K Number
K231071

Validate with FDA (Live)

Date Cleared
2024-01-19

(280 days)

Product Code
Regulation Number
880.5440
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Mix2Vial® Transfer Device is intended for transferring drugs contained in two vials.

Device Description

The subject device, Mix2Vial® Transfer Device (M2V), is a single-use, gamma sterilized, nonpyrogenic device intended for the transfer of drugs contained in two vials. The M2V consists of two vial adapter bodies with piercing spike and wings, connected back-to-back via a Luer connection, with an in-line filter.

The female vial adapter (clear) connects to the drug vial and the male vial adapter (blue) connects to the diluent vial. The vacuum present in the lyophilized drug vial draws in the contents of the diluent vial.

The female Luer connection interfaces to a syringe and has an in-line 5um filter that strains the drug or solution when aspirated out of the vial. The male connection has a "Tight Grip" that provides a secure connection to the diluent vial having a 20mm neck diameter.

Puncturing the elastomeric closure or "stopper" of a drug or diluent vial is achieved by means of an integral polycarbonate cannulated spike located in the center of each Vial Adapter (VA) body of the M2V. Each side of the device (diluent vial side and drug vial side) is a polycarbonate molded part containing a Luer port.

After reconstitution, the drug can be administered by disconnecting the diluent vial and connecting a needleless syringe to the female Luer Lock of the female Vial Adapter. The device does not contain any medicinal substances and can be used with drug vials/diluent vials with a neck diameter of 20mm.

AI/ML Overview

I am sorry, but the provided text describes a 510(k) premarket notification for a medical device (Mix2Vial® Transfer Device) and does not contain any information about a study involving an AI/Machine Learning device or its acceptance criteria within the context of AI performance metrics like sensitivity, specificity, or AUC.

The document focuses on demonstrating substantial equivalence to a predicate device, which is a regulatory pathway for non-AI medical devices. The performance data discussed relates to physical and chemical properties of the device (e.g., fragmentation, particulate, Luer stability, sterility, biocompatibility, force measurements), not the performance of an AI algorithm.

Therefore, I cannot fulfill your request for:

  1. A table of acceptance criteria and reported device performance related to AI.
  2. Sample sizes used for a test set (in the context of AI).
  3. Data provenance for an AI test set.
  4. Number of experts establishing ground truth for AI.
  5. Adjudication method for an AI test set.
  6. MRMC comparative effectiveness study results.
  7. Standalone AI performance.
  8. Type of ground truth (for AI).
  9. Training set sample size (for AI).
  10. Ground truth establishment for a training set (for AI).

The document explicitly states: "Clinical trials were not performed for the Mix2Vial® Transfer Device." This further confirms that there are no clinical performance studies, let alone AI performance studies, described in this regulatory submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo includes the Department of Health & Human Services logo on the left. To the right of that is the FDA logo in blue. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

January 19, 2024

% Lauren Tiller Senior Regulatory Affairs Specialist West Pharmaceutical Services, Inc. 530 Hermon O. West Drive Exton, Pennsylvania 19341

Re: K231071

Trade/Device Name: Mix2Vial® Transfer Device Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: LHI Dated: December 21, 2023 Received: December 21, 2023

Dear Lauren Tiller:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510/k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations. please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Porsche Bennett

Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K231071

Device Name Mix2Vial® Transfer Device

Indications for Use (Describe) The Mix2Vial® Transfer Device is intended for transferring drugs contained in two vials.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K231071 - 510(K) SUMMARY

SUBMITTER

Applicant:

West Pharma. Services IL, Ltd. 4 Hasheizaf St. Ra'anana, Israel 4366411 Facility Establishment Registration Number: 3000223297

Manufacturer:

West Pharma. Services IL, Ltd. 4 Hasheizaf St. Ra'anana, Israel 4366411 Facility Establishment Registration Number: 3000223297

Applicant & Correspondent Contact Person:

Lauren Tiller Senior Regulatory Affairs Specialist Phone: 484-753-6859 Fax: 610-717-0668 E-mail: Lauren.Tiller@westpharma.com

