The NEST Disposable Sampler (Viral Transport Medium) is intended for the collection and transport of upper respiratory clinical specimens to the laboratory for standard diagnostic or identification techniques. The Viral Transport Medium can be used in the laboratory to perform culture, isolation and detection of upper respiratory viruses including Influenza A, Rhinovirus, and Respiratory Syncytial Virus (RSV).
Device Story
Device consists of preservation tubes containing Viral Transport Media (VTM), optionally kitted with flocked nylon fiber swabs on ABS rods. Used for collection/transport of upper respiratory clinical specimens to laboratories. Operates by stabilizing viral samples in a chemical solution (salts, buffers, antibiotics, antifungal agents, BSA) to maintain viability for culture/isolation. Used in clinical settings by healthcare providers; samples processed in laboratories using standard diagnostic techniques. Benefits patient by ensuring sample integrity during transport, enabling accurate viral detection.
Clinical Evidence
Bench testing only. Viral recovery studies performed using three lots of preservation solution spiked with Influenza A (H1N1), Rhinovirus Type 16, and RSV Type A. Samples stored at 23-25°C and 2-8°C for 0, 24, and 48 hours. Viability measured via plaque assay on host cell lines. Results demonstrate successful viral stabilization over 48 hours, meeting performance specifications.
Indicated for collection and transport of upper respiratory clinical specimens from patients suspected of viral infection to a laboratory for culture, isolation, and detection of Influenza A, Rhinovirus, and RSV.
Regulatory Classification
Identification
A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.
Predicate Devices
Copan Universal Transport Medium (utm-rt) System (K042970)
Related Devices
K201674 — Cultura Collection and Transport System · Merit Medical Systems, Inc. · Mar 3, 2021
K212743 — MedSchenker Smart Transport Medium (STM15-A/STM20-A/STM30-A/SCS30-A) System · Medschenker, Inc. · May 3, 2023
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K211256
B Applicant
Wuxi Nest Biotechnology Co., Ltd.
C Proprietary and Established Names
Disposable Sampler Viral Transport Media
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JSM | Class I, reserved | 21 CFR 866.2390 - Transport Culture Medium | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain substantial equivalence determination for the NEST Disposable Sampler (Viral Transport Media) device for the collection, transport, and storage of viral specimens for laboratory culture and downstream testing.
B Measurand:
Not Applicable
C Type of Test:
Non-propagating Transport Device with culture medium.
## III Intended Use/Indications for Use:
A Intended Use(s):
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
See Indications for Use below.
## B Indication(s) for Use:
The NEST Disposable Sampler (Viral Transport Medium) is intended for the collection and transport of upper respiratory clinical specimens to the laboratory for standard diagnostic or identification techniques. The Viral Transport Medium can be used in the laboratory to perform culture, isolation and detection of upper respiratory viruses including Influenza A, Rhinovirus, and Respiratory Syncytial Virus (RSV).
## C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
## D Special Instrument Requirements:
None.
## IV Device/System Characteristics:
### A Device Description:
The Viral Transport Media is composed of media tube filled with VTM (Viral Transport Media), with or without swabs, depending on the kit configuration.
The Viral Transport Media is composed of Sodium chloride, Disodium hydrogen phosphate dodecahydrate, Potassium chloride, Potassium dihydrogen phosphate, Magnesium sulfate heptahydrate, glucose, HEPES, Sodium bicarbonate, Fluconazole, Gentamicin sulfate, Griseofulvin, Polymyxin sulfate, Sodium hydroxide, Calcium chloride, BSA, L-cysteine, and with or without Phenol red.
The preservation tube is made of medical-grade polypropylene materials. The pre filled VTM tubes are provided as 5 mL tubes filled with 2.5 mL VTM or 10 mL tubes filled with 3 mL VTM.
