K Number
K222152

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-09-23

(65 days)

Product Code
Regulation Number
880.2800
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Use the 3M™ Attest™ Hydrogen Peroxide Indicator Tape 1228 to secure packs and as an external pack process indicator to differentiate unprocessed items processed in the following sterilizers:

STERRAD 100® Sterilization System

STERRAD 100S® Sterilization System

STERRAD NX® Sterilization System (Standard and Advanced cycles)

STERRAD 100NX® Sterilization System (Standard, Flex, Express and Duo cycles)

STERRAD NX® with ALLClear® Technology Sterilization System (Standard and Advanced cycles) STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express and Duo cycles)

V-PRO® 1 Low Temperature Sterilization System (Lumen cycle)

V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles)

V-PRO® maX Low Temperature Sterilization System (Lumen, and Flexible cycles)

V-PRO® 60 Low Temperature Sterilization System (Lumen, Non Lumen and Flexible cycles)

V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Flexible, and Fast Non Lumen cycles)

V-PRO® s2 Low Temperature Sterilization System (Lumen, Flexible, and Fast cycles) STERIZONE® VP4 Sterilizer (Cycle 1)

The tape is suitable for use on non-woven disposable wraps and peel pouches. The chemical indicator stripes turn from blue toward pink after exposure to vaporized hydrogen peroxide.

Device Description

The 3M™ Attest™ Hydrogen Peroxide Indicator Tape 1228 consists of a non-cellulosic plastic backing material with a pressure-sensitive adhesive on one side and chemical indicator stripes on the other side. The chemical indicator stripes turn from blue toward pink after exposure to vaporized hydrogen peroxide. 3MTM Attest™ Hydrogen Peroxide Indicator Tape 1228 is a Type 1 (Category e1) Process Indicator as categorized by ISO 11140-1:2014.

AI/ML Overview

The provided text describes the 3M™ Attest™ Hydrogen Peroxide Indicator Tape 1228 and its substantial equivalence to a predicate device. It includes nonclinical testing to support its performance. Here's an analysis of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

Test NamePurposeAcceptance CriteriaReported Device Performance
Absence of hydrogen peroxide (ISO 11140-1:2014 Type 1)Verify device requires the presence of vaporized hydrogen peroxide to reach endpoint.45 Min at 50°C: Device does not reach endpoint color.Pass
Hydrogen peroxide (ISO 11140-1:2014 Type 1)Confirm device meets the Type 1 process indicator for vaporized hydrogen peroxide requirements.7 Sec at 50°C and gas concentration of 2.3 mg/l: Device does not reach endpoint color. 6 Min at 50°C and gas concentration of 2.3 mg/l: Device reaches endpoint color.Pass
Off-set/transference (ISO 11140-1:2014)Confirm device meets the Type 1 process indicator requirements.After exposure to 6 Min at 50°C and gas concentration of 2.3 mg/l: Ink does not transfer to another surface.Pass

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the exact sample size used for each individual test. It mentions "nonclinical testing" was conducted.
The data provenance is from nonclinical laboratory testing conducted by 3M Company, likely in the US (given the FDA submission). It is retrospective in the sense that the results are reported after the tests were performed to support the 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not applicable as the "ground truth" for chemical indicators is defined by physical/chemical reactions to specific controlled environmental conditions (e.g., presence/absence of hydrogen peroxide, specific temperature, time, and concentration). Human experts are not used to establish this type of ground truth for these device tests.

4. Adjudication Method for the Test Set:

This information is not applicable for these types of physical/chemical tests. The outcome (color change or lack thereof, ink transfer or lack thereof) is objectively observed and does not typically require adjudication by multiple parties.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

No, an MRMC comparative effectiveness study was not done. This type of study is more common for diagnostic imaging or screening devices where human interpretation is a critical component. The 3M™ Attest™ Hydrogen Peroxide Indicator Tape 1228 is a chemical indicator that shows a direct color change based on exposure to a sterilant, not a device requiring complex human interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Yes, the testing described is effectively "standalone" in that it assesses the device's inherent physical and chemical properties and its reaction to controlled conditions. There is no "human-in-the-loop" component for the performance of the indicator itself, beyond observing its color change.

