Incontinence Probe
K210441 · Shenzhen Med-Link Electronics Tech Co., Ltd. · KPI · Sep 17, 2021 · Gastroenterology, Urology
Device Facts
| Record ID | K210441 |
| Device Name | Incontinence Probe |
| Applicant | Shenzhen Med-Link Electronics Tech Co., Ltd. |
| Product Code | KPI · Gastroenterology, Urology |
| Decision Date | Sep 17, 2021 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.5320 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Incontinence Probes are intended to provide electromyographic musculature or electrical stimulation to pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control during the treatment of urinary incontinence.
Device Story
Incontinence Probes (models PE0001-PE0010) are passive, stand-alone electrodes for vaginal or rectal insertion. Devices function as conduits, transmitting electrical stimulation signals from external stimulators to pelvic musculature or capturing EMG signals from pelvic floor muscles for biofeedback. Operated by patients in home or clinical settings; compatible with low-frequency (<1000Hz) stimulators/EMG devices using 2.0 DIN connections. Probes facilitate muscle rehabilitation and neuromuscular control restoration. Healthcare providers use the resulting biofeedback or stimulation to guide incontinence treatment plans. Benefit includes non-invasive pelvic floor muscle strengthening.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing: biocompatibility (ISO 10993: cytotoxicity, sensitization, vaginal/rectal irritation), shelf-life validation, electrical safety (IEC 60601-1), and performance bench testing (visual inspection, electrical output verification, cable dielectric strength, contact resistance, and lead wire testing per ANSI/AAMI EC53).
Technological Characteristics
Materials: S304 stainless steel and Acrylonitrile-Butadiene-Styrene (ABS). Sensing/Actuation: Passive electrode for EMG signal acquisition or electrical stimulation delivery. Connectivity: 2.0 DIN connector for low-frequency (<1000Hz) host devices. Form factor: Vaginal and rectal probes. Sterility: Non-sterile; reusable (except PE0003, which is disposable). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ANSI/AAMI EC53.
Indications for Use
Indicated for adult patients requiring pelvic floor muscle rehabilitation and neuromuscular control restoration for urinary incontinence treatment via electromyographic feedback or electrical stimulation.
Regulatory Classification
Identification
A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).
Predicate Devices
- Fuji Dynamics Incontinence Stimulation Electrode (K161055)
Related Devices
- K222528 — Well-Life Probe Electrode for Stimulation/EMG Probe (Model: SA), Well-Life Incontinence Stimulation/EMG Electrode Probe, Well-Life Probe Electrode · Well-Life Healthcare Limited · Dec 29, 2022
- K231166 — Levina, Levina Incontinence Stimulation Electrodes (Sterile), Pelvic Floor Muscle Probe Electrodes · Zmi Electronics , Ltd. · Jan 18, 2024
- K233576 — Konmed Incontinence Stimulation Electrode (Vaginal Probe: KM-503, KM-504, KM-505, KM-506, KM-507, KM-5013; Rectal Probe: KM-502, KM-5018) · Shenzhen Konmed Technology Co., Ltd. · Feb 23, 2024
- K974036 — PATHWAY VAGINAL/RECTAL PERINEOMETER PROBE · The Prometheus Group · Dec 22, 1997
- K140265 — ELECTRODE FOR URINARY INCONTINENCE · Guangzhou Finecure Medical Equipment Co.,Ltd · Jan 23, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left and consists of a symbol and text. The U.S. Food & Drug Administration logo is on the right and consists of the FDA acronym in a blue box, followed by the full name of the agency in blue text. The word "ADMINISTRATION" is written in a smaller font size than "U.S. FOOD & DRUG".
September 17, 2021
Shenzhen Med-link Electronics Tech Co, Ltd Yongfen Yang Regulatory Affairs Specialist 4th And 5th Floor, Building Two, Hualian Industrial Zone, Xinshi Community Dalang Street Shenzhen, Guangdong 518109 China
K210441 Trade/Device Name: Incontinence Probe Regulation Number: 21 CFR& 876.5320 Regulation Name: Nonimplanted Electrical Continence Device Regulatory Class: II Product Code: KPI, HIR Dated: August 10, 2021 Received: August 13, 2021
Dear Yongfen Yang:
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jessica K. Nguyen -S
Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210441
Device Name Incontinence Probe
Indications for Use (Describe)
The Incontinence Probes are intended to provide electromyographic musculature or electrical stimulation to pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control during the treatment of urinary incontinence.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;"><b>X</b></span> |
|----------------------------------------------|------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span></span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for MedLinket. The logo is in blue and orange. The word "MedLinket" is written in a sans-serif font, with the "M" in orange and the rest of the letters in blue.
