ApolloVue S100 Image System
K201552 · Apollo Medical Optics, Ltd. · NQQ · Sep 2, 2020 · Radiology
Device Facts
| Record ID | K201552 |
| Device Name | ApolloVue S100 Image System |
| Applicant | Apollo Medical Optics, Ltd. |
| Product Code | NQQ · Radiology |
| Decision Date | Sep 2, 2020 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1560 |
| Device Class | Class 2 |
Intended Use
Apollo Vue S100 Image System is intended to be used as a non-invasive imaging tool in the evaluation of external human tissue microstructure by providing two-dimensional, cross-sectional and en-face real-time depth visualization for assessment by physicians to support in forming a clinical judgment.
Device Story
ApolloVue S100 is a cart-mounted optical coherence tomography (OCT) system for non-invasive imaging of external human tissue microstructure. It utilizes a single-crystal fiber light source (near-infrared) and a robotic arm probe containing an interferometer. The system employs full-field OCT (FF-OCT) with a camera for parallel detection, enabling high-resolution B-scan and en-face imaging without lateral scanning. The device is operated by clinicians in a clinical setting. It provides real-time depth visualization of tissue microstructure, which physicians review on the system interface to support clinical judgment. The system includes a reusable scanning window cap for skin contact. Benefits include high-resolution (micron-scale) visualization of tissue layers to assist in diagnostic evaluation.
Clinical Evidence
Bench testing only. Performance testing included electrical safety, electromagnetic compatibility, biocompatibility (ISO 10993), disinfection, mechanical safety, and laser safety (IEC 60825-1). Imaging performance was evaluated by capturing B-scan and en-face images of human skin microstructure from various body sites to demonstrate the ability to visualize typical skin patterns.
Technological Characteristics
OCT imaging system; near-infrared light source (750 nm); Class 1 laser (IEC 60825-1). Features cart-mounted console, robotic arm probe, and reusable scanning window cap (biocompatible per ISO 10993 for limited skin contact). Provides 1.0 µm lateral and 1.5 µm axial resolution. Connectivity includes image export capabilities. Software-based image processing for real-time B-scan and en-face visualization.
Indications for Use
Indicated for non-invasive imaging of external human tissue microstructure to provide 2D, cross-sectional, and en-face real-time depth visualization for physician assessment and clinical judgment support.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
- VivoSight Topical OCT System (K093520)
Related Devices
- K093520 — VIVOSIGHT TOPICAL OCT SYSTEM · Michelson Diagnostics, Ltd. · Jan 5, 2010
- K133209 — FORESEE (4C) IMAGING SYSTEM · Diagnostic Photonics, Inc. · Mar 17, 2014
- K120057 — ENVISU SPECTRAL DOMAIN OPHTHALMIC IMAGING SYSTEM (SDOIS) · Bioptigen, Inc. · May 11, 2012
- K153283 — VivoSight Dx Topical OCT System · Michelson Diagnostics, Ltd. · Aug 11, 2016
- K132800 — SKINTELL · Agfa Healthcare N.V. · Mar 3, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Apollo Medical Optics, Ltd. % Feng-Yu Lee Principal Regulatory Consultant Elite BioMedical Consulting, Inc. 29122 Rancho Viejo Rd., Suite 212 San Juan Capistrano, California 92675
September 2, 2020
Re: K201552
Trade/Device Name: ApolloVue S100 Image System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic Pulsed Echo Imaging System Regulatory Class: Class II Product Code: NOO Dated: June 9, 2020 Received: June 10, 2020
Dear Feng-Yu Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Neil R.P. Ogden, MS Assistant Director, THT4A4 DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201552
Device Name ApolloVue S100 Image System
### Indications for Use (Describe)
Apollo Vue S100 Image System is intended to be used as a non-invasive imaging tool in the evaluation of external human tissue microstructure by providing two-dimensional, cross-sectional and en-face real-time depth visualization for assessment by physicians to support in forming a clinical judgment.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <span style="color: black;"> <span style="font-weight: normal;"> <span style="font-style: normal;"> <span style="text-decoration: none;"> </span> </span> </span> </span> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows a logo with the letters 'An' in a stylized, cursive font. The letters are a gradient of blue, with the 'A' being a lighter shade and the 'n' transitioning to a darker blue. To the right of the 'n', there is a circular shape that is also a dark blue color, with a bright white star-like effect in the center, giving the impression of a planet or celestial body.
