Disposable Medical Surgical Face Mask
K203200 · Guangdong Haiou Medical Apparatus Co., Ltd. · FXX · Jun 16, 2021 · General, Plastic Surgery
Device Facts
| Record ID | K203200 |
| Device Name | Disposable Medical Surgical Face Mask |
| Applicant | Guangdong Haiou Medical Apparatus Co., Ltd. |
| Product Code | FXX · General, Plastic Surgery |
| Decision Date | Jun 16, 2021 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4040 |
| Device Class | Class 2 |
Intended Use
Disposable Medical Surgical Face Mask is indicated as a protective nose and mouth covering for healthcare workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of exposure to microorganisms and body fluids.
Device Story
Disposable, single-use, three-layer, flat-folded surgical face mask. Inner/outer layers: spun-bond polypropylene; middle layer: melt-blown polypropylene filter. Features malleable polyethylene/aluminum nose piece and elastic ear loops (latex-free). Used in clinical settings by healthcare workers and patients to provide a physical barrier against microorganisms and body fluids. Device is non-sterile. Performance meets ASTM F2100 Level 3 standards, providing protection against fluid penetration, bacterial/particulate filtration, and flammability risks.
Clinical Evidence
Bench testing only. Testing included fluid resistance (ASTM F1862, 160 mm Hg), bacterial filtration efficiency (ASTM F2100-19, ≥98%), differential pressure (ASTM F2100-19, <6.0 mm H2O/cm2), particulate filtration efficiency (ASTM F2100-19, ≥98%), and flammability (21 CFR 1610, Class 1). Biocompatibility testing per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization) confirmed the device is non-cytotoxic, non-irritating, and non-sensitizing.
Technological Characteristics
Three-layer construction: spun-bond polypropylene (inner/outer), melt-blown polypropylene (filter). Malleable polyethylene/aluminum nose piece. Elastic polyester ear loops. Dimensions: 17.5cm x 9.5cm. Pleated design. Non-sterile. ASTM F2100 Level 3 performance. Biocompatibility per ISO 10993-5 and ISO 10993-10.
Indications for Use
Indicated for healthcare workers and patients as a protective nose and mouth covering during medical and surgical procedures where there is a risk of exposure to microorganisms and body fluids.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
- ASEPT® Surgical Face Mask (K201137)
Related Devices
- K212097 — Disposable medical surgical mask · Hebei Titans Hongsen Medical Technology Co., Ltd. · Aug 18, 2021
- K211105 — Ear-Friendly Mask · Ray Co., Ltd. · May 17, 2021
- K213136 — CPL Surgical Face Mask · Control Print Limited · Dec 24, 2021
- K222651 — Surgical Face Mask · Xiantao Sanda Industrial Co., Ltd. · Dec 13, 2022
- K211042 — Disposable Procedural Ear-loop Face Mask · Changzhou Dsb Medical Co., Ltd. · Jul 15, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol representing the Department of Health & Human Services on the left, and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue on the right. The text is arranged in three lines, with "FDA" in a larger font size and a blue square behind it.
June 16, 2021
Guangdong Haiou Medical Apparatus Co., Ltd. % Salon Chen System Engineer IMD Medical & Drug technology service institutions Tianbao office room 225,Sha Tai Road No.209 ShenZhen city, Guangdong Province 518117 China
Re: K203200
Trade/Device Name: Disposable Medical Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: May 13, 2021 Received: May 13, 2021
Dear Salon Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203200
Device Name
Disposable Medical Surgical Face Mask
### Indications for Use (Describe)
Disposable Medical Surgical Face Mask is indicated as a protective nose and mouth covering for healthcare workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of exposure to microorganisms and body fluids.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
## 510(k) Summary
### 1. Submitter's Information
- > Company Name: Guangdong Haiou Medical Apparatus Co., Ltd.
- > Establishment Registration Number: 3010596027
- > Address:Nanyuan Industrial Zone, North Liusha, Puning City, Guangdong, China
- > Phone:+86-663-2758666
- > Fax: +86-663-2905999
- Contact Person(Title):Zhang Zhuo Xuan(General Manager) >
- ア E-mail: hoyl@haiou.net.cn
- Device Information 2.
