SIRONA 100 Therapy Laser System (OEM's Model EVANLAS)

K231997 · Cure Point Therapeutics, Inc. · ILY · Oct 15, 2024 · Physical Medicine

Device Facts

Record IDK231997
Device NameSIRONA 100 Therapy Laser System (OEM's Model EVANLAS)
ApplicantCure Point Therapeutics, Inc.
Product CodeILY · Physical Medicine
Decision DateOct 15, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5500
Device ClassClass 2
AttributesTherapeutic

Intended Use

The SIRONA 100 Therapy Laser System emits energy in the visible and near-infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm; the temporary increase in local blood circulation; and the temporary relaxation of muscle. The SIRONA 100 Therapy Laser System is also indicated for use for the temporary increase of clear nail in patients with onychomycosis including dermatophyton rubrum and T. mentagrophytes, and/ or yeasts Candida albicans, etc.

Device Story

SIRONA 100 Therapy Laser System is a laser surgical instrument; emits 810 nm and 980 nm wavelengths. Device provides topical heating to elevate tissue temperature; intended for temporary relief of minor muscle/joint pain, stiffness, arthritis, muscle spasm, and increased local blood circulation. 980 nm wavelength indicated for temporary increase of clear nail in onychomycosis patients. Operated by clinicians in professional settings; prescription use only. Device delivers laser energy to target tissues; clinical benefit derived from thermal effects and potential therapeutic interaction with fungal pathogens in nail beds.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Laser surgical instrument; emits 810 nm and 980 nm wavelengths; visible and near-infrared spectrum. Class II device. Prescription use.

Indications for Use

Indicated for patients with minor muscle/joint pain, stiffness, arthritis pain, muscle spasm, or onychomycosis (including dermatophyton rubrum, T. mentagrophytes, and Candida albicans).

Regulatory Classification

Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 15, 2024 Cure Point Therapeutics, Inc. % Alexander Henderson Regulatory Consultant - Official Correspondent BraunSolutions Regulatory Group 2298 Americus Blvd. East, Unit 14 Clearwater, Florida 82001 Re: K231997 Trade/Device Name: SIRONA 100 Therapy Laser System (OEM's Model EVANLAS) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: PDZ, ILY Dated: September 7, 2024 Received: September 9, 2024 Dear Alexander Henderson: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). {2}------------------------------------------------ Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@;tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Digitally signed by TANISHA L. HITHE -S TANISHA L. HITHE -S Date: 2024.10.15 Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K231997 Device Name SIRONA 100 Therapy Laser System Indications for Use (Describe) Product Code ILY - 810 and 980 nm wavelength: The SIRONA 100 Therapy Laser System emits energy in the visible and near-infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm; the temporary increase in local blood circulation; and the temporary relaxation of muscle. Product Code PDZ - 980 nm wavelength: The SIRONA 100 Therapy Laser System is also indicated for use for the temporary increase of clear nail in patients with onychomycosis including dermatophyton rubrum and T. mentagrophytes, and/ or yeasts Candida albicans, etc. | Type of Use ( <i>Select one or both, as applicable</i> ) | | |-----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------| | <span style="padding-right: 5px;"> ☒ </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="padding-right: 5px;"> □ </span> Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...