K Number
K233897

Validate with FDA (Live)

Date Cleared
2024-02-06

(57 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

C510S, G510S, CB5MP, MB5MP, C516S and G516S are intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners or certified personnel. They're intended to be used in digital mammography PACS, digital breast tomosynthesis and modalities including FFDM.

Device Description

5MP LCD Monitors C510S, G510S, CB5MP, MB5MP, C516S, G516S are 21.3-inch TFT LCD monitors, which are specifically designed to provide the high definition image output for general radiography. These products have been strictly calibrated so that they can meet DICOM Part 3.14 and other standards. They use the latest generation of LED backlight panel, supporting resolution 2560 x 2048. The built-in brightness stabilization control circuit makes sure the brightness of these monitors is stable and the calibration is continuous, so these products meet the demand of high precision medical imaging. The anti-glare screen can prevent display from reflection under highlight conditions, make the image and display clearer.

AI/ML Overview

The provided text describes a 510(k) premarket notification for medical display monitors and does not contain information about the acceptance criteria and study proving device meets the criteria. This document focuses on demonstrating substantial equivalence to a predicate device through comparison tables and performance testing rather than defining acceptance criteria and presenting a study with specific performance metrics against those criteria.

Therefore, I cannot fulfill your request to describe acceptance criteria and a study that proves the device meets the acceptance criteria based on the provided text.

Here's why the requested information is not available:

  1. Acceptance Criteria and Reported Device Performance: While the document presents detailed technical specifications of the proposed devices (e.g., resolution, contrast ratio, maximum luminance, DICOM calibrated luminance) and compares them to a predicate device, it does not explicitly state these specifications as "acceptance criteria" against which a study's results are evaluated. The document focuses on showing that the proposed device's characteristics are comparable to or within an acceptable range of the predicate device's characteristics. There is no specific table labeled "acceptance criteria" and "reported device performance."

  2. Sample Size and Data Provenance for Test Set: Performance testing was conducted, but the document does not mention a "test set" in the context of clinical images or patient data. The tests are bench tests on the display monitors themselves (e.g., measuring luminance, uniformity, response time). Therefore, there is no sample size for an image-based test set or data provenance (country of origin, retrospective/prospective).

  3. Number and Qualifications of Experts for Ground Truth: Since the performance testing comprises bench tests on the display's physical and optical characteristics, there is no "ground truth" derived from expert interpretation of medical images. The evaluation is based on technical measurements.

  4. Adjudication Method: Not applicable, as there are no expert interpretations requiring adjudication.

  5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: The document explicitly states: "No animal or clinical testing is needed for C510S, CB5MP, MB5MP, C516S, G516S." Therefore, no MRMC study or human reader performance evaluation (with or without AI assistance) was performed.

  6. Standalone (Algorithm Only) Performance: This is a medical display monitor, not an AI algorithm. Therefore, standalone algorithm performance is not applicable.

  7. Type of Ground Truth Used: The ground truth for the performance testing consists of technical standards and measurements (e.g., DICOM Part 3.14, AAPM TG18 guideline measurements) for display characteristics. It is not expert consensus, pathology, or outcomes data.

  8. Sample Size for Training Set: Not applicable. These are display monitors, not an AI model that requires a training set.

  9. How Ground Truth for Training Set was Established: Not applicable, as there is no training set.

The document primarily demonstrates substantial equivalence to a predicate device based on technical specifications and adherence to relevant industry standards for display performance in diagnostic radiology, as outlined in the bench tests described.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Shenzhen Beacon Display Technology Co., Ltd. Li Yafei Official Correspodent 15F, Building 6, Hengda Shishang Huigu (East) Fulong Road, Dalang Subdistrict, Longhua Shenzhen, Guangdong 518109, China

February 6, 2024

Re: K233897

Trade/Device Name: 5MP Color LCD Monitors/ 5MP Monochrome LCD Monitors (C510S, G510S. CB5MP, MB5MP, C516S, G516S) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: PGY Dated: December 11, 2023 Received: December 11, 2023

Dear Li Yafei:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb Assistant Director Imaging Software Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K233897

Device Name

5MP Color LCD Monitors C510S, CB5MP, C516S. 5MP Monochrome LCD Monitors G510S, MB5MP, G516S.

