K Number
K203680

Validate with FDA (Live)

Date Cleared
2022-07-22

(582 days)

Product Code
Regulation Number
866.3920
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BioFire JI Control Panel M420 is intended for use as an external positive assayed quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of gram-positive and gram-negative bacteria: Anaerococcus prevotii/vaginalis, Clostridium perfringens, Cutibacterium avidum/granulosum, Enterococcus faecalis, Enterococcus faecium, Finezoldia magna, Parvimonas micra, Peptoniphilus, Peptostreptococcus anaerobius, Staphylococcus aureus, Staphylococcus lugdunensis, Streptococcus spp., Streptococcus agalactiae, Streptococus pneumoniae, Streptococcus pyogenes, Bacteroides fragilis, Citrobacter cloacae complex, Escherichia coli, Haemophilus influenzae, Kingella kingae, Klebsiella aerogenes, Klebsiella pneumoniae group, Morganella morganii, Neisseria gonorrhoeae, Proteus spp., Pseudomonas aeruginosa, Salmonella spp., and Serratia marcescens, antimicrobial resistance genes: CTX-M, IMP, KPC, mecA/C and MREJ (MRSA), NDM, OXA-48-like, vanA/B, VIM, and yeast pathogens: Candida albicans on the BioFire JI Panel assay on the BioFire systems. BioFire II Control Panel M420 is composed of synthetic DNA specifically designed for and intended to be used solely with the BioFire II Panel assay. This product is not intended to replace manufacturer controls provided with the device.

Device Description

BioFire JI Control Panel M420, P/N M420, is a quality control panel consisting of 2 controls, BioFire JI Positive Control, P/N M42118, and JI Negative Control, P/N M42218. The Positive Control contains non-infectious surrogate control material; a solution of synthetic DNA in buffers, stabilizers and preservatives. The DNA in the Positive Control carries nucleic acid corresponding to the genome segments of all the pathogens and antimicrobial resistance genes detected and identified by the BioFire® Joint Infection (JI) Panel assay (see Table 1. below) on the BioFire® systems. The Negative Control contains buffers, stabilizers and preservatives. Each liquid control of BioFire JI Control Panel M420 is processed separately according to the BioFire JI Panel assay manufacturer's Instructions for Use for patient samples (synovial fluid) obtained from individuals suspected of infection. Each tube of control contains sufficient liquid for a single use.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the BioFire JI Control Panel M420, a quality control material. The study demonstrates the reproducibility of this control panel.

Here's the breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Overall Correct Result Rate100%
Correct Positive Control Result Rate100%
Correct Negative Control Result Rate100%

2. Sample Size and Data Provenance

  • Test Set Sample Size:
    • Combined Reproducibility Study: 301 tests (1 Invalid result was repeated and not included in correct analysis, leading to 300 correct results analyzed).
    • External Site Testing: 181 tests (1 Invalid result was not included in percent correct analysis, leading to 180 valid results analyzed).
    • Internal Site Testing (MMQCI): 120 tests.
  • Data Provenance: The study involved both internal (MMQCI, Saco, Maine, USA) and external sites (3 CLIA-certified clinical sites, likely within the US given the FDA submission context). The data is prospective, generated specifically for this performance evaluation.

3. Number of Experts and Qualifications for Ground Truth

The document does not mention the use of experts to establish a ground truth for the test set. Instead, the performance of the control panel is evaluated against its expected positive or negative detection for each target, as designed by the manufacturer (Maine Molecular Quality Controls, Inc.). The BioFire JI Control Panel M420 itself is intended to be an external control for an in vitro laboratory nucleic acid testing procedure (BioFire JI Panel assay).

4. Adjudication Method for the Test Set

No adjudication method is described for the test set. The results are straightforward positive or negative detections as determined by the BioFire JI Panel assay. "Correct" results are based on whether the assay detected what was designed to be in the positive control or did not detect anything in the negative control.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study being performed, nor any effect size of human readers improving with or without AI assistance. This study focuses on the performance of a quality control material and not on human interpretation or AI assistance.

