LUXEN 5G

K241151 · Dentalmax Co., Ltd. · EIH · Sep 30, 2024 · Dental

Device Facts

Record IDK241151
Device NameLUXEN 5G
ApplicantDentalmax Co., Ltd.
Product CodeEIH · Dental
Decision DateSep 30, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6660
Device ClassClass 2
AttributesTherapeutic

Intended Use

LUXEN 5G is indicated for the production of all-ceramic inlays, multi-units bridges, onlays, and veneers without zirconium dioxide frameworks.

Device Story

LUXEN 5G is a yttria-stabilized zirconia ceramic block used for fabricating customized dental restorations (inlays, onlays, veneers, multi-unit bridges). The device is processed via CAD/CAM milling systems in dental clinics or laboratories. It serves as a restorative material for patients needing dental prosthetics. The healthcare provider uses the milled ceramic to restore tooth structure, potentially improving patient oral function and aesthetics. The device is non-sterile and intended for long-term contact with oral mucosa, enamel, and dentin.

Clinical Evidence

Bench testing only. Performance testing included visual inspection, dimensions, packaging, uniformity, freedom from extraneous materials, chemical solubility, flexural strength, and linear thermal expansion per ISO 6872. Biocompatibility testing conducted per ISO 10993-5, 10, 11, and 23. Pyrogen testing performed per USP <151>.

Technological Characteristics

Ceramic block composed of yttria-stabilized zirconia (3-4 mol% TZP). Conforms to ISO 6872:2015 Type 2, Class 5. Flexural strength >800 MPa. Thermal expansion coefficient 11.0 x 10^-6 K^-1. Non-sterile. CAD/CAM millable form factor (Block, Wieland, D95). Biocompatibility per ISO 10993-1.

Indications for Use

Indicated for the production of all-ceramic inlays, multi-units bridges, onlays, and veneers without zirconium dioxide frameworks in patients requiring dental restorations.

