Intended Use

The Sterile Hypodermic Syringe for Single Use with/without needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body.

Device Description

1 ml Sterile Hypodermic Syringe is 3-piece design. These 3 pieces are piston, plunger and barrel. The piston is made from isoprene rubber. Barrel is made from polypropylene. Plunger is made from polypropylene and polyethylene combination. Needle is made from stainless steel 304 grade and hub of needle is made from polypropylene. There are with or without needle configurations within this Special 510k submission. There are luer and luer-lock configurations within this special 510k submission. This special 510k submission only covers the 1ml syringe but several needle dimensions including 21Gx5/8", 23Gx1", and 25Gx1". The barrel marking scale interval is 0.1 ml and legible for all configurations covered by this Special 510k submission. Luer and Luer-lock have a 6% conical design. The barrel is transparent. The length (without needle), outside diameter are 99.30/100.2mm, 6.9/6.75 and 4.70/4.70, respectively for luer and luer-lock configurations. Intended users are medical professionals and trained caregivers. It is prescription use only.

It is EtO sterile and a single use device. Packaging combination is maintained from a transparent sterilization film and Tyvek paper appropriate to EtO sterilization.

The plunger of syringe can be pulled and pushed along inside the barrel, allowing the syringe to take in and expel the fluids through the connector to the patient.

The products covered in this special 510k are in compliance with ISO 10993-1 and ISO 7886-1.

AI/ML Overview

This is a 510(k) premarket notification for a medical device (Sterile Hypodermic Syringe for Single Use) and not a study on an AI/ML device. Therefore, the requested information regarding acceptance criteria, study design for AI/ML, expert adjudication, MRMC studies, standalone performance, ground truth establishment, and training set details are not applicable as they relate to AI/ML device evaluations.

The document indicates that the device's performance was evaluated against recognized international standards. It states that the device complies with:

  • ISO 7886-1: Sterile hypodermic syringes for single use - Part 1: Syringes for manual use
  • ISO 7864: Sterile hypodermic needles for single use — Requirements and test methods
  • ISO 9626: Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
  • ISO 10993-4: Biological Evaluation of Medical Devices — Part 4: Selection of Tests for Interactions with Blood
  • ISO 10993-5: Biological Evaluation of Medical Devices — Part 5: Tests for In Vitro Cytotoxicity
  • ISO 10993-10: Biological Evaluation of Medical Devices — Part 10: Tests for Irritation and Skin Sensitization
  • ISO 10993-11: Biological Evaluation of Medical Devices — Part 11: Tests for Systemic Toxicity

The purpose of this submission is to demonstrate substantial equivalence to a predicate device (K201284) by adding a new size (1ml syringe). The conclusion states that the differences do not raise new questions of safety or effectiveness because the modified devices were tested in accordance with ISO 7886-1.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

Set Medikal Sanayi Ve Ticaret Anonim Sirketi % Mehmet Ormeci Consultant Medcer Uluslararasi Medikal Belgelendirme Anonim Sirketi Taspinar Mahallesi 2800. Caddesi A-2 Apt. No:6 B/49 Ankara. 06830 Turkey

Re: K210200

Trade/Device Name: 1 ml Sterile Hypodermic Syringe For Single Use. With/Without Needle. Luer/Luer-Lock Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, FMF, FMI Dated: April 7, 2021 Received: April 16, 2021

Dear Mehmet Ormeci:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Rumi Young Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K210200

Device Name

Sterile Hypodermic Syringe for Single use, with/without needle; luer/luer-lock

Indications for Use (Describe)

The Sterile Hypodermic Syringe for Single Use with/without needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Ser Medikal. The logo consists of a blue square with a white "S" shape inside of it. To the right of the square is the text "ser medikal" in a sans-serif font.

