Oocyte Flushing & Retrieval Medium; Gamete Buffer
K234023 · Gimbo Medical Technology Shenzhen Co., Ltd. · MQL · May 10, 2024 · Obstetrics/Gynecology
Device Facts
| Record ID | K234023 |
| Device Name | Oocyte Flushing & Retrieval Medium; Gamete Buffer |
| Applicant | Gimbo Medical Technology Shenzhen Co., Ltd. |
| Product Code | MQL · Obstetrics/Gynecology |
| Decision Date | May 10, 2024 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 884.6180 |
| Device Class | Class 2 |
Intended Use
Oocyte Flushing & Retrieval Medium is intended for use during and oocyte collection procedures for use in in vitro fertilization procedures. Gamete Buffer is indicated for use in short-term handling gametes and embryos, including washing and intracytoplasmic sperm injection (ICSI).
Device Story
Oocyte Flushing & Retrieval Medium and Gamete Buffer are reproductive media used in clinical IVF laboratories. These solutions provide a controlled environment for oocyte retrieval, gamete/embryo handling, and ICSI procedures. They are used by embryologists or clinicians to maintain viability of gametes and embryos during laboratory manipulation. The media support the transition of samples from the patient to the incubator environment, facilitating successful fertilization and embryo development.
Indications for Use
Indicated for use in in vitro fertilization (IVF) procedures, specifically for oocyte collection (Oocyte Flushing & Retrieval Medium) and short-term handling, washing, and intracytoplasmic sperm injection (ICSI) of gametes and embryos (Gamete Buffer).
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Related Devices
- K153290 — Sydney IVF Follicle Flush Buffer, Sydney IVF Fertilization Medium, Sydney IVF Cleavage Medium, Sydney IVF Blastocyst Medium · William A. Cook Australia Pty. , Ltd. · Jan 27, 2016
- K200408 — VitaVitro® Fertilization Medium, VitaVitro® Gamete Buffer Medium, VitaVitro® Flushing Buffer Medium · Shenzhen Vitavitro Biotech Co., Ltd. · Oct 16, 2020
- K251637 — Dewin Reproductive Media (Dewin Gamete Buffer [with HSA and without HSA]; Dewin Follicle Flushing Solution, Dewin One Step medium [with HSA and without HSA]) · Donnevie Medical Technology (Shanghai) Co. , Ltd. · Aug 27, 2025
- K170498 — Gems Oocyte Retrieval Buffer ORB-20, ORB-50 · Genea Biomedx Pty, Ltd. · Sep 28, 2017
- K242640 — FertiCult Flushing medium; FertiCult Flushing medium with phenol red and gentamicin · Fertipro NV · May 30, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.
May 10, 2024
Gimbo Medical Technology Shenzhen Co., Ltd. Leo Guo General Manager 601, Building C5, No. 459 Qiaokai Road, Fenghuang Street, Guangming District Shenzhen. Guangdong Province 518107 China
Re: K234023
Trade/Device Name: Oocyte Flushing & Retrieval Medium; Gamete Buffer Regulation Number: 21 CFR\$ 884.6180 Regulation Name: Reproductive Media and Supplements Regulatory Class: II Product Code: MQL Dated: December 20, 2023 Received: April 9, 2024
Dear Leo Guo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Michael T. Bailey -S
For
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K234023
Device Name
Oocyte Flushing & Retrieval Medium; Gamete Buffer
Indications for Use (Describe)
Oocyte Flushing & Retrieval Medium is intended for use during and oocyte collection procedures for use in in vitro fertilization procedures.
Gamete Buffer is indicated for use in short-term handling gametes and embryos, including washing and intracytoplasmic sperm injection (ICSI).
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."