Intense Pulsed Light Treatment System
K192521 · Shangdong Huamei Technology Co., Ltd. · ONF · Dec 6, 2019 · General, Plastic Surgery
Device Facts
| Record ID | K192521 |
| Device Name | Intense Pulsed Light Treatment System |
| Applicant | Shangdong Huamei Technology Co., Ltd. |
| Product Code | ONF · General, Plastic Surgery |
| Decision Date | Dec 6, 2019 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4810 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Intense Pulsed Light Treatment System (Model: HM-IPL-B8) is indicated for use in surgical and aesthetic applications in permanent hair removal, reduction of benign pigmented lesions and benign vascular lesions. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen.
Device Story
Intense Pulsed Light (IPL) system (Model: HM-IPL-B8) delivers pulsed light (430-1200nm) to target chromophores via selective thermolysis; pulse width limited below skin thermal relaxation time to spare normal tissue. Device used in clinical settings by professionals for hair removal and lesion reduction. System utilizes sapphire crystal handpiece for energy delivery. Output parameters (wavelength, energy density, pulse width/delay) are adjusted based on treatment indication. Clinical benefit includes long-term stable hair reduction and cosmetic improvement of skin lesions.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), performance requirements for non-laser light sources (IEC 60601-2-57), and biocompatibility (ISO 10993-5, ISO 10993-10).
Technological Characteristics
IPL system; 430-1200nm wavelength range; sapphire crystal delivery system; 2000W max power; 10-50J/cm² energy density; 1-20ms pulse width; 5-50ms pulse delay. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for permanent hair removal, reduction of benign pigmented lesions, and reduction of benign vascular lesions in surgical and aesthetic applications.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
- Intense Pulsed Light (IPL) Systems (K122995)
Related Devices
- K242440 — Intense Pulsed Light Treatment System (LK-PT) · Nanjing Bestview Laser S&T Co., Ltd. · Nov 14, 2024
- K122995 — INTENSE PULSED LIGHT (IPL) SYSTEMS · Beijing Kes Biology Technology Co., Ltd. · Jan 22, 2013
- K251545 — Intense Pulsed Light Therapy Device (MMABM-1) · Micowey Medical Equipment (Guangxi) Co., Ltd. · Sep 4, 2025
- K101901 — IPL THERAPY SYSTEM · Beijing Starlight Science & Technology Development · Mar 8, 2011
- K200746 — IPL Treatment Systems · Shanghai Apolo Medical Technology Co., Ltd. · May 15, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 6, 2019
Shangdong Huamei Technology Co., Ltd. % Ray Wang Official Correspondent Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd. FangShan District Beijing, 102401 CN
Re: K192521
Trade/Device Name: Intense Pulsed Light Treatment System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: ONF Dated: September 11, 2019 Received: September 13, 2019
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Purva Pandva Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K192521
Device Name Intense Pulsed Light Treatment System
The Intense Pulsed Light Treatment System (Model: HM-IPL-B8) is indicated for use in surgical and aesthetic applications in permanent hair removal, reduction of benign pigmented lesions and benign vascular lesions. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen.
| Type of Use (Select one or both, as applicable) | <span> <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D) <input checked="" type="checkbox"/> </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) <input type="checkbox"/> </span> |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Paperwork Reduction Act (PRA) Staff
PRAStaff(@fda.hhs.gov
==============================================================================================================================================================================
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K192521
- 1. Date of Preparation: 11/25/2019
- 2. Sponsor Identification
## Shangdong Huamei Technology Co., Ltd.
No. 588, Changning Street, High-tech District, WeiFang, ShanDong, China, 261205
Contact Person: Xu QingHua Position: General Manager Tel: +86-536-2110001 Fax: +86-536-2109823 Email: xuqh@yeah.net
- 3. Designated Submission Correspondent
Mr. Ray Wang
## Beijing Believe-Med Technology Service Co., Ltd.
Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing City, China, 102401 Tel: +86-18910677558 Fax: +86-10-56335780 Email: ray.wang@believe-med.com
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: Intense Pulsed Light Treatment System Common Name: Powered Light Based Non-Laser Surgical Instrument With Thermal Effect Model(s): HM-IPL-B8
Regulatory Information:
Classification Name: Powered Light Based Non-Laser Surgical Instrument With Thermal Effect Classification: II; Product Code: ONF; Regulation Number: 21 CFR 878.4810; Review Panel: General & Plastic Surgery;
Indication For Use Statement:
The Intense Pulsed Light Treatment System (Model: HM-IPL-B8) is indicated for use in surgical and aesthetic applications in permanent hair removal, reduction of benign pigmented lesions and benign vascular lesions.
Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen.
### Device Description:
The HM-IPL-B8 device is an intense pulsed light system which delivers intense pulsed light at wavelengths ranging from 430mm-1200mm. Intense Pulsed Light (IPL) systems work on the principles of selective thermolysis. That is, causing thermal damage to target chromophores by using light of appropriate wavelength in pulses that exceeds the chromophores' thermal relaxation time but sparing normal skin by limiting the pulse width below the thermal relaxation time for skin.
IPL systems are different from lasers in that they deliver many wavelengths in each pulse of light instead of just one wavelength. Generally, IPL enhances penetration without using excessive energy levels and enables targeting of specific chromophores.
- 5. Identification of Predicate Device(s)
510(k) Number: K122995 Predicate Device Name: Intense Pulsed Light (IPL) Systems Manufacturer: Beijing KES Biology Technology Co., Ltd.
#### 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
{5}------------------------------------------------
a) IEC 60601-1-2:2014, Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests.
b) IEC 60601-1:2005/A1:2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
c) IEC60601-2-57:2011, Medical electrical equipment -- Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
d) ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro cytotoxicity.
e) ISO 10993-10:2010 Standard, Biological Evaluation of Medical Device, Part 10-Test for irritation and delay-type hypersensitivity.
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
{6}------------------------------------------------
#### Substantially Equivalent (SE) Comparison 8.
| Table 7-1 General Comparison | | | |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| ITEM | Proposed Device | Predicate Device | Remark |
| Product Code | ONF | ONF | SE |
| Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 | SE |
| Class | II | II | SE |
| Intended Use | The Intense Pulsed Light Treatment<br>System (Model: HM-IPL-B8) are<br>indicated for use in surgical and<br>aesthetic applications in permanent<br>hair removal, reduction of benign<br>pigmented lesions and benign<br>vascular lesions.<br>Permanent hair reduction is defined<br>as the long-term, stable reduction in<br>the number of hairs regrowing<br>when measured at 6, 9, and 12<br>months after the completion of a<br>treatment regimen. | Intense Pulsed Light (IPL) System<br>s (inclusive of the handpiece used<br>to deliver pulsed-light energy) are<br>indicated for use in surgical,<br>aesthetic and cosmetic applications<br>in permanent hair removal,<br>reduction of pigmented lesions,<br>acne therapy, freckles, vascular<br>lesions and facial blemish. | SE |
## Table 7-1 General Comparison
{7}------------------------------------------------
| ITEM | Proposed Device | Predicate Device<br>Intense Pulsed Light (IPL)<br>Systems(K122995) | Remark |
|----------------|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|------------|
| Light source | Intense pulsed light | Intense pulsed light | SE |
| Wavelength | 430-1200nm, 530-1200nm,<br>640-1200nm,<br>Optional: 480-1200nm,<br>560-1200nm, 590-1200nm,<br>690-1200nm, 750 -1200nm | 430-1200nm, 530-1200nm,<br>640-1200nm,<br>Optional: 480-1200nm,<br>560-1200nm, 590-1200nm,<br>690-1200nm, 750 -1200nm | SE |
| Deliver system | Sapphire | Sapphire | SE |
| Energy density | 10-50J/cm² | 10-60J/cm² | Analysis 1 |
| Pulse Delay | 5-50ms | 5 - 50ms | SE |
| Pulse Width | 1-20ms | 1-20ms | SE |
| Max. Power | 2000W | 2000 W | SE |
| Spot size | 15mm×50mm; 80mm×40mm; | MED-210: 15mmX50mm<br>(optional: 12mmX33mm,<br>15mmX35mm )<br>MED-230:<br>A: 12mm X33mm;<br>B: 15mmX50mm (optional:<br>15mmX35mm) | Analysis 2 |