Indications for Use (Describe)

C510S, G510S, CB5MP, MB5MP, C516S and G516S are intended to be used in displaying and viewing medical images for diagnosis by trained medical practified personnel. They're intended to be used in digital mammography PACS, digital breast tomosynthesis and modalities including FFDM.

Type of Use (Select one or both, as applicable)

× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510 (k) Summary

{4}------------------------------------------------

510(k) Summary

[As required by 21 CFR 807.92]

1. Date Prepared [21 CFR807.92 (a) (1)]

Feb 04, 2024

2. Submitter's Information [21 CFR807.92 (a) (1)]

Name of Sponsor:Shenzhen Beacon Display Technology Co., Ltd.
Address:15F, Building 6, Hengda Shishang Huigu(East), Fulong Road, DalangSubdistrict, Longhua, Shenzhen, 518109 China
Contact Name:Li Yafei
Telephone No.:+86-024-88087610
Fax No.:+86-024-88087629
Email Address:liyf@beacon-display.cn

3. Trade Name, Common Name, Classification [21 CFR807.92 (a) (2)]

● Trade Name/Model:5MP Color LCD Monitors C510S, CB5MP, C516S.5MP Monochrome LCD Monitors G510S, MB5MP, G516S.
● Common Name:5MP LCD Monitors C510S, G510S, CB5MP, MB5MP, C516S,G516S
● Classification Name:Medical image management and processing system
● Regulation Number:21 CFR 892.2050
● Product code:PGY
● Classification Panel:Radiology
● Device Class:II

4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]

{5}------------------------------------------------

The identified predicate within this submission is as follows:

JVC Kenwood Corporation, 5MP Color LCD Monitor CL-S500, 5MP Monochrome LCD Monitor MS-S500 has been cleared by FDA through 510(k) No. K191137 (Decision Date -July 2, 2019).

5. Description of the Device [21 CFR 807.92(a) (4)]

5MP LCD Monitors C510S, G510S, CB5MP, MB5MP, C516S, G516S are 21.3-inch TFT LCD monitors, which are specifically designed to provide the high definition image output for general radiography.

These products have been strictly calibrated so that they can meet DICOM Part 3.14 and other standards. They use the latest generation of LED backlight panel, supporting resolution 2560 x 2048. The built-in brightness stabilization control circuit makes sure the brightness of these monitors is stable and the calibration is continuous, so these products meet the demand of high precision medical imaging. The anti-glare screen can prevent display from reflection under highlight conditions, make the image and display clearer.

C510S, CB5MP, C516S have the same color LCD panel and similar functionality. The difference between C510S and CB5MP is only different customers which results in the two models. For C510S, CB5MP, C516S, the difference is the main board and design principle.

G510S, MB5MP, G516S have the same monochrome LCD panel and similar functionality. The difference between G510S and MB5MP is only different customers which results in the two models. For G510S, MB5MP, G516S, the difference is the main board and design principle.

6. Indications for use

C510S, G510S, CB5MP, MB5MP, C516S and G516S are intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners or certified personnel. They're intended to be used in digital mammography PACS, digital breast tomosynthesis and modalities including FFDM.

{6}------------------------------------------------

7. Technological Characteristics [21 CFR 807.92(a)(6)]

Panel characteristics:
Screen technology21.3", TFT, color, LCD screen, anti-glare, hard coating
Active area (H x V)422.4 x 337.92 mm
Pixel pitch0.165 mm (H) x 0.165 mm (V)
Resolution2560 x 2048
Contrast ratio2000:1 (typical), 1600:1 (minimal)
Viewing angle(CR > 10)Horizontal: 178° (Typical)Vertical: 178° (Typical)
Screen brightness1150 cd/m² (typical)
Refresh rate50 Hz
BacklightingLEDs
Lifetime of backlight50,000 hours
Response time (TR+TF)11 + 14 ms (typical)
Color support1.07G
Power supply:
Line voltageDC 24 V
Current consumption5.0 A
Power consumption< 100 W
Standby≤ 0.5 W
Control and connection:
Front1 operation LED, 6 functional keys
BackPower switch1 DVI-D1 DP1 USB Upstream Port2 USB Downstream Port*4