Regulatory Classification

Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text. September 30, 2024 Dentalmax Co., Ltd % Priscilla Chung Official Correspondent LK Consulting Group USA, Inc. Contact Address Re: K241151 Trade/Device Name: LUXEN 5G Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: INOTE: Use date of most recent supplementl Received: April 25, 2024 Dear Priscilla Chung: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, # Bobak Shirmohamm adi -S For Michael E. Adjodha, M. Ch.E., RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use Submission Number (if known) K241151 Device Name LUXEN 5G Indications for Use (Describe) LUXEN 5G is indicated for the production of all-ceramic inlays, multi-units bridges, onlays, and veneers without zirconium dioxide frameworks. Type of Use (Select one or both, as applicable) > | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # K241151 - 510(k) Summary This summary of 510(K) information is being submitted in accordance with requirements of 21 CFR Part 807.92. Date: 2/1/2024 # 1. Submitter DENTALMAX Co., Ltd. 50-7, Pungsesandan 2-Ro, Pungse-Myeon, Dongnam-Gu Cheonan-Si, Chungnam, South Korea, 31217 # 2. U.S Agent/Contact Person Priscilla Chung LK Consulting Group USA, Inc. 18881 Von Karman Ave STE 160, Irvine CA 92612 Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c(@lkconsultinggroup.com ### 3. Device - Trade Name: LUXEN 5G - Common Name: Dental Frame Material for Dental Prosthesis - Classification Name: Porcelain Powder for Clinical Use ● - Product Code: EIH - . Classification regulation: 21 CFR 872.6660 ### 4. Predicate Device: LUXEN ZR by DENTALMAX Co., Ltd. (K171585) ### 5. Description: LUXEN 5G is ceramic block used to create customized dental restorations using CAD/CAM system. LUXEN 5G block are available in three shapes (Block, Wieland, D95) and various thickness sizes. LUXEN 5G is composed of yttria-stabilized zirconia and 3~4 mol % TZP. The weight of yttria is classified in accordance with ISO 6872:2015 Type2, class 5 for 3Y-TZP Zirconia. {5}------------------------------------------------ ### 6. Indication for use: LUXEN 5G is indicated for the production of all-ceramic inlays, multi-units bridges, onlays, and veneers without zirconium dioxide frameworks. # 7. Basis for Substantial Equivalence #### 1. Comparison Chart | Characteristics | Subject Device | Predicate Device | Summary | | |--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|--| | Device Name | LUXEN 5G | LUXEN ZR | None | | | Manufacturer | DENTALMAX CO.,<br>LTD | DENTALMAX CO.,<br>LTD | None | | | 510(k) Number | - | K171585 | None | | | Classification Name | Porcelain powder for<br>clinical use | Porcelain powder for<br>clinical use | Same | | | Product Code | EIH | EIH | Same | | | Device Class | II | II | Same | | | Indications for Use | LUXEN 5G is indicated<br>for the production of<br>allceramic inlays, multi<br>units bridges, onlays,<br>and veneers without<br>zirconium dioxide<br>frameworks. | LUXEN Zr is indicated<br>for the production of<br>allceramic inlays, multi<br>units bridges, onlays, and<br>veneers without<br>zirconium dioxide<br>frameworks. | Same | | | Principles of Operation | Zirconia oxide blank is<br>a material for crowns<br>and bridges fabricated<br>by CAD/CAM milling. | Zirconia oxide blank is<br>a material for crowns<br>and bridges fabricated<br>by CAD/CAM milling. | Same | | | Chemical Composition | Zirconia Powder Zpex<br>ZrO2+HfO2+Y2O3<br>+ various color powders | Zirconia Powder Zpex<br>ZrO2+HfO2+Y2O3<br>+ various color powders | Similar | | | Physical Characteristics & Mechanical Properties | | | | | | Standard Conformed | ISO 6872:2015 | ISO 6872:2015 | Same | | | Classification | Type II Class 5 | Type II Class 5 | Same | | | Flexural Strength (Mpa) | >800Mpa | 1038 ± 135 Mpa | Similar | | | Thermal expansion<br>coefficient(20-500°C) | 11.0 X 10-6 K-1 | 10.7 X 10-6 K-1 | Similar | | | Coefficient of thermal<br>expansion | | | | | | Chemical solubility(µg/cm²) | < 100 µg/cm² | < 100 µg/cm² | Same | | | Radioactivity | ≤ 1.0 Bq⋅g⁻¹ | ≤ 1.0 Bq⋅g⁻¹ | Same | | | Sterile | Non-sterile | Non-sterile | Same | | | Biocompatibility | | | | | | Standard Conformed | ISO 10993-1 | ISO 10993-1 | Same | | | Classification | Externally<br>communicating device<br>in oral mucosa, enamel,<br>and dentin; and contact<br>duration of C-long term<br>(>30d) | Externally<br>communicating device in<br>oral mucosa, enamel, and<br>dentin; and contact<br>duration of C-long term<br>(>30d) | Same | | {6}------------------------------------------------ #### 2. Substantial Equivalence Discussion The subject device has the same indications for use, principle of operation, and similar chemical compositions. The main components (ZrO2 + Hfo2 + Y2O3) and their content are similar to the comparator device, but the content of some elements is slightly different. However, the added elements are in very small amounts (<1%). We performed the biocompatibility and performance testing and the test results support that this difference does not raise a question in safety and performance. The devices also have similar mechanical characteristics. Although the flexural strength and thermal expansion coefficient values are slightly different but both conform ISO 6872:2016 Class 5 requirements (>800Mpa). Based on the information submitted herein, we determined that the subject device is substantially equivalent to the predicate device. # 8. Non-Clinical Testing - Performance Tests including Visual, Dimensions, Packaging, Uniformity, Freedom ● from extraneous materials, Chemical Solubility, Flexural Strength, Linear Thermal Expansion, Flexural Strength in accordance with ISO 6872 - Biocompatibility tests in accordance with ISO 10993-5, 10, 11, and 23 ● - Pyrogen Test in accordance with USP <151> ● # 9. Conclusion The subject device and the predicate device have the same intended use and have the same technological characteristics. Based on the similarities and the test results, we conclude that the subject device is substantially equivalent to the predicate device.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%