TENS and EMS Stimulation (OTC)
K230174 · Changsha Yuwen Medical Equipment Co., Ltd. · NUH · Mar 20, 2023 · Neurology
Device Facts
| Record ID | K230174 |
| Device Name | TENS and EMS Stimulation (OTC) |
| Applicant | Changsha Yuwen Medical Equipment Co., Ltd. |
| Product Code | NUH · Neurology |
| Decision Date | Mar 20, 2023 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 882.5890 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
TENS: To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities. EMS: To be used for stimulation healthy muscles in order to improve or facilitate muscle performance.
Device Story
TENS and EMS Stimulation (OTC) is a transcutaneous electrical nerve stimulator used for pain relief and muscle stimulation. Device delivers electrical impulses to muscles via electrodes; intended for over-the-counter use by patients in home settings. TENS function targets sore/aching muscles; EMS function targets healthy muscles to improve performance. Device provides non-invasive, drug-free pain management and muscle conditioning. Users operate device to select stimulation parameters; output affects muscle activity and sensory nerve perception to provide relief or facilitate performance.
Clinical Evidence
No clinical data provided; bench testing only.
Indications for Use
Indicated for adults seeking temporary relief of muscle pain in the shoulder, waist, back, neck, arms, legs, abdomen, and bottom caused by exercise or household work (TENS), and for stimulation of healthy muscles to improve or facilitate muscle performance (EMS).
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Related Devices
- K230163 — TENS and EMS Stimulation (OTC) · Changsha Yuwen Medical Equipment Co., Ltd. · Mar 20, 2023
- K212184 — TENS and EMS Stimulation (OTC) · Changsha Yuwen Medical Equipment Co., Ltd. · Apr 1, 2022
- K223825 — Tens & Ems Device · Shenzhen Jian Feng Electronic Technology Co., Ltd. · May 9, 2023
- K143268 — TENS AND POWERED MUSCLE STIMULATOR · Shenzhen As-Tec Technology Co., Ltd. · Jul 21, 2015
- K230164 — TENS and EMS Stimulation (OTC) · Changsha Yuwen Medical Equipment Co., Ltd. · Apr 10, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".
March 20, 2023
Changsha Yuwen Medical Equipment Co., Ltd Chen Dahui CEO Room 2404, Kairui Building, No. 245 Laodong West Road Tianxin District Changsha, Hunan 430103 China
Re: K230174
Trade/Device Name: TENS and EMS Stimulation (OTC) Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NGX Dated: January 20, 2023 Received: January 20, 2023
Dear Chen Dahui:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
## Lauren E. Woodard -S
for Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230174
Device Name TENS and EMS Stimulation (OTC)
Indications for Use (Describe) TENS:
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities.
EMS:
To be used for stimulation healthy muscles in order to improve or facilitate muscle performance.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."