C510S, CB5MP Color LCD monitors

{7}------------------------------------------------

Mechanical characteristics:
Housing componentsPlastic
Ventilation openingsNatural heat radiation
Degree of protectionIP20
Climatic conditions:
Working temperature0°C - 40°C
Working humidity15% - 85% Relative humidity, no condensation
Transport and storage temperature-20°C - 60°C
Transport and storage humidity10% - 90% Relative humidity, no condensation
Air pressure700hPa - 1060hPa
Safety regulations:
Safety standardsIEC 60601-1, IEC 60601-1-2EN 60601-1, EN 60601-1-2ANSI/AAMI ES60601-1: 2005 + A2(R2012) + A1CAN/CSA-C22.2 NO. 60601-1:14
ConformityCCC, CE, FCC, TUV, CB, China energy
Dimension:
With packing (W x H x D)483 x 321 x 627 mm
Weight:
Gross weightApproximate 12.6 ± 0.5 kg
Net weightApproximate 9.1 ± 0.5 kg

G510S, MB5MP Monochrome LCD Monitors

Panel characteristics:
Screen technology21.3", TFT, grayscale, LCD screen, anti-glare, hard coating
Active area (H x V)422.4 x 337.92 mm
Pixel pitch0.165 mm (H) x 0.165 mm (V)

{8}------------------------------------------------

Resolution2560 x 2048
Contrast ratio2000:1 (typical), 1600:1 (minimal)
Viewing angle(CR > 10)Horizontal: 178° (Typical)Vertical: 178° (Typical)
Screen brightness3000 cd/m² (typical)
Refresh rate50 Hz
BacklightingLEDs
Lifetime of backlight50,000 hours
Response time (TR+TF)11 + 14 ms (typical)
Power supply:
Line voltageDC 24 V
Current consumption5.0 A
Power consumption< 100 W
Standby≤ 0.5 W
Control and connection:
Front1 operation LED, 6 functional keys
Back● Power switch1● DVI-D1● DP1● USB Upstream Port2● USB Downstream Port*4
Mechanical characteristics:
Housing componentsPlastic
Ventilation openingsNatural heat radiation
Degree of protectionIP20
Climatic conditions:
Working temperature0°C - 40°C
Working humidity15% - 85% Relative humidity, no condensation
Transport and storage temperature-20°C - 60°C
Transport and storage humidity10% - 90% Relative humidity, no condensation
Air pressure700hPa - 1060hPa
Safety regulations :
IEC 60601-1, IEC 60601-1-2
EN 60601-1, EN 60601-1-2
Safety standardsANSI/AAMI ES60601-1: 2005 + A2(R2012) + A1
CAN/CSA-C22.2 NO. 60601-1:14
ConformityCCC, CE, FCC, TUV, CB, China energy
Dimension:
With packing (W x H x D)483 x 321 x 627 mm
Weight :
Gross weightApproximate 12.6 ± 0.5 kg
Net weightApproximate 9.1 ± 0.5 kg

{9}------------------------------------------------

C516S Color LCD monitor

Panel characteristics:
Screen technology21.3", TFT, color, LCD screen, anti-glare, hard coating
Active area(H x V)422.4 x 337.92 mm
Pixel pitch0.165 mm (H) x 0.165 mm (V)
Resolution2560 x 2048
Contrast ratio2000:1 (typical), 1600:1 (minimal)
Viewing angle(CR > 10)Horizontal: 178° (Typical)
Vertical: 178° (Typical)
Screen brightness1150 cd/m² (typical)
Refresh rate50 Hz
BacklightingLEDs
Lifetime of backlight50,000 hours
Response time (TR+TF)11 + 14 ms (typical)
Color support1.07G
Power supply:
Power connectionAdapter
Output+24VDC, 5.0A
Power consumption:
Normal Mode100W (Maximum)
Standby Mode0.5 W (Maximum)
Control and connection :
Front1 operation LED, 6 functional keys
BackPower switch1 DC Input1 DVI-D1 DP2 DP-OUT1 USB-A1 USB Upstream Port2 USB Downstream Port4
Mechanical characteristics :
Housing componentsPlastic
Ventilation openingsNatural heat radiation
Degree of protectionIP20
Climatic conditions :
Working temperature0°C - 40°C
Working humidity15% - 85% Relative humidity, no condensation
Transport and storage temperature-20°C - 60°C
Transport and storage humidity10% - 90% Relative humidity, no condensation
Air pressure700hPa - 1060hPa
Safety regulations :
Safety standardsIEC 60601-1, IEC 60601-1-2EN 60601-1, EN 60601-1-2ANSI/AAMI ES60601-1: 2005 + A2(R2012) + A1CAN/CSA-C22.2 NO. 60601-1:14
ConformityCCC, CE, FCC, cTUVus, Energy ConservationCertification
Dimension:
With packing(W x H x D)483 x 321 x 627 mm
Weight:
Gross weightApproximate $13.3 \pm 0.5$ kg
Net weightApproximate $9.3 \pm 0.5$ kg

