Inessa System

K230545 · Triple Jump Israel , Ltd. · QFG · Dec 20, 2023 · Clinical Chemistry

Device Facts

Record IDK230545
Device NameInessa System
ApplicantTriple Jump Israel , Ltd.
Product CodeQFG · Clinical Chemistry
Decision DateDec 20, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5730
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The Inessa System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The Inessa System is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices. The Inessa System is intended for single patient, home use and requires a prescription. The Inessa System is indicated for use in individuals 6 years of age and greater.

Device Story

Inessa System is an Alternate Controller Enabled (ACE) infusion pump for subcutaneous insulin delivery. System comprises a handheld controller and a skin-adhered patch pump (reusable motor/electronics unit + sterile disposable cartridge with reservoir, cannula, and needle). Controller communicates with pump via Bluetooth Low Energy (BLE). Integrated Bolus Calculator uses user-inputted blood glucose, carbohydrate intake, and insulin parameters (duration, correction factor, ratio) to calculate correction boluses, meal boluses, and 'Insulin on Board'. Used in home environment by patients; requires prescription. Device receives commands from compatible automated insulin dosing software. Provides basal and bolus delivery; includes occlusion detection and data logging. Benefits patients by automating insulin delivery and facilitating interoperability with other diabetes management devices.

Clinical Evidence

No clinical efficacy trials were conducted. Evidence consists of bench verification and validation, including software testing, delivery volume accuracy, occlusion detection, and insulin stability. Human factors and usability validation testing confirmed that use-related risks are mitigated for both pediatric (6+) and adult populations. On-body adhesive testing confirmed performance throughout the intended wear life.

Technological Characteristics

Wearable patch pump; reusable pump unit with disposable cartridge. Materials biocompatible per ISO 10993-1. Connectivity via BLE. Power: rechargeable Li-ion battery. Software-controlled basal/bolus delivery. Standards: ES 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 11137 (sterilization), ASTM 1980-21 (shelf life).

Indications for Use

Indicated for subcutaneous insulin delivery in patients 6 years and older with diabetes mellitus requiring insulin. Contraindicated for patients unable to perform 4+ blood glucose tests daily, maintain HCP contact, follow instructions, or recognize pump alarms/signals. Not for use at low atmospheric pressures (<700hPA) or extreme temperatures.

Regulatory Classification

Identification

An alternate controller enabled infusion pump (ACE pump) is a device intended for the infusion of drugs into a patient. The ACE pump may include basal and bolus drug delivery at set or variable rates. ACE pumps are designed to reliably and securely communicate with external devices, such as automated drug dosing systems, to allow drug delivery commands to be received, executed, and confirmed. ACE pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of drug delivery.

