Semi-Automatic<br>Core Biopsy Instrument. It is not intended<br>for use in bone. | Intended for use in obtaining biopsy<br>samples from soft tissues such as from the<br>lung, liver, spleen, kidney, prostate, lymph<br>nodes, breast, thyroid, and various soft<br>tissue tumors. It is not intended for use in<br>bone. | Same |
| Material of patient content<br>component | Stainless Steel, TPE | Stainless Steel | Difference 1 |
| Body contact site | Soft tissue such as lung, liver, spleen,<br>kidney, prostate, lymph nodes, breast,<br>thyroid and various soft tissue tumors | lung, liver, spleen, kidney, prostate, lymph<br>nodes, breast, thyroid, and various soft<br>tissue tumors. | Same |
| Structure | ● The Disposable Coaxial Biopsy<br>Needle consists of Inner stylet, Cutting | ● The BARD® TRUGUIDE®<br>Disposable Coaxial Biopsy Needle is a | Difference 2 |
| Comparison Elements | Subject Device | Predicate Device…