Zhenjiang Assure Mechanical Wheelchair,model:K1

K231320 · Zhenjiang Assure Medical Equipment Co., Ltd. · IOR · Jun 5, 2023 · Physical Medicine

Device Facts

Record IDK231320
Device NameZhenjiang Assure Mechanical Wheelchair,model:K1
ApplicantZhenjiang Assure Medical Equipment Co., Ltd.
Product CodeIOR · Physical Medicine
Decision DateJun 5, 2023
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 890.3850
Device ClassClass 1

Intended Use

The device is intended for medical purpose to provide mobility to persons restricted to a sitting position.

Device Story

Zhenjiang Assure Mechanical Wheelchair (Model K1) is a manual, foldable, indoor/outdoor mobility device. Constructed from SPCC steel pipe with standard sling-type upholstery. Operated by the user or a caregiver to provide mobility for individuals restricted to a sitting position. Device features four wheels (front casters and rear wheels), armrests, backrest, and footplates. The device is manually propelled. It is designed to be disassembled for transport. Clinical benefit is provided by enabling independent or assisted mobility for patients with physical limitations.

Clinical Evidence

Bench testing only. No clinical data provided. Performance validated against ISO 7176 series standards, including static stability (ISO 7176-1), brake effectiveness (ISO 7176-3), dimensions/maneuvering (ISO 7176-5), test dummies (ISO 7176-11), friction (ISO 7176-13), documentation (ISO 7176-15), and ignition resistance (ISO 7176-16).

Technological Characteristics

Manual mechanical wheelchair. Materials: SPCC steel pipe frame, flame-retardant upholstery. Dimensions: 8" front casters, 24" rear wheels. Foldable frame. Compliant with ISO 7176 series standards for stability, braking, and safety. No electronic components or software.

Indications for Use

Indicated for elderly and disabled persons restricted to a sitting position requiring mobility assistance.

