Upper Arm Electronic Blood Pressure Monitor

K202372 · Conmo Electronic Company Limited · DXN · Dec 30, 2020 · Cardiovascular

Device Facts

Record IDK202372
Device NameUpper Arm Electronic Blood Pressure Monitor
ApplicantConmo Electronic Company Limited
Product CodeDXN · Cardiovascular
Decision DateDec 30, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1130
Device ClassClass 2

Intended Use

Upper Arm Electronic blood pressure monitor is intended to measure the blood pressure and pulse rate of adult at household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia).

Device Story

Oscillometric blood pressure monitor; uses occluding cuff, air pump, and pressure sensor. Cuff inflates above systolic pressure, then slowly deflates; device detects pressure pulsations caused by heart ventricle contractions. Pulsations increase in amplitude until mean arterial pressure (MAP) is reached, then decrease. Device calculates systolic/diastolic pressure and pulse rate; displays results on LCD. Used in household or medical center settings by adults. Provides digital readout of blood pressure and pulse; includes irregular heartbeat, body movement, and cuff wrapping detection features. Helps users monitor cardiovascular health.

Clinical Evidence

Comparative clinical study conducted per ISO 81060-2:2013 comparing subject device, predicate device, and mercury sphygmomanometer. Results demonstrate significant correlation in performance between subject and predicate devices. Additional non-clinical bench testing verified compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 80601-2-30.

Technological Characteristics

Oscillometric measurement; automatic inflation/deflation. Powered by 4x1.5V batteries. LCD display. Biocompatible materials (ISO 10993-1, -5, -10). Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home use), IEC 80601-2-30 (performance).