Both the 5mL and the 10mL size tubes filled with the volumes noted above are sold alone, kitted with either an oropharyngeal (OP) swab, a nasopharyngeal (NP) swabs or kitted with both OP and NP swabs. The OP and NP swabs are flocked nylon fiber with the swab shaft made of ABS (acrylonitrile butandiene styrene).
### B Principle of Operation:
The Viral Transport Media consists of a universal transport medium that can sustain the viability of clinically important viruses. The media is intended to be used by trained health care professionals. The VTM contains proteins for stabilization, antibiotics to minimize bacterial and fungal growth, a buffer to maintain a neutral pH and a pH indicator (phenol red) which is optional.
### V Substantial Equivalence Information:
K211256 - Page 2 of 8
{2}
A Predicate Device Name(s):
Copan Universal Transport Medium (utm-rt) System
B Predicate 510(k) Number(s):
K042970
C Comparison with Predicate(s):
| Device & Predicate Device(s): | Device: K211256 | Predicate: K042970 |
| --- | --- | --- |
| Device Trade Name | NEST Disposable Sampler Viral Transport Media | Copan universal transport medium (UTM-RT) system |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The NEST Disposable Sampler (Viral Transport Medium) is intended for the collection and transport of upper respiratory clinical specimens to the laboratory for standard diagnostic or identification techniques. The Viral Transport Medium can be used in the laboratory to perform culture, isolation and detection of upper respiratory viruses including Influenza A, Rhinovirus, and Respiratory Syncytial Virus (RSV). | Copan Universal Transport Medium (UTM-RT) System is intended for the collection and transport of clinical specimens containing viruses, chlamydiae, mycoplasma or ureaplasma from the collection site to the testing laboratory. UTM-RT can be processed using standard clinical laboratory operating procedures for viral, chlamydial, mycoplasma and ureaplasma culture. |
| Device Product Code and Classification | JSM, Class I | JSM, Class I |
| Shelf Life | 12 months | 12 months |
| pH | pH 7.3 ± 0.2 at 25°C | pH 7.3 ± 0.2 at 25°C |
| General Device Characteristic Differences | | |
| Media formulation | HANK's Balanced Salts Solution, HEPES, Sodium bicarbonate, Fluconazole, | HANK's Balanced Salts, BSA, L-cysteine, gelatin, sucrose, L-glutamic acid, HEPES |
| | fluconazole, L-glutamic acid, HEPES, and L-glutamic acid, HEPES, and L-glutamic acid, HEPES, and L-glutamic acid, HEPES, and L-glutamic acid, HEPES, and L-glutamic acid, HEPES, and L-glutamic acid, HEPES, and L-glutamic acid, HEPES, and L-glutamic acid, | |
K211256 - Page 3 of 8
{3}
K211256 - Page 4 of 8
| | Gentamicin sulfate, Griseofulvin, Polymyxin sulfate, Phenol red, Sodium hydroxide, glucose, BSA and L-cysteine. | buffer, vancomycin, amphotericin B, colistin, phenol red |
| --- | --- | --- |
| Vial Specification | 5 mL vial: 2.5 mL VTM
10 mL vial: 3.0 mL VTM | 1 mL UTM in 12x80 mm tube
3 mL UTM in 16x100 mm tube
10 mL UTM in 25x90 mm tube |
| Supported claims to perform culture, isolation and detection of: | Influenza A, Rhinovirus, and Respiratory Syncytial Virus (RSV) | chlamydiae, mycoplasma or ureaplasma and viruses |
| pH indicator | Optional | None |
VI Standards/Guidance Documents Referenced:
1. ISO 10993-5(2009) Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.
2. ISO 10993-10 (2010) Biological Evaluation of Medical Devices – Tests for irritation and skin sensitization.
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Not Applicable
2. Linearity:
Not Applicable
3. Analytical Specificity/Interference:
Not Applicable
4. Assay Reportable Range:
Not Applicable
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
{4}
Shelf-life: The shelf-life of the Viral Transport Media was determined in real-time to be 12 months when stored at 2-8°C and 25 ± 3°C. To assess the integrity of the kit, the following properties of the kit were assessed, at 5 replicates per timepoint, and had the following results. Three lots were assessed for the following categories:
a. The media was assessed for appearance. The Viral Transport Media remained a clear red color when assessed at 2-8°C and 25 ± 3°C at various time intervals out to 12 months.
