da Vinci SP Surgical System (SP1098)

K211595 · Intuitive Surgical, Inc. · NAY · Jun 23, 2021 · Gastroenterology, Urology

Device Facts

Record IDK211595
Device Nameda Vinci SP Surgical System (SP1098)
ApplicantIntuitive Surgical, Inc.
Product CodeNAY · Gastroenterology, Urology
Decision DateJun 23, 2021
DecisionSESU
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Intended Use

Da Vinci SP Surgical System, Model SP1098: The Intuitive Surgical Endoscopic Instrument Control System (da Vinci SP Surgical System, Model SP1098) is intended to assist in the accurate control of Intuitive Surgical EndoWrist SP Instruments during urologic surgical procedures that are appropriate for a single port approach and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use. EndoWrist SP Instruments: Intuitive Surgical EndoWrist SP Instruments are controlled by the da Vinci SP Surgical System, Model SP1098, and include flexible endoscopes, blunt and sharp endoscopic dissectors, scissors, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single port. The system is indicated for urologic surgical procedures that are appropriate for a single port approach and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Device Story

Computer-controlled surgical system enabling minimally invasive surgery via single port. Components: Surgeon Console (hand controls, foot pedals, 3D viewer), Vision Cart (image processing, networking), Patient Cart (instrument arm with four drives). Surgeon operates console to manipulate instruments/camera; patient-side assistant manages instrument installation. System transforms surgeon inputs into instrument movement. Modification: updated control algorithms to damp structural vibrations, reducing settling time during mode transitions. Used in operating rooms by trained physicians. Benefits: improved stabilization of endoscope view and instruments; enhanced surgical precision.

Clinical Evidence

Bench testing only. Regression testing verified embedded software and user interface. Control algorithms verified via unit testing, surgeon control mode accuracy, fault distance testing, and joint controller performance. Animal model validation confirmed system met acceptance criteria for safe and effective use.

Technological Characteristics

Computer-controlled surgical system; 3D vision system; electromechanical instrument drives. Software-based control algorithms for vibration damping. Connectivity via Vision Cart networking hardware. Sterilization: compatible with EndoWrist SP instruments.

Indications for Use

Indicated for adult patients undergoing urologic surgical procedures via single port approach or transoral otolaryngology procedures in the oropharynx for benign tumors and malignant tumors (T1/T2).