- > Trade Name: Disposable Medical Surgical Face Mask
- > Common Name: Mask,Surgical
- > Classification Name: Surgical Apparel
- > Model: HO-KZ01
- > 510(k) Number: K203200
- Classification 3.
- > Classification Product Code: FXX
- > Regulation Number: 21 CFR 878.4040
- > Classification: Class II
- > Review Panel: General Hospital
- Predicate Device Information: 4.
{4}------------------------------------------------
- 510(k) Number: K201137 >
- > Predicate Device Name: ASEPT® Surgical Face Mask
- > Manufacturer: PFM Medical, Inc.
- > This predicate has not been subject to a design-related recall
- > No reference devices were used in this submission.
#### 5. Application Correspondent
- > Company Name: IMD Medical & Drug technology service institutions
- > Phone: +86-18613190779
- > Fax: +86-755-62809168
- > Contact Person(Title): Salon Chen (System engineer)
- > E-mail: 33999439@qq.com
#### 6. Device Description
Disposable Medical Surgical Face Mask is a single-use, three layer, flat-folded mask with ear loops and nose piece. The inner and outer layers are constructed of spun-bond polypropylene and the middle layer is constructed of melt blown polypropylene filter. The mask is held in place over the mouth and nose by two elastic loops welded to the facemask. The elastic loops are not made with natural rubber latex. The nose piece is made of malleable polyethylene with aluminum wire and allows the user to fit the facemask around their nose. Disposable Medical Surgical Face Mask is sold non-sterile and is intended to be a single use, disposable device.
### 7. Indications for Use Statement
Disposable Medical Surgical Face Mask is indicated as a protective nose and mouth covering for healthcare workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of exposure to microorganisms and body fluids.
#### Comparison to the Predicate Device 8.
{5}------------------------------------------------
| Elements of Comparison | Proposed Device | Predicate Device | Judgment |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| Company Name | Guangdong Haiou Medical<br>Apparatus Co., Ltd.. | PFM Medical, Inc. | / |
| Device Name | Disposable Medical<br>Surgical Face Mask | ASEPT® Surgical Face Mask | / |
| 510(k) Number | K203200 | K201137 | / |
| Classification Product Code | FXX | FXX | Same |
| Regulation | 21 CFR 878.4040 | 21 CFR 878.4040 | Same |
| Classification Name | Surgical Apparel | Surgical Apparel | Same |
| Class | 2 | 2 | Same |
| Prescription or OTC | OTC | OTC | Same |
| Intended Use & Indications for Use | Disposable Medical<br>Surgical Face Mask is<br>indicated as a protective<br>nose and mouth covering<br>for healthcare workers and<br>patients involved in<br>medical and surgical<br>procedures. The masks<br>are indicated in any<br>procedure or situation<br>where there is a risk of<br>exposure to<br>microorganisms and body<br>fluids. | The ASEPT® Surgical<br>Face Mask is indicated as<br>a protective nose and<br>mouth covering for<br>healthcare workers and<br>patients involved in<br>medical and surgical<br>procedures. The masks<br>are indicated in any<br>procedure or situation<br>where there is a risk of<br>exposure to<br>microorganisms and body<br>fluids. | Same |
{6}------------------------------------------------
| | | Inner and Outer<br>Layers | Spun-bond polypropylene | Spun-bond polypropylene | |
|-------------------------------------------------|------------------------------------|---------------------------|------------------------------------------------------------|------------------------------------------------------------|----------------|
| Materials | Middle Layer | | Melt blown polypropylene<br>filter | Melt blown polypropylene<br>filter | Same |
| | Ear Loops | | Polyester | Polyester | |
| | Nose Piece | | Malleable polyethylene<br>with aluminum wire | Malleable polyethylene<br>with aluminum wire | |
| | | | | | |
| Dimension | | | 17.5cm length x 9.5cm<br>height | 17.5cm length x 9.5cm<br>height | Same |