Date Prepared: January 19, 2024

DEVICE

Trade Name: Mix2Vial® Transfer Device Common/Usual Name: I.V. Fluid Transfer Set Regulation Name: Intravascular Administration Set Product Code: LHI Regulation No.: 880.5440 Class: II Panel Identification: General Hospital Panel Predicate Device: Mix2vial Transfer Device – K031861

{4}------------------------------------------------

DEVICE DESCRIPTION

Device Overview

The subject device, Mix2Vial® Transfer Device (M2V), is a single-use, gamma sterilized, nonpyrogenic device intended for the transfer of drugs contained in two vials. The M2V consists of two vial adapter bodies with piercing spike and wings, connected back-to-back via a Luer connection, with an in-line filter.

The female vial adapter (clear) connects to the drug vial and the male vial adapter (blue) connects to the diluent vial. The vacuum present in the lyophilized drug vial draws in the contents of the diluent vial.

The female Luer connection interfaces to a syringe and has an in-line 5um filter that strains the drug or solution when aspirated out of the vial. The male connection has a "Tight Grip" that provides a secure connection to the diluent vial having a 20mm neck diameter.

Puncturing the elastomeric closure or "stopper" of a drug or diluent vial is achieved by means of an integral polycarbonate cannulated spike located in the center of each Vial Adapter (VA) body of the M2V. Each side of the device (diluent vial side and drug vial side) is a polycarbonate molded part containing a Luer port.

After reconstitution, the drug can be administered by disconnecting the diluent vial and connecting a needleless syringe to the female Luer Lock of the female Vial Adapter. The device does not contain any medicinal substances and can be used with drug vials/diluent vials with a neck diameter of 20mm. The figure below depicts the components of the device.

{5}------------------------------------------------

Image /page/5/Figure/14 description: The image shows a diagram of a vial adapter component with labels. The diagram shows two main parts, a male vial adapter component (VA) labeled as '1' and a female VA labeled as '3'. The in-line filter is labeled as '2', the spike is labeled as '4', and the other parts are labeled as '5' and '6'.

Mix2Vial® Transfer Device

    1. Spike
    1. Luer Connector
  • Skirt Wings 6.

The M2V has a 5-year shelf life.

The M2V device is intended for use by a patient (adolescents and adults) or caregivers for Home Use (HU) and by Healthcare Professionals (HCPs) in a clinical, hospital, or other healthcare environment. The subject device is available by prescription use only and has no known contraindications.

The M2V primary packaging consists of a polyethylene terephthalate glycol (PETG) blister that is sealed with a Tyvek® lid on top of the blister pack.

Principle of Operation

The M2V is operated by a manual process. The device is connected to a diluent vial (supplied by the Drug Manufacturer) having a neck diameter of 20mm via the male Vial Adapter portion of the device. Then, the device is connected to a drug vial (supplied by the Drug Manufacturer) having a neck diameter of 20mm via the female Vial Adapter portion of the device.

The vial adapters contain an integral plastic cannulated spike located in the center of the Vial Adapter components, intended to puncture the vial stopper membrane allowing access to the vial contents. This piercing cannulated spike then facilitates the transfer of the drug between the diluent vial and the drug vial.

{6}------------------------------------------------

The device is swirled to mix the drug per the Drug Manufacturer's package insert.

The male Vial Adapter (with diluent bottle still attached) is unscrewed from the female Vial Adapter and discarded.

An empty syringe is pressurized then connected to the female Luer Lock (port on female Adapter). Once connection is secure, the air from the syringe is injected into the vial.

The device is then flipped, and the plunger of the syringe is pulled downward to withdraw the desired amount of the reconstituted drug. The device is flipped back with the vial side down, and the syringe is removed.

The drug can now be administered through the syringe.

INDICATION FOR USE

The Mix2Vial® Transfer Device is intended for transferring drugs contained in two vials.

TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

The Mix2Vial® Transfer Device is substantially equivalent in its intended use, design/construction, technology/principle of operation, materials, and performance to the predicate device Mix2vial Transfer Device, cleared under K031861.