6. Standalone (Algorithm Only) Performance

This study is inherently a standalone performance evaluation of the BioFire JI Control Panel M420 in combination with the BioFire JI Panel assay. The control panel itself is a diagnostic device (quality control material), not an algorithm. The performance described is the control material's ability to consistently elicit the expected results from the BioFire JI Panel assay.

7. Type of Ground Truth Used

The ground truth used is the expected composition of the BioFire JI Control Panel M420. The Positive Control is "composed of synthetic DNA specifically designed for" and contains "nucleic acid corresponding to the genome segments of all the pathogens and antimicrobial resistance genes detected and identified by the BioFire® Joint Infection (JI) Panel assay." The Negative Control is designed to contain "buffers, stabilizers and preservatives" but no target nucleic acid. The "correct" result for each test is predefined based on this engineered composition.

8. Sample Size for the Training Set

This document describes the performance evaluation of a quality control material and does not mention a "training set" in the context of an algorithm or machine learning model.

9. How the Ground Truth for the Training Set Was Established

As noted above, there is no "training set" in this context. The control panel is a manufactured product with a defined composition that serves as its inherent ground truth for validation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".

Maine Molecular Quality Controls, Inc. Joan Gordon President 23 Mill Brook Road Saco, Maine 04072

July 22, 2022

Re: K203680

Trade/Device Name: BioFire JI Control Panel M420 Regulation Number: 21 CFR 866.3920 Regulation Name: Assayed Quality Control Material For Clinical Microbiology Assays Regulatory Class: Class II Product Code: PMN Dated: December 15, 2020 Received: June 14, 2022

Dear Joan Gordon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Noel J. Gerald, Ph.D. Branch Chief Bacterial Respiratory and Medical Countermeasures Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203680

Device Name BioFire JI Control Panel M420

Indications for Use (Describe)

BioFire JI Control Panel M420 is intended for use as an external positive assayed quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of gram-positive and gram-negative bacteria: Anaerococcus prevotii/vaginalis, Clostridium perfringens, Cutibacterium avidum/granulosum, Enterococcus faecalis, Enterococcus faecium, Finezoldia magna, Parvimonas micra, Peptoniphilus, Peptostreptococcus anaerobius, Staphylococcus aureus, Staphylococcus lugdunensis, Streptococcus spp., Streptococcus agalactiae, Streptococus pneumoniae, Streptococcus pyogenes, Bacteroides fragilis, Citrobacter cloacae complex, Escherichia coli, Haemophilus influenzae, Kingella kingae, Klebsiella aerogenes, Klebsiella pneumoniae group, Morganella morganii, Neisseria gonorrhoeae, Proteus spp., Pseudomonas aeruginosa, Salmonella spp., and Serratia marcescens, antimicrobial resistance genes: CTX-M, IMP, KPC, mecA/C and MREJ (MRSA), NDM, OXA-48-like, vanA/B, VIM, and yeast pathogens: Candida albicans on the BioFire JI Panel assay on the BioFire systems. BioFire II Control Panel M420 is composed of synthetic DNA specifically designed for and intended to be used solely with the BioFire II Panel assay. This product is not intended to replace manufacturer controls provided with the device.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo and contact information for Maine Molecular Quality Controls, Inc. The logo consists of the letters "MMQCI" in a stylized font, with the words "MAINE MOLECULAR QUALITY CONTROLS, INC." written below. The contact information includes the address "23 Mill Brook Road, Saco, Maine 04072", the website "www.mmqci.com", the phone number "207-885-1072", and the fax number "207-885-1079".