Indications for Use

Indicated for blood pressure and pulse rate measurement in adults in household or medical settings. Contraindicated for neonates, pregnant patients, or patients with pre-eclampsia.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. December 30, 2020 Conmo Electronic Company Limited % Charlie Mack Principal Engineer International Regulatory Consultants 2950 E Lindrick Drive Chandler, Arizona 85249 Re: K202372 Trade/Device Name: Upper Arm Electronic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: November 23, 2020 Received: December 1, 2020 Dear Charlie Mack: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for LT Stephen Browning Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K202372 Device Name Upper Arm Electronic Blood Pressure Monitor, BM100-EN Indications for Use (Describe) Upper Arm Electronic blood pressure monitor is intended to measure the blood pressure and pulse rate of adult at household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia). | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the word "CONMO" in blue letters. The word is surrounded by two curved lines, one above and one below, that form an oval shape. The background is white. #### 510(k) Summary This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92. ## Date: August 13, 2020 - 1. Company and Correspondent submitting this 510(k): Name – Conmo Electronic Company Limited Address – Room 1402, Building No.10 Shenzhen Biomedicine Innovations Industrial Park Jinhui Road No.14, Kengzi Street Pingshan District Shenzhen City, Guangdong, China 518126 Tel: +86-755-28397469 Fax: +86-755-82922769 Contact - Mrs. Alice Qi General Manager US Agent and Correspondent Mr. Charles Mack Principal Engineer IRC USA 2950 E Lindrick Drive, Chandler, Arizona 85249 USA Tel: 931-6254938 Email: charliemack@irc-us.com - 2. Device: | Trade/proprietary name: | Upper Arm Electronic Blood Pressure Monitor<br>Model BM100-EN | | |-------------------------|---------------------------------------------------------------|--| | Common Name: | Non-invasive blood pressure measurement system | | | Classification Name: | System, measurement, blood-pressure, non-invasive | | | Product Code: | DXN | | | Regulation Number: | 21CFR870.1130 | | | Device Class: | 2 | | {4}------------------------------------------------ - 3. Predicate Device : | Manufacturer | Predicate Device | 510(k)<br>Number | |-----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|------------------| | Jiangsu Yuyue<br>Medical<br>Equipment&<br>Supply Co., Ltd | Upper Arm Type<br>Electronic Blood<br>Pressure Monitor<br>Series, Electronic<br>Blood Pressure<br>Monitor: YE670D | K170605 | ## 4. Reason for Submission: New Device ## 5. Device Description: The Upper Arm Electronic Blood Pressure Monitor, Model BM100-EN operation is based on the oscillometric method. This method takes advantage of the pressure pulsations taken during measurements. An occluding cuff is placed on the left arm and is connected to an air pump and a pressure sensor. The cuff is inflated until a pressure higher than the typical systolic value is reached, then the cuff is slowly deflated. As the cuff deflates, when systolic pressure value approaches, pulsations start to appear. These pulsations represent the pressure changes due to heart ventricle contraction and can be used to calculate the heartbeat rate. Pulsations grow in amplitude until mean arterial pressure (MAP) is reached, then decrease until they disappear. The blood pressure information is displayed as digital information on the device's screen. {5}------------------------------------------------ - 6. Indication for use: The Upper Arm Electronic Blood Pressure Monitor is intended to measure the blood pressure and pulse rate of adults at a household or medical center. (Not suitable for neonate, pregnancy or pre-eclampsia). - 7. Comparison with the predicate device: Conmo Electronic Company Limited believes that the Upper Arm Electronic Blood Pressure Monitor, BM100-EN is substantially equivalent to the Jiangsu Yuyue Medical Equipment& Supply Co., Ltd Upper Arm Type Electronic Blood Pressure Monitor Series, Electronic Blood Pressure Monitor: YE670D, FDA 510(k) clearance K170605. {6}------------------------------------------------ | Characteristics | Submitted Device | Predicate Device | Difference Discussion | |------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------| | Device Name and model | Upper Arm Electronic Blood Pressure<br>Monitor, BM100-EN | Upper Arm Type Electronic Blood Pressure<br>Monitor YE670D | N/A | | Manufacturer | CONMO ELECTRONIC COMPANY<br>LIMITED | Jiangsu Yuyue Medical Equipment & Supply<br>Co., Ltd | N/A | | 510(K) Number | Pending | K170605 | N/A | | Indication for Use | The Upper Arm Electronic Blood Pressure<br>Monitor is intended to measure the blood<br>pressure and pulse rate of adults at a<br>household or medical center. (Not suitable<br>for neonate, pregnancy or pre-eclampsia). | Electronic blood pressure monitor is<br>intended to measure the blood pressure and<br>pulse rate of adult at household or medical<br>center. (Not suitable for neonate, pregnancy<br>or pre-eclampsia). | Identical | | Clinical Use | household or medical center | household or medical center | Identical | | Patient Population | Adult | Adult | Identical | | Measurement Type | Upper arm | Upper arm | Identical | | Measurement Method | Non-invasive, Sociometric | Non-invasive, Sociometric | Identical | | Cuff Circumference | 220mm ~ 320mm | 220mm ~ 320mm | Identical | | Power Source | 4x1.5V | 4x1.5V | Identical | | Measurement Pressure<br>Range | 0~295mmHg (0-39.3kPa) | 0~280mmHg (0-37.3kPa) | The Pressure Range of BM100-EN is wider<br>than Predicate Devices, but still complies<br>with IEC 80601-2-30 | | Range Accuracy | ±3 mmHg<br>(±0.4kPa) | ±3 mmHg<br>(±0.4kPa) | Identical | | Measurement Pulse<br>Range | 40 ~ 180 beats/min | 40 ~ 200 beats/min | The Pulse Range of BM100-EN is narrower<br>than Predicate Devices, but still complies with<br>IEC 80601-2-30 | | Pulse Rate Measurement<br>Accuracy | ±5% of reading<br>value | ±5% of reading<br>value | Identical | | Number of Users | 2 | 2 | Identical | | Memory Space | 99 | 60 | The difference does not raise any new<br>questions of safety and effectiveness. | | Number of buttons | 3 | 3 | Identical | | Characteristics | Submitted Device | Predicate Device | Difference Discussion | | Controls | User 1/ Confirm<br>Guest/Setting<br>User 2/Plus | SET Button,<br>Plus/Minus Button,<br>START/PULSE Button | The control difference does not raise any new<br>questions of safety and effectiveness. | | Display Type | LCD | LCD | Identical | | Pressurization Mode | Automatic Inflation | Automatic Inflation | Identical | | Deflation Mode | Automatic