| Mask Style | | | Pleated | Pleated | Same |
| Design Features | | | Malleable nosepiece, flat<br>pleated,<br>elastic ear loops | Malleable nosepiece, flat<br>pleated,<br>elastic ear loops | Same |
| Sterility | | | Non-sterile | Non-sterile | Same |
| Use | | | Single Use,Disposable | Single Use,Disposable | Same |
| Color | | | Blue and White | Blue and White | Same |
| ASTM F2100 level | | | Level 3 | Level 2 | Different<br>1 |
| Resistance to Penetration by<br>Synthetic Blood | | | Level 3: 160 mm Hg | Level 2: 120 mm Hg | |
| Differential Pressure | | | Level 3: < 6.0 | Level 2: <6.0 | Same |
| Flammability | | | Class 1 | Class 1 | Same |
| Bacterial Filtration Efficiency<br>(BFE) | | | Level 3: ≥ 98 | Level 2: ≥ 98 | Same |
| Sub-Micron Particulate Filtration | | | Level 3: ≥ 98 | Level 2: ≥ 98 | Same |
| Biocompatibility | Cytotoxicity,ISO<br>10993-5:2009 | Non-cytotoxic | Non-cytotoxic | Same | |
| | Irritation,ISO<br>10993-10:2002 | Non-irritating | Non-irritating | | |
| | Sensitization,ISO<br>10993010:2002 | Non-sensitizing | Non-sensitizing | | |
{7}------------------------------------------------
### Analysis:
Different 1:According to ASTM F2100 standard and test result, Haiou's Disposable Medical Surgical Face Mask ratings is level 3, the level 3 performance is the better than the level 2 performance in ASTM F2100 standard,thus,Haiou's Disposable Medical Surgical Face Mask is safe.
## 9. Non-Clinical Test Conclusion
### Non-clinical data:
Per FDA document Guidance for Industry and FDA Staff: Surgical Masks – Premarket Notification [510(k)] Submissions, the below testing has been completed on the subject device:
| Item | Standard | Acceptance Criteria | Results |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------|--------------------------|--------------------------------------------------------------|
| Fluid Resistance<br>Performance | ASTM F1862:<br>Standard Test Method<br>for Resistance of<br>Medical Face Masks to<br>Penetration by<br>Synthetic Blood | Level 3:160 mm Hg | All samples met the<br>predetermined<br>acceptance criteria. |
| Bacterial Filtration<br>Efficiency | ASTM F2100-19:<br>Standard Specification<br>for Performance of<br>Materials Used in<br>Medical Face Masks | $\ge$ 98% | All samples met the<br>predetermined<br>acceptance criteria. |
| Differential Pressure | ASTM F2100-19: | < 6.0 mm H2 O/cm2 | All samples met the |
| | | | |
| (Delta P) | Standard Specification | | predetermined |
| | for Performance of | | acceptance criteria. |
| | Materials Used in | | |
| | Medical Face Masks | | |
| | ASTM F2100-19: | | All samples met the |
| Particulate Filtration | Standard Specification | > 98% | predetermined |
| Efficiency | for Performance of | | acceptance criteria. |
| | Materials Used in | | |
| | Medical Face Masks | | |
| | | | All samples met the |
| Flammability | 21 CFR 1610 | Class I, Does not Ignite | predetermined |
| | | | acceptance criteria. |
| | | | |
| | | | |
| | ISO 10993-5: | | |
| | Biological evaluation | | All samples met the |
| Cytotoxicity | of medical devices | Non-cytotoxic | predetermined |
| | — Part 5: Tests | | acceptance criteria. |
| | for in vitro | | |
| | cytotoxicity | | |
| | | | |
| | ISO 10993-10: | | All samples met the |
| Irritation | Biological | Non-irritating | |
| | evaluation of medical | | predetermined |
| | | | acceptance criteria. |
| | devices — | | |
| | Part 10: Tests for | | |
| | irritation and<br>skin sensitization | | |
| Sensitization | ISO 10993-10:<br>Biological<br>evaluation of medical devices —<br>Part 10: Tests for<br>irritation and<br>skin sensitization | Non-sensitizing | All samples met the<br>predetermined<br>acceptance criteria. |
{8}------------------------------------------------
{9}------------------------------------------------
# 10. Conclusion
The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(k) Disposable Medical Surgical Face Mask is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K201137.