Section 5: 510 (k) Summary

510(k) Number:

Purpose for submission: New product

Applicant Information:

Applicant:Maine Molecular Quality Controls, Inc
Address:23 Mill Brook Road
Saco, Maine 04072
Contact Person:Joan Gordon, President MMQCI
Phone:207-885-1072 extension 201
Fax:207-885-1079
Email Address:jgordon@mmqci.com
Preparation Date:December 15, 2020

Device

Device Trade Name:BioFire JI Control Panel M420, P/N M420
Device Common Name:Quality Control Material for Microbiology Assays
Device Type:Assayed quality control material for clinical microbiology
assays
Class:Class II (Special controls)
Regulation:21 CFR 866.3920
Panel:Microbiology - 83
Product code:PMN

Predicate Device

K200010: FilmArray BCID2 Control Panel M416. P/N M416. Maine Molecular Quality Controls, Inc.

Device Description

BioFire JI Control Panel M420, P/N M420, is a quality control panel consisting of 2 controls, BioFire JI Positive Control, P/N M42118, and JI Negative Control, P/N M42218. The Positive Control contains non-infectious surrogate control material; a solution of synthetic DNA in buffers, stabilizers and preservatives. The DNA in the Positive Control carries nucleic acid corresponding to the genome segments of all the pathogens and antimicrobial resistance genes detected and identified by the BioFire® Joint Infection (JI) Panel assay (see Table 1. below) on the BioFire® systems. The Negative Control contains buffers, stabilizers and preservatives. Each liquid control of BioFire JI Control Panel M420 is processed separately according to the BioFire JI Panel assay manufacturer's Instructions for Use for patient samples (synovial fluid) obtained from individuals suspected of infection. Each tube of control contains sufficient liquid for a single use.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the contact information for MMQCI, or Maine Molecular Quality Controls, Inc. The address is 23 Mill Brook Road, Saco, Maine 04072. The website is www.mmqci.com, the phone number is 207-885-1072, and the fax number is 207-885-1079.

Table 1. Pathogens and antimicrobial resistance genes found in BioFire JI Control Panel M420, detected by BioFire JI Panel assay.

Positive ControlNegative Control
Gram Positive Bacteria
Anaerococcus prevotii/vaginalisDetectedNot Detected
Clostridium perfringensDetectedNot Detected
Cutibacterium avidum/granulosumDetectedNot Detected
Enterococcus faecalisDetectedNot Detected
Enterococcus faeciumDetectedNot Detected
Finegoldia magnaDetectedNot Detected
Parvimonas micraDetectedNot Detected
PeptoniphilusDetectedNot Detected
Peptostreptococcus anaerobiusDetectedNot Detected
Staphylococcus aureusDetectedNot Detected
Staphylococcus lugdunensisDetectedNot Detected
Streptococcus spp.DetectedNot Detected
Streptococcus agalactiaeDetectedNot Detected
Streptococcus pneumoniaeDetectedNot Detected
Streptococcus pyogenesDetectedNot Detected
Gram Negative Bacteria
Bacteroides fragilisDetectedNot Detected
CitrobacterDetectedNot Detected
Enterobacter cloacae complexDetectedNot Detected
Escherichia coliDetectedNot Detected
Haemophilus influenzaeDetectedNot Detected
Kingella kingaeDetectedNot Detected
Klebsiella aerogenesDetectedNot Detected
Klebsiella pneumoniae groupDetectedNot Detected
Morganella morganiiDetectedNot Detected
Neisseria gonorrhoeaeDetectedNot Detected
Proteus spp.DetectedNot Detected
Pseudomonas aeruginosaDetectedNot Detected
Salmonella spp.DetectedNot Detected
Serratia marcescensDetectedNot Detected
Antimicrobial resistance genes
CTX-MDetectedN/A
IMPDetectedN/A
KPCDetectedN/A
mecA/C and MREJ (MRSA)DetectedN/A
NDMDetectedN/A
OXA-48-likeDetectedN/A
vanA/BDetectedN/A
VIMDetectedN/A
Yeast
CandidaDetectedNot Detected
Candida albicansDetectedNot Detected

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo and contact information for Maine Molecular Quality Controls, Inc. The address is 23 Mill Brook Road, Saco, Maine 04072. The website is www.mmqci.com, the phone number is 207-885-1072, and the fax number is 207-885-1079.