7. The Type of Ground Truth Used:

The ground truth used is based on controlled experimental conditions (e.g., specific concentrations of hydrogen peroxide, temperature, and time) as defined by the ISO 11140-1:2014 standard. The "truth" is whether the indicator performs as expected under these known conditions (e.g., changes color when exposed, does not change color when not exposed).

8. The Sample Size for the Training Set:

This information is not applicable. The 3M™ Attest™ Hydrogen Peroxide Indicator Tape 1228 is a chemical indicator, not an AI/ML-enabled device. Therefore, it does not involve a "training set" in the context of machine learning.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

September 23, 2022

3M Company Yumi Wackerfuss Senior Regulatory Affairs Associate 2510 Conway Avenue, Bldg. 275-5W-06 St. Paul, Minnesota 55144-1000

Re: K222152

Trade/Device Name: 3MTM Attest™ Hydrogen Peroxide Indicator Tape 1228 Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: July 14, 2022 Received: July 20, 2022

Dear Yumi Wackerfuss:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

Submission Number (if known)

K222152

Device Name

3M™ Attest™ Hydrogen Peroxide Indicator Tape 1228

Indications for Use (Describe)

Use the 3M™ Attest™ Hydrogen Peroxide Indicator Tape 1228 to secure packs and as an external pack process indicator to differentiate unprocessed items processed in the following sterilizers:

STERRAD 100® Sterilization System

STERRAD 100S® Sterilization System

STERRAD NX® Sterilization System (Standard and Advanced cycles)

STERRAD 100NX® Sterilization System (Standard, Flex, Express and Duo cycles)

STERRAD NX® with ALLClear® Technology Sterilization System (Standard and Advanced cycles) STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express and Duo cycles)

V-PRO® 1 Low Temperature Sterilization System (Lumen cycle)

V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles)

V-PRO® maX Low Temperature Sterilization System (Lumen, and Flexible cycles)

V-PRO® 60 Low Temperature Sterilization System (Lumen, Non Lumen and Flexible cycles)

V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Flexible, and Fast Non Lumen cycles)

V-PRO® s2 Low Temperature Sterilization System (Lumen, Flexible, and Fast cycles) STERIZONE® VP4 Sterilizer (Cycle 1)

The tape is suitable for use on non-woven disposable wraps and peel pouches. The chemical indicator stripes turn from blue toward pink after exposure to vaporized hydrogen peroxide.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the 3M logo. The logo is in red and features the number "3" followed by the letter "M". The font is bold and sans-serif. The logo is simple and recognizable.

Image /page/3/Picture/3 description: The image shows the 3M logo, which is a well-known brand. The logo consists of the number "3" and the letter "M" in a bold, sans-serif font. The color of the logo is a vibrant red, and it is set against a white background. The logo is simple, yet recognizable.

510(k) Summary for 3M™ Attest™ Hydrogen Peroxide Indicator Tape 1228 K222152

Sponsor Information:

3M Company 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144-1000

Contact: Yumi Wackerfuss Senior Regulatory Affairs Associate Phone Number: (651) 733-3556 Fax Number: (651) 737-5320

Date of Summary: July 14th, 2022

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the 3M logo. The logo is red and features the number "3" stacked on top of the letter "M". The font is bold and sans-serif.

1. Device Name and Classification:

Common or Usual NameChemical Indicator
Proprietary Name3M™ Attest™ Hydrogen Peroxide Chemical Indicator 1228
Classification NameIndicator, physical/chemical sterilization process
Device ClassificationClass II, 21 CFR § 880.2800
Product CodeJOJ

2. Predicate Device:

K212081, 3M™ Comply™ Hydrogen Peroxide Indicator Tape 1228

3. Description of Device:

The 3M™ Attest™ Hydrogen Peroxide Indicator Tape 1228 consists of a non-cellulosic plastic backing material with a pressure-sensitive adhesive on one side and chemical indicator stripes on the other side. The chemical indicator stripes turn from blue toward pink after exposure to vaporized hydrogen peroxide. 3MTM Attest™ Hydrogen Peroxide Indicator Tape 1228 is a Type 1 (Category e1) Process Indicator as categorized by ISO 11140-1:2014.