Special Controls

Alternate controller enabled infusion pumps must comply with the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following: (i) Evidence demonstrating that device infusion delivery accuracy conforms to defined user needs and intended uses and is validated to support safe use under actual use conditions. (A) Design input requirements must include delivery accuracy specifications under reasonably foreseeable use conditions, including ambient temperature changes, pressure changes ( *e.g.,* head-height, backpressure, atmospheric), and, as appropriate, different drug fluidic properties.(B) Test results must demonstrate that the device meets the design input requirements for delivery accuracy under use conditions for the programmable range of delivery rates and volumes. Testing shall be conducted with a statistically valid number of devices to account for variation between devices. (ii) Validation testing results demonstrating the ability of the pump to detect relevant hazards associated with drug delivery and the route of administration ( *e.g.,* occlusions, air in line, etc.) within a clinically relevant timeframe across the range of programmable drug delivery rates and volumes. Hazard detection must be appropriate for the intended use of the device and testing must validate appropriate performance under the conditions of use for the device.(iii) Validation testing results demonstrating compatibility with drugs that may be used with the pump based on its labeling. Testing must include assessment of drug stability under reasonably foreseeable use conditions that may affect drug stability ( *e.g.,* temperature, light exposure, or other factors as needed).(iv) The device parts that directly or indirectly contact the patient must be demonstrated to be biocompatible. This shall include chemical and particulate characterization on the final, finished, fluid contacting device components demonstrating that risk of harm from device-related residues is reasonably low. (v) Evidence verifying and validating that the device is reliable over the ACE pump use life, as specified in the design file, in terms of all device functions and in terms of pump performance. (vi) The device must be designed and tested for electrical safety, electromagnetic compatibility, and radio frequency wireless safety and availability consistent with patient safety requirements in the intended use environment. (vii) For any device that is capable of delivering more than one drug, the risk of cross-channeling drugs must be adequately mitigated. (viii) For any devices intended for multiple patient use, testing must demonstrate validation of reprocessing procedures and include verification that the device meets all functional and performance requirements after reprocessing. (2) Design verification and validation activities must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following: (i) Risk control measures shall be implemented to address device system hazards and the design decisions related to how the risk control measures impact essential performance shall be documented. (ii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design. (3) The device shall include validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following: (i) Secure authentication (pairing) to external devices. (ii) Secure, accurate, and reliable means of data transmission between the pump and connected devices. (iii) Sharing of necessary state information between the pump and any digitally connected alternate controllers ( *e.g.,* battery level, reservoir level, pump status, error conditions).(iv) Ensuring that the pump continues to operate safely when data is received in a manner outside the bounds of the parameters specified. (v) A detailed process and procedure for sharing the pump interface specification with digitally connected devices and for validating the correct implementation of that protocol. (4) The device must include appropriate measures to ensure that safe therapy is maintained when communications with digitally connected alternate controller devices is interrupted, lost, or re-established after an interruption ( *e.g.,* reverting to a pre-programmed, safe drug delivery rate). Validation testing results must demonstrate that critical events that occur during a loss of communications (*e.g.,* commands, device malfunctions, occlusions, etc.) are handled appropriately during and after the interruption.(5) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the system must, at a minimum, include: (i) A record of all drug delivery (ii) Commands issued to the pump and pump confirmations (iii) Device malfunctions (iv) Alarms and alerts and associated acknowledgements (v) Connectivity events ( *e.g.,* establishment or loss of communications)(6) Design verification and validation must include results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use. (7) Device labeling must include the following: (i) A prominent statement identifying the drugs that are compatible with the device, including the identity and concentration of those drugs as appropriate. (ii) A description of the minimum and maximum basal rates, minimum and maximum bolus volumes, and the increment size for basal and bolus delivery, or other similarly applicable information about drug delivery parameters. (iii) A description of the pump accuracy at minimum, intermediate, and maximum bolus delivery volumes and the method(s) used to establish bolus delivery accuracy. For each bolus volume, pump accuracy shall be described