Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs, with Gastric Acid and Fentanyl
K230681 · Central Medicare Sdn Bhd · LZA · Oct 2, 2023 · General Hospital
Device Facts
| Record ID | K230681 |
| Device Name | Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs, with Gastric Acid and Fentanyl |
| Applicant | Central Medicare Sdn Bhd |
| Product Code | LZA · General Hospital |
| Decision Date | Oct 2, 2023 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 880.6250 |
| Device Class | Class 1 |
Intended Use
Blue Non Sterile Powder Free Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs, with Gastric Acid and Fentanyl is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs, with gastric acid and fentanyl citrate in accordance with ASTM D6978-05.
Device Story
Disposable nitrile examination gloves; worn by healthcare personnel to prevent cross-contamination between patient and examiner. Device provides a physical barrier; tested for permeation resistance against various chemotherapy drugs, gastric acid, and fentanyl citrate per ASTM D6978-05. Used in clinical settings; provides protection during handling of hazardous substances. Warning: Not for use with Carmustine or Thiotepa due to insufficient breakthrough times.
Clinical Evidence
Bench testing only. Permeation testing performed per ASTM D6978-05 to determine breakthrough detection times for various chemotherapy drugs, gastric acid, and fentanyl citrate.
Technological Characteristics
Nitrile material; powder-free; non-sterile; disposable. Tested per ASTM D6978-05 for chemical permeation resistance.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand to prevent cross-contamination between patient and examiner. Tested for use with specific chemotherapy drugs, gastric acid, and fentanyl citrate.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
- K232614 — Non-sterile, Powder-Free Nitrile Examination Glove Tested for use with Chemotherapy Drugs and Fentanyl (White, Black, and Violet Blue) · Sri Trang Gloves (Thailand) Public Company Limited · May 13, 2024
- K202536 — Non-sterile, Powder Free Nitrile Examination Aloe Vera Gloves, Low Dermatitis Potential, Tested for use with Chemotherapy Drugs, And Non-Sterile, Powder Free Nitrile Examination Gloves, Tested for use with Chemotherapy Drugs and Fentanyl Citrate · Yty Industry (Manjung) Sdn Bhd · Aug 11, 2021
- K220228 — Powder Free Nitrile Examination Gloves, Blue and Violet Blue, Test For Use With Chemotherapy Drugs · Shandong Blue Sail Health Technology Co., Ltd. · Mar 18, 2022
- K213325 — Disposable Medical Nitrile Examination Gloves- Tested for Use With Chemotherapy Drugs · Anhui Xinyisheng Medical Technology Co.,Ltd · Jan 28, 2022
- K202003 — Nitrile Powder Free Examination Glove, Blue Tested for Use with 32 Chemotherapy Drugs and Fentanyl Permeation Resistance Claim · Gmp Medicare Sdn. Bhd. · Feb 24, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 2, 2023
Central Medicare Sdn. Bhd. Chua Kah Ying Product Executive PT 2609-2620, Batu 8, Jalan Changkat Jong Teluk Intan, Perak 36000 Malaysia
Re: K230681
Trade/Device Name: Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs, with Gastric Acid and Fentanyl Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, ODO, OPJ Dated: August 16, 2023 Received: August 21, 2023
Dear Chua Kah Ying:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE bv email (DICE@tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
# Bifeng Qian -S
Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230681
#### Device Name
Blue Non Sterile Powder Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs, with Gastric Acid and Fentanyl
#### Indications for Use (Describe)
Blue Non Sterile Powder Free Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs, with Gastric Acid and Fentanyl is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs, with gastric acid and fentanyl citrate in accordance with ASTM D6978-05.