Special 510(k) Summary

510(k) Submitter NameSET MEDIKAL SANAYI VE TICARET ANONIM SIRKETI
510(k) Submitter AddressOsmangazi Mahallesi Maresal Fevzi Cakmak Caddesi No:18 Esenyurtİstanbul/TURKEY
510(k) Submitter Telephone No+90-212-622-0400
Primary Correspondent Contact InformationMehmet Fatih Ormeci (Consultant)+90-532-397-3043mformeci@gmail.comFax Number: +90 (212) 689 4 828
Summary Preparation Date05/12/2021
Trade orProprietary NameSterile Hypodermic Syringe for Single use, with/without needle; luer/luer-lock
Common NameHypodermic Syringe
Classification NamePiston Syringe
Regulation Number21 CFR 880.5860
Product CodeFMF, FMI
Subject Device Special510k NoPredicate Device 510kNoPredicate DeviceManufacturer
Sterile HypodermicSyringe for Single use,with/without needle;luer/luer-lockK210200/S002K201284SET MEDIKAL SANAYIVE TICARET ANONIMSIRKETI

Indications for Use:

The Sterile Hypodermic Syringe for Single use with/without needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for SER Medikal. The logo consists of a blue square with a white "S" inside of it on the left side of the image. To the right of the square is the text "SER MEDIKAL" in a gray, sans-serif font. The text is slightly stylized, with the letters having a modern, geometric appearance.

Special 510k Summary Piston Syringe: K210200/S002

Syringe SizeGaugeLengthWall TypeApproval Number
1mL215/8"TWK210200 CurrentSubmission
1mL231"TW
1mL251"RW
2mL201 1/2"TW
2mL215/8"TW
2mL221 1/4"TW
2mL231 1/4"TWK201284
2mL251"TW
2mL272"RW
2mL271 1/2"RW
2,5mL211 1/2"TW
2,5mL215/8"TW
2,5mL221 1/2"TW
2,5mL221 1/4"TW
2,5mL231 1/4"TWK201284
2,5mL251"TW
2,5mL251"RW
2,5mL271 1/2"RW
2,5mL272"RW
3mL211 1/2"TW
3mL215/8TW
3mL221 1/4"TW
3mL221 1/2"TWK201284
3mL231 1/4"TW
3mL231"TW
3mL251"RW
5mL181 1/2"TW
5mL201 1/2"TW
5mL211 1/2"TW
5mL215/8"TW
5mL221 1/4"TW
5mL221 1/2"TW
5mL231 1/4"TWK201284
5mL231"TW
5mL251"RW
Syringe SizeGaugeLengthWall TypeApproval Number
10 ml181 1/2"TWK201284
201 1/2"TW
215/8"TW
211 1/2"TW
221 1/4"TW
221 1/2"TW
231"TW
231 1/4"TW
20 ml181 1/2"TWK201284
201 1/2"TW
215/8"TW
211 1/2"TW
221 1/4"TW
221 1/2"TW
231"TW
231 1/4"TW
50 ml151"TWK201284
161"TW
181 1/2"TW
191 1/2"TW
201 1/2"TW
215/8"TW
211 1/2"TW
231 1/4"TW

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for SER Medikal. The logo consists of a blue diamond shape with a white "S" inside of it. To the right of the diamond is the text "SER medikal" in a gray, sans-serif font.

Special 510k Summary Piston Syringe: K210200/S002

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains a logo for "Ser Medikal". The logo consists of a blue square with a stylized "S" inside it, followed by the text "Ser Medikal" in a gray, sans-serif font. The "S" in the square is also stylized and appears to be made up of multiple lines.

Sterile Hypodermic Syringe for Single use, with/without needle; luer/luer-lock
ITEMLegally Marketed Device /Predicate DeviceK201284Modified Device(Adding New Size, 1ml)Comparison
Product nameSterile Hypodermic Syringefor Single use, with/withoutneedle; luer/luer-lockSterile Hypodermic Syringefor Single use, with/withoutneedle; luer/luer-lockSame
Product codeFMFFMFSame
Regulation No.21 CFR 880.586021 CFR 880.5860Same
ClassIIIISame
Indications for useThe Sterile HypodermicSyringe for Single Usewith/without needle isintended to be used formedical purposes to injectfluid into or withdraw fluidfrom body.The Sterile HypodermicSyringe for Single Usewith/without needle isintended to be used formedical purposes to injectfluid into or withdraw fluidfrom body.Same
ConfigurationPistonPlungerBarrelPistonPlungerBarrelSame
Sterility conditionEO SterilizedEO SterilizedSame
Environment of usePrescription usePrescription useSame
Intended usersMedical professionals andtrained care giversMedical professionals andtrained care giversSame
Single useYesYesSame
Operation modeThe plunger of syringe can bepulled and pushed along insidethe barrel, allowing the syringeto take in and expel the fluidsthrough the connector to thepatient.The plunger of syringe canbe pulled and pushed alonginside the barrel, allowingthe syringe to take in andexpel the fluids through theconnector to the patient.Same
Label/labelingComplied with 21 CFR part801Complied with 21 CFR part801Same
ProductperformanceComplied withISO 7886-1ISO 7864ISO 9626Complied withISO 7886-1ISO 7864ISO 9626Same
Volume2ml, 2,5ml, 3ml, 5ml, 6ml, 10ml,12ml, 20ml, 24ml, 50ml1mlDifference 1
Sizes14G,15G,16G,18G,19G,20G,21G,22G,23G,24G,25G,26G,27G,28G,29G,30G21G, 23G, 25GSame
Needle Lengths4-50 mm4-50 mmSame
Sterile Hypodermic Syringe for Single use, with/without needle; luer/luer-lock
ITEMLegally Marketed Device /Predicate DeviceK201284Modified Device(Adding New Size, 1ml)Comparison
Patient contact component and material
PistonIsoprene RubberIsoprene RubberSame
BarrelPPPPSame
PlungerPP/PEPP/PESame
NeedlePP, SS304PP, SUS304Same
Sterilization
SAL10-610-6Same
Endotoxin Limit20 EU per device20 EU per deviceSame