A summary of equivalence and differences between the subject device and the predicate device are provided in the table below.

Areas forComparisonSubject DeviceMix2Vial® Transfer DevicePredicate Device (K031861)Mix2vial Transfer DeviceComparison
Indications for UseTransferring drugs contained intwo vialsTransferring and mixing ofdrugs contained in two vialsDifference #1: The subjectdevice has had a reductionof "mixing" from theindications for use.
ContraindicationsNone knownNone knownIdentical
Intended UserPopulationIntended for use by patients(adolescents and adults),caregivers, and by HealthcareProfessionals (HCPs)Intended for use by HealthcareProfessionals (HCPs)Difference #2: The subjectdevice is intended to beused by HealthcareProfessionals (HCPs) andby patients / lay users(adolescents and adults)and caregivers for HomeUse.
Areas forComparisonSubject DeviceMix2Vial® Transfer DevicePredicate Device (K031861)Mix2vial Transfer DeviceComparison
Intended UseEnvironmentIntended for Home Use and inhospitals, outpatient nursingunits and other suitable clinicalenvironmentsIntended for use in hospitals,outpatient nursing units andother suitable clinicalenvironmentsDifference #3: The subjectdevice is intended to beused for Home Use and inhospitals, outpatientnursing units and othersuitable clinicalenvironments.
Device Class &ClassificationNameClass II, I.V. Fluid TransferSetClass II, I.V. Fluid TransferSetIdentical
RegulationNumber / Name21CFR 880.5440Intravascular AdministrationSet21CFR 880.5440Intravascular AdministrationSetIdentical
Product CodeLHILHIIdentical
Prescription UseYesYesIdentical
Single UseYesYesIdentical
Shelf life5 years5 yearsIdentical
Design
OperationPrincipleManualManualIdentical
Direction of FlowTwo-wayTwo-wayIdentical
Design/ConstructionFemale 20mm Vial Adapterwith wings and piercing shortspike, containing an in-linefilter intended to be attached toa standard drug vial with aneck diameter of 20mm.Female Luer Lock fitting iscompliant with ISO 80369-7for attachment to a standardaccessory such as a syringe(not supplied).Male 20mm Vial Adapter with"tight grip" wings and piercingshort spike intended to beattached to a standard diluentvial with a neck diameter of20mm.Female 20mm Vial Adapterwith wings and piercing spike,containing an in-line filterintended to be attached to astandard drug vial with a neckdiameter of 20mm. FemaleLuer Lock fitting is compliantwith ISO 594-1 and ISO 594-2for attachment to a standardaccessory such as a syringe(not supplied).Male 20mm Vial Adapter withwings and piercing spikeintended to be attached to astandard diluent vial with aneck diameter of 20mm.Difference #4: The subjectdevice female vial adapteris compliant with ISO80369-7, both vial adaptershave a 1.5mm shorterspike, and the male vialadapter has "tight grip"wings.
Female Luer LockConnectorCompliant withISO 80369-7:2021Compliant withISO 594-1 and ISO 594-2Difference #5: The subjectdevice complies with themost current Luer standardrevision.
Compatible VialSize20mm20mmIdentical
Areas forComparisonSubject DeviceMix2Vial® Transfer DevicePredicate Device (K031861)Mix2vial Transfer DeviceComparison
Female VialAdapter WingDiameter30.4mm to accommodate20mm standard vials30.4mm to accommodate20mm standard vialsIdentical
Male Vial AdapterWing Diameter30.1mm to accommodate20mm standard vials30.4mm to accommodate20mm standard vialsDifference #6: The malevial adapter of the subjectdevice has a Tight Grip(TG) 30.1mm externalbody diameter, a .3mmreduction.
Piercing SpikeSingle lumen11.25mm short spikeSingle lumen12.75mm standard spikeDifference #7: The subjectdevice has a 11.25mmshort spike, a 1.5mmreduction.
Vial Adapter FitSnap fit to vialSnap fit to vialIdentical
MaterialVial Adapters: PolycarbonateIn-Filter: Polyethylene +Hydrophilic MeshVial Adapters: PolycarbonateIn-Filter: Polyethylene +Hydrophilic MeshIdentical
BiocompatibilityISO 10993-1:2018External Communicating,Prolonged Indirect Contactwith Patient Blood Path(24 hours to 30 days)USP Biological ReactivityExternal Communicating,Transient Indirect Contact withPatient Blood Path(≤ 24 hours)Difference #8: The subjectdevice is compliant withthe latest biocompatibilitystandard and has ProlongedIndirect Contact withPatient Blood Path (24hours to 30 days).
Non-pyrogenicYesYesIdentical
Sterilization
SterilitySterileSterileIdentical
SterilizationMethodGammaGamma or ETODifference #9: The subjectdevice is gamma sterilized.
Sterility AssuranceLevelSAL of 10-6SAL of 10-6Identical
Packaging
PackagingSterile Barrier packagematerials: PETG blister withTyvek® sealSterile Barrier packagematerials: PETG blister withTyvek® sealIdentical