Tested chemotherapy drugs and breakthrough detection time (minutes) are as follows:
| Chemotherapy Drug | Concentration | Breakthrough Detection Time in Minutes |
|----------------------------|---------------|----------------------------------------|
| Arsenic Trioxide | 1.0 mg/ml | > 240 minutes |
| Azacitidine (Vidaza) | 25.0 mg/ml | > 240 minutes |
| Bendamustine HCl | 5.0 mg/ml | > 240 minutes |
| Bleomycin Sulfate | 15.0mg/ml | > 240 minutes |
| Busulfan | 6.0 mg/ml | > 240 minutes |
| Carboplatin | 10.0 mg/ml | > 240 minutes |
| Carmustine (BCNU) | 3.3 mg/ml | 55.1 minutes |
| Carfilzomib | 2.0 mg/ml | > 240 minutes |
| Cetuximab | 2.0 mg/ml | > 240 minutes |
| Chloroquine | 50.0 mg/ml | > 240 minutes |
| Cisplatin | 1.0 mg/ml | > 240 minutes |
| Cladribine | 1.0 mg/ml | > 240 minutes |
| Cyclophosphamide (Cytoxan) | 20.0 mg/ml | > 240 minutes |
| Cyclosporin A | 100.0 mg/ml | > 240 minutes |
| Cytarabine | 100.0 mg/ml | > 240 minutes |
| Cytovene (Ganciclovir) | 10.0 mg/ml | > 240 minutes |
| Dacarbazine | 10.0 mg/ml | > 240 minutes |
| Daunorubicin | 5.0 mg/ml | > 240 minutes |
| Decitibine | 5.0 mg/ml | > 240 minutes |
| Docetaxel | 10.0 mg/ml | > 240 minutes |
| Doxorubicin Hydrochloride | 2.0 mg/ml | > 240 minutes |
| Epirubicin (Ellence) | 2.0 mg/ml | > 240 minutes |
| Etoposide (Toposar) | 20.0 mg/ml | > 240 minutes |
| Fludarabine | 25.0 mg/ml | > 240 minutes |
| Fluorouracil | 50.0 mg/ml | > 240 minutes |
| Fulvestrant | 50.0 mg/ml | > 240 minutes |
| Gemcitabine (Gemzar) | 38.0 mg/ml | > 240 minutes |
| Idarubicin | 1.0 mg/ml | > 240 minutes |
| Ifosfamide | 50.0 mg/ml | > 240 minutes |
| Irinotecan | 20.0 mg/ml | > 240 minutes |
| Mechlorethamine HCl | 1.0 mg/ml | > 240 minutes |
| Melphalan | 5.0 mg/ml | > 240 minutes |
| Mesna | 50.0 mg/ml | > 240 minutes |
| Methotrexate | 25.0 mg/ml | > 240 minutes |
| Mitomycin C | 0.5 mg/ml | > 240 minutes |
| Mitoxantrone | 2.0 mg/ml | > 240 minutes |
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
EF
{4}------------------------------------------------
| Oxaliplatin | 2.0 mg/ml | > 240 minutes |
|-----------------------------|------------|---------------|
| Paclitaxel | 6.0 mg/ml | > 240 minutes |
| Paraplatin | 10.0 mg/ml | > 240 minutes |
| Pemetrexed | 25.0 mg/ml | > 240 minutes |
| Pertuzumab | 30.0 mg/ml | > 240 minutes |
| Propofol | 10.0 mg/ml | > 240 minutes |
| Raltitrexed | 0.5 mg/ml | > 240 minutes |
| Retrovir | 10.0 mg/ml | > 240 minutes |
| Rituximab | 10.0 mg/ml | > 240 minutes |
| Temsirolimus | 25.0 mg/ml | > 240 minutes |
| Thiotepa | 10.0 mg/ml | 98.7 minutes |
| Topotecan HCl | 1.0 mg/ml | > 240 minutes |
| Trastuzumab | 21.0 mg/ml | > 240 minutes |
| Triclosan | 2.0 mg/ml | > 240 minutes |
| Trisenox (Arsenic Trioxide) | 1.0 mg/ml | > 240 minutes |
| Velcade (Bortezomib) | 1.0 mg/ml | > 240 minutes |
| Vinblastine | 1.0 mg/ml | > 240 minutes |
| Vincristine Sulfate | 1.0 mg/ml | > 240 minutes |
| Vinorelbine | 10.0 mg/ml | > 240 minutes |
| Zoledronic Acid | 0.8 mg/ml | > 240 minutes |
Under the testing conditions of ASTM D6978-05, Fentanyl Citrate Injection (100mcg/2mL) was found to have no breakthrough detected up to 240 minutes.
Under the testing conditions of ASTM D6978-05, Gastric Acid was found to have no breakthrough detected up to 240 minutes.
CAUTION: Testing showed breakthrough time of 55.1 minutes with Carmustine and 98.7 minutes with Thiotepa.
WARNING: Do not use with Carmustine and Thiotepa.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."