4. Indications for Use

Use the 3M™ Attest™ Hydrogen Peroxide Indicator Tape 1228 to secure packs and as an external pack process indicator to differentiate unprocessed items from items processed in the following sterilizers:

STERRAD 100® Sterilization SystemSubmission Device: K222152Predicate Device (K212081)Comparison
STERRAD 100S® Sterilization SystemFeature3MTM AttestTM Hydrogen PeroxideIndicator Tape 12283MTM ComplyTM Hydrogen PeroxideIndicator Tape 1228
STERRAD NX® Sterilization System (Standard and Advanced cycles)Use the 3MTM AttestTM Hydrogen PeroxideIndicator Tape 1228 to secure packs and as anexternal pack process indicator to differentiateunprocessed items from items processed in thefollowing sterilizers:Use the 3MTM ComplyTM Hydrogen PeroxideIndicator Tape 1228 to secure packs and as anexternal pack process indicator to differentiateunprocessed items from items processed in thefollowing sterilizers:Similar. Product brandname is changed fromComplyTM to AttestTM.
STERRAD 100NX® Sterilization System (Standard, Flex, Express and Duo cycles)STERRAD 100® Sterilization SystemSTERRAD 100® Sterilization System
STERRAD NX® with ALLClear® Technology Sterilization System (Standard and Advanced cycles)STERRAD 100S® Sterilization SystemSTERRAD 100S® Sterilization System
STERRAD 100NX® with ALLClear® Technology Sterilization System (Standard, Flex, Express and Duo cycles)STERRAD NX® Sterilization System(Standard and Advanced cycles)STERRAD NX® Sterilization System(Standard and Advanced cycles)
V-PRO® 1 Low Temperature Sterilization System (Lumen cycle)STERRAD 100NX® Sterilization System(Standard, Flex, Express and Duo cycles)STERRAD 100NX® Sterilization System(Standard, Flex, Express and Duo cycles)
V-PRO® 1 Plus Low Temperature Sterilization System (Lumen and Non Lumen cycles)STERRAD NX® with ALLClear®Technology Sterilization System (Standardand Advanced cycles)STERRAD NX® with ALLClear®Technology Sterilization System (Standardand Advanced cycles)
V-PRO® maX Low Temperature Sterilization System (Lumen, Non Lumen, and Flexible cycles)STERRAD 100NX® with ALLClear®Technology Sterilization System (Standard,Flex, Express and Duo cycles)STERRAD 100NX® with ALLClear®Technology Sterilization System (Standard,Flex, Express and Duo cycles)
V-PRO® 60 Low Temperature Sterilization System (Lumen, Non Lumen and Flexible cycles)V-PRO® 1 Low Temperature SterilizationSystem (Lumen cycle)V-PRO® 1 Low Temperature SterilizationSystem (Lumen cycle)
V-PRO® maX 2 Low Temperature Sterilization System (Lumen, Non Lumen, Flexible, and Fast Non Lumen cycles)V-PRO® 1 Plus Low TemperatureSterilization System (Lumen and NonLumen cycles)V-PRO® 1 Plus Low TemperatureSterilization System (Lumen and NonLumen cycles)
V-PRO® s2 Low Temperature Sterilization System (Lumen, Non Lumen, Flexible, and Fast Non Lumen cycles)V-PRO® maX Low TemperatureSterilization System (Lumen, Non Lumen,and Flexible cycles)V-PRO® maX Low TemperatureSterilization System (Lumen, Non Lumen,and Flexible cycles)
V-PRO® 60 Low Temperature SterilizationSystem (Lumen, Non Lumen and Flexiblecycles)V-PRO® 60 Low Temperature SterilizationSystem (Lumen, Non Lumen and Flexiblecycles)
Indications foruse
Submission Device: K222152Predicate Device (K212081)Comparison
Feature3M™ Attest™ Hydrogen Peroxide3M™ Comply™ Hydrogen Peroxide
Indicator Tape 1228Indicator Tape 1228
V-PRO® maX 2 Low TemperatureV-PRO® maX 2 Low Temperature
Sterilization System (Lumen, Non Lumen,Sterilization System (Lumen, Non Lumen,
Flexible, and Fast Non Lumen cycles)Flexible, and Fast Non Lumen cycles)
V-PRO® s2 Low Temperature SterilizationV-PRO® s2 Low Temperature Sterilization
System (Lumen, Non Lumen, Flexible, andSystem (Lumen, Non Lumen, Flexible, and
Fast cycles)Fast cycles)
STERIZONE® VP4 Sterilizer (Cycle 1)STERIZONE® VP4 Sterilizer (Cycle 1)
The tape is suitable for use on non-wovenThe tape is suitable for use on non-woven
disposable wraps and peel pouches. Thedisposable wraps and peel pouches. The
chemical indicator stripes turn from bluechemical indicator stripes turn from blue
toward pink after exposure to vaporizedtoward pink after exposure to vaporized
hydrogen peroxide.hydrogen peroxide.
SubstrateNon-cellulosic plasticNon-cellulosic plasticSame
Type ofType 1 process indicatorNo claim for indicator typeNew claim for Type 1process indicator
The exposure to health care professionals isThe exposure to health care professionals isSame
Biocompatibilityminimal and well below any identified toxicminimal and well below any identified toxic
thresholds for the compounds.thresholds for the compounds.
Color ChangeBlue toward pinkBlue toward pinkSame
DetectionHydrogen PeroxideHydrogen PeroxideSame
Stability of theSame
endpointSix (6) monthsSix (6) months
reaction
Shelf lifeEighteen (18) monthsEighteen (18) monthsSame