Device Intended Use

BioFire JI Control Panel M420 is intended for use as an external positive and negative assayed quality control to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Gram positive and Gram negative bacteria: Anaerococcus prevotii/vaginalis, Clostridium perfringens, Cutibacterium avidum/granulosum, Enterococcus faecalis, Enterococcus faecium, Finegoldia magna, Parvimonas micra, Peptoniphilus, Peptostreptococcus anaerobius, Staphylococcus aureus, Staphylococcus lugdunensis, Streptococcus spp., Streptococcus agalactiae, Streptococcus pneumoniae, Streptococcus pyogenes, Bacteroides fragilis, Citrobacter, Enterobacter cloacae complex, Escherichia coli, Haemophilus influenzae, Kingella kingae, Klebsiella aerogenes, Klebsiella pneumoniae group, Morganella morganii, Neisseria gonorrhoeae, Proteus spp., Pseudomonas aeruginosa, Salmonella spp., and Serratia marcescens, antimicrobial resistance genes: CTX-M, IMP, KPC, mecA/C and MREJ (MRSA), NDM, OXA-48-like, vanA/B, VIM, and yeast pathogens: Candida and Candida albicans on the BioFire JI Panel assay on the BioFire systems. BioFire II Control Panel M420 is composed of synthetic DNA specifically designed for and intended to be used solely with the BioFire JI Panel assay. This product is not intended to replace manufacturer controls provided with the device.

Substantial Equivalence

Maine Molecular Quality Controls, Inc. (MMQCI) proposes that BioFire JI Control Panel M420 is substantially equivalent to currently marketed medical device FilmArray BCID2 Control Panel M416 (K200010). BioFire JI Control Panel M420 is manufactured using the same processes and formulation as for MMQCI's quality control, FilmArray BCID2 Control Panel M416.

CharacteristicCandidate Device: BioFireJI Control Panel M420Predicate Device:FilmArray BCID2 ControlPanel M416 (K200010)
Intended UseExternal assayed qualitycontrol to monitor in vitrolab nucleic acid testSame
Physical formatReady-to-Use LiquidSame
Directions for UseProcess like patient sample(Pipette from synovial fluid)Process like patient sample(Pipette from blood culture)
CompositionSynthetic DNASame
Assay Steps MonitoredAmplification, detection,identificationSame
Number of targetsmonitored in one assayMultiple, >30 targetsSame, >30 targets

{6}------------------------------------------------

Image /page/6/Picture/0 description: This image contains the contact information for Maine Molecular Quality Controls, Inc. The address is listed as 23 Mill Brook Road, Saco, Maine 04072. The phone number is 207-885-1072, and the fax number is 207-885-1079. The website is www.mmqci.com.

Summary Performance Data

Three lots of BioFire JI Control M420, consisting of BioFire JI Positive Control and BioFire JI Negative Control, were manufactured by MMQCI. The lots were manufactured and tested such that routine variables including multiple lots of key manufacturing components, different operators, pouch lots, instruments, test sites, and testing over time were incorporated to challenge performance. External and internal studies were performed by testing the 3 lots with the BioFire® JI Panel assay on BioFire FilmArray Systems. The external performance study was performed by testing 10 samples of each of the 3 lots of BioFire JI Control Panel M420 on different days at 3 CLIA-certified clinical sites using 3 FilmArray pouch lots, incorporating multiple operators for a total of 181 tests (1 Invalid result was repeated). The internal reproducibility study was performed by testing 20 samples of each of the 3 lots at MMQCI (Saco, Maine) on different days with 3 pouch lots by 3 operators for a total of 120 tests.

Results and Conclusions:

Of the total 301 tests performed for the Reproducibility study, there were 300 correct results, positive and negative as listed in Table 2, and 1 Invalid result. The sample that gave an Invalid result was re-tested according to BioFire instructions and was not included in the Percent Correct analysis in the table below.