支股份有限公司 lo Medical Optics, Ltd.
## 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is:
- 1. Submitter Identification:
Apollo Medical Optics, Ltd 2F., No.45, Ln. 188, Ruiguang Rd., Neihu Dist., Taipei City 114, Taiwan
c/o Elite BioMedical Consulting, Inc. Contact Person: Mrs. Feng-Yu Lee Address: Rm. 5, 5F., No. 233, Fuxing 2nd Road, Zhubei City, Hsinchu County 30271, Taiwan 29122 Rancho Viejo Road, Suite 212, San Juan Capistrano, CA 92675 (US Branch) Phone: (949) 218-0929 Fax: (949) 218-0928
Date Summary Prepared: June 9, 2020
- 2. Name of the Device:
ApolloVue S100 Image System
- 3. Common or Usual Name: System, Imaging, Optical Coherence Tomography (OCT)
| Product Code | Classification | Regulation Section | Panel |
|----------------------------------------------------------------|----------------|--------------------|------------------------------|
| NQQ; System, Imaging,<br>Optical Coherence<br>Tomography (OCT) | Class II | 21 CFR 892.1560 | General & Plastic<br>Surgery |
#### 4. Device Description:
The ApolloVue S100 Image System consists of a cart-mounted imaging console and robotic arm imagine probe with reusable non-sterile scanning window cap. The light source module, main control board, power supply and personal computer are configured inside the cart. The probe contains interferometer comprise of optics and electronics. The system user interface allows the viewing, capture, review and export of images.
With AMO single-crystal fiber light source, the ApolloVue S100 Image System provides nearinfrared broadband optical output and cellular images with an axial resolution around one micron. The ApolloVue S100 Image System employs full-field OCT (FF-OCT) utilizing a camera for parallel detection to avoid latera scanning and thus increases the scanning speed. The ApolloVue S100 Image System provides both high resolution B-scan and en face imaging with decent scanning speed. The cross-sectional image can be shown in real-time without reconstruction after whole volume is scanned. In addition, by scanning an en-face image plane with the coherence and confocal
{4}------------------------------------------------
# K201552
Page 2 of 4
生技股份有限公司
Apollo Medical Optics, Ltd.
gates matched, the ApolloVue S100 Image System does not suffer from depth-of-field limitations present in standard OCT and can achieve micron scale transverse image resolutions. With a simple optical switch, user can switch between two modes to improve the efficiency of lesion examination and gather more structure information.
#### ડ. Indications for Use:
ApolloVue S100 Image System is intended to be used as a non-invasive imaging tool in the evaluation of external human tissue microstructure by providing two-dimensional, cross-sectional and en-face real-time depth visualization for assessment by physicians to support in forming a clinical judgment.
#### 6. Predicate Device Information:
ApolloVue S100 Image System is substantially equivalent to the following device: Name: VivoSight Topical OCT System Device Company: Michelson Diagnostics Ltd 510(K) Number: K093520
#### 7. Comparison to Predicate Device:
As shown in the following comparison table, the ApolloVue S100 Image System has the same intended use and similar technical characteristics as its predicate device (K093520). The minor technical differences between the ApolloVue S100 Image System and its predicate do not rise new or different questions of safety and effectiveness.
| Specification | Subject Device -<br>ApolloVue S100 Image System | Predicate Device -<br>VivoSight Topical OCT System<br>(K093520) |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Similarities & Differences | | |
| Regulation Number | 21 CFR 892.1560 | Same |
| Class | II | Same |
| Product Code | NQQ | Same |
| Device Class/Name | System, Imaging, Optical Coherence<br>Tomography (OCT) | Same |
| Indication for Use | ApolloVue S100 Image System is<br>intended to be used as a non-<br>invasive imaging tool in the<br>evaluation of external human tissue<br>microstructure by providing two-<br>dimensional, cross-sectional and en-<br>face real-time depth visualization<br>for assessment by physicians to<br>support in forming a clinical<br>judgment. | VivoSight is a Multi-Beam Optical<br>Coherence Tomography (OCT)<br>system indicated for use in the two-<br>dimensional, cross-sectional, real-<br>time imaging of external tissues of the<br>human body. |
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows a logo with the letters 'An' in a stylized, cursive font. The 'A' is in a lighter blue color, while the 'n' transitions to a darker blue, almost navy, color. The end of the 'n' has a small, bright star-like effect, adding a touch of visual interest to the logo.
