(93 days)
The Linear Health Science™ Orchid Safety Release Valve™ is a tension-activated accessory for single patient use and placed between the existing IV administration set and IV extension. The Orchid SRV™ is intended for use with electronic IV pumps in peripheral IV catheter applications where tension may act on the IV tubing. The Orchid SRVTM is designed to allow flow to an IV catheter. When excessive tension acts on the line, the Orchid SRV™ separates and closes the flow path in both directions. The Orchid SRV™ can be used during intermittent infusion and continuous infusion.
The Orchid SRV™ is intended to aid in reduction of peripheral IV mechanical complications requiring IV replacement.
The Orchid SRVTM is for use with patients two (2) weeks of age and older.
The Orchid Safety Release Valve™ or Orchid SRV™ connects via standard luer-locking connection, allowing flow during IV therapy. The Orchid SRV™ is designed to allow the device to separate into two halves when longitudinal tension exceeds the SRV tension window, automatically closing the flow path to both IV extension set and IV administration set. Following separation, a component of the Orchid SRV™ is left attached to each side of the infusion system to protect the intraluminal pathway. Upon separation, replacement of the SRV™ is necessary. Follow institutional policy to replace the SRV™, or at least every seven (7) days.
This document does not contain acceptance criteria or study results that demonstrate the device meets those criteria.
The information provided confirms that the Orchid Safety Release Valve™ (K230266) is substantially equivalent to its predicate device, also named Orchid Safety Release Valve™ (K212064). The only difference between the proposed device and the predicate is an update to the Indications for Use to include pediatric patients two (2) weeks of age and older, whereas the predicate was for patients eighteen (18) years of age and older.
The justification for this expanded indication is that "Pediatric extrapolation results supported safety and effectiveness of the Orchid SRV in pediatric patients." Additionally, the submission states that "The device performance of the subject and predicate device are identical and compatible with IV therapy in adult and pediatric patients. The results of pediatric extrapolation supported the similarities in risk profiles across all patient populations. Therefore, the change in indication does not questions of safety or effectiveness."
Here's what can be extracted based on your request, though limited by the document's content:
1. A table of acceptance criteria and the reported device performance:
The document does not explicitly state specific acceptance criteria in a numerical or measurable format for performance, nor does it provide a table reporting on these. It generally asserts that the device is identical in performance to its predicate. The only performance-related characteristic mentioned is:
| Characteristic | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Separation Force | The document states the predicate had a separation force of "1-4.2 lbf" and that the proposed device has "same". | "1-4.2 lbf" |
| Safety and Effectiveness | Demonstrated through pediatric extrapolation, implying comparability to adult use. | Supported by pediatric extrapolation results, with no new or different questions of safety or effectiveness. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
The document does not specify a sample size or data provenance for any test set. It mentions "Pediatric extrapolation results," but no details on how these results were obtained (e.g., sample size, type of study).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
This information is not provided in the document. The concept of "ground truth" and expert involvement in a test set as typically found in AI/imaging studies is not relevant to this device's submission, which relies on demonstrating substantial equivalence to a predicate.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not provided as there is no described test set or adjudication process in the context of this 510(k) submission.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This is not applicable as the device (Orchid Safety Release Valve™) is a physical, tension-activated accessory, not an AI or imaging device involving human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This is not applicable as the device is not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
The concept of "ground truth" in the context of an algorithm's performance is not applicable here. The safety and effectiveness for the expanded pediatric indication were supported by "pediatric extrapolation results," implying an analysis of existing data or principles rather than a new data collection with a "ground truth" as typically defined for AI/diagnostic studies.
8. The sample size for the training set:
There is no mention of a training set in this document as the device is not an AI algorithm.