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for SER Medikal. The logo consists of a blue square tilted at a 45-degree angle with a white stylized letter "S" inside. To the right of the square is the text "SER MEDIKAL" in a modern, sans-serif font, with the letters in a light gray color.

Device Description

1 ml Sterile Hypodermic Syringe is 3-piece design. These 3 pieces are piston, plunger and barrel. The piston is made from isoprene rubber. Barrel is made from polypropylene. Plunger is made from polypropylene and polyethylene combination. Needle is made from stainless steel 304 grade and hub of needle is made from polypropylene. There are with or without needle configurations within this Special 510k submission. There are luer and luer-lock configurations within this special 510k submission. This special 510k submission only covers the 1ml syringe but several needle dimensions including 21Gx5/8", 23Gx1", and 25Gx1". The barrel marking scale interval is 0.1 ml and legible for all configurations covered by this Special 510k submission. Luer and Luer-lock have a 6% conical design. The barrel is transparent. The length (without needle), outside diameter are 99.30/100.2mm, 6.9/6.75 and 4.70/4.70, respectively for luer and luer-lock configurations. Intended users are medical professionals and trained caregivers. It is prescription use only.

It is EtO sterile and a single use device. Packaging combination is maintained from a transparent sterilization film and Tyvek paper appropriate to EtO sterilization.

The plunger of syringe can be pulled and pushed along inside the barrel, allowing the syringe to take in and expel the fluids through the connector to the patient.

The products covered in this special 510k are in compliance with ISO 10993-1 and ISO 7886-1.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains a logo for "ser medikal". The logo consists of a blue square with a stylized "S" inside it on the left side. To the right of the square is the text "ser medikal" in a gray, sans-serif font.

Special 510k Summary

Piston Syringe: K210200/S002

Product NameSterile Hypodermic Syringe for Single use, with/without needle; luer/luer lock
Instructions for useThe Sterile Hypodermic Syringe for Single Use with/without needle is intended to be used for medical purposes to inject fluid into or
withdraw fluid from body.
Volume1ml
Barrel Marking Specs (Scale interval)0.1ml
Tip typeLuer
Luer lock
ConfigurationBarrel, plunger
Piston, plunger, barrel
BarrelTransparencyTransparent
UV Resistant-
Opaque-
Needle (Needle configurations are mentioned in thehypodermic needle product description table)With or without needle
Nozzle TypeLuerLuer Lock
Length (without needle)99.30100.2
Diameter, Outside6.906.75
Diameter, Inner4.704.70
Plunger MaterialPE
Barrel MaterialPP
Piston MaterialIsoprene rubber
Nozzle TypeLuer, Luer lock, 6% conical, 1,2mm
Lubricant CompositionSilicone
Sterility ConditionEO sterilized
Single UseYes
Environment of usePrescription Use
Intended UsersMedical professionals and trained care givers
Operation ModeThe plunger of syringe can be pulled and pushed along inside the barrel, allowing the syringe to take in and expel the fluids through the connector to the patient.
Graduation LegibilityLegible
Endotoxin Limit20 EU per device
BiocompatibilityComplied with ISO 10993-4/5/10/11 (Cytotoxicity, Irritation, Sensitization Acute Systemic Toxicity, Hemolytical effect)
SAL10-6
Product PerformanceComplied with ISO 7886-1

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains the logo for "ser medikal". The logo consists of a blue diamond shape with a stylized "S" inside it on the left. To the right of the diamond is the text "ser medikal" in a gray, sans-serif font.