Regulatory Classification

Identification

A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 5, 2023 Zhenjiang Assure Medical Equipment Co.,Ltd. Eric Shi QA & QC Manager No.297, Chuqiao road, Zhenjiang city, Jiangsu province China Re: K231320 Trade/Device Name: Zhenjiang Assure Mechanical Wheelchair, Model:K1 Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I, reserved Product Code: IOR Dated: May 5, 2023 Received: May 8, 2023 Dear Eric Shi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Tushar Bansal -S Tushar Bansal, PhD Acting Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K231320 #### Device Name Zhenjiang Assure Mechanical Wheelchair,model:K1 Indications for Use (Describe) The device is intended for medical purpose to provide mobility to persons restricted to a sitting position. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510k Summary As per 21 CFR 807.92 ## 1.Submitter Information Company Name:Zhenjiang Assure Medical Equipment Co.,Ltd. Address: No.297, Chuqiao road, Zhenjiang city, Jiangsu province,China. Phone:+86-18621771627 Contact Person (including title):Eric Shi (QA&QC Manager) E-mail:eric-shi@isosh.com Subject Device Information - � Type of 510(k) submission:Special - � Common Name:Mechanical Wheelchair - � Proprietary Name:Zhenjiang Assure Mechanical Wheelchair,model:K1 - Regulation Name:Wheelchair, Mechanical � - � Product Code:IOR - � Regulation Number:890.3850 - � Regulation Class:1 ### 2.Predicate Device Information - � Sponsor:Zhenjiang Assure Medical Equipment Co., Ltd - � Common Name: Mechanical Wheelchair - � Proprietary Name:Zhenjiang Assure Mechanical Wheelchair, model:A227 - � Regulation Name:Wheelchair, Mechanical - 510(k) number:K112816 � - � Product Code:IOR - � Regulation Number:890.3850 - � Regulation Class:1 ### 3.Device Description The Zhenjiang Assure Mechanical Wheelchair,model:K1 is indoor/outdoor wheelchair that has a base with four-wheeled with a seat.The device can be disassembled for transport and it is foldable easily. The device can be disassembled for transport and it is foldable easily. The device is {4}------------------------------------------------ consistent with the ISO 7176 series standards and uses a standard sling type back and seat, the upholstery fabric meets the flame retardant test. | Elements of<br>Comparison | Subject Device: K231320 | Predicate Device: K112816 | Comment | |-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | Proprietary Name | Zhenjiang Assure Mechanical<br>Wheelchair,model:K1 | Zhenjiang Assure Mechanical<br>Wheelchair, model:A227 | N/A | | General Comparison | | | | | 510(k) Number | K231320 | K112816 | N/A | | Regulation Name | Wheelchair, Mechanical | Wheelchair, Mechanical | Same | | Regulation Number | 890.3850 | 890.3850 | Same | | Product Code | IOR | IOR | Same | | Indications for Use | The device is intended for<br>medical purpose to provide<br>mobility to persons restricted<br>to a sitting position. | The device is intended for medical<br>purpose to provide mobility to<br>persons restricted to a sitting<br>position. | Same | | Model | K1 | A227 | | | Over-The-Counter<br>Use | Yes | Yes | | | Where used | Indoor/outdoor | Indoor/outdoor | Same | | Target population | The elderly and disabled | The elderly and disabled | Same | | Frame Material | high-quality SPCC steel pipe | high-quality SPCC steel pipe | Same | | Framework | foldable | foldable | Same | | Casters | 8" | 5" ~ 8" | | | Rear wheel | 24" | 12 1/2" ~ 22" | Note 1 | | Seat length | 16"~18" | 10~17" | | | Seat height | 19.5" | 21" | | | Elements of<br>Comparison | Subject Device | Predicate Device | Comment | | Seat width | 16"~20" | 10"~18" | | | Max loading | 136 kg/300 lbs | 100 kg/220 lbs | | | Materials | high-quality SPCC steel pipe | high-quality SPCC steel pipe | Same | | Components of<br>Mechanical<br>Wheelchair | Handle | Handle | Note 2 | | | Footplate | Footplate | | | | Backrest | Backrest | | | | Armrest | Armrest | | | | Front caster | Front caster | | | | Front fork | Front fork | | | | Rear wheel | Rear wheel | | | Handle size | 103*34mm | 103*34mm | Same | | Footplate size | 197*161mm | 197*161mm | Same | | Backrest size | 16"~20" | 16"~20" | Same | | Armrest size | 10.12" x 2.2" x 1.42" | 10.12" x 2.2" x 1.42" | Same | | Front caster size | 8"*1.25" | 8"*1" | Same | | Rear wheel size | 24"*1" | 24"*1" | Same | | Handle size | 103*34mm | 103*34mm | Same | | Performance | compatible as requirement of<br>ISO7176-1:1999,ISO7176-3:<br>2003,ISO7176-5:2008,ISO7176<br>6-11:1992,ISO7176-13:1989,I<br>SO7176-15:1996,ISO7176-16:<br>1997 | compatible as requirement of<br>ISO7176-1:1999,ISO7176-3:<br>2003,ISO7176-5:2008,ISO7176-11:<br>1992,ISO7176-13:1989,ISO7176-1<br>5:1996,ISO7176-16:1997 | Same | ## 4.Summary of Comparison and Technological Characteristics {5}------------------------------------------------ Note 1: The K1 is a new specification for predicate device.The parameters of Casters,Rear wheel,Seat length,Seat height,Seat width and Max loading differ from those of predicate devices. {6}------------------------------------------------ Note 2: The K1 is a new specification for predicate device.The K1 is also composed of Handle,Footplate,Backrest,Armrest,Front caster,Front fork and Rear wheel.The design of front fork is differ from those of predicate devices. #### 6.Summary of Verification and Validation The manufacturer has performed non-clinical performance testing based on its risk assessment utilizing Failure Mode Effect Analysis (FMEA). Following Quality System processes,required testing was conducted to validate the cumulative modifications made to the subject devices. Performance test | No. | FDA<br>recognition<br>number | Standards Development Organization (SDO), Designation<br>Number-Year, and Title | |-----|------------------------------|--------------------------------------------------------------------------------------------------------------------| | 1 | 16-195 | ISO 7176-1:2014 Wheelchairs-Part 1: Determination of static stability. | | 2 | 16-192 | ISO 7176-3:2012 Wheelchairs-Part 3: Determination of effectiveness of<br>brakes | | 3 | 16-163 | ISO 7176-5:2008 Wheelchairs-Part 5:Determination of overall<br>dimensions, mass and maneuvering space | | 4 | 16-190 | ISO 7176-11:2012 Wheelchairs-Part 11: Test dummies | | 5 | 16-25 | ISO 7176-13:1989 Wheelchairs-Part 13:Determination of coefficient of<br>friction of test surfaces | | 6 | 16-27 | ISO 7176-15:1996 Wheelchairs-Part 15:Requirements for information<br>disclosure, documentation and labelling. | | 7 | 16-191 | ISO 7176-16:2012 Wheelchairs-Part 16:Resistance to ignition of<br>upholstered parts, Requirements and test methods | #### 7.Conclusion The conclusions drawn from the nonclinical and clinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed predicate device ## 8.Prepared Date 6 May 2023 {7}------------------------------------------------
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%