Auto Disable Syringe

K210464 · Anhui Tiankang Medical Technology Co., Ltd. · FMF · Aug 24, 2021 · General Hospital

Device Facts

Record IDK210464
Device NameAuto Disable Syringe
ApplicantAnhui Tiankang Medical Technology Co., Ltd.
Product CodeFMF · General Hospital
Decision DateAug 24, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic

Intended Use

The AUTO DISABLE SYRINGE is intended for use in the suction of vaccine for medical purposes. Additionally , after injection to the body, the plunger can be automatically locked by the triggered mechanism to prevent the re-use of this syringe.

Device Story

Auto Disable Syringe is a sterile, single-use hypodermic syringe with an integrated fixed needle. Device consists of a calibrated polypropylene barrel, movable plunger, polyisoprene rubber stopper, and steel clip mechanism. Input is manual aspiration of vaccine; device transforms input via plunger movement to deliver fixed dose; output is vaccine injection. Upon completion of injection, internal steel clip triggers to lock plunger, preventing reuse. Used in clinical settings by healthcare professionals for immunization. Benefits include prevention of needle reuse and associated blood-borne pathogen transmission. Device is provided sterile via ethylene oxide.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including functional performance (ISO 7886-1, ISO 7886-3, ISO 7864, ISO 9626), biocompatibility (ISO 10993-1), and sterility/shelf-life validation (ISO 11135, ASTM D4169, ASTM F1980).

Technological Characteristics

Materials: Polypropylene (barrel/plunger), polyisoprene rubber (piston), stainless steel (needle). Energy: Manual. Dimensions: 0.05ml-1.0ml capacity; 23G-30G needles. Sterilization: Ethylene Oxide (10^-6 SAL). Standards: ISO 7886-1/3, ISO 7864, ISO 9626, ISO 10993, ISO 11135, ASTM F88, ASTM F1886, ASTM F1929, ASTM F1980.