Substantial Equivalence Comparison Table

{7}------------------------------------------------

{8}------------------------------------------------

The results of performance, biocompatibility, and sterility testing for the subject device, demonstrate that the differences between the subject device and the predicate device do not raise new or different questions of safety and effectiveness. See details as follows:

  • Difference #1: The "mixing" claim has been removed from the device labeling, including ● the Indications for Use statement as the drug mixing requirements are defined by the Drug

{9}------------------------------------------------

Manufacturer, as indicated in the device Instructions for Use. Therefore, West Pharmaceutical Services, Inc does not maintain verification or validation data within the device Design History File to substantiate the mixing claim.

This change does not raise new or different questions of safety and effectiveness.

  • Difference #2: The subject device is intended to be used by Healthcare Professionals . (HCPs) which is identical to the predicate: however, the subject device is also intended for use by patients / lay users (adolescents and adults) and caregivers for Home Use. This change does not raise new or different questions of safety and effectiveness.
  • Difference #3: The subject device is intended to be used in hospitals, outpatient nursing . units, and other suitable clinical environments, identical to the predicate; however, the subject device is also intended for Home Use. This change does not raise new or different questions of safety and effectiveness.
  • Difference #4: The subject device female vial adapter is compliant with ISO 80369-7, both ● vial adapters have a 1.5mm shorter spike, and the male vial adapter has "tight grin" wings. The predicate device Luer fitting is compliant with ISO 594-1 and ISO 594-2. Both ISO 594-1 and ISO 594-2 were superseded with ISO 80369-7.

This change does not raise new or different questions of safety and effectiveness.

  • Difference #5: The subject device complies with the most current Luer standard revision. . This change does not raise new or different questions of safety and effectiveness, as demonstrated by bench testing.
  • . Difference #6: The subject device male Vial Adapter has "Tight Grip" which is a .3mm diameter reduction from the predicate, while still accommodating a standard 20mm drug vial. This change does not raise new or different questions of safety and effectiveness, as demonstrated by bench testing.
  • Difference #7: The subject device has 11.25mm "short spikes," which is a 1.5mm . reduction from the predicate. This change does not raise new or different questions of safety and effectiveness, as demonstrated by bench testing.
  • . Difference #8: The subject device contact duration has been extended to prolonged (24 hours to 30 days) and complies with the most current biocompatibility standards. This change does not raise new or different questions of safety and effectiveness, as demonstrated by biocompatibility testing.
  • . Difference #9: The subject device is only sterilized by Gamma Irradiation, while the predicate device had the option to be either EtO or Gamma sterilized. This change does not raise new or different questions of safety and effectiveness, as demonstrated by sterilization validation.

For these reasons, the Mix2Vial® Transfer Device, subject of this Traditional 510(k), is substantially equivalent in its intended use, design/construction, technology/principle of operation, materials, and performance to the predicate device, Mix2vial Transfer Device, which is cleared under K031861.

{10}------------------------------------------------

PERFORMANCE DATA

The following non-clinical performance data were provided in support of the substantial equivalence determination.