{10}------------------------------------------------

510(k) Summary

{11}------------------------------------------------

G516S Monochrome LCD Monitor

Panel characteristics:
Screen technology21.3", TFT, grayscale, LCD screen, anti-glare, hard coating
Active area(H x V)422.4 x 337.92 mm
Pixel pitch0.165 mm(H) x 0.165 mm(V)
Resolution2560 x 2048
Contrast ratio2000:1 (typical), 1600:1 (minimal)
Viewing angle(CR > 10)Horizontal: 178°(Typical)
Vertical: 178°(Typical)
Screen brightness3000 cd/m²(typical)
Refresh rate50 Hz
BacklightingLEDs
Lifetime of backlight50,000 hours
Response time(TR+TF)11 + 14 ms(typical)
Power supply:
Power connectionAdapter
Output+24VDC, 5.0A
Power consumption:
Normal Mode100W (Maximum)
Standby Mode0.5 W (Maximum)
Control and connection :
Front1 operation LED, 6 functional keys
BackPower switch1 DC Input1 DVI-D1 DP2 DP-OUT1 USB-A1 USB Upstream Port2 USB Downstream Port4
Mechanical characteristics :
Housing componentsPlastic
Ventilation openingsNatural heat radiation
Degree of protectionIP20
Climatic conditions :
Working temperature0°C - 40°C
Working humidity15% - 85% Relative humidity, no condensation
Transport and storage temperature-20°C - 60°C
Transport and storage humidity10% - 90% Relative humidity, no condensation
Air pressure700hPa - 1060hPa
Safety regulations :
Safety standardsIEC 60601-1, IEC 60601-1-2EN 60601-1, EN 60601-1-2
ANSI/AAMI ES60601-1: 2005 + A2(R2012) + A1CAN/CSA-C22.2 NO. 60601-1:14
ConformityCCC, CE, FCC, cTUVus, Energy ConservationCertification
Dimension:
With packing(W x H x D)483 x 321 x 627 mm
Weight:
Gross weightApproximate 13.3 ± 0.5 kg
Net weightApproximate 9.3 ± 0.5 kg

{12}------------------------------------------------

510(k) Summary

{13}------------------------------------------------

{14}------------------------------------------------

8. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92]