# 510(k) Summary
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
### 1. Submitter Information:
| Name and Address: | Shenzhen Med-link Electronics Tech Co., Ltd.<br>4th and 5th Floor, Building Two,<br>Hualian Industrial Zone, Xinshi Community,<br>Dalang Street, Longhua District, Shenzhen,<br>Guangdong 518109 .<br>People's Republic of China |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact: | Yongfen Yang (Regulatory Affairs Specialist)<br>E-mail: mdl012@medlinket.com<br>Tel: 0086-755-61568825<br>Fax: 0086-755-61120055 |
| Data of Preparation: | September 15, 2021 |
# 2. Device Information:
| Trade Name: | Incontinence Probe |
|------------------------|-------------------------------------------|
| Common Name: | Incontinence Probe |
| Classification Number: | 21 CFR 876.5320 |
| Classification Name: | Nonimplanted Electrical Continence Device |
| Product Code: | KPI, HIR |
| Regulatory Class: | II |
| Review Panel: | Gastroenterology/Urology |
# 3. Identification of the Predicate Device:
| Device Name | Common Name | Manufacture | 510(k) |
|--------------------------------------------------------|-----------------------|-----------------------|---------|
| Fuji Dynamics<br>Incontinence<br>Stimulation Electrode | Incontinence<br>Probe | Fuji Dynamics<br>Ltd. | K161055 |
{4}------------------------------------------------
#### 4. Device Description:
The Incontinence Probe models PE0001, PE0002, PE0004, PE0005, PE0006, PE0007 , PE0008, PE0009, and PE0010 are stand-alone probes/electrodes, which can be placed inside the adult rectum or female vagina to transmit electrical stimulation signals or pelvic floor EMG signals when connected to pelvic electrical stimulators or EMG biofeedback devices during the treatment of urinary incontinence . All the models are reusable for a single patient except PE0003, which is a disposable vaginal probe. The probes are not packaged with specific floor muscle stimulators and EMG/biofeedback equipment but are compatible with low-frequency (less than 1000Hz working frequency) stimulators or EMG/biofeedback devices that meet the standards of IEC 60601-1, IEC 60601-1-2 and IEC 60601-2-10 and use 2.0 DIN connection. The probes are made of S304 stainless steel and Acrylonitrile-Butadiene-Styrene (ABS). The probes are designed for use in home or clinic by a single user. The probes do not require sterilization but need cleaning for reuse according to the cleaning instructions provided in the instructions for use manual.
### 5. Indication for Use:
The Incontinence Probes are intended to provide electromyographic feedback from pelvic musculature or electrical stimulation to pelvic musculature for the purpose of rehabilitation of weak pelvic floor muscles and restoration of neuromuscular control during the treatment of urinary incontinence.
| Characteristic | Vaginal Probe | | Rectal Probe | |
|------------------------|---------------------------------------------------------------|-------------------------------------------------------|---------------------------------------------------------|----------------------------------------------------|
| | Predicate device | Subject device | Predicate device | Subject device |
| 510(K)<br>Submitter | Fuji Dynamics Ltd. | Shenzhen Med-link<br>Electronics Tech<br>Co. Ltd. | Fuji Dynamics Ltd. | Shenzhen Med-link<br>Electronics Tech<br>Co., Ltd. |
| Model | Fuji Dynamics<br>Fuji-01/ 02 / 03/ 04 /<br>05/ 06/ 07/ 14/ 15 | PE0002,PE0003,PE0<br>005,PE0007,PE000<br>9,<br>PE0010 | Fuji Dynamics<br>Fuji- 08/ 09/ 10/ 11/ 12<br>13/ 14/ 15 | PE0001,PE0004<br>PE0006,PE0008 |
| Number of<br>Electrode | 2-Stimulation / EMG | Same | 2-Stimulation / EMG | Same |
# 6. Comparison of the Technological Characteristics with Predicate Device:
{5}------------------------------------------------
| Usage<br>Conditions | Reusable - single<br>patient | Except the model<br>PE0003 which is<br>disposable, the other<br>models are<br>reusable-single use | Reusable - single<br>patient | Same |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|------|
| Electrode<br>Material | Stainless Steel & ABS | Same | Stainless Steel & ABS | Same |
| Electrode<br>Placement | Vagina | Same | Rectum | Same |
| Contact<br>Duration | Intermittent mucosal<br>contact < 30<br>min/session -Stim <<br>1 hour/session -<br>EMG not exceeding<br>hr combined<br>Stim/EMG | Same | Intermittent mucosal<br>contact < 30<br>min/session -Stim < 1<br>hour/session - EMG<br>not exceeding hr<br>combined Stim/EMG | Same |
| Sterility | Non-sterile | Same | Non-sterile | Same |
In addition to the differences noted in the above table, there are differences in dimensions between the subject and the predicate device. However, these differences do not raise new questions of safety or effectiveness and the testing mentioned below showed that the subject is substantially equivalent with the predicate.
# 7. Non-clinical Testing:
The sponsor provided the following performance testing to support substantial equivalence:
- Biocompatibility per ISO 10993 Cytotoxicity, Sensitization, Vaginal irritation, and ● Rectal irritation
- Shelf-life validation -functional testing ●
- Electrical Safety testing per IEC 60601-1
- Performance Bench Testing .
- On the probes visual testing, electrical output verification testing after o connecting the probe with a compatible host, cable dielectric strength testing, and lead wire plug to instrument connector contact resistance testing
- On the lead wires per ANSI/AAMI EC53 o
The protocol and results of the provided non-clinical testing are acceptable.
{6}------------------------------------------------
# 8. Conclusion:
Based on the comparison and analysis in this submission, it can be concluded that Incontinence Probe is substantially equivalent to the predicate device.