支股份有限公司 ollo Medical Optics, Ltd.
| Measurement Technique | Optical Coherence Tomography | Optical Coherence Tomography |
|-------------------------------------------|-----------------------------------|------------------------------|
| Near-Infrared Wavelength<br>(700-1400 nm) | Yes | Yes |
| Light Source Center<br>Wavelength | 750 nm | 1300 nm |
| Lateral Scanning Range | 0.5 mm | 5.0 mm |
| Axial Scanning Range | 0.35 mm | Up to 2.0 mm |
| Lateral Resolution | 1.0 µm | < 7.5 µm |
| Axial Resolution | 1.5 µm | < 5 um |
| Scan Time | En face: 6 fps<br>B-scan: 0.4 fps | > 6 fps |
| Optical Radiation Safety | Class 1 Laser | Same |
| Power Supply | 100-240 V<br>50-60 Hz | 110-132 V<br>50-60 Hz |
#### 8. Technology Characteristics:
The ApolloVue S100 Image System has the same intended use and similar technological characteristics as the predicate, VivoSight Topical OCT System (K093520). Both the ApolloVue S100 Image System and the predicate device utilize Optical Coherence Tomography to create high resolution cross-sectional and en face real-time images of tissue microstructure. The ApolloVue S100 Image System and the predicate device both use near infrared light to produce OCT images with comparable axial and lateral resolution. The comparison of the ApolloVue S100 Image System and its predicate device shows that their technological characteristics are similar and the minor differences do not raise new or different question of safety and effectiveness.
#### 9. Discussion of Performance Data for Determination of Substantial Equivalence are as follows:
The ApolloVue S100 Image System is designed and tested to be in compliance with international standards concerning safety for electrical safety, electromagnetic compatibility, biocompatibility, disinfection, mechanical safety, and laser safety. A series of tests were conducted to assess the performance and safety of the ApolloVue S100 Image System. All acceptance criteria were met, supporting substantial equivalence for the subject device. A brief summary of the performance testing is described below.
The system uses a laser that is limited in power to meet the requirements of a class 1 laser device in accordance with IEC 60825-1. The handheld probe has been designed to be a reusable component and to support cleaning and disinfection. The reusable probe tip (scanning window cap) is designed using materials that adhere to the requirements for skin surface limited contact (< 24 hours) and to be replaced after 25 times of cleaning and disinfection. The reusable scanning window cap has been tested for biocompatibility per ISO 10993 based on the intended use.
The system image performance has been tested to evaluate the performance of the ApolloVue S100 Image System to demonstrate the imaging capabilities for the visualization of skin microstructure in vivo. Images were taken using both B-scan and En face modes of the ApolloVue S100 Image
{6}------------------------------------------------
安盟生技股份有限公司 Apollo Medical Optics, Ltd.
Image /page/6/Picture/2 description: The image shows a logo with the letters 'An' in a stylized, cursive font. The letters are in a gradient of blue, transitioning from a lighter shade to a darker shade. To the right of the letters, there is a dark blue sphere with a bright white star-like burst in the center, giving the impression of a celestial body or a stylized representation of a star.
System from different body sites of human subjects. Typical patterns of skin can be linked to image patterns visible in the B-scan- and En face-images of the ApolloVue S100 Image System.
#### 10. Substantial Equivalence Conclusions:
The ApolloVue S100 Image System has the same intended use and similar indications, technological characteristics, and principles of operation as the predicate device, VivoSight Topical OCT System (K093520). The minor technical differences between the ApolloVue S100 Image System and the predicate device do not rise new or different questions of safety and effectiveness. Performance data demonstrate that the ApolloVue S100 Image System is substantially equivalent to the predicate device.