9. How the ground truth for the training set was established:
This is not applicable as there is no training set mentioned.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 4, 2023
Linear Health Sciences, LLC % Jessica Czamanski Managing Director MedTactics, LLC 19924 NE 19th PL Miami, Florida 33179
Re: K230266
Trade/Device Name: Orchid Safety Release Valve™ Regulation Number: 21 CFR 880.5220 Regulation Name: Intravenous Catheter Force-Activated Separation Device Regulatory Class: Class II Product Code: QOI Dated: January 20, 2023 Received: January 31, 2023
Dear Jessica Czamanski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K230266
Device Name Orchid Safety Release Valve™
Indications for Use (Describe)
The Linear Health Science™ Orchid Safety Release Valve™ is a tension-activated accessory for single patient use and placed between the existing IV administration set and IV extension. The Orchid SRV™ is intended for use with electronic IV pumps in peripheral IV catheter applications where tension may act on the IV tubing. The Orchid SRVTM is designed to allow flow to an IV catheter. When excessive tension acts on the line, the Orchid SRV™ separates and closes the flow path in both directions. The Orchid SRV™ can be used during intermittent infusion and continuous infusion.
The Orchid SRV™ is intended to aid in reduction of peripheral IV mechanical complications requiring IV replacement.
The Orchid SRVTM is for use with patients two (2) weeks of age and older.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| × Prescription Use (Part 21 CFR 801 Subpart D) | □ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Linear Health Sciences. The word "LINEAR" is in large, blue, sans-serif font, with a small "TM" symbol in the upper right corner. Below the word "LINEAR" is the phrase "HEALTH SCIENCES" in a smaller, gray, sans-serif font. A thin gray line extends from the left of the word "LINEAR" to the right.
K230266 - 510(k) SUMMARY
This 510(k) Summary is provided per the requirements of section 21 CFR 807.92.
l. Submitter
| Submitter's Name: | Linear Health Sciences, LLC |
|---|---|
| Contact Person: | Mr. Daniel ClarkCEO |
| Address: | 5333 Wisteria DriveOklahoma City, OK 73142 |
| Telephone: | (310) 721-6222 |
| Email: | dan.clark@linearsciences.com |
| Date Preparation: | May 4, 2023 |
II. Application Correspondent
| Contact's Name: | MedTactics, LLC |
|---|---|
| Contact Person: | Jessica CzamanskiManaging Director |
| Address: | 19924 NE 19TH PLMiami, FL 33179 |
| Telephone: | (754) 422-9101 |
| Email: | iczamanski@medtactics.us |
III. Device
| Trade Name: | Orchid Safety Release Valve™ |
|---|---|
| Common Name: | Quick Disconnect Accessory |
| Classification Name: | Intravenous Catheter Force-Activated Separation Device |
| Product Classification: | Class II |
| Regulation Number: | 21 CFR § 880.5220 |
| Product Code: | QOI |
IV. Predicate Device
| Manufacturer: | Linear Health Sciences, LLC |
|---|---|
| Device Name: | Orchid Safety Release Valve™ |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Linear Health Sciences. The word "LINEAR" is in blue, with a vertical line in the middle of the first letter. The words "HEALTH SCIENCES" are in gray below the word "LINEAR". There is a horizontal line on either side of the words "HEALTH SCIENCES".
| 510(k) Number: | K212064 |
|---|---|
| Product Classification: | Class II |
| Regulation Number: | 21 CFR § 880.5220 |
V. Device Description
The Orchid Safety Release Valve™ or Orchid SRV™ connects via standard luer-locking connection, allowing flow during IV therapy. The Orchid SRV™ is designed to allow the device to separate into two halves when longitudinal tension exceeds the SRV tension window, automatically closing the flow path to both IV extension set and IV administration set. Following separation, a component of the Orchid SRV™ is left attached to each side of the infusion system to protect the intraluminal pathway. Upon separation, replacement of the SRV™ is necessary. Follow institutional policy to replace the SRV™, or at least every seven (7) days.
VI. Intended Use
The Linear Health Sciences™ Orchid Safety Release Valve™ is a tension-activated accessory. provided sterile and for single use, in line with an IV administration set and peripheral IV extension set on a patient. The Orchid SRV™ provides a quick separation feature that allows the device to quickly separate into two halves upon tension, closing the flow path to prevent leakage. The device is intended to reduce the risk of IV catheter failure, requiring IV catheter replacement.
VII. Indications for Use
The Linear Health Sciences™ Orchid Safety Release Valve™ is a tension-activated accessory for single patient use and placed between the existing IV administration set and IV extension set connection. The Orchid SRV™ is intended for use with electronic IV pumps in peripheral IV catheter applications where tension may act on the IV tubing. The Orchid SRV™ is designed to allow flow to an IV catheter. When excessive tension acts on the line, the Orchid SRV™ separates and closes the flow path in both directions. The Orchid SRV™ can be used during intermittent infusion and continuous infusion.