Performance Testing

Performance testing was conducted to confirm the Mix2Vial® Transfer Device meets all applicable design and performance requirements throughout its defined shelf life, conforms to the applicable external and internal standards, and demonstrates substantial equivalence to the predicate device. The table below provides a list of non-clinical bench performance tests that were completed on the device and provided within this submission.

TestTest Method/ Standard
Fragmentation TestISO 8536-2:2010 section 6.2.2
Particulate TestingUSP <788>
Internal Diameter Upper SkirtISO 8362-6:2010 Section 4.2
Luer Gauging TestISO 594-1:1986 and ISO 594-2:1998
Luer Stability and compliance to ISO 80369-7ISO 80369-7:2021
Luer Stability and compliance to ISO 80369-7ISO 80369-20:2015, Annex B & Annex C forthe leakage reference connector (fluidleakage)
Luer Stability and compliance to ISO 80369-7ISO 80369-20:2015, Annex D & Annex C forthe leakage reference connector (air leakage)
Luer Stability and compliance to ISO 80369-7ISO 80369-20: 2015, Annex E & Annex C forthe stress cracking reference connector
Luer Stability and compliance to ISO 80369-7ISO 80369-20: 2015, Annex F & Annex C forthe axial load reference connector
Luer Stability and compliance to ISO 80369-7ISO 80369-20: 2015, Annex G & Annex C forthe resistance separation from unscrewingreference connector
Luer Stability and compliance to ISO 80369-7ISO 80369-20: 2015, Annex G & Annex C forthe overriding reference connector
Luer Stability and compliance to ISO 80369-7ISO 80369-7:2021 Table B.2 and B.5(compliance to dimensions)

Summary of Performance Testing

{11}------------------------------------------------

TestTest Method/ Standard
Residual VolumeIn-house test method
Device LeakageIn-house test method
Device Total Penetration ForceIn-house test method
Vial Adapter Detachment ForceIn-house test method
Product Retention in BlisterIn-house test method
Vacuum TestIn-house test method
Device Removal Force from BlisterIn-house test method
Tyvek Total Peel Test ForceIn-house test method
Functionality according to IFUIn-house test method
Leakage under Normal UseIn-house test method
Product Skirt Position on VialIn-house test method
Injection ForceIn-house test method
Aspiration ForceIn-house test method
Label LegibilityIn-house test method
Opening Torque TestIn-house test method

Performance testing and risk management review indicate all product design requirements are verified and the residual risk level is acceptable based on the test results. Together, objective evidence satisfies the product requirements for performance, safety, and effectiveness, and the results support a determination of substantial equivalence to the predicate device.

Biocompatibility Testing

The biocompatibility evaluation for the Mix2Vial® Transfer Device was conducted in accordance with, 2020 FDA Guidance: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process. and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.

In accordance with ISO 10993-1:2018, the subject device Mix2Vial® Transfer Device is classified as an externally communicating device, having prolonged indirect contact with the patient blood path (24 hours to 30 days).

{12}------------------------------------------------

Sterilization

The subject device is terminally sterilized using a Gamma irradiation sterilization method, validated in accordance with standard BS EN ISO 11137-1:2015 & A2:2019 Sterilization of health care products – Radiation Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices and BS EN ISO 11137-2:2015 Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose, and ISO 13004 -Sterilization of health care products – Radiation – Substantiation of a selected sterilization dose: Method VDmaxSD. The sterilization method of Gamma irradiation provides a sterility assurance level (SAL) of 10-6.

Bacterial Endotoxin Testing by limulus amebocyte lysate (LAL) was also performed on the same batch of product used for sterility dose verification, which passed with acceptable levels, further ensuring the safety of the device.

CLINICAL DATA

Clinical trials were not performed for the Mix2Vial® Transfer Device.

CONCLUSION

In summary, the subject device, Mix2Vial® Transfer Device is determined to be substantially equivalent to the predicate device (K031861) in technology, principle of operation, materials, and performance in comparison to the predicate device, when used as intended use environment by the intended users.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.