Endophys Pressure Sensing Sheath

K190098 · Endophys Holdings, LLC · DYB · Mar 7, 2019 · Cardiovascular

Device Facts

Record IDK190098
Device NameEndophys Pressure Sensing Sheath
ApplicantEndophys Holdings, LLC
Product CodeDYB · Cardiovascular
Decision DateMar 7, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1340
Device ClassClass 2

Intended Use

The Endophys Pressure Sensing Sheath Kit (Endophys Pressure Sensing Sheath and vessel dilator) is intended to facilitate the introduction of diagnostic and interventional devices into the vasculature and to continuously measure blood pressure during the procedure when used with the Endophys Blood Pressure Monitor. Together, the Pressure Sensing Sheath and the Blood Pressure Monitor comprise the Endophys Pressure Sensing Access System (PSAS™).

Device Story

Device is an introducer sheath with integrated fiber optic pressure transducer and vessel dilator; used to facilitate vascular access and provide continuous, high-resolution blood pressure monitoring. Operates by connecting to the Endophys Blood Pressure Monitor (BPM) to display hemodynamic pulse wave data. Used in clinical settings by physicians during interventional procedures. Output is continuous blood pressure waveform and numerical data displayed on the BPM; assists clinicians in real-time hemodynamic monitoring and clinical decision-making during vascular procedures. Benefits include integrated access and monitoring, reducing need for separate arterial lines.

Clinical Evidence

Bench testing only. Performance data included dimensional analysis, sensor accuracy, device compatibility, flexibility (tip deflection, kink resistance), sensor functionality, pouch peel strength, separation force, and biocompatibility (L929 MEM elution, Neutral Red Uptake, Rabbit Pyrogen, Rabbit Blood Hemolysis).

Technological Characteristics

Introducer sheath with integrated fiber optic pressure transducer; includes vessel dilator. Materials include braided wire (encapsulated), hemostasis valve, and primary/secondary lumen liners. Sterilized via ethylene oxide (EO). Connectivity: wired connection to Endophys Blood Pressure Monitor. Technical specs: 8F size.