The Orchid SRV™ is intended to aid in reduction of peripheral IV mechanical complications requiring IV replacement.
The Orchid SRV™ is for use with patients two (2) weeks of age and older.
VIII.Comparison of Technological Characteristics with the Predicate Devices
There are no differences in technological characteristics of the device. Only differences are present in the labeling of the Orchid Safety Release Valve, specifically the instructions for use and quick reference auide.
The following table (Table 5-1) provides an overview of general technological characteristics in comparison to the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Linear Health Sciences. Below the logo is the text 'HEALTH SCIENCES' in a smaller font size. The image also contains the text 'Table 5-1: General Technological Characteristics Comparison'.
| Table 5-1: General Technological Characteristics Comparison | |||
|---|---|---|---|
| Product Features | ProposedLinear Health Sciences'Orchid Safety ReleaseValve™ | PredicateLinear Health Sciences'Orchid Safety ReleaseValve™(K212064) | SubstantialEquivalenceDetermination |
| Classification | Class II | -same- | |
| Product Code | QOI | -same- | |
| RegulationNumber | §880.5220 | -same- | |
| DeviceClassificationName | Intravenous CatheterForce-ActivatedSeparation Device | -same- | |
| Intended Use | The Linear HealthSciences™ Orchid SafetyRelease Valve™ is atension-activatedaccessory, providedsterile and for single use,in line with an IVadministration set andperipheral IV extensionset on a patient. TheOrchid SRV™ provides aquick separation featurethat allows the device toquickly separate into twohalves upon tension,closing the flow path toprevent leakage. Thedevice is intended toreduce the risk of IVcatheter failure, requiringIV catheter replacement. | -same- | |
| Indications for Use | The Linear HealthSciences™ Orchid SafetyRelease Valve™ is atension-activatedaccessory for singlepatient use and placedbetween the existing IVadministration set and IVextension set connection.The Orchid SRV™ isintended for use withelectronic IV pumps inperipheral IV catheterapplications wheretension may act on the IVtubing. The Orchid SRV™is designed to allow flowto an IV catheter. Whenexcessive tension acts onthe line, the Orchid | The Linear HealthSciences™ Orchid SafetyRelease Valve™ is atension-activatedaccessory for singlepatient use and placedbetween the existing IVadministration set and IVextension set connection.The Orchid SRV™ isintended for use withelectronic IV pumps inperipheral IV catheterapplications wheretension may act on the IVtubing. The Orchid SRV™is designed to allow flowto an IV catheter. Whenexcessive tension acts onthe line, the Orchid | Indications for use arebeing updated toinclude pediatricpatients 2 weeks old to17 years old. Thedevice performance ofthe subject andpredicate device areidentical andcompatible with IVtherapy in adult andpediatric patients. Theresults of pediatricextrapolationsupported thesimilarities in riskprofiles across allpatient populations.Therefore, the changein indication does not |
| Table 5-1: General Technological Characteristics Comparison | |||
| ProposedLinear Health Sciences'Orchid Safety ReleaseValve™ | PredicateLinear Health Sciences'Orchid Safety ReleaseValve™(K212064) | SubstantialEquivalenceDetermination | |
| Product Features | closes the flow path inboth directions. TheOrchid SRV™ can beused during intermittentinfusion and continuousinfusion.The Orchid SRV™ isintended to aid inreduction of peripheral IVmechanical complicationsrequiring IV replacement.The Orchid SRV™ is foruse with patients two (2)weeks of age and older. | closes the flow path inboth directions. TheOrchid SRV™ can beused during intermittentinfusion and continuousinfusion.The Orchid SRV™ isintended to aid inreduction of peripheral IVmechanical complicationsrequiring IV replacement.The Orchid SRV™ is foruse with patientseighteen (18) years ofage and older. | questions of safety oreffectiveness. |
| Materials | Polycarbonate andsilicone | -same- | |
| Environment ofUse | Hospital | -same- | |
| Provided Sterile | Yes | -same- | |
| Principle ofOperation | The Orchid SafetyRelease Valve™ has luerlock connections that willlock the device in placeduring use. The femaleluer connects to anadministration set whilethe male luer connects toa vascular access devicehub or extension set.Once connected thedevice allows forcontinuous flow. TheOrchid SRV will separateinto the male and femalesubassemblies, upon atension event,automatically closing theflow path, whilemaintaining sterility andpreventing fluid leakageform the device. | -same- | |