Product NameSterile Hypodermic needle for Single use
Sizes21G23G25G
Needle Lengths$58$ "1"1"
(*): with 15 micron filter
Needle Bond StrengthMin.44NMin.34NMin.22N
Needle Cover Dimensions$8×3$$8×3$$8×31$
Needle Cover StrengthMin. 2N - max. 20N
Needle Cover ColorTransparent
Needle Tip ConfigurationBack bevel needle
ConfigurationProtective cap, Needle, Adhesives, Needle hub
Sterility ConditionEO sterilized
Single UseYes
Environment of usePrescription Use
Intended UsersMedical professionals and trained care givers
Operation ModeThe plunger of syringe can be pulled and pushed along inside the barrel, allowing the syringe to take in and expel the fluids through the connector to the patient.
Graduation LegibilityLegible
Protective Cap MaterialPP
Adhesive MaterialEpoxy resin
Needle Hub MaterialPP
Needle MaterialPP (hub and cover), SUS304 Stainless Steel
Product PerformanceComplied with ISO 7864, ISO 9626
Endotoxin Limit20 EU per device
BiocompatibilityComplied with ISO 10993-4/5/10/11 (Cytotoxicity, Irritation, Sensitization Acute Systemic Toxicity, Hemolytical effect)
SAL10-6

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains the logo for SER Medikal. The logo consists of a blue diamond shape with a white "S" inside of it. To the right of the diamond is the company name, "SER Medikal", in a gray, sans-serif font.

Modification Discussion

In this Special 510(k) submission, a new size: 1ml syringe luer/luer lock with or without needle is covered. The intended use, raw material, manufacturing site, sterilization method, sterilization site, packaging material is same with cleared device K201284.

Performance Testing

The sterile, hypodermic syringe described in this summary were tested and demonstrated to be in

conformance with the following FDA recognized standards:

  • . ISO 9626 Second edition: Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
  • ISO 7886-1 Second edition 2017-05 Sterile hypodermic syringes for single use - Part 1: Syringes for manual use
  • ISO 7864:2016 Fourth Edition: Sterile hypodermic needles for single use — Requirements and test methods
  • . ISO 10993-4:2017 Biological Evaluation of Medical Devices — Part 4: Selection of Tests for Interactions with Blood
  • ISO 10993-5:2009 Biological Evaluation of Medical Devices — Part 5: Tests for In Vitro Cytotoxicity
  • ISO 10993-10:2010 Biological Evaluation of Medical Devices — Part 10: Tests for Irritation and Skin Sensitization
  • . ISO 10993-11:2017 Biological Evaluation of Medical Devices — Part 11: Tests for Systemic Toxicity

The sterile, hypodermic needle described in this summary were tested and demonstrated to be in

conformance with the following FDA recognized standards:

  • . ISO 9626 Second edition: Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
  • ISO 7864:2016 Fourth Edition: Sterile hypodermic needles for single use — Requirements and test methods
  • . ISO 10993-4:2017 Biological Evaluation of Medical Devices — Part 4: Selection of Tests for Interactions with Blood

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image contains the logo for "ser medikal". The logo consists of a blue, stylized letter "S" enclosed in a diamond shape, followed by the text "ser medikal" in a gray, sans-serif font. The logo is simple and modern, with a focus on the company's name.

  • ISO 10993-5:2009 Biological Evaluation of Medical Devices — Part 5: Tests for In Vitro Cytotoxicity
  • ISO 10993-10:2010 Biological Evaluation of Medical Devices — Part 10: Tests for Irritation and Skin Sensitization
  • . ISO 10993-11:2017 Biological Evaluation of Medical Devices — Part 11: Tests for Systemic Toxicity

Conclusion

The differences between the cleared (predicate) and the modified device do not raise any new or different questions of safety or effectiveness since modified devices are tested in accordance with ISO 7886-1 standard. The subject devices are substantially equivalent to the predicate devices with respect to the indications for use, target populations, treatment method, and technological characteristics.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).