Indications for Use

Indicated for vaccine administration in patients requiring hypodermic injection. Intended for single-use only; features automatic locking mechanism to prevent reuse.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 24, 2021 Anhui Tiankang Medical Technology Co.,Ltd. Bai Baodong General Manager No.228 Weiyi Road, Economic Development Zone Tianchang, Anhui 239300 China Re: K210464 Trade/Device Name: Auto Disable Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: July 23, 2021 Received: July 26, 2021 Dear Bai Baodong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For CAPT. Alan M. Stevens Assistant Director Injection Devices Team DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K210464 Device Name AUTO DISABLE SYRINGE #### Indications for Use (Describe) The AUTO DISABLE SYRINGE is intended for use in the suction of vaccine for medical purposes. Additionally , after injection to the body, the plunger can be automatically locked by the triggered mechanism to prevent the re-use of this syringe. Type of Use (Select one or both, as applicable) | <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | |----------------------------------------------------------------------------------------------| | <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### K210464 510(K) Summary #### Date prepared: August 24, 2021 #### Submitter Name and Address: | Submitter Name: | Anhui Tiankang Medical Technology Co., Ltd. | |-----------------|-------------------------------------------------------------------------------| | Address: | No.228 Weiyi Road, Economic Development Zone, Tianchang City.<br>Anhui, China | | Contact Person: | Bai Baodong, General Manager | | TEL: | +86-550-7309187 | | FAX: | +86-550-7308901 | | E-mail: | tkquality@126.com | ### Submission Devices Information: Trade/Proprietary Name: AUTO DISABLE SYRINGE Classification name: Syringe, piston Class: II. Product codes: FMF Regulation number: 21CFR880.5860, Submission Type: 510(K) #### Predicate Device Information: Company Name: Becton, Dickinson, and Company Device name: BD SoloShot Mini Syringe/ BD Auto Disable Syringe 510(K) Number: K201234 #### Device Description 1) Indication for use: The AUTO DISABLE SYRINGE is intended for use in the suction and injection of vaccine for medical purposes. Additionally, after injection to the plunger can be automatically locked by the triggered mechanism to prevent the re-use of this syringe. 2) Design of syringe: The AUTO DISABLE SYRINGE is designed for the hypodermic injection of the intended micro {4}------------------------------------------------ scale solution. It has the automatic function in the structure of the syringe which after injection of an intended fixed dose locks the syringe to prevent re-use of the syringe. It can be widely used for injection of vaccine for the immunization. 3) Structure of syringe (with needle): The syringe consists of a calibrated hollow barrel, a movable plunger, a rubber stopper (piston) assembled at end of the plunger, and a steel clip installed between the barrel and the plunger that functions to prevent re-use of the syringe. At the end of the barrel, a fixed an unmovable needle is installed, and a needle cap covers the needle. The zero line and nominal capacity line are printed outside of barrel. The main materials are PP, latex free (polyisoprene rubber) and stainless steel. The AUTO DISABLE SYRNGE is for Prescription (Rx) use only ### Product models: | Description | Needle Length | |--------------------------|----------------------------| | 1cc(ml), with needle:23G | 3/8", 1/2", 3/4", 7/8", 1" | | 1cc(ml), with needle:24G | 3/8", 1/2", 3/4", 7/8", 1" | | 1cc(ml), with needle:25G | 3/8", 1/2", 3/4", 7/8", 1" | | 1cc(ml), with needle:26G | 3/8" and 1/2" | | 1cc(ml), with needle:27G | 3/8" and 1/2" | | 1cc(ml), with needle:28G | 3/8" and 1/2" | | 1cc(ml), with needle:29G | 3/8" and 1/2" | | 1cc(ml), with needle:30G | 3/8" and 1/2" | | Description | Needle Length | |----------------------------|----------------------------| | 0.5cc(ml), with needle:23G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.5cc(ml), with needle:24G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.5cc(ml), with needle:25G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.5cc(ml), with needle:26G | 3/8" and 1/2" | | 0.5cc(ml), with needle:27G | 3/8" and 1/2" | | 0.5cc(ml), with needle:28G | 3/8" and 1/2" | | 0.5cc(ml), with needle:29G | 3/8" and 1/2" | | 0.5cc(ml), with needle:30G | 3/8" and 1/2" | | Description | Needle Length | |----------------------------|----------------------------| | 0.4cc(ml), with needle:23G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.4cc(ml), with needle:24G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.4cc(ml), with needle:25G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.4cc(ml), with