b. pH was assessed using a pH meter to demonstrate that the Viral Transport Media maintain a pH between 7.3-7.5 when assessed at 2-8°C and 25 ± 3°C for 12 months.
c. Sterility was assessed by plating 100 μL of Viral Transport Media stored at 2-8°C and 25 ± 3°C at various time intervals out to 12 months, on PCA plates. The media showed no microbial growth at any of the time points tested.
d. The integrity of the antimicrobial components in the medium was assessed via a bacterial growth inhibition (bacteriostasis) study. The study demonstrated that media was bacteriostatic when the media was stored at 2-8°C and 25 ± 3°C for 12 months. The following organisms E. coli, S. aureus, B. subtilis, C. albicans and A. niger, were assessed and culture growth was determined to be inhibited through its claimed shelf life.
Sterilization:
The Viral Transport Media tube with media are not sold as sterile nor are they intended to be sterilized by the user. These vials are single use devices that do not require cleaning by the operator.
Swabs are sterilized in the final packaging using Irradiation Sterilization. Irradiation sterilization was performed using ISO 11137-1: 2015 and ISO 11137-2: 2016. A bioburden study was conducted, and it was determined that a 14.8 kGy dose of electron beam irradiation was required to reach a sterility assurance level of 10⁻⁶.
6. Detection Limit:
Viral Recovery:
Culture-Based Viral Recovery Studies: Performance of the Viral Transport Media was evaluated for virus viability at different incubation times, temperatures, and 3 lots of media.
Viability Assay Layout. Strains of Respiratory syncytial virus Type A (RSV-A), Influenza A (A/PR/8/34 H1N1) virus, and Rhinovirus Type 16 (HRV-16) were used for media validation. Each virus was first mixed with nasopharyngeal matrix, transferred with the collection swab and then added to the Viral Transport Media to yield a final concentration of 10⁵ TCID₅₀/mL. The mixtures were then incubated at 4°C and 25°C in triplicates. Aliquots of each replicate were recovered at 0, 24, or 48 hours and serially diluted. A 100 μL volume of each dilution was then inoculated in duplicate into the susceptible host cell line. Host cells (HEp-2 for RSV-A, MDCK for H1N1, and HeLa for HRV-16) were plated at a suitable density in microwell plates for 2-3 days prior to evaluation. The following controls were run: a viability control containing no virus to ensure mammalian cell integrity and absence of contamination; a virus control containing virus stock to ensure infectivity of virus; a cytotoxicity control containing VTM only to observe any cytotoxic effects from the VTM on the mammalian cells; and a recovery control containing virus in culture medium stored and tested under the same conditions as the test samples. Host cells and
K211256 - Page 5 of 8
{5}
diluted virus were incubated for 2-3 days, followed by fixation of the cells with Formalin for 1 hour and finally staining with methylene blue for 15 minutes. Viability was determined by assessing for plaque formation in triplicate for each time point tested. Tables 1, 2, and 3 show the mean PFU/mL values for each virus tested at time 0 and different times and temperatures for viral recovery. The data supported the claim for storage times of 24 hours and 48 hours and temperatures of $4^{\circ}\mathrm{C}$ and $25^{\circ}\mathrm{C}$ .