8.1 Indication for use:

IDComparison ItemProposed Device5MP Color LCD Monitors C510S, CB5MP,C516S.5MP Monochrome LCD Monitors G510S,MB5MP, G516S.Predicate Device5MP Color LCD Monitor CL-S500,5MP Monochrome LCD Monitor MS-S500
1Indication for useC510S, G510S, CB5MP, MB5MP, C516S and G516S are intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners or certified personnel. They're intended to be used in digital mammography PACS, digital breast tomosynthesis and modalities including FFDM.CL-S500 and MS-S500 are intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners or certified personnel. They're intended to be used in digital mammography PACS, digital breast tomosynthesis and modalities including FFDM.
IDComparisonItemProposed Device5MP Color LCD MonitorsC510S, CB5MPPredicate Device5MP Color LCD MonitorCL-S500Explanation of Differences
2Display Performance/Specifications
2.1Screen TechnologyTFT Color LCD Panel (IPS)TFT Color LCD Panel (IPS)-
2.2Viewing angle (H, V)CR>10Horizontal: Typ.178Vertical: Typ.178CR>50Horizontal: Typ.178Vertical: Typ.178Different screen provided by thedifferent manufacturer
2.3Resolution ormatrix size5MP(2560 x 2048)5MP(2048 X 2560)-
2.4Aspect ratio5:44:5-
2.5Display areaHorizontal: 422.4mmVertical: 337.92mmHorizontal: 337.92mmVertical: 422.4mm-
2.6Pixel pitchHorizontal: 0.165mmVertical: 0.165mmHorizontal: 0.165mmVertical: 0.165mm-
2.7Response time (Typ.)25ms (On/Off)25ms (On/Off)-
2.8Maximum luminanceMin. 920cd/m²Typ. 1,150cd/m²Min. 920cd/m²Typ. 1,150cd/m²-
2.9DICOM calibratedluminance450cd/m²500cd/m²Difference complies the DICOM3.14.
2.10Contrast ratioMin. 1600:1Typ. 2000:1Min. 1600:1Typ. 2000:1-
2.11BacklightingLEDLED-
2.12Grayscale Tones10-bit (DisplayPort): 1.073billion1024 from a palette of 16,384tones8-bit(DVI): 16.77 million 256from a palette of 16,384 tones10-bit (DisplayPort): 1.073billion1024 from a palette of 16,369tones8-bit(DVI): 16.77 million 256from a palette of 16,369 tonesTone between the predicatedevice and our proposeddevices are different. But theypass the exams in AAPM-TG184.3 "Luminance Response".Therefore, they are equivalentto the predicate device.
2.13Luminancenon-uniformitycompensation-Digital Uniformity CorrectionSystemDifferent design scheme
3Video Signals
3.1Input video signalsDVI-D x1DisplayPort x1DVI-D x1DisplayPort x1-
3.2Scanning Frequency(H, V)Landscape:Horizontal: 104.852KHzVertical: 50HzPortrait:Horizontal: 129.1KHzVertical: 50HzLandscape:Horizontal: 103.5KHzVertical: 50HzDifferent design scheme
3.3Dot Clock285.2 MHz285 MHz-
4Power Related Specifications
4.1Power RequirementsDC 24V, 5AAC 100-240V, 50/60HzDifference between built-inpower supply and built-outpower supply
4.2Power Consumption /Save Mode100WLess than 0.5W95WLess than 1WCompared with the predicatedevice, the proposed deviceconsumes more power in theoperating mode and less powerin the power saving mode.
4.3Power ManagementVESA DPMS and EPA powersaving,DisplayPort 1.2DVI DMPM,DisplayPort 1.2aDifferent design scheme
5Miscellaneous Features/Specifications
5.1Calibration toolsBeacon Monitor Manage/Calibration Feedback SystemQA Medivisor/ MedivisorNXF-CALDifferent design scheme
5.2SensorsBuilt-in Front SensorBuilt-in Ambient Light SensorBuilt-in Backlight SensorBuilt-in Front SensorBuilt-in Ambient Light SensorDifferent design scheme
5.3USB Ports/Standard2 upstream,4 downstream / Rev. 2.01 upstream,2 downstream / Rev. 2.0
5.4Dimensions withstand (W x H x D)360.5 x 513.49/613.49 x 224mm361.5 x 517/612 x 196.5 mmDifferent housing design due tothe different panel size