STERIZONE® VP4 Sterilizer (Cycle 1)

The tape is suitable for use on non-woven disposable wraps and peel pouches. The chemical indicator stripes turn from blue toward pink after exposure to vaporized hydrogen peroxide.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the 3M logo. The logo is red and features the number "3" followed by the letter "M". The logo is simple and recognizable.

5. Summary of Technological Characteristics compared to Predicate Device

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the 3M logo. The logo is red and features the number "3" followed by the letter "M". The font is bold and sans-serif. The logo is simple and recognizable.

The 3M™ Attest™ Hydrogen Peroxide Indicator Tape 1228 is the same design as the previously cleared device of the same model number (the predicate) which is sold under the tradename 3M™ Comply™ Hydrogen Peroxide Indicator Tape 1228 (K212081). No changes have been made to the device materials or fundamental technology.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the 3M logo. The logo is in red and features the number "3" followed by the letter "M". The logo is simple and recognizable, and it is often used on 3M products and marketing materials. The logo is bold and eye-catching.

6. Nonclinical Comparison to the Predicate Device

3M conducted nonclinical testing to support the product performance in accordance with :

  • FDA Guidance for Industry and FDA Staff: Premarket Notification [510(k)] Submissions for ● Chemical Indicators, issued December 19, 2003
  • . ISO 11140-1:2014 Sterilization of healthcare products-Chemical Indicators-Part 1: General requirements for Type 1 (e1) Process Indicators.

Reference Table 6.1 for summary of nonclinical testing.

There is no change for the subject device from predicate device regarding the type 1 process indicator claim.

Test NamePurposeAcceptance CriteriaResult
Absence of hydrogenperoxide(ISO 11140-1:2014Type 1)Verify devicerequires thepresence ofvaporizedhydrogenperoxide to reachendpoint.45 Min at 50°CDevice doesnot reachendpoint color.Pass
Hydrogen peroxide(ISO 11140-1:2014Type 1)Confirm devicemeets the Type 1process indicatorfor vaporizedhydrogenperoxiderequirements.7 Sec at 50°C and gasconcentration of 2.3mg/l6 Min at 50°C and gasconcentration of 2.3mg/lDevice doesnot reachendpoint color.Device reachesendpoint color.Pass
Off-set/transference(ISO 11140-1:2014)Confirm devicemeets the Type 1process indicatorrequirements.After exposure to 6 Minat 50°C and gasconcentration of 2.3mg/lInk does nottransfer toanothersurface.Pass

Table 6.1 Summary of Nonclinical Testing

7. Conclusion

The conclusions drawn from the nonclinical testing demonstrate that the subject device, the 3M™ Attest™ Hydrogen Peroxide Indicator Tape 1228 is as safe, as effective, and performs as well as or better than the legally marketed predicate device, 3M™ Attest™ Hydrogen Peroxide Indicator Tape 1228 cleared under K212081, Class II (21 CFR 880.2800), product code JOJ.

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).