Table 1: Viral recovery for Influenza A at various incubation times and tempearatures.
| Test Virus | Test samples (lot #) | Test conditions | Mean Virus Titer (x104PFU/mL) | Percent decrease (%) |
| --- | --- | --- | --- | --- |
| Influenza A (A/PR/8/34 H1N1) | 080921ES1 | 0h | 83.4 | - |
| | | 4°C, 24h | 70.5 | 15.47 |
| | | 4°C, 48h | 61.3 | 26.50 |
| | | | | |
| | | 25°C, 24h | 66.4 | 20.38 |
| | | 25°C, 48h | 44.7 | 46.40 |
| | | | | |
| | 040121PS | 0h | 79.2 | - |
| | | 4°C, 24h | 71.8 | 9.34 |
| | | 4°C, 48h | 63.7 | 19.57 |
| | | | | |
| | | 25°C, 24h | 66.9 | 15.53 |
| | | 25°C, 48h | 44.4 | 43.94 |
| | | | | |
| | 101020E01 | 0h | 92.3 | - |
| | | 4°C, 24h | 79.1 | 14.30 |
| | | 4°C, 48h | 73.5 | 20.37 |
| | | | | |
| | | 25°C, 24h | 74.3 | 19.50 |
| | | 25°C, 48h | 60.2 | 34.78 |
| | | | | |
Table 2: Viral recovery for Rhinovirus Type 16 at various incubation times and tempearatures.
| Test Virus | Test samples (lot #) | Test conditions | Mean Virus Titer (x104PFU/mL) | Percent decrease (%) |
| --- | --- | --- | --- | --- |
| Rhinovirus Type 16 (HRV-16) | 080921ES1 | 0h | 231.7 | - |
| | | 4°C, 24h | 213.8 | 7.73 |
| | | 4°C, 48h | 204.5 | 11.74 |
| | | | | |
| | | 25°C, 24h | 168.5 | 27.28 |
| | | 25°C, 48h | 124.7 | 46.18 |
K211256 - Page 6 of 8
{6}
Table 3: Viral recovery for Respiratory Syncytial Virus Type A at various incubation times and temperatures.
| Test Virus | Test samples (lot #) | Test conditions | Mean Virus Titer (x104PFU/mL) | Percent decrease (%) |
| --- | --- | --- | --- | --- |
| Respiratory Syncytial Virus Type A (RSV-A) | 080921ES1 | 0h | 26.8 | - |
| | | 4°C, 24h | 22.0 | 17.91 |
| | | 4°C, 48h | 20.0 | 25.37 |
| | | | | |
| | | 25°C, 24h | 23.2 | 13.43 |
| | | 25°C, 48h | 21.0 | 21.64 |
| | | | | |
| | 040121PS | 0h | 31.0 | - |
| | | 4°C, 24h | 24.0 | 22.58 |
| | | 4°C, 48h | 21.2 | 31.61 |
| | | | | |
| | | 25°C, 24h | 24.8 | 20.00 |
| | | 25°C, 48h | 16.2 | 47.74 |
| | | | | |
| | 101020E01 | 0h | 33.5 | - |
| | | 4°C, 24h | 26.3 | 21.49 |
| | | 4°C, 48h | 25.7 | 23.28 |
| | | | | |
| 25°C, 24h | | 23.8 | 28.96 | |
| 25°C, 48h | | 26.7 | 20.30 | |
Results of Viability Assay. The Viral Transport Media demonstrated virus viability of Flu A (H1N1), RSV, Rhinovirus for all replicates in all lots, incubation times, and storage
K211256 - Page 7 of 8
{7}
temperatures. The studies demonstrate that the VTM maintains the viability of all viruses tested for 48 hours at refrigerated (4°C) and room temperature (25°C).
7. Assay Cut-Off:
Not Applicable
## B Comparison Studies:
1. Method Comparison with Predicate Device:
Not Applicable
2. Matrix Comparison:
Not Applicable
## C Clinical Studies:
1. Clinical Sensitivity:
Not Applicable
2. Clinical Specificity:
Not Applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not Applicable
## D Clinical Cut-Off:
Not Applicable
## E Expected Values/Reference Range:
Not Applicable
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K211256 - Page 8 of 8
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.