8.2 Comparison table

{15}------------------------------------------------

Table 2 General Comparison of C510S, CB5MP

{16}------------------------------------------------

{17}------------------------------------------------

{18}------------------------------------------------

IDComparisonItemProposed Device5MP Monochrome LCDMonitors G510S, MB5MPPredicate Device5MP Monochrome LCDMonitor MS-S500Explanation of Differences
2Display Performance/Specifications
2.1Screen TechnologyTFT Monochrome LCD Panel(IPS)TFT Monochrome LCD Panel(IPS)-
2.2Viewing angle (H, V)CR>10Horizontal: Typ.178Vertical: Typ.178CR>50Horizontal: Typ.178Vertical: Typ.178Different screen provided by thedifferent manufacturer
2.3Resolution ormatrix size5MP (2560 X 2048)5MP (2048 X 2560)-
2.4Aspect ratio5: 44:5-
2.5Display AreaHorizontal: 422.4mmVertical: 337.92mmHorizontal: 337.92mmVertical: 422.4mm-
2.6Pixel pitchHorizontal: 0.165mmVertical: 0.165mmHorizontal: 0.165mmVertical: 0.165mm-
2.7Response Time (Typ.)25ms (On/Off)25ms (On/Off)-
2.8Maximum luminanceMin. 2,400cd/m²Typ. 3,000cd/m²Min. 2,400cd/m²Typ. 3,000cd/m²-
2.9DICOM calibratedluminance450cd/m²1,000cd/m²Difference complies the DICOM3.14.
2.10Contrast ratioMin. 1600:1Typ. 2000:1Min. 1600:1Typ. 2000:1-
2.11BacklightingLEDLED-
2.12Grayscale Tones10-bit (DisplayPort): 1,024from a palette of 16,384 tones8-bit (DVI): 256 from a paletteof 16,384 tones10-bit (DisplayPort): 1,024from a palette of 16,369 tones8-bit (DVI): 256 from a paletteof 16,369 tonesTone between the predicatedevice and our proposeddevices are different. But theypass the exams in AAPM-TG184.3 "Luminance Response".Therefore, they are equivalentto the predicate device.
2.13Luminancenon-uniformitycompensation-Digital Uniformity CorrectionSystemDifferent design scheme
3Video Signals
3.1Input video signalsDVI-D x1DisplayPort x1DVI-D x1DisplayPort x1-
3.2Scanning Frequency(H, V)Landscape:Horizontal: 104.852KHzVertical: 50HzPortrait:Horizontal: 129.1KHzVertical: 50HzLandscape:Horizontal: 103.5KHzVertical: 50HzDifferent design scheme
3.3Dot Clock285.2 MHz285 MHz-
4Power Related Specifications
4.1Power RequirementsDC 24V, 5.0AAC 100-240V, 50/60HzDifference between built-inpower supply and built-outpower supply
4.2Power Consumption /Save Mode100WLess than 0.5W80WLess than 1WCompared with the predicatedevice, the proposed deviceconsumes more power in theoperating mode and less powerin the power saving mode.
4.3Power ManagementVESA DPMS and EPA powersaving,DisplayPort 1.2DVI DMPM,DisplayPort 1.2aDifferent design scheme
5Miscellaneous Features/Specifications
5.1Calibration toolsBeacon Monitor Manage/Calibration Feedback SystemQA Medivisor/ MedivisorNXF-CALDifferent design scheme
5.2SensorsBuilt-in Front SensorBuilt-in Ambient Light SensorBuilt-in backlight sensorBuilt-in Front SensorBuilt-in Ambient Light SensorDifferent design scheme
5.3USB Ports/Standard2 upstream,4 downstream / Rev. 2.01 upstream,2 downstream / Rev. 2.0-
5.4Dimensions withstand (W x H x D)360.5 x 513.49/613.49 x 224mm361.5 x 517/612 x 196.5 mmDifferent housing design due tothe different panel size

