ReliOn Premier BLU Blood Glucose Monitoring System
K230625 · I-Sens, Inc. · NBW · Aug 10, 2023 · Clinical Chemistry
Device Facts
| Record ID | K230625 |
| Device Name | ReliOn Premier BLU Blood Glucose Monitoring System |
| Applicant | I-Sens, Inc. |
| Product Code | NBW · Clinical Chemistry |
| Decision Date | Aug 10, 2023 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 862.1345 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
The ReliOn Premier BLU Blood Glucose Monitoring System is intended for single patient use only.
Device Story
ReliOn Premier BLU is a blood glucose monitoring system consisting of a meter and test strips. It measures glucose in fresh capillary whole blood from fingertips. The device features Bluetooth Low Energy for wireless smartphone communication. A new target range indicator function allows users to set a target range with a healthcare professional; the meter displays results on a colored backlight (red, green, blue) to indicate if the result is below, within, or above the target. The device is intended for single-patient home use. Healthcare providers use the output to assist patients in monitoring diabetes control. The system is powered by two 3.0V lithium batteries.
Clinical Evidence
No clinical data. Performance was established through non-clinical verification and validation testing, including software function tests, power consumption, battery lifetime, memory, EMC, electrical safety, and a human factors study.
Technological Characteristics
Blood glucose meter with Bluetooth Low Energy. Features a colored backlight LCD (red, green, blue) for target range indication. Powered by two 3.0V CR2032 lithium batteries. Includes C-Clips for PCB connection. Measurement range: 20-600 mg/dL. Sample size: 0.5 µL. Test time: 5 seconds. Memory: 1,000 results.
Indications for Use
Indicated for single-patient use for the quantitative measurement of glucose in fresh capillary whole blood from the fingertip, palm, or forearm. Intended for use by people with diabetes at home as an aid to monitor the effectiveness of diabetes control programs.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
- CareSens N Premier BT Blood Glucose Monitoring System (K170614)
Related Devices
- K231679 — CONTOUR® PLUS BLUE Blood Glucose Monitoring System · Ascensia Diabetes Care · Feb 23, 2024
- K223293 — CONTOUR® NEXT GEN Blood Glucose Monitoring System · Ascensia Diabetes Care U.S., Inc. · Nov 18, 2022
- K193407 — Contour® next GEN Blood Glucose Monitoring System · Ascensia Diabetes Care · Dec 23, 2021
- K241787 — CONTOUR® PLUS BLUE Blood Glucose Monitoring System; CONTOUR® NEXT GEN Blood Glucose Monitoring System · Ascensia Diabetes Care U.S., Inc. · Aug 27, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "HUMAN SERVICES-USA" around the edge. To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
August 10, 2023
i-SENS, Inc. H.S. Yoo Regulatory Affairs Specialist /Manager 43, Banpo-Daero 28 Gil Seocho-Gu, Seoul 06646 South Korea
Re: K230625
Trade/Device Name: ReliOn Premier BLU Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW Dated: July 12, 2023 Received: July 14, 2023
Dear H.S. Yoo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Joshua Balsam -S
Joshua M. Balsam, PhD. Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230625
Device Name
ReliOn Premier BLU Blood Glucose Monitoring System
#### Indications for Use (Describe)
The ReliOn Premier BLU Blood Glucose Monitoring System is comprised of the ReliOn Premier BLU Blood Glucose Meter and the ReliOn Premier Blood Glucose Test Strips.
The ReliOn Premier BLU Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips.
The ReliOn Premier BLU Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.
The system is intended to be used by a single person and should not be shared.
The system is not intended for use on neonates and is not for the diagnosis or screening of diabetes.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"> <div style="margin-right:5px;"></div>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"> <div style="margin-right:5px;">☑</div>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for i-sens. The logo is in blue, with a green dot above the "i". The text is in a sans-serif font and is slightly italicized. The logo is simple and modern.
i-SENS, Inc. 43, Banpo-Daero 28-Gil, Seocho-Gu, Seoul, Korea, 06646 Tel.: +82-2-916-6191 | Fax.: +82-2-942-2514 | www.i-sens.com
# k230625 510(k) Summary
Prepared in accordance with the requirements of 21 CFR 807.92 to provide an understanding of the basis for a substantial equivalence determination.