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 23, 2021 Intuitive Surgical, Inc. Connor McCarty Sr. Regulatory Engineer 1266 Kifer Road Sunnyvale, California 94086 Re: K211595 Trade/Device Name: da Vinci SP Surgical System (SP1098) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: NAY Dated: May 20, 2021 Received: May 24, 2021 Dear Connor Mccarty: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdth/cfdocs/cfpmn/pmn.cfm identifies.combination.product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. The OHT4: Office of Surgical and Infection Control Devices has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the device's labeling: The safety and effectiveness of this device for use in the performance of general laparoscopic surgery procedures have not been established. This device is only intended to be used for single port {1}------------------------------------------------ urological procedures and for transoral otolaryngology surgical procedures in the oropharynx for benign tumors and malignant tumors classified as T1 and T2 with the da Vinci EndoWrist SP Instruments and the da Vinci SP Surgical System (SP1098). Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling. Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical {2}------------------------------------------------ devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Binita Ashar, M.D., M.B.A., F.A.C.S. Director OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use Submission Number (if known) K211595 Device Name da Vinci SP Surgical System (SP1098) Indications for Use (Describe) Da Vinci SP Surgical System, Model SP1098: The Intuitive Surgical Endoscopic Instrument Control System (da Vinci SP Surgical System, Model SP1098) is intended to assist in the accurate control of Intuitive Surgical EndoWrist SP Instruments during urologic surgical procedures that are appropriate for a single port approach and transoral otolaryngology surgical procedures in the oropharvnx restricted to benign tumors and malignant. tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use. # EndoWrist SP Instruments: Intuitive Surgical EndoWrist SP Instruments are controlled by the da Vinci SP Surgical System, Model SP1098, and include flexible endoscopes, blunt and sharp endoscopic dissectors, scissors, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single port. The system is indicated for urologic surgical procedures that are appropriate for a single port approach and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary [As Required by 21 CFR 807.92(c)] June 14, 2021 | Submitter: | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086 | |-------------------|---------------------------------------------------------------------------------------| | Official Contact: | Connor McCarty<br>Sr. Regulatory Engineer<br>Phone: 805-798-4205<br>Fax: 408-523-8907 | | Trade Name: | Da Vinci SP Surgical System | | Common Name: | Endoscope and accessories | | Classification: | System, surgical, computer controlled instrument | | Predicate Device: | Da Vinci SP Surgical System (K202968) | #### Device Description: The da Vinci SP Surgical System is designed to enable complex surgery using a minimally invasive approach. The system consists of a Surgeon Console, a Vision Cart, and a Patient Cart and is used with a camera, instruments, and accessories. The surgeon seated at the Surgeon Console controls all movement of the instruments and camera by using two hand controls and a set of foot pedals. The surgeon views the camera image on a three-dimensional (3D) viewer, which provides a view of patient anatomy and instrumentation, along with icons and other user interface features. The Vision Cart includes supporting electronic equipment, such as the camera light source, video and image processing, and the networking hardware. The Vision Cart also has a touchscreen to view the camera image and adjust system settings. The Patient Cart is the operative component of the da Vinci SP Surgical System. Its primary function is to support the positioning of the surgical port and to manipulate the surgical instruments and camera. The Patient Cart is positioned at the operating room table and contains an instrument arm that is positioned with respect to the target patient anatomy. The instrument arm contains four instrument drives that hold up to three surgical instruments and the camera. The patient-side assistant installs and removes the camera and instruments intra-operatively. The design modifications included in this submission are limited to updated system software, specifically changes to the control algorithms to damp system structural vibrations. {5}------------------------------------------------ #### Intended Use/Indications for Use: Da Vinci SP Surgical System, Model SP1098: The Intuitive Surgical Endoscopic Instrument Control System (da Vinci SP Surgical System, Model SP1098) is intended to assist in the accurate control of Intuitive Surgical EndoWrist SP Instruments during urologic surgical procedures that are appropriate for a single port approach and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors and malignant tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use. #### EndoWrist SP Instruments: Intuitive Surgical EndoWrist SP Instruments are controlled by the da Vinci SP Surgical System, Model SP1098, and include flexible endoscopes, blunt and sharp endoscopic dissectors, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, and suturing through a single port. The system is indicated for urologic surgical procedures that are appropriate for a single port approach and transoral otolaryngology surgical procedures in the oropharynx restricted to benign tumors classified as T1 and T2. The system is indicated for adult use. It is intended for use by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use. #### Indications for Use Comparison: The indications for use are unchanged from the previously-cleared version of the system (K202968). ### Technological Comparison: The technological differences from the most recent clearance (K202968) are limited to the system software, specifically changes to the control algorithms to damp system structural vibrations. The enhanced control algorithms decrease the settling time of the surgeon transitions between control modes, so that the endoscope view and instruments stabilize more quickly after moving. The algorithm does not impact the physical feel of the surgeon's manipulations at the Surgeon Console, nor the feel of a user clutching the Instrument Arm on the Patient Cart. These modifications do not affect the substantial equivalence of the subject device to the predicate, as verification and validation testing have established there are no new issues of safety or effectiveness. {6}------------------------------------------------ ### Non-Clinical and/or Clinical Tests Summary & Conclusions: Regression testing of the software was performed to verify that the embedded software and user interface continued to meet specifications. The system control algorithms were verified by unit testing, the verification of surgeon control mode accuracy and fault distances, and the verification of joint controller performance. Validation testing using an animal model was performed to confirm that the system continued to meet its acceptance criteria for safe and effective use. Verification and validation testing on the subject device confirmed that no issues of safety or effectiveness and no additional unexpected risks were identified. The subject device met the same acceptance criteria as the predicate device. Therefore, the test results demonstrate that the subject device is substantially equivalent to its predicate device. Image /page/6/Picture/6 description: The image shows the word "INTUITIVE" in blue font. There is a small blue circle above the letter "U". The font is sans-serif and the letters are evenly spaced. The word is centered in the image.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%