Table 3 General Comparison of G510S, MB5MP

{19}------------------------------------------------

{20}------------------------------------------------

Table 4 General Comparison of C516S

IDComparisonItemProposed Device5MP Color LCD MonitorsC516SPredicate Device5MP Color LCD MonitorCL-S500Explanation of Differences
2Display Performance/Specifications
2.1Screen TechnologyTFT Color LCD Panel (IPS)TFT Color LCD Panel (IPS)-
2.2Viewing angle (H, V)CR>10Horizontal: Typ.178Vertical: Typ.178CR>50Horizontal: Typ.178Vertical: Typ.178Different screen provided by thedifferent manufacturer
2.3Resolution ormatrix size5MP(2560 x 2048)5MP(2048 X 2560)-
2.4Aspect ratio5:44:5-
2.5Display areaHorizontal: 422.4mmVertical: 337.92mmHorizontal: 337.92mmVertical: 422.4mm-
2.6Pixel pitchHorizontal: 0.165mmVertical: 0.165mmHorizontal: 0.165mmVertical: 0.165mm-
2.7Response time (Typ.)25ms (On/Off)25ms (On/Off)-
2.8Maximum luminanceMin. 920cd/m²Typ. 1,150cd/m²Min. 920cd/m²Typ. 1,150cd/m²-
2.9DICOM calibratedluminance450cd/m²500cd/m²Difference complies the DICOM3.14.
2.10Contrast ratioMin. 1600:1Typ. 2000:1Min. 1600:1Typ. 2000:1-
2.11BacklightingLEDLED-
2.12Grayscale Tones10-bit (DisplayPort): 1.073 billion1024 from a palette of 16,384 tones8-bit(DVI): 16.77 million 256 from a palette of 16,384 tones10-bit (DisplayPort): 1.073 billion1024 from a palette of 16,369 tones8-bit(DVI): 16.77 million 256 from a palette of 16,369 tonesTone between the predicate device and our proposed devices are different. But they pass the exams in AAPM-TG18 4.3 "Luminance Response" . Therefore, they are equivalent to the predicate device.
2.13Luminance non-uniformity compensation-Digital Uniformity Correction SystemDifferent design scheme
3Video Signals
3.1Input video signalsDVI-D x1DisplayPort x2DVI-D x1DisplayPort x1Different design scheme
3.2Scanning Frequency (H, V)Landscape:Horizontal: 104.852KHzVertical: 50HzPortrait:Horizontal: 129.1KHzVertical: 50HzLandscape:Horizontal: 103.5KHzVertical: 50HzDifferent design scheme
3.3Dot Clock285.2 MHz285 MHz-
4Power Related Specifications
4.1Power RequirementsDC 24V, 5AAC 100-240V, 50/60HzDifference between built-in power supply and built-out power supply
4.2Power Consumption /Save Mode100WLess than 0.5W95WLess than 1WCompared with the predicatedevice, the proposed deviceconsumes more power in theoperating mode and less powerin the power saving mode.
4.3Power ManagementVESA DPMS and EPA powersaving,DisplayPort 1.2DVI DMPM,DisplayPort 1.2aDifferent design scheme
5Miscellaneous Features/Specifications
5.1Calibration toolsBeacon Monitor Manage/Calibration Feedback SystemQA Medivisor/ MedivisorNXF-CALDifferent design scheme
5.2SensorsBuilt-in Front SensorBuilt-in Ambient Light SensorBuilt-in Backlight SensorBuilt-in Presence SensorGravity SensorBuilt-in Front SensorBuilt-in Ambient Light SensorDifferent design scheme
5.3USB Ports/Standard2 upstream,4 downstream / Rev. 2.01 USB-A/ Rev. 2.01 upstream,2 downstream / Rev. 2.0
5.4Dimensions withstand (W x H x D)360.5 x 513.49/613.49 x 224mm361.5 x 517/612 x 196.5 mmDifferent housing design due tothe different panel size

{21}------------------------------------------------

{22}------------------------------------------------

{23}------------------------------------------------

{24}------------------------------------------------

IDComparisonItemProposed Device5MP Monochrome LCDMonitors G516SPredicate Device5MP Monochrome LCDMonitor MS-S500Explanation of Differences
2Display Performance/Specifications
2.1Screen TechnologyTFT Monochrome LCD Panel(IPS)TFT Monochrome LCD Panel(IPS)-
2.2Viewing angle (H, V)CR>10Horizontal: Typ.178Vertical: Typ.178CR>50Horizontal: Typ.178Vertical: Typ.178Different screen provided by thedifferent manufacturer
2.3Resolution ormatrix size5MP (2560 X 2048)5MP (2048 X 2560)-
2.4Aspect ratio5: 44:5-
2.5Display AreaHorizontal: 422.4mmVertical: 337.92mmHorizontal: 337.92mmVertical: 422.4mm-
2.6Pixel pitchHorizontal: 0.165mmVertical: 0.165mmHorizontal: 0.165mmVertical: 0.165mm-
2.7Response Time (Typ.)25ms (On/Off)25ms (On/Off)-
2.8Maximum luminanceMin. 2,400cd/m²Typ. 3,000cd/m²Min. 2,400cd/m²Typ. 3,000cd/m²-
2.9DICOM calibratedluminance450cd/m²1,000cd/m²Difference complies the DICOM3.14.
2.10Contrast ratioMin. 1600:1Typ. 2000:1Min. 1600:1Typ. 2000:1-
2.11BacklightingLEDLED-
2.12Grayscale Tones10-bit (DisplayPort): 1,024from a palette of 16,384 tones8-bit (DVI): 256 from a paletteof 16,384 tones10-bit (DisplayPort): 1,024from a palette of 16,369 tones8-bit (DVI): 256 from a paletteof 16,369 tonesTone between the predicatedevice and our proposeddevices are different. But theypass the exams in AAPM-TG184.3 "Luminance Response"Therefore, they are equivalentto the predicate device.
2.13Luminancenon-uniformitycompensation-Digital Uniformity CorrectionSystemDifferent design scheme
3Video Signals
3.1Input video signalsDVI-D x1DisplayPort x2DVI-D x1DisplayPort x1-
3.2Scanning Frequency(H, V)Landscape:Horizontal: 104.852KHzVertical: 50HzPortrait:Horizontal: 129.1KHzVertical: 50HzLandscape:Horizontal: 103.5KHzVertical: 50HzDifferent design scheme
3.3Dot Clock285.2 MHz285 MHz-
4Power Related Specifications
4.1Power RequirementsDC 24V, 5.0AAC 100-240V, 50/60HzDifference between built-inpower supply and built-outpower supply
4.2Power Consumption /Save Mode100WLess than 0.5W80WLess than 1WCompared with the predicatedevice, the proposed deviceconsumes more power in theoperating mode and less powerin the power saving mode.
4.3Power ManagementVESA DPMS and EPA powersaving,DisplayPort 1.2DVI DMPM,DisplayPort 1.2aDifferent design scheme
5Miscellaneous Features/Specifications
5.1Calibration toolsBeacon Monitor Manage/Calibration Feedback SystemQA Medivisor/ MedivisorNXF-CALDifferent design scheme
5.2SensorsBuilt-in Front SensorBuilt-in Ambient Light SensorBuilt-in Backlight SensorBuilt-in Presence SensorGravity SensorBuilt-in Front SensorBuilt-in Ambient Light SensorDifferent design scheme
5.3USB Ports/Standard2 upstream,4 downstream / Rev. 2.01 USB-A/ Rev. 2.01 upstream,2 downstream / Rev. 2.0
5.4Dimensions withstand (W x H x D)360.5 x 513.49/613.49 x 224mm361.5 x 517/612 x 196.5 mmDifferent housing design due tothe different panel size

