LUXEN CL Dental Liquid
K212548 · Dentalmax Co., Ltd. · EIH · Dec 6, 2021 · Dental
Device Facts
| Record ID | K212548 |
| Device Name | LUXEN CL Dental Liquid |
| Applicant | Dentalmax Co., Ltd. |
| Product Code | EIH · Dental |
| Decision Date | Dec 6, 2021 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.6660 |
| Device Class | Class 2 |
Intended Use
LUXEN CL Dental Liquid is a liquid used for the complete or partial coloration of milled porcelain substructures before sintering.
Device Story
Water-based liquid containing inorganic pigments; used by dental technicians to stain milled zirconia frameworks/restorations prior to final sintering; applied via brushing or immersion; enables color matching to natural teeth; output is a colored dental substructure; benefits patient by providing aesthetic, customized dental restorations.
Clinical Evidence
Bench testing only. Testing included shelf life, cytotoxicity (ISO 10993-5), sensitization and irritation (ISO 10993-10), oral mucosa irritation (ISO 10993-11), and bench assessments of appearance, volume, packaging, color stability, and extraction.
Technological Characteristics
Water-based solution with inorganic pigments; non-sterile; liquid form; available in 10ml, 30ml, 50ml, and 100ml volumes. Complies with ISO 10993 standards for biocompatibility.
Indications for Use
Indicated for complete or partial coloration of milled porcelain substructures prior to sintering in dental patients requiring zirconia restorations.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
Predicate Devices
- Aadva Zr Coloring Liquid (K092020)
Related Devices
- K173769 — DMAX Coloring Liquid; Chang's Liquid; Confident Coloring Liquid; CAMeleon Coloring Liquid · Dmax Co., Ltd. · Mar 8, 2018
- K220017 — Coloring Liquid · Dongguan Xiangtong Co., Ltd. · Jun 29, 2022
- K192723 — Coloring Liquid · Aidite (Qinhuangdao) Technology Co., Ltd. · Aug 18, 2020
- K232682 — Pre-Sintered Zirconia Coloring Liquid · Chengdu Besmile Medical Technology Co., Ltd. · Mar 19, 2024
- K233885 — rainbow Paste Stain SE · Genoss Co., Ltd. · Jun 21, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue.
December 6, 2021
DENTALMAX Co., Ltd. Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 18881 Von Karman Ave. STE 160 Irvine, California 92612
Re: K212548
Trade/Device Name: LUXEN CL Dental Liquid Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: September 20, 2021 Received: October 7, 2021
Dear Priscilla Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
## K212548
Device Name LUXEN CL Dental Liquid
### Indications for Use (Describe)
LUXEN CL Dental Liquid is a liquid used for the complete or partial coloration of milled porcelain substructures before sintering.
Type of Use (Select one or both, as applicable)
| <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|-----------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K212548 510(k) Summary
This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date:
#### 1. Applicant / Submitter:
DENTALMAX Co., Ltd. 50-7, Pungsesandan 2-Ro, Pungse-Myeon, Dongnam-Gu Cheonan-Si, Chungcheongnamdo, Republic of Korea 31217
#### 2. Submission Correspondent:
Priscilla Chung LK Consulting Group USA, Inc. 18881 Von Karman Ave. STE 160 Irvine CA 92612 Phone: 714-202-5789 Fax: 714-409-3357 Email: juhee.c(@lkconsultinggroup.com
#### 3. Device:
| Proprietary Name: | LUXEN CL Dental Liquid |
|------------------------------|----------------------------------------------|
| Common Name: | Liquid Stain for Dental Zirconium Prosthesis |
| Classification Name: | Porcelain Powder for Clinical Use |
| Classification: | Class II, 21 CFR 872.6660 |
| Classification Product Code: | EIH |
#### 4. Predicate Device:
Aadva Zr Coloring Liquid (K092020) by GC AMERICA Inc.
#### 5. Device Description:
The subject device is water-based solution. It is used for the individual staining of dental zirconia frameworks and restorations prior to the final sintering. It enables trained dental technicians to adjust the restoration to match the natural color of the patient's teeth.
{4}------------------------------------------------
The liquid has 100 shades are available in 10ml, 30ml, 50ml, 50ml and 100ml volumes.
Luxen CL-Diluent and Luxen CL-Pontic Liquid are used to dilute the color of other products. Luxen CL-Pontic and Luxen-CL Diluent is used to dilute Base liquid and special color liquid.
#### Indications for Use: 6.
LUXEN CL Dental Liquid is a liquid used for the complete or partial coloration of milled porcelain substructures before sintering.
#### 7. Performance Data(Non-Clinical):
The following properties were tested based on the referenced standards. All the test results support substantial equivalence to the predicate devices.
- = Shelf Life Test
- ISO 10993-5 Cytotoxicity
- ISO 10993-10 Sensitization & Irritation
- ISO 10993-11 Oral mucosa irritation
- Other bench testing Appearance, Volume, Packaging, Color Stability, Extraction
#### 8. Substantial Equivalence
The subject device is compared to the predicate device, Aadva Zr Coloring Liquid (K092020), in the table below.
| | Subject device | Predicate device |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | LUXEN CL Dental Liquid | Aadva Zr Coloring Liquid |
| 510(k) | - | K092020 |
| Product Code | EIH | EIH |
| Manufacturer | DENTALMAX Co., Ltd. | GC AMERICA Inc. |
| Technology | Water-based with inorganic pigments | Water-based with inorganic pigments |
| Indication for Use | LUXEN CL Dental Liquid is a liquid used for the complete or partial coloration of milled porcelain substructures before sintering. | Aadva Zr Coloring Liquid is a liquid used for the complete or partial coloration of milled porcelain substructures before sintering. |
| Principles of Operation | Brush or immerse zirconia ceramic with coloring liquid before sintering | Brush or immerse zirconia ceramic with coloring liquid before sintering |
| Prescription Use | Prescription only | Prescription only |
{5}------------------------------------------------
| Target population | General, mostly adults | General, mostly adults |
|-----------------------|------------------------|------------------------|
| Type of Packaging | Liquid container | Liquid container |
| Packaging Volume(ml) | 30ml, 50ml and 100ml | - |
| Shade | Various | Various |
| General Physical Form | Liquid | Liquid |
| Sterility | Non-Sterile | Non-Sterile |
The subject device is the same as the predicate device in the indications for use, principle of operation, and device characteristics. Both devices are water-based with inorganic pigments and employs brushing or immersing method for use.
The major difference would be chemical compositions, but we have performed the biocompatibility and other performance tests and the test results support that the subject device is substantially equivalent to the predict device.
#### 9. Conclusion:
Based on the testing results, DENTALMAX Co., Ltd. concludes that the subject device is substantially equivalent to the predicate device.