in terms of the number of bolus doses measured to be within a given range as compared to the commanded volume. An acceptable accuracy description (depending on the drug delivered and bolus volume) may be provided as follows for each bolus volume tested, as applicable: Number of bolus doses with volume that is <25 percent, 25 percent to <75 percent, 75 percent to <95 percent, 95 percent to <105 percent, 105 percent to <125 percent, 125 percent to <175 percent, 175 to 250 percent, and >250 percent of the commanded amount. (iv) A description of the pump accuracy at minimum, intermediate, and maximum basal delivery rates and the method(s) used to establish basal delivery accuracy. For each basal rate, pump accuracy shall be described in terms of the amount of drug delivered after the basal delivery was first commanded, without a warmup period, up to various time points. The information provided must include typical pump performance, as well as worst-case pump performance observed during testing in terms of both over-delivery and under-delivery. An acceptable accuracy description (depending on the drug delivered) may be provided as follows, as applicable: The total volume delivered 1 hour, 6 hours, and 12 hours after starting delivery for a typical pump tested, as well as for the pump that delivered the least and the pump that delivered the most at each time point. (v) A description of delivery hazard alarm performance, as applicable. For occlusion alarms, performance shall be reported at minimum, intermediate, and maximum delivery rates and volumes. This description must include the specification for the longest time period that may elapse before an occlusion alarm is triggered under each delivery condition, as well as the typical results observed during performance testing of the pumps. (vi) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device. (vii) For any infusion pumps intended for multiple patient reuse, instructions for safely reprocessing the device between uses.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 20, 2023 Triple Jump Israel Ltd. Liron Hadar Regulatory Affairs Director 5 HaCarmel St., P.O.B. 205 Yokneam Illit, 2069203 Israel Re: K230545 Trade/Device Name: Inessa System Regulation Number: 21 CFR 880.5730 Regulation Name: Alternate Controller Enabled Infusion Pump Regulatory Class: Class II Product Code: QFG, NDC Dated: November 20, 2023 Received: November 20, 2023 Dear Liron Hadar: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Joshua Balsam -S Joshua M. Balsam, Ph.D. Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ 510(k) Number (if known) K230545 Device Name #### Inessa System Indications for Use (Describe) The Inessa System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The Inessa System is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, and confirm commands from these devices. The Inessa System is intended for single patient, home use and requires a prescription. The Indicated for use in individuals 6 years of age and greater. Type of Use (Select one or both, as applicable) 区Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. {3}------------------------------------------------ # 510(k) SUMMARY Triple Jump Israel Ltd.'s Inessa System K230545 #### Submitter Triple Jump Israel Ltd. 5 HaCarmel St., P.O.B. 205 Yokneam Illit 2069203 Phone: +972 73 707 9070 Facsimile: +972 4 822 3787 Contact Person: Dr. Liron Hadar | Date Prepared: | November 15 , 2023 | |-----------------------|--------------------------------------------| | 510(k) Number: | K230545 | | Name of Device: | Inessa System | | Common or Usual Name: | Insulin Pump | | Classification Name: | Alternate Controller Enabled Infusion Pump | | Regulatory Class: | 21 CFR 880.5730 - Class II | | Product Code: | QFG | #### Predicate Device Insulet Corporation Omnipod DASH™ Insulin Management System with interoperable technology (K191679) #### Device Description . The Inessa System ("System") is intended for subcutaneous delivery of insulin at set and variable rates, bolus or basal. The Inessa System includes a skin-adhered Patch Pump that is programmed and controlled wirelessly by a handheld Controller. The System main components include: - Patch Pump: a skin adhered, syringe pump type, designed for insulin delivery at set and variable basal and/or bolus doses. The Patch Pump includes two parts: - Pump: a reusable part that includes motor, electronics, drive mechanism, and o rechargeable battery. Two Pumps are provided, one is charged (P2) while the other is in use (P1). - Cartridge: a sterile disposable part that includes insulin Reservoir. o {4}------------------------------------------------ - Controller: The System user interface is a handheld, Alternate Controller Enabled (ACE), providing instructions to the Pump and receiving information from the Pump using wireless Bluetooth Low Energy (BLE) communication. The System includes the following accessories: - Inserter: disposable, for insertion of Soft Cannula and retraction of Insertion needle. - Filling kit: disposable, for filling of the Cartridge. It includes Filling needle, Filling syringe and Vial adaptor - . Charger: reusable, for charging the Pump and the Controller. - Docker: reusable, docking station for pumps for charging and protecting the Pump when not in use. The System also includes a Bolus Calculator, accessible through the System Controller. Based on user inputs of blood glucose (current and targeted), carbohydrate intake (meals), patient's insulin characteristics (i.e., Insulin Duration