Pressure Release Valve | Automatic Pressure Release Valve | Identical | | Display Content | Cuff Pressure, Pulse,<br>Date,<br>Time,<br>Systolic/Diastolic<br>Pressure, error message,<br>Measurements result in memory,<br>Irregular Heart Beat<br>Feature, Body<br>movement detection<br>Cuff Wrapping Detection | Cuff Pressure, Pulse,<br>Date,<br>Time,<br>Systolic/Diastolic<br>Pressure, error message,<br>Measurements result in memory,<br>Irregular Heart Beat<br>Feature, Body<br>movement detection<br>Cuff Wrapping Detection | Identical | | Operation Environments | Temperature: +10 °C~ +40 °C<br>Humidity: 15% RH~ 85% RH(no<br>condensation) | Temperature: +10 °C~ +40 °C<br>Humidity: 15% RH~ 90% RH(no<br>condensation) | The operating environment difference does<br>not raise any new questions of safety and<br>effectiveness. | | Storage Environments | Temperature: -20 °C~ +55 °C<br>Humidity: 15% RH~ 85% RH(no<br>condensation) | Temperature: -20 °C~ +55 °C<br>Humidity: 15% RH~ 90% RH(no<br>condensation) | The storage environment difference does not<br>raise any new questions of safety and<br>effectiveness. | | Performance | ANSI/AAMI/ISO 81060-2<br>IEC 80601-2-30 | ANSI/AAMI/ISO 81060-2<br>IEC 80601-2-30 | Identical | | Biocompatibility | ISO 10993-1, FDA Guidance, Tests<br>included Cytotoxicity, Sensitization and<br>Intracutaneous Reactivity | ISO 10993-1, FDA Guidance, Tests included<br>Cytotoxicity, Sensitization and<br>Intracutaneous Reactivity | Identical | | Electrical Safety | IEC 60601-1 | IEC 60601-1 | Identical | | EMC | IEC 60601-1-2 | IEC 60601-1-2 | Identical | | Home Use | IEC 60601-1-11 | IEC 60601-1-11 | Identical | | Prescription or OTC | OTC | OTC | Identical | {7}------------------------------------------------ {8}------------------------------------------------ ## Testing Summary: The following performance data is provided in support of the substantial equivalence determination. ## Non-Clinical Study: Non-clinical tests were conducted to verify that the proposed device meets the same design specifications as the predicate Yuyue Medical Equipment device. The test results demonstrate that the proposed device complies with the following standards: ## Safety and EMC Testing was performed to verify the basic safety and essential performance of the Upper Arm Electronic Blood Pressure Monitor. BM100-EN. The following tests were performed: - IEC 60601-1: 2005+A1: 2012 Medical electrical equipment Part 1: - General requirements for basic safety and essential performance - IEC 60601-1-2: 2014 Medical devices part 1-2: General requirements for basic safety and essential performance – Collateral standards: electromagnetic compatibility - Test and requirements - IEC 60601-1-11 Edition 2.0 2015-01, medical electrical equipment part 1--11: General requirements for basic safety and essential performance collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. The test results demonstrate the device complies with the necessary safety and essential performance standards requirements. {9}------------------------------------------------ ## Performance Data: The subject Upper Arm Electronic Blood Pressure Monitor, BM100-EN was subjected to the following tests and passed all test criteria: - IEC 80601-2-30 Edition 2.0 Medical electrical equipment Part 2-30: Particular Performance 2018-03 requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers - ISO 81060-2 Non-invasive sphygmomanometers Part 2: Clinical validation NIBP Performance Second edition 2013-05-01 of automated measurement type - FDA Guidance Non-Invasive Blood Pressure (NIBP) Monitor Guidance - ## Software Verification and Validation Software documentation, including verification & validation, was provided following FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices for software with a moderate level of concern. ## Cleaning Validation The device was cleaned following the procedure defined in the User's Manual. The device was checked for performance following the cleaning and met all performance requirements following the cleaning. ## Shelf Life: The Upper Arm Blood Pressure Monitor BM100-EN, is not subject to the shelf life, as the device doesn't contain any sterile or degradable components. ## Biocompatibility The subject Upper Arm Electronic Blood Pressure Monitor, BM100-EN, uses similar material to the predicate Yuyue Medical Equipment device. Biocompatibility testing was performed to demonstrate compliance with the same biocompatibility standards as performed by the predicate device. {10}------------------------------------------------ The Upper Arm Blood Pressure Monitor, Model BM100-EN device is classified per ISO 10993-1: 2009 Annex 1 Biological evaluation tests as follows: Surface Device - Contact Skin - Contact Duration <24h Based on this classification, the following cytotoxicity, skin irritation and sensitization tests were conducted per the following standards: - ISO 10993-5: 2009 Biological evaluation of medical devices Part 5 Tests । for In Vitro Cytotoxicity - ISO 10993-10: 2010 Biological evaluation of Medical Devices Part 10: । Tests for Irritation and Delayed-Type Hypersensitivity The test results confirm compliance with the requirements of the standards. ## Sterility Information: The device is not delivered sterile, nor is it required to be sterilized for use. #### Package and Shelf Life: The Upper Arm Blood Pressure Monitor, Model BM100-EN, is not subject to the shelf life, as the device contains no sterile or degradable components. ## Clinical Study: We conducted a comparative clinical study to verify the performance of the subject device and predicate device as well as a mercury sphygmomanometer according to ISO 81060-2:2013. The study demonstrates a correlation in the performance of subject device and predicate device. was significantly correlated The design of the submitted device specifications is substantially the same as the predicate. All the labeling and characteristics of the submitted Upper Arm Electronic Blood Pressure Monitor, BM100-EN, is the same as the predicate device, and most normal blood pressure monitors currently on the market. The submitted device and predicate both use similar measuring methodologies and components to achieve the measurements. {11}------------------------------------------------ - 10. Conclusions: Following the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, Conmo Electronic Company Limited concludes that the Upper Arm Electronic Blood Pressure Monitor, BM100-EN is substantially equivalent to predicate Yuyue Medical Equipment, Inc's device as described herein. Conmo Electronic Company Limited will update and include in a summary any other information deemed reasonably necessary by the FDA. END
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%