Matrix HA PEEK Cervical Implant System
K192316 · Sapphire Medical Group · ODP · Oct 17, 2019 · Orthopedic
Device Facts
| Record ID | K192316 |
| Device Name | Matrix HA PEEK Cervical Implant System |
| Applicant | Sapphire Medical Group |
| Product Code | ODP · Orthopedic |
| Decision Date | Oct 17, 2019 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 888.3080 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Matrix HA PEEK Cervical Implant System is intended for spinal fusion procedures at one level (C2 to T1) in skeletally mature patients with degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are intended to be packed with autogenous bone graft. The Matrix HA PEEK Cervical Implant System is intended to be used with a supplemental fixation system.
Device Story
Matrix HA PEEK Cervical Implant System is an intervertebral body fusion device; used in cervical spine surgery to facilitate fusion at one level (C2-T1). Device consists of PEEK implants packed with autogenous bone graft; requires supplemental fixation system. Used by surgeons in clinical settings to treat degenerative disc disease. Provides structural support to disc space during fusion process; benefits patient by stabilizing cervical spine segment.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and intended use.
Technological Characteristics
Intervertebral body fusion device; material: PEEK (Polyetheretherketone) with HA (Hydroxyapatite) coating; form factor: cervical implant; intended for use with autogenous bone graft and supplemental fixation; Class II, Product Code ODP.
Indications for Use
Indicated for skeletally mature patients with cervical degenerative disc disease (discogenic back pain confirmed by history/radiography) requiring one-level spinal fusion (C2-T1). Must be used with autogenous bone graft and supplemental fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Related Devices
- K181115 — CxHA PEEK Cervical IBF System · Innovasis, Inc. · Jul 16, 2018
- K212715 — Vy Spine ClariVy Cervical IBF System · Vy Spine, LLC · Dec 20, 2021
- K162531 — STERISPINE CC · Safe Orthopaedics · Dec 29, 2016
- K161809 — ShurFit CpTi-HA ACIF Interbody Fusion System · Precision Spine, Inc. · Dec 6, 2016
- K173077 — Cavetto-SA Cervical Cage System · Neurostructures, Inc. · Feb 23, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
Sapphire Medical Group % Marshall McCarty Consultant Innovasis. Inc. 614 E 3900 South Salt Lake City, Utah 84107 October 17, 2019
## Re: K192316
Trade/Device Name: Matrix HA PEEK Cervical Implant System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP Dated: September 19, 2019 Received: September 20, 2019
Dear Mr. McCarty:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Ronald P. Jean, Ph.D. Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192316
Device Name Matrix HA PEEK Cervical Implant System
#### Indications for Use (Describe)
The Matrix HA PEEK Cervical Implant System is intended for spinal fusion procedures at one level (C2 to T1) in skeletally mature patients with degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) of the cervical spine. Implants are intended to be packed with autogenous bone graft. The Matrix HA PEEK Cervical Implant System is intended to be used with a supplemental fixation system.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;"> <svg height="12" width="12"> <rect height="12" style="fill:rgb(0,0,0);stroke-width:1;stroke:rgb(0,0,0)" width="12"></rect> <line style="stroke:rgb(255,255,255);stroke-width:2" x1="0" x2="12" y1="0" y2="12"></line> <line style="stroke:rgb(255,255,255);stroke-width:2" x1="0" x2="12" y1="12" y2="0"></line> </svg> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <svg height="12" width="12"> <rect height="12" style="fill:rgb(255,255,255);stroke-width:1;stroke:rgb(0,0,0)" width="12"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."