Estacore-Pro Automatic Biopsy Needle, Geotek Semiautomatic Biopsy Needle, Maxicore-M Automatic Biopsy Gun Needle, Maxicore-M Automatic Biopsy Needle and Instrument
K223300 · Geotek Medikal Ltd Sti · KNW · Apr 21, 2023 · Gastroenterology, Urology
Device Facts
| Record ID | K223300 |
| Device Name | Estacore-Pro Automatic Biopsy Needle, Geotek Semiautomatic Biopsy Needle, Maxicore-M Automatic Biopsy Gun Needle, Maxicore-M Automatic Biopsy Needle and Instrument |
| Applicant | Geotek Medikal Ltd Sti |
| Product Code | KNW · Gastroenterology, Urology |
| Decision Date | Apr 21, 2023 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1075 |
| Device Class | Class 2 |
Intended Use
Geotek Disposable Core Biopsy Needles (Three Models: Estacore-Pro Automatic Biopsy Needle, Geotek Semiautomatic Biopsy Needle, and Maxicore-M Automatic Biopsy Needle and Instrument) are intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors. It is not Intended for use in bone.
Device Story
Geotek Disposable Core Biopsy Needles consist of three models: Estacore-Pro Automatic, Geotek Semiautomatic, and Maxicore-M Automatic. Devices are used by clinicians to obtain tissue samples from soft organs and tumors. Operation involves manual or spring-loaded mechanical action to advance a needle into target tissue to excise a core sample. Devices are single-use, disposable instruments. They provide diagnostic tissue specimens for pathological examination, aiding in disease diagnosis and clinical decision-making. No electronic components, software, or power sources are involved.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and design characteristics.
Technological Characteristics
Mechanical biopsy needles; manual or spring-loaded actuation. Single-use, disposable. Materials include medical-grade stainless steel and plastics. No software, electronics, or connectivity. Sterilization via standard methods.
Indications for Use
Indicated for obtaining core biopsies from soft tissues including liver, kidney, prostate, spleen, lymph nodes, and soft tissue tumors. Contraindicated for use in bone.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Related Devices
- K040893 — EASY CORE BIOPSY SYSTEM · Boston Scientific Corp · May 6, 2004
- K133948 — BARD(R) MONOPTY(R) DISPOSABLE CORE BIOPSY INSTRUMENT, BARD(R) MAX-CORE(R) DISPOSABLE CORE INSTRUMENT · Bard Peripheral Vascular, Inc. · Feb 21, 2014
- K210946 — Promisemed Automatic Biopsy Needles, Promisemed Co-Axial Biopsy Devices, Promisemed Semi-automatic Biopsy Needles · Promisemed Hangzhou Meditech Co., Ltd. · Aug 12, 2021
- K051361 — EASY CORE DETACHABLE BIOPSY SYSTEM · Boston Scientific Corp · Jun 16, 2005
- K222865 — M Biopsy /SureCore Automatic Disposable Biopsy Needle, M Biopsy /SureCore Semi-Automatic Disposable Biopsy Needle, M Biopsy /SureAim Coaxial Biopsy Needle · Canyon Medical, Inc. · Apr 3, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 21, 2023
GEOTEK Medikal Ltd sti % Mr. Raymond J Kelly IV Chief Regulatory Officer Licensale Inc 3422 Leonardo Lane New Smyrna Beach, Florida 32168
Re: K223300
Trade/Device Name: Geotek Disposable Core Biopsy Needles (Three Models: Estacore-Pro Automatic Biopsy Needle, Geotek Semiautomatic Biopsy Needle, and Maxicore-M Automatic Biopsy Needle and Instrument)
Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW Dated: October 11, 2022 Received: October 27, 2022
Dear Mr. Kelly IV:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
| Mark<br>Trumbore -S | Digitally signed by<br>Mark Trumbore -S<br>Date: 2023.04.21<br>08:24:24 -04'00' |
|---------------------|---------------------------------------------------------------------------------|
|---------------------|---------------------------------------------------------------------------------|
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223300
#### Device Name
Geotek Disposable Core Biopsy Needles (Three Models: Estacore-Pro Automatic Biopsy Needle, Geotek Semiautomatic Biopsy Needle, and Maxicore-M Automatic Biopsy Needle and Instrument)
### Indications for Use (Describe)
Geotek Disposable Core Biopsy Needles (Three Models: Estacore-Pro Automatic Biopsy Needle, Geotek Semiautomatic Biopsy Needle, and Maxicore-M Automatic Biopsy Needle and Instrument) are intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors. It is not Intended for use in bone.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | □ Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."