| Table 7-2 Performance Comparison |
|----------------------------------|
|----------------------------------|
| ITEM | Proposed Device | Predicate Device<br>Intense Pulsed Light (IPL)<br>Systems(K122995) | Remark |
|---------------------------------|-----------------------------|--------------------------------------------------------------------|--------|
| <b>permanent hair reduction</b> | | | |
| Wavelength Range<br>(nm) | 640-1200/690-1200/ 750-1200 | 640-1200/690-1200/ 750-1200 | SE |
| Energy Range<br>(J/cm2) | 10-44 | 10-44 | |
| Pulse Width (ms) | 3-14 | 3-14 | |
| Pulse Delay (ms) | 16-32 | 16-32 | |
| Spot Size (mm) | 15mm×50mm; 80mm×40mm; | 12mm X33mm; 15mmX50mm<br>15mmX35mm | |
| <b>pigmented lesions</b> | | | |
| Wavelength Range<br>(nm) | 480-1200/530-1200/560-1200 | 480-1200/530-1200/560-1200 | SE |
| Energy Range<br>(J/cm2) | 12-44 | 12-44 | |
| Pulse Width (ms) | 3-9 | 3-9 | |
Table 7-3 Setting Comparison of Specified Indication for Use
{8}------------------------------------------------
| Pulse Delay (ms) | 16-32 | 16-32 | |
|--------------------------|--------------------------------|------------------------------------|----|
| Spot Size (mm) | 15mm×50mm; 80mm×40mm; | 12mm X33mm; 15mmX50mm<br>15mmX35mm | |
| vascular lesions | | | |
| Wavelength Range<br>(nm) | 530-1200/560-1200/<br>590-1200 | 530-1200/560-1200/<br>590-1200 | |
| Energy Range<br>(J/cm2) | 10-42 | 10-42 | SE |
| Pulse Width (ms) | 3-8 | 3-8 | |
| Pulse Delay (ms) | 16-32 | 16-32 | |
| Spot Size (mm) | 15mm×50mm; 80mm×40mm; | 12mm X33mm; 15mmX50mm<br>15mmX35mm | |
## Analysis 1:
The energy density range of the proposed device is within the energy density range of the predicate for the indications for use, and the slight difference does not result in a negative effect on safety and effectiveness.
Therefore, the slight difference is considered to have no negative effect on effectiveness and safety.
The spot size of the proposed device is different from that of predicate device, but the difference of the spot size is only for different treatment area, and it does not negatively affect the output parameter for treatment. Therefore, the slight difference is considered to have no negative effect on effectiveness and safety.
{9}------------------------------------------------
| Table 7-4 Safety Comparison | | | |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|--------|
| ITEM | Proposed Device<br>HM-IPL-B8 | Predicate Device<br>Intense Pulsed Light (IPL)<br>Systems(K122995) | Remark |
| Power supply | $110V \pm 10\% \ 60Hz$ | $220V \pm 20V \ 50Hz$ or<br>$110V \pm 20V \ 60Hz$ | SE |
| Electrical<br>Safety | The proposed devices were tested<br>to demonstrated to comply with<br>IEC 60601-1 | The proposed devices were tested<br>to demonstrated to comply with<br>IEC 60601-1 | SE |
| EMC | The proposed devices were tested<br>to demonstrated to comply with<br>IEC 60601-1-2 | The proposed devices were tested<br>to demonstrated to comply with<br>IEC 60601-1-2 | SE |
| Patient Contact<br>Material | Handpiece (Sapphire Crystal) | Handpiece (Sapphire Crystal) | SE |
| Biocompatibility | | | |
| Cytotoxicity | No toxicity (ISO 10993-5) | No toxicity (ISO 10993-5) | SE |
| Irritation | Applied sample did not induce<br>irritation to skin. (ISO 10993 -10) | Applied sample did not induce<br>irritation to skin. (ISO 10993 -10) | SE |
| Sensitization | The test article showed no<br>signification evidence of causing<br>skin sensitization in the guinea<br>pig .(ISO 10993-10) | The test article showed no<br>signification evidence of causing<br>skin sensitization in the guinea<br>pig .(ISO 10993-10) | SE |
#### 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.