Reproducibility of BioFire JI Control Panel M420 at 4 sites, across 3 pouch lots, 3 control lots, on multiple instruments by multiple operators is acceptable according to predetermined criteria with an overall correct result rate of 100%. Test results demonstrate robust performance across 4 testing sites.

Summary of Reproducibility Test Results: MMQCI and 3 Clinical Laboratory Sites
TotalTestsInvalid*CorrectPositiveControlResultIncorrectPositiveControlResultPercentCorrectPositiveControlCorrectNegativeControlResultIncorrectNegativeControlResultPercentCorrectNegativeControlTotalPercentCorrect
30111500100%1500100%100%

Table 2. BioFire JI Control Panel M420: Reproducibility results, all sites.

*The Invalid sample was re-tested according to BioFire instructions and was not included in the Percent Correct analysis.

External Site Testing Details

Of the total 301 reproducibility tests. 181 BioFire JI Control Panel M420 samples were tested at 3 clinical sites. Three lots of BioFire JI Positive Control (D19DEC19A, D23DEC19A, E27DEC19A) and 3 lots of BioFire JI Negative Control (C28NOV18F, A14JUN19G, F29OCT19D) were tested on 3 BioFire JI Panel pouch lots across the 3 sites, incorporating multiple operators and instruments. The laboratories were instructed to run 3 Positive and 3 Negative controls, for each lot, per day on 10 different days. Valid results were obtained for 180 controls. Of the 181 controls tested, 1 gave an Invalid result and was retested according to BioFire Instructions for Use. Correct results were obtained for the retest. The Invalid result was not included in the Percent Correct analysis in the table below.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the contact information for MMQCI, or Maine Molecular Quality Controls, Inc. The address is 23 Mill Brook Road, Saco, Maine 04072. The website is www.mmqci.com, the phone number is 207-885-1072, and the fax number is 207-885-1079.

Results and Conclusion:

Out of 180 tests with valid pouch controls, the correct analytes were detected (for positive controls) or not detected (for negative controls) in all 180 tests, for an overall success rate of 100% (Table 3). Predetermined acceptance criteria for reproducibility were met. BioFire JI Control Panel M420 performed robustly at the 3 external test sites across 3 pouch lots on 19 BioFire FilmArray 2.0 and Torch instruments/ modules, incorporating multiple operators.

Table 3. BioFire JI Control Panel M420: Summary of Reproducibility Results for 3 Control Lots at 3 Clinical Sites

Summary of External Results for 3 Control Lots of BioFire JI Control Panel M420
SiteTotalTestsInvalidCorrectPositiveControlResultIncorrectPositiveControlResultPercentCorrectPositiveControlCorrectNegativeControlResultIncorrectNegativeControlResultPercentCorrectNegativeControlTotalPercentCorrect
1611*300100%300100%100%
2600300100%300100%100%
3600300100%300100%100%
AllSites1811*900100%900100%100%

*Invalid result was not included in percent correct.

Internal Site Testing Details

Run-to-run Precision Testing at MMOCI: Of the total 301 reproducibility tests, 120 BioFire II Control Panel M420 samples were tested over February and March of 2020 at MMQCI's facility. Each control was tested 20 times on 20 different days. Three lots of Positive Control (D19DEC19A, D23DEC19A, E27DEC19A) and 3 lots of Negative Control (C28NOV18F, A14JUN19G, F29OCT19D) were tested on 3 BioFire JI pouch lots, incorporating 3 operators and 2 FilmArray 2.0 instruments (Table 4).

Results and conclusions:

All Controls gave correct results for an overall correct result rate of 100%. Predetermined acceptance criteria for reproducibility were met. BioFire JI Control Panel M420 performed robustly on 20 different days at MMQCI across 3 pouch lots on 2 BioFire FilmArray 2.0 instruments, incorporating 3 operators.

