IPL Hair Removal Device (UI06 PN, UI06 PL, UI06 JL, UI06 BR, UI06 DB, UI06 PR, UI06 OG, UI06 RD,UI06 WH, UI06 MG, UI06 OB, UI06 LY,UI06 IG,)
K233664 · Shenzhen Ulike Smart Electronics Co., Ltd. · OHT · Jan 11, 2024 · General, Plastic Surgery
Device Facts
| Record ID | K233664 |
| Device Name | IPL Hair Removal Device (UI06 PN, UI06 PL, UI06 JL, UI06 BR, UI06 DB, UI06 PR, UI06 OG, UI06 RD,UI06 WH, UI06 MG, UI06 OB, UI06 LY,UI06 IG,) |
| Applicant | Shenzhen Ulike Smart Electronics Co., Ltd. |
| Product Code | OHT · General, Plastic Surgery |
| Decision Date | Jan 11, 2024 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4810 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
IPL Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Device Story
IPL Hair Removal Device uses intense pulsed light (IPL) technology to target and remove unwanted hair; intended for over-the-counter use by consumers for home-based hair removal; device emits light pulses absorbed by hair follicles to inhibit regrowth; provides long-term, stable reduction in hair regrowth; operation involves user-directed application to skin areas; benefits include convenient, non-invasive hair removal and reduction.
Clinical Evidence
No clinical data provided in the document.
Technological Characteristics
Intense Pulsed Light (IPL) hair removal device; intended for over-the-counter use; Class II device under 21 CFR 878.4810.
Indications for Use
Indicated for removal of unwanted hair and permanent reduction in hair regrowth in individuals seeking hair removal.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Related Devices
- K233887 — IPL Hair Removal Device (UI06PN, UI06 PL, UI06 JL, UI06 BR, UI06 DB, UI06 PR, UI06 OG, UI06 RD, UI08 BR, UI08 DB, UI08 BL, UI08 GP, UI08 PN, UI08 RE, UI08 WH, UI08 PW, UI07 BR, UI07 DB, UI07 BL, UI07 GP, UI07 PN, UI07 RE, UI07 WH, UI07 PW, UI20 BR, UI20 DB, UI20 BL, UI20 GP, UI20 PN, UI20 RE, UI20 WH, UI20 PW) · Shenzhen Ulike Smart Electronics Co., Ltd. · Jan 22, 2024
- K233679 — IPL Hair Removal Device (UI04 MK,UI04 BN,UI04 BU,UI04 CB,UI04 GY,UI04 LG,UI04 DG,UI04 PL,UI04 PN,UI04 SD,UI04 SG,UI04 MG) · Shenzhen Ulike Smart Electronics Co., Ltd. · Jan 5, 2024
- K240282 — IPL Hair Removal Device (model: KCA450, KCA446, KCA504, KCA505, KCA506, KCA507, KCA508, KCA509, KCA501, KCA423, KCA448, KCA449, KCA502, KCA521) · Dongguan Boyuan Intelligent Technology Co.,Ltd · Apr 3, 2024
- K120080 — SHASER HRS2 HAIR REMOVAL SYSTEM · Shaser, Inc. · May 25, 2012
- K230473 — IPL Hair Removal Device Models: MLY-T069 · Miley Technology (Hebei) Co., Ltd. · Jun 15, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 11, 2024
Shenzhen Ulike Smart Electronics Co.,Ltd Blue Yang Registration Director 810, Building 1, Xunmei Science and Technology Plaza No.8 Keyuan Road, Science Park Community, Yuehai Sub-District Shenzhen, Nanshan District, Guangdong 518000 China
Re: K233664
Trade/Device Name: IPL Hair Removal Device (UI06 PN, UI06 JL, UI06 JL, UI06 BR, UI06 DB, UI06 PR, UI06 OG, UI06 RD,UI06 WH, UI06 MG, UI06 OB, UI06 LY,UI06 IG.) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: November 15, 2023 Received: November 15, 2023
Dear Blue Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tanisha L. Tanisha L. Hithe -2024.01.11 Hithe -S 15:45:11 -05'00' Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K233664
Device Name
IPL Hair Removal Device (U106 PN, U106 PL, U106 JL, U106 BR, U106 DB, U106 PR, U106 OG, U106 RD, U106 WH, U106 MG, U106 OB, U106 LY, U106 IG, )
Indications for Use (Describe)
IPL Hair Removal Device is indicated for the removal of unwanted hair. The device is also indicated for the permanent reduction in hair regrowth, defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."