| User Profile | Physician or clinicalpersonnel with clearanceto administer IV sets andrelated products | -same- | |
| Separation Force | 1-4.2 lbf | -same- | |
| Product Features | ProposedLinear Health Sciences'Orchid Safety ReleaseValve™ | PredicateLinear Health Sciences'Orchid Safety ReleaseValve™(K212064) | SubstantialEquivalenceDetermination |
| Vascular AccessCatheter Type | Peripheral IV Catheters | -same- | |
| For Use withElectronic Pumps | Yes | -same- | |
| Single Use | Yes | -same- | |
| Continuous andIntermittentInfusion | Yes | -same- | |
| Sterilization | Ethylene Oxide | -same- | |
| SAL | 10-6 | -same- | |
| Shelf-Life | 2 years | -same- |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Linear Health Sciences. The logo is blue and has the words "LINEAR" and "HEALTH SCIENCES" on it. Below the logo is the text "Table 5-1: General Technological Characteristics Comparison".
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Linear Health Sciences. The word "LINEAR" is in large, blue, sans-serif font, with a horizontal line extending from the left side of the "L" and a trademark symbol next to the "R". Below "LINEAR" in a smaller, gray, sans-serif font are the words "HEALTH SCIENCES."
echnological Characteristics Comparison
IX. Performance Data
The following performance data was considered in support of the substantial equivalence determination.
Performance Testing - Bench
There is no change in performance data since there are no changes to the device or its manufacturing processes.
Biocompatibility
There is no change in biocompatibility since there are no changes to the device or its manufacturing processes.
Electrical Safety and Electromagnetic Compatibility (EMC)
There are no electrical or metal components in the proposed Orchid Safety Release Valve™; therefore, the proposed device does not require EMC and Electrical Safety evaluation.
Software Verification and Validation Testing
The Orchid Safety Release Valve™ does not contain software; therefore, the proposed device does not require software verification and validation testing.
Performance Testing - Animal
This submission does not include any animal performance testing. It was determined that no such testing was required to demonstrate substantial equivalence.
Clinical Data
Pediatric extrapolation results supported safety and effectiveness of the Orchid SRV in pediatric patients.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Linear Health Sciences. The word "LINEAR" is in blue, with a thin horizontal line extending from the left side of the "L" and the right side of the "R". The words "HEALTH SCIENCES" are in gray below the word "LINEAR".
X. Conclusion
The Orchid Safety Release Valve™ has the same intended use, environment, operating principle and fundamental technology, manufacturing, and materials as the predicate device. The conducted pediatric extrapolation assessment demonstrates that the differences in the indications for use do not raise any new or different questions of safety or efficacy. The information provided in this submission demonstrates that the device is substantially equivalent to its predicate.
§ 880.5220 Intravenous catheter force-activated separation device.
(a)
Identification. An intravenous catheter force-activated separation device is placed in-line with an intravenous (IV) catheter and an intravascular administration set, including any administration set accessories. It separates into two parts when a specified force is applied. The device is intended to reduce the risk of IV catheter failure(s) requiring IV catheter replacement.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Performance data must be provided to demonstrate clinically acceptable performance for the intended use of the device.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Separation force testing;
(ii) Validation of anti-reconnect features;
(iii) Air and liquid leakage testing, both before and after separation;
(iv) Luer connection testing;
(v) Flow rate testing;
(vi) Particulate testing; and
(vii) Microbial ingress testing.
(3) The device must be demonstrated to be biocompatible.
(4) Performance testing must demonstrate that the device is sterile and non-pyrogenic.
(5) Performance testing must support the shelf life of the device by demonstrating continued sterility and device functionality over the identified shelf life.
(6) Device labeling must include:
(i) Instructions for use; and
(ii) A discussion of catheter dressings intended to be used with the device.