Summary of Precision Testing for 3 JI Control Lots at MMQCI
ControlControl Lot #No. ofTestsInvalidCorrectResultsIncorrectResultsPercentCorrect
BioFire JI Negative ControlC28NOV18F200200100%
BioFire JI Negative ControlA14JUN19G200200100%
BioFire JI Negative ControlF29OCT19D200200100%
BioFire JI Positive ControlD19DEC19A200200100%
BioFire JI Positive ControlD23DEC19A200200100%
BioFire JI Positive ControlE27DEC19A200200100%
TOTAL12001200100%

Table 4. FilmArray BCID2 Control Panel M416: Reproducibility results at MMOCI.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the contact information for MMQCI (Maine Molecular Quality Controls, Inc.). The address is 23 Mill Brook Road, Saco, Maine 04072. The website is www.mmqci.com, the phone number is 207-885-1072, and the fax number is 207-885-1079.

Matrix Effects

Human Synovial fluid used for the study: Single Donor Human Synovial Fluid, Part Number IRHUSYNS1ML, Company: Innovative Research

The matrix of the BioFire JI Control Panel M420 is synthetic, which allows for reproducible manufacturing and stability, provides a non-infectious material for the laboratory staff and carries the control DNA through the extraction process. Since the matrix is not identical to that of the routine JI assay sample, synovial fluid, a test was performed to investigate the effect of the matrix on the assay. Equal volumes of the same concentration of gDNA Streptococcus pneumoniae, strain TCH8431 (BEI Resources, P/N HM-145D) were spiked into 270uL of BioFire JI Control Panel M420 matrix and 270uL of contrived patient matrix (Human Synovial Fluid; Innovative Research), then tested on the BioFire JI Panel assay in triplicate. Criteria to be met are correct calls for the spiked simulated clinical sample and spiked matrix.

Results and Conclusions:

The same expected calls were obtained for the spiked human synovial fluid (simulated clinical sample) and spiked JI matrix (Detected for Streptococcus spp. and S. pneumoniae, Not Detected for all other assays)

As indicated in the summary data table below (Table 11), the BioFire JI Control Panel M420 matrix has no effect on the assay.

PathogenAnalyteSpiked BEI S.pneumoniae Crossing Point (Cp)
MMQCI JI MatrixMean CpHuman Synovial Fluid (SF)Mean Cp
Streptococcus10.810.412.711.311.411.811.711.6
Spneumoniae9.39.110.39.610.110.310.210.2

Table 5. Matrix Testing for BioFire JI Control Panel M420 with Human Synovial fluid

§ 866.3920 Assayed quality control material for clinical microbiology assays.

(a)
Identification. An assayed quality control material for clinical microbiology assays is a device indicated for use in a test system to estimate test precision or to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. This type of device consists of single or multiple microbiological analytes intended for use with either qualitative or quantitative assays.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate:
(i) Analyte concentration;
(ii) Expected values;
(iii) Analyte source;
(iv) Base matrix;
(v) Added components;
(vi) Safety and handling information; and
(vii) Detailed instructions for use.
(2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including:
(i) Description of the process for value assignment and validation.
(ii) Description of the protocol(s) used to establish stability.
(iii) Line data establishing precision/reproducibility.
(iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance.
(v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method.
(vi) Where applicable, detailed documentation related to studies for surrogate controls.
(3) Premarket notification submissions must include an adequate mitigation (e.g., real-time stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling.
(4) Your 21 CFR 809.10 compliant labeling must include the following:
(i) The intended use of your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following:
(A) Assayed control material analyte(s);
(B) Whether the material is intended for quantitative or qualitative assays;
(C) Stating if the material is a surrogate control; and
(D) The system(s), instrument(s), or test(s) for which the quality control material is intended.
(ii) The intended use in your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following statement: “This product is not intended to replace manufacturer controls provided with the device.”
(iii) A limiting statement that reads “Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.”