Date prepared: July 28, 2023
## 1. Applicant Information
| Name | : i-SENS, Inc. |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Address | : 43, Banpo-daero 28-gil, Seocho-gu, Seoul, South Korea 06646 |
| Applicant Contact<br>Correspondent | : H.S. Yoo (Regulatory Affairs Specialist / Manager)<br>i-SENS, Inc.,<br>43, Banpo-daero 28-gil, Seocho-gu, Seoul, South Korea 06646 |
| E-mail | : registration@i-sens.com |
# 2. Medical Device Information
| Device Name | : ReliOn Premier BLU Blood Glucose Monitoring System |
|-------------------|------------------------------------------------------|
| Common name | : Blood Glucose Monitoring System |
| Regulation number | : 21 CFR 862.1345 |
| Class | : Class II |
| Product Code | : NBW |
| Submission Type | : Special 510(k) |
# 3. Predicate Device Information
CareSens N Premier BT Blood Glucose Monitoring System (k170614)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for i-sens. The "i" is blue with a green dot above it. The rest of the word, "sens", is in blue. The logo is simple and modern.
# 4. Device Description
The ReliOn Premier BLU Blood Glucose Monitoring System is a blood glucose meter with Bluetooth Low Energy technology that the meter communicates to smartphones wirelessly. The ReliOn Premier BLU BGMS has a target range indicator that upon consulting with a healthcare professional, users have the option to activate the function that displays blood glucose measurement results on a colored backlight display. By default, this indicator function is set to OFF. However, it is recommended that users consult with a healthcare professional to determine the target range for their individual needs.
## 5. Intended Use
The ReliOn Premier BLU Blood Glucose Monitoring System is comprised of the ReliOn Premier BLU Blood Glucose Meter and the ReliOn Premier Blood Glucose Test Strips.
The ReliOn Premier BLU Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips.
The ReliOn Premier BLU Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.
The system is intended to be used by a single person and should not be shared.
The system is not intended for use on neonates and is not for the diagnosis or screening of diabetes.
| Feature | Predicate device<br>CareSens N Premier BT<br>(k170614) | Candidate device<br>ReliOn Premier BLU Blood<br>Glucose Monitoring System |
|------------------------|---------------------------------------------------------|-----------------------------------------------------------------------------------|
| Measurement range | 20-600 mg/dL | Same |
| Sample size | Minimum 0.5 µL | Same |
| Test time | 5 seconds | Same |
| Sample type | Fresh capillary whole blood | Same |
| Calibration | Plasma-equivalent | Same |
| Battery life | 1,000 tests | Same |
| Power | Two 3.0V lithium batteries<br>(disposable, type CR2032) | Same |
| Memory | 1,000 test results | Same |
| Target range indicator | N/A | Red, Green, Blue colored<br>screens for below, within, and<br>above target range. |
### 6. Technological characteristics compared to the Predicate device
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for i-sens. The logo is composed of the letter "i" in blue, with a green dot above it. The word "sens" is also in blue and is connected to the "i". The logo is simple and modern.
# 7. Modification from the predicate device
The modifications made are as follows:
- The inclusion of a target range function with the red, green, and blue backlight screen. a.
- Change in the LCD specifications, which consists of three backlight colors and the addition of b. C-Clips for connecting to the PCB.
- Amendments to the PCB to enable connection of the LCD to other components of the blood C. glucose meter.
- Updates to the labeling to feature information on the target range function and the method for d. setting the target range.
## 8. Performance Testing (Non-Clinical Testing)
The verification and validation activities carried out based on the adoption of the target range function included the following tests: Software function tests, power consumption, battery lifetime, memory, EMC, electrical safety, and human factors study.
The results of these tests indicated that the ReliOn Premier BLU Blood Glucose Monitoring System, with the proposed changes outlined in this document, performed as intended and fulfilled the specifications.
### 9. Conclusion
The verification and validation results demonstrated that the ReliOn Premier BLU Blood Glucose Monitoring System has been shown to be substantially equivalent to the predicate device, CareSens N Premier BT Blood Glucose Monitoring System (K170614).