of Action, Insulin Correction Factor, Insulin-to-Carbs Ratio), this feature calculates suggested and estimated values for: - Correction Bolus (amount of insulin needed to correct elevated blood glucose (BG) ● level): - . Meal Bolus (amount of insulin needed to cover carbohydrates in an upcoming meal); and - "Insulin on Board" or "Bolus on Board" (estimation of how much Active Insulin (Al) . remains in the body from previous boluses). # Intended Use / Indications for Use The Inessa System is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The lnessa System is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, and confirm commands from these devices. The Inessa System is intended for single patient, home use and requires a prescription. The Inessa System is indicated for use in individuals 6 years of age and greater. #### Summary of Technological Characteristics The subject device and predicate device use similar operating principles to achieve the intended therapeutic effect. The subject device and predicate device are both insulin infusion pumps that include a software-controlled, programmable pump capable of both basal and bolus delivery of insulin. Both devices are wearable units, which are secured to the patient's skin with an adhesive patch. Both devices include bolus calculators with identical calculations and equations and similar operational specifications. The differences between subject Inessa System and the predicate Omnipod device do not raise any different questions about safety and effectiveness. Therefore, the Inessa System is substantially equivalent to its predicate. {5}------------------------------------------------ | | Inessa System (Subject Device) | Omnipod DASH TM (K191679) | Comparison | |------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | | Device Description | | | | Regulation and<br>Classification | 21 CFR 880.5730; Product Code QFG;<br>Class II | 21 CFR 880.5730; Product Code QFG; Class II | Identical | | Intended Use /<br>Indications for<br>Use | The Inessa System is intended for the<br>subcutaneous delivery of insulin, at set<br>and variable rates, for the management of<br>diabetes mellitus in persons requiring<br>insulin. The Inessa System is able to<br>reliably and securely communicate with<br>compatible, digitally connected devices,<br>including automated insulin dosing<br>software, to receive, execute, and confirm<br>commands from these devices. The<br>Inessa System is intended for single<br>patient, home use and requires a<br>prescription. The Inessa System is<br>indicated use in individuals 6 years of age<br>and greater. | The Omnipod DASH Insulin Management System<br>(the Pump) with interoperable technology is<br>intended for subcutaneous delivery of insulin at set<br>and variable rates for the management of diabetes<br>mellitus in persons requiring insulin. The Pump is<br>able to reliably and securely communicate with<br>compatible, digitally connected devices, including<br>automated insulin dosing software, to receive,<br>execute and confirm commands from these<br>devices. The Pump is intended for single patient,<br>home use and requires a prescription. The Pump is<br>indicated for use with NovoLog®, Humalog®,<br>Admelog®, or Apidra® U-100 insulin. | Similar | | Prescription<br>Use | Yes | Yes | Identical | | Insulin Type | NovoLog or Humalog U-100 insulin | NovoLog or Humalog U-100 insulin | Identical | | Pump Design | Reusable pump with disposable fillable<br>reservoir, patient-activated cannula<br>insertion system, and wireless controller | Single-use, on-body linear piston pump with<br>integrated reservoir, patient-activated cannula<br>insertion system, and wireless controller | Similar | | Principles of<br>Operation | Delivery of insulin (Bolus and Basal)<br>programmed by the patient based on<br>health care provider recommendations | Delivery of insulin (Bolus and Basal)<br>programmed by the patient based on health care<br>provider recommendations | Identical | | Accessories | Filling kit (sterile)<br>• Inserter/needle retractor (pre-<br>attached, sterile, single-use)<br>• Docker<br>• Charger | Filling kit (sterile)<br>• Inserter/needle retractor (built-in, sterile,<br>single-use)<br>• Controller Charger | Similar | | | Technological Characteristics | | | | Insulin<br>Cartridge<br>Volume | 60-210 U | 85-200 U | Similar | | Flow Rates and<br>Profiles | • Basal: 0.05 – 30U/h in 0.05U<br>increments<br>• Bolus: 0.05 - 30U in 0.05U<br>increments<br>• Extended Bolus: 30 minutes to 8<br>hours in 30-minute increments<br>• 8 basal programs<br>• 12 basal segments per program | • Basal: 0.05 – 30U/h in 0.05U increments<br>• Bolus: 0.05 - 30U in 0.05U increments<br>• Extended Bolus: 30 minutes to 8 hours in<br>30-minute increments<br>• 7 basal programs<br>• 24 basal segments per program | Similar | | Insulin<br>Delivery<br>Accuracy | Bolus:<br>• ± 5% for amounts ≥ 0.2U<br>• ± 8% for amounts < 0.2U<br>Basal*:<br>• ± 5% at rates ≥ 1.0 U/h | Bolus:<br>• ± 5% for amounts ≥ 1.0U<br>• ± 0.05 units for amounts < 1.0U<br>Basal:<br>• ± 5% at all rates | Similar | | | Inessa System (Subject Device) | Omnipod DASH™ (K191679) | Comparison | | | * Basal delivery accuracy shall be<br>assessed using the basal flow rate<br>averaged over 4h observation windows. | | | | Occlusion<br>Detection | Yes | Yes | Similar | | Cartridge and<br>Cannula<br>Lifespan | 72 hours | 72 hours | Identical | | Temporary<br>Basal Option | Yes | Yes | Identical | | Bolus<br>Calculator | Inputs:<br>- Current and Target Blood<br>Glucose (BG)<br>- Insulin Correction Factor (ICF)<br>- Carb Intake<br>- Insulin-to-Carbs Ratio (ICR)<br>- Duration of Insulin Action (DIA) | Inputs:<br>- Current and Target Blood Glucose (BG)<br>- Insulin Correction Factor (ICF)<br>- Carb Intake<br>- Insulin-to-Carbs Ratio (ICR)<br>- Duration of Insulin Action (DIA) | Identical | | | Calculations:<br>- Prelim Correction Bolus<br>- Prelim Meal Bolus<br>- Final Correction Bolus<br>- Final Meal Bolus<br>- Total Bolus | Calculations:<br>- Prelim Correction Bolus<br>- Prelim Meal Bolus<br>- Final Correction Bolus<br>- Final Meal Bolus<br>- Total Bolus | Identical | | | Optional Features:<br>- Reverse Correction<br>- Extension of Meal and Correction<br>Boluses<br>- Bolus Limiting (Max Bolus) | Optional Features:<br>- Reverse Correction<br>- Extension of Meal and Correction Boluses<br>- Bolus Limiting (Max Bolus) | Identical | | | Specifications:<br>- Max BG Input: 600 mg/dL<br>- Min BG Input : 20 mg/dL<br>- Duration of unavailable Active<br>Insulin after RESET or unknown<br>bolus: based on DIA (2-8 hrs) | Specifications:<br>- Max BG Input: 600 mg/dL<br>- Min BG Input : 50-70 mg/dL<br>- Duration of unavailable Active Insulin after<br>RESET or unknown bolus: 8.5 hrs | Similar | | Power Source | Rechargeable Lithium-Ion | Rechargeable Lithium-Ion for controller; silver oxide<br>for Pod | Similar | | Controller<br>Battery Life | ~ 2 days | ~ 2 days | Identical | | User Interface<br>(UI) | Handheld touchscreen password-locked<br>mobile device | Handheld touchscreen password locked mobile<br>device | Identical | | UI Visual<br>Indicators | - Time and date<br>- Battery level - Controller<br>- Battery level - Pump<br>- Insulin level<br>- Active basal rate<br>- Bolus in progress<br>- Delivery status<br>- Alert and alarm<br>- Communication status | - Time and date<br>- Battery (PDM)<br>- Insulin level<br>- Active basal rate<br>- Bolus in progress<br>- Delivery status<br>- Alert and alarm<br>- Communication status | Similar | | Alarms | Audible, visual, & vibration | Audible, visual, & vibration | Similar | {6}------------------------------------------------ {7}------------------------------------------------ #### Safety and Performance Data #### Safety Assurance Case: A comprehensive Safety Assurance Case (SAC) was developed for the Inessa System in full conformance to AAMI TIR38:2019 - Medical device safety assurance case guidance. This SAC was performed to address the applied risk management process addressing all reasonably foreseeable hazards / hazardous situations have been effectively mitigated, providing evidence that the mitigation are adequate and that they will remain effective over the product's lifetime. # Sterility and Shelf Life Testing: Sterilization has been fully validated per ISO 11137-1:2015/A2:2019, ISO 11137-2:2013, AAMI TIR33:2005, ISO 11737-1:2018 and ISO 11737-2:2019. The Inessa System's shelf life of 6 months has been fully validated through accelerated aging testing per ASTM 1980-21. #### Biocompatibility Testing: Full biocompatibility testing for the Inessa System was conducted in accordance with ISO 10993-1:2018 and FDA's Biocompatibility Guidance "Use of international standard ISO 10993-1, biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process, Guidance for Industry and Food and Drug Administration Staff, September 2020." # Electrical Safety and EMC: All appropriate testing for electrical safety, electromagnetic compatibility, immunity to radio frequency identification readers (RFID) immunity, and wireless coexistence was performed in accordance to ES 60601-1,IEC 60601-1-2, RTCA DO-160 Edition G:2010, AIM 7351731, and IEEE/ANSI C63.27-2021. #### Pre-Clinical Bench Testing: Complete bench verification and validation testing has been performed to demonstrate that the Inessa System's performance specifications are all adequately met. Specifically, the following tests were performed: - -Software Verification and Validation - -Delivered Volume Accuracy Testing - -Occlusion Testing - Insulin Stability and Bioidentity Testing - - -Lifetime Verification Testing - -Adhesive Peeling Force Testing - Particulate Matter Testing - - -Unintended Insulin Delivery Testing - -Pump Battery Empty Alarm Testing - System Reliability Verification - {8}------------------------------------------------ #### Clinical Testing: On-Body Adhesive Testing and Human Factors and Usability Validation Testing were conducted. Adhesive testing results supported the safety and performance of the device's adhesion to patients throughout its intended use life. Human Factors and Usability Validation Testing demonstrated that all use-related risks are adequately mitigated across both child and adult patient user populations. # Special Controls The Inessa System conforms to all special controls as required under 21 CFR 880.5730. # Conclusions The Inessa System is substantially equivalent to the cleared Omnipod DASH™ predicate device (K191679). The subject device has the same intended use and substantially similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not alter the intended therapeutic use of the device and do not raise different questions of safety and effectiveness. In addition, the minor technological differences between the Inessa System and its predicate device do not raise different questions of safety or effectiveness. Performance data demonstrate that the subject device is as safe and effective as the predicate device. Thus, the Inessa System is substantially equivalent.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%