Table 5 General Comparison of G516S

{25}------------------------------------------------

{26}------------------------------------------------

{27}------------------------------------------------

The technological characteristics differences discussed above do not affect the safety and the effectiveness of the C510S, G510S, CB5MP, MB5MP, C516S, G516S.

8.3 Performance Testing

According to the instructions in" Guidance for Industry and Food and Drug Administration Staff Display Devices for Diagnostic Radiology", the bench tests were performed on C510S, G510S, CB5MP, MB5MP, C516S, G516S as below.

  • Verify the conformance to DICOM GSDF in accordance with Assessment of Display Performance for Medical Imaging Systems by AAPM Task Group 18 (TG18 quideline).
  • Measure the calibrated luminance according to the requirements of the "Luminance Response" test in TG18 guideline.
  • Measure the luminance non-uniformity characteristics of the display screen in accordance with TG18 guideline.
  • Measure the luminance stability and chromaticity response with the temperature 0 ℃, 25°C and 40°C on Luminance response by AAPM-TG18.
  • Measure the chromaticity non-uniformity characteristics of the display screen in accordance with TG18 quideline.
  • Measure the luminance at the angles of 30% and 45º in diagonal, horizontal and vertical directions at center and four corners by AAPM-TG18.
  • Measure the temporal response using the typical data provided by the panel manufacturer.
  • Visually check the presence or absence of miscellaneous artifacts on the display screen in accordance with TG18 guideline.
  • Measure the spatial noise by noise power spectrum.
  • Measure the reflection coefficient with specular reflection and diffuse reflection by TG18.
  • Measure the veiling glare of small-spot contrast performing veiling glare test by TG-18.
  • Measure the spatial resolution expressed as modulation transfer function (MTF)
  • Maximum number allowed for each type of pixel defects/faults
  • Measurement of Color tracking and Gray tracking

{28}------------------------------------------------

  • Measure pixel fill factor like pixel structure and aperture ratio etc.
    The test results showed that C510S, G510S, CB5MP, MB5MP, C516S, G516S are with display characteristics equivalent to those of the predicate devices, MS-S500, CL-S500 except some items, each of which was determined that it would not affect observer's performance.

No animal or clinical testing is needed for C510S, CB5MP, MB5MP, C516S, G516S.

9. Conclusion [21 CFR 807.92(b) (3)]

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, Shenzhen Beacon Display Technology Co., Ltd. concludes that:

  • The intended use of C510S, G510S, CB5MP, MB5MP, C516S, G516S is totally same as that of the predicate devices.
  • The technological characteristics differences between C510S, G510S, CB5MP, MB5MP, C516S, G516S and the predicate devices do not affect the safety and effectiveness, so no new risk is raised.
  • Demonstrated by the bench tests, the display characteristics of C510S, G510S, CB5MP, MB5MP, C516S, G516S are equivalent to those of the predicate devices.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).