CUPID Brand Male Natural Rubber Latex Condoms (Plain (banana) , Plain (strawberry), Dotted , Multi-textured (Dotted+Contoured + Ribbed))

K201067 · Cupid Limited · HIS · Aug 7, 2020 · Obstetrics/Gynecology

Device Facts

Record IDK201067
Device NameCUPID Brand Male Natural Rubber Latex Condoms (Plain (banana) , Plain (strawberry), Dotted , Multi-textured (Dotted+Contoured + Ribbed))
ApplicantCupid Limited
Product CodeHIS · Obstetrics/Gynecology
Decision DateAug 7, 2020
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 884.5300
Device ClassClass 2

Intended Use

CUPID Brand Male Natural Rubber Latex Condoms are used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections).

Device Story

CUPID Brand Male Natural Rubber Latex Condoms are physical barrier devices designed for contraception and the prevention of sexually transmitted infections. The device is a sheath made of natural rubber latex, available in various textures including plain, dotted, and multi-textured (dotted, contoured, and ribbed). It is intended for over-the-counter use by individuals. The device functions by providing a mechanical barrier during sexual intercourse to prevent the exchange of bodily fluids and pregnancy.

Technological Characteristics

Natural rubber latex condom; available in plain, dotted, and multi-textured (dotted, contoured, ribbed) configurations. Class II device under 21 CFR 884.5300.

Indications for Use

Indicated for use by individuals for contraception and the prevention of sexually transmitted infections.

Regulatory Classification

Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. August 7, 2020 CUPID Limited % Manoj Zacharias Consultant Liberty Management Group Ltd. 75 Executive Drive, Suite 114 Aurora, IL 60504 Re: K201067 > Trade/Device Name: CUPID Brand Male Natural Rubber Latex Condoms (Plain (banana), Plain (strawberry), Dotted, Multi-textured (Dotted+Contoured + Ribbed)) Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: HIS Dated: July 5, 2020 Received: July 8, 2020 Dear Manoj Zacharias: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Monica D. Garcia, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) K201067 ## Device Name CUPID Brand Male Natural Rubber Latex Condoms (Plain (strawberry), Dotted, Multi-textured (Dotted+Contoured + Ribbed)) Indications for Use (Describe) CUPID Brand Male Natural Rubber Latex Condoms are used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted infections). Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Image /page/2/Picture/9 description: The image shows a checkbox with an X inside, followed by the text "Over-The-Counter Use (21 CFR 801 Subpart C)". This indicates that the product or item being referred to is intended for over-the-counter use, as defined by the Code of Federal Regulations (CFR) Title 21, Section 801, Subpart C. The checkbox suggests that this is a declaration or confirmation of the product's intended use. Comments of Children - ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Image /page/2/Picture/16 description: This image shows the header of a document. The document is labeled "FORM FDA 3881 (7/17)". It also indicates "Page 1 of 1". The header also includes the text "PSC Publishing Services (301) 443-6740 EF".
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%