needle:26G | 3/8" and 1/2" | | 0.4cc(ml), with needle:27G | 3/8" and 1/2" | | 0.4cc(ml), with needle:28G | 3/8" and 1/2" | | 0.4cc(ml), with needle:29G | 3/8" and 1/2" | | 0.4cc(ml), with needle:30G | 3/8" and 1/2" | {5}------------------------------------------------ | Description | Needle Length | |----------------------------|----------------------------| | 0.3cc(ml), with needle:23G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.3cc(ml), with needle:24G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.3cc(ml), with needle:25G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.3cc(ml), with needle:26G | 3/8" and 1/2" | | 0.3cc(ml), with needle:27G | 3/8" and 1/2" | | 0.3cc(ml), with needle:28G | 3/8" and 1/2" | | 0.3cc(ml), with needle:29G | 3/8" and 1/2" | | 0.3cc(ml), with needle:30G | 3/8" and 1/2" | | Description | Needle Length | |-----------------------------|----------------------------| | 0.25cc(ml), with needle:23G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.25cc(ml), with needle:24G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.25cc(ml), with needle:25G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.25cc(ml), with needle:26G | 3/8" and 1/2" | | 0.25cc(ml), with needle:27G | 3/8" and 1/2" | | 0.25cc(ml), with needle:28G | 3/8" and 1/2" | | 0.25cc(ml), with needle:29G | 3/8" and 1/2" | | 0.25cc(ml), with needle:30G | 3/8" and 1/2" | | Description | Needle Length | |----------------------------|----------------------------| | 0.2cc(ml), with needle:23G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.2cc(ml), with needle:24G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.2cc(ml), with needle:25G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.2cc(ml), with needle:26G | 3/8" and 1/2" | | 0.2cc(ml), with needle:27G | 3/8" and 1/2" | | 0.2cc(ml), with needle:28G | 3/8" and 1/2" | | 0.2cc(ml), with needle:29G | 3/8" and 1/2" | | 0.2cc(ml), with needle:30G | 3/8" and 1/2" | | Description | Needle Length | |----------------------------|----------------------------| | 0.1cc(ml), with needle:23G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.1cc(ml), with needle:24G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.1cc(ml), with needle:25G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.1cc(ml), with needle:26G | 3/8" and 1/2" | | 0.1cc(ml), with needle:27G | 3/8" and 1/2" | | 0.1cc(ml), with needle:28G | 3/8" and 1/2" | | 0.1cc(ml), with needle:29G | 3/8" and 1/2" | | 0.1cc(ml), with needle:30G | 3/8" and 1/2" | | Description | Needle Length | |-----------------------------|----------------------------| | 0.05cc(ml), with needle:23G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.05cc(ml), with needle:24G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.05cc(ml), with needle:25G | 3/8", 1/2", 3/4", 7/8", 1" | | 0.05cc(ml), with needle:26G | 3/8" and 1/2" | | 0.05cc(ml), with needle:27G | 3/8" and 1/2" | | 0.05cc(ml), with needle:28G | 3/8" and 1/2" | | 0.05cc(ml), with needle:29G | 3/8" and 1/2" | | 0.05cc(ml), with needle:30G | 3/8" and 1/2" | {6}------------------------------------------------ ## Comparison of technological characteristics with the predicate: | Elements of<br>comparison | Subject device | Predicate device: | Comments | |--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|------------| | AUTO DISABLE<br>SYRINGE | BD SoloShot Mini<br>Syringe/ BD Auto Disable<br>Syringe (K201234) | | | | Indications for<br>Use | The AUTO DISABLE<br>SYRINGE is intended<br>for use in the suction<br>and injection of vaccine<br>for medical purposes.<br>Additionally, after<br>injection to the body,<br>the plunger can be<br>automatically locked by<br>the triggered<br>mechanism to prevent<br>the re-use of this<br>syringe. | BD SoloShot Mini<br>Syringe/ BD Auto<br>Disable Syringe is<br>intended for aspiration<br>and injection of fluids. | Comment 1. | | Principle of<br>operation | Normal | Normal | Identical | | Needle length | 3/8", 1/2", 1" | 3/8", 5/8", 3/4", 1" | Comment 2. | | Needle gauge | 23G, 24G, 25G, 26G,<br>27G, 28G, 29G, 30G | 23G, 24G, 25G, 27G | Comment 3 | | Syringe<br>capacity | 0.05ml, 0.10ml, 0.20ml,<br>0.25ml, 0.30ml, 0.40ml,<br>0.50ml, 1.0ml | 0.05ml, 0.10ml, 0.50ml, | Comment 4 | | Reuse<br>Prevention<br>(Safety)<br>Feature | Auto-disabled, prevents<br>syringe re-use | Auto-disabled,<br>prevents syringe<br>re-use | Identical | | Lubricant for<br>barrel | Silicone oil | Silicone oil | Identical | | Barrel<br>transparency | Transparent and clear | Transparent and clear | Identical | | Gradations<br>legibility | Legible | Legible | Identical | | Material | Barrel: PP | Barrel: PP | Comment 5 | | | Plunger: PP | Plunger: PP | | | | Piston: Latex Free | Piston: rubber | | | | (Polyisoprene Rubber) | Needle: stainless steel | | | | Needle: stainless steel | | | | Performance | Conform to ISO7886-3<br>and ISO7886-1,<br>ISO9629, ISO7864 | Conform to ISO7886-3<br>and ISO7886-1,<br>ISO9629, ISO7864 | Identical | | Biocompatibility | Conform to ISO10993 | Conform to ISO10993 | Identical | | Labeling | Meet the requirements<br>of 21CFR Part 801 | Meet the requirements<br>of 21CFR Part 801 | Identical | | Sterility | EO sterilization 10-6 | EO sterilization 10-6 | Identical | {7}------------------------------------------------ Discussions of differences in technological characteristics ## Comment 1: The indications for use are similar for both devices (to the extent that both devices are intended for aspiration and injection of fluids for medical purposes). The predicate device defines that the device is for general use. The subject device is specifically used for vaccine administration, which is included in the scope of general use. In addition, the subject device includes a statement regarding the locking mechanism to prevent re-use. This mechanism is included in the predicate design, but is not included in the indications for use statement. The differences to the indications for use statement between the predicate and the subject device are acceptable. ## Comment 2, 3 and 4: The specifications of the proposed device and predicate device are different; however, both of them comply with the same recognized standards. The different sizes of needle are addressed through performance testing. ## Comment 5: The piston used the new technology material to replace the traditional rubber material for less toxicity (Polyisoprene Rubber). The safety testing report of biocompatibility and function performance testing report demonstrates that the change does not negatively impact safety and performance. {8}------------------------------------------------ ## Non-Clinical Testing ## Functional Performance Testing The AUTO DISABLE SYRINGE described in this summary were tested and demonstrated to be in conformance with the following FDA recognized standards: ISO 7864:2016, Sterile hypodermic needles for single use - Requirements and test methods - . Fragmentation test - . Determination of flow rate - . Lubricant: Penetration force and drag force - . Bonding strength ISO 7886-3, sterile hypodermic syringes for single use - Part 3: Auto-disabled syringes for fixed-dose immunization and ISO7886-1, sterile hypodermic syringes for single use - Part 1: Syringes for manual use - . Limits for acidity and alkalinity - . Limits for extractable metals - . Lubricant: Penetration force and drag force - . Tolerance on nominal capacity - Graduated scale ● - . Barrel- dimensions and flanges - Plunger stopper and assembly - fitness of stopper /plunger in barrel - Integrated needle - . Sharps protection features - . Dead space - Freedom from air and liquid leakage ● - Auto-disable syringe feature . ISO9626, Stainless steel needle tubing for the manufacture of medical devices -Requirements and test methods - Limits for acidity and alkalinity - designated metric size ● - Stiffness . - Resistance to breakage ● - Resistance to corrosion . ## Sterility, Shipping and Shelf-Life - . EO sterilization validation and package validation conducted in accordance with ISO11135:2014, ISO11607-1:2019, ISO11607:2019 {9}------------------------------------------------ - . Sterilant residuals were evaluated per ISO10993-7:2008. Average daily dose to patient of EO less than 4mg/device and ECH less than 9mg/device - . Package integrity testing, after environmental conditioning and simulated transportation in accordance with ASTM D4169-16, was conducted on the final, packaged, and sterile devices. All packaging deemed acceptable for protection of product and sterility maintenance. - . Sterile Barrier Packaging Testing performed on the proposed device: - o Seal strength ASTM F88/F88-15 - o Visual Inspection ASTM F 1886-16 - o Dye penetration ASTM F1929-15 - . Shelf life of 5 years is validated using the FDA recognized standard ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices ## Biocompatibility Testing (ISO10993-1:2018) In accordance with ISO 10993-1, the syringe is classified as: Externally Communicating Device, Blood Path Indirect, Limited Contact (<24 hours). The following testing was conducted: - . Cytotoxicity (per ISO10993-5:2009) - . Sensitization (per ISO10993-10:2014) - Irritation (per ISO10993-10:2017) . - Acute Systemic Toxicity (per ISO10993-11:2017) . - Pyrogenicity (per ISO10993-4:2017) ● - Hemocompatibility (per ISO10993-4:2017) . - . Particulate matter ( per USP<788>) ## Conclusions The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The AUTO DISABLE SYRINGE is substantially equivalent to BD SoloShot Mini Syringe/ BD Auto Disable Syringe with respect to the indications for use and technological characteristics.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%