Indications for Use

Indicated for patients requiring vascular access for diagnostic or interventional procedures where continuous blood pressure monitoring is desired.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font. March 7, 2019 Endophys Holdings LLC % Gregory Sachs President Sachs & Associates, Inc 5116 Birch Road Minnetonka, Minnesota 55345 Re: K190098 Trade/Device Name: Endophys Pressure Sensing Sheath Kit - 8F Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB, DXO Dated: January 18, 2019 Received: January 22, 2019 Dear Gregory Sachs: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shawn W. Forrest -S 2019.03.07 22:26:06 -05'00' for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K190098 Device Name Endophys Pressure Sensing Sheath Kit- 8F ### Indications for Use (Describe) The Endophys Pressure Sensing Sheath Kit (Endophys Pressure Sensing Sheath and vessel dilator) is intended to facilitate the introduction of diagnostic and interventional devices into the vasculature and to continuously measure blood pressure during the procedure when used with the Endophys Blood Pressure Monitor. Together, the Pressure Sensing Sheath and the Blood Pressure Monitor comprise the Endophys Pressure Sensing Access System (PSAS™). | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------| | <span style="text-align:left;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="text-align:left;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summarv Endophys Pressure Sensing Sheath Kit #### I. SUBMITTER Endophys Holdings, LLC Thanksgiving Tower, Suite 3500 1601 Elm Street Dallas, TX 75201 Contact person: Phillip D. Purdy, M.D. Phone: 214-801-0578 Fax: 214-377-0709 Date prepared: January 18, 2019 #### II. DEVICE Name of the device: Endophys Pressure Sensing Sheath Kit Common of usual name: Introducer Catheter and accessories Classification name: Catheter Introducer Regulatory Class: 2 Product Code: DYB (classification); DXO (subsequent) #### III. PREDICATE DEVICE Endophys Pressure Sensing Sheath Kit (K173399) This predicate has not been subject to a design-related recall No reference devices were used in this submission. #### IV. DEVICE DESCRIPTION The Endophys Pressure Sensing Sheath Kit is comprised of the Endophys Pressure Sensing Sheath and vessel dilator. The Endophys Pressure Sensing Sheath and accessories are provided sterile (EO). The Pressure Sensing Sheath ("PSS") is an introducer sheath with an integrated fiber optic pressure transducer. The PSS is provided with a dilator which together make up the Endophys Pressure Sensing Sheath Kit ("PSS Kit"). The PSS Kit is intended to be used only with the Endophys Blood Pressure Monitor ("BPM"), which connects to the PSS and displays the blood pressure measurements. Together with the BPM, the PSS is used to continuously monitor patient blood pressure during procedures requiring vascular access. The PSS introducer sheath is used to assist vascular access in the same manner as do standard catheter introducers. The PSS introducer sheath has an integrated pressure sensor that provides high resolution, continuous and instantaneous pulse wave hemodynamics to monitor blood pressure. {4}------------------------------------------------ Together, the Pressure Sensing Sheath and the Blood Pressure Monitor comprise the Endophys Pressure Sensing Access System (PSASTM). The vessel dilator is used to facilitate insertion of the PSS, and is packaged with the Endophys PSS Kit. Once the PSS has been properly positioned in the vessel, the dilator is removed. #### V. INDICATION FOR USE The Endophys Pressure Sensing Sheath Kit (Endophys Pressure Sensing Sheath and vessel dilator) is intended to facilitate the introduction of diagnostic and interventional devices into the vasculature and to continuously measure blood pressure during the procedure when used with the Endophys Blood Pressure Monitor. Together, the Pressure Sensing Sheath and the Blood Pressure Monitor comprise the Endophys Pressure Sensing Access System (PSAS"). ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE Introducer catheters, by nature, are tubular structures intended to provide a pathway for therapeutic and/or diagnostic devices into the human vasculature. Specialized introducer catheters have been developed which incorporate sensors into these devices to allow active monitoring of blood pressure during interventional procedures. Blood pressure monitoring using a fiber optic transducer is the technological principle for both the subject and predicate devices. The technology requires use of a compatible blood pressure monitor to provide this data to the clinician. At a high level, the subject and predicate device are based on the following same technological elements: | • Product Code | • Pressure Sensor Type | |----------------------------------------|-------------------------------------| | • Device Classification Name | • Sensor Location | | • 21 CFR Regulation Number | • Pressure Range | | • Principle of Operation | • Accuracy of Pressure Reading | | • Target Population | • Calibration | | • Anatomical Site | • Power Source | | • Where Used | • Sensor Fiber | | • Contraindications | • Sensor Diaphragm, Pedestal | | • Human Factors | • 2-Part Epoxy | | • Placement | • Hemostasis Valve Adapter Assembly | | • Not Resterilized | • 3-Way Stopcock | | • Sterilization Method | • Dilator Sheath Connectivity | | • Method of Blood Pressure measurement | • Dilator Size | Page 3 {5}------------------------------------------------ | • Instructions For Use | • Dilator Tube/Extrusions | |---------------------------------|---------------------------------| | • Labeling | • Dilator Hub Resin | | • Biocompatibility Test Results | • Dilator Hub Ink | | • Sheath Taper Length | • Dilator Hub Colorant | | • Inside/outside diameter | • Sterile Barrier System | | • Suture Ring | • Pouch/Shelf/Shipper Box Label | | • Radiopaque Marker | • Outer Jacket | | • Molded Y-Hub | • Packaged Sterile | | • Hemostasis Valve | • Braided Wire (Encapsulated) | | • Primary Lumen Liner | • Sheath Size | | • Secondary Lumen Liner | • UV Adhesive | The following technological differences exist between the subject and predicate device. Please note that a guidewire is not supplied with the proposed device. | • Indications for Use | • Guidewire Wire and Core Material | |-------------------------------|------------------------------------| | • Components | • Guidewire Hoop and Clip | | • Sheath Overall Length | • Guidewire Luer Lock | | • Sheath Working Length | • Shelf Life | | • Dilator Working Length | • Primary Barrier Height | | • Guidewire length | • Die-Cut Card Height | | • Guidewire width/diameter | • Shelf Carton Height | | • Guidewire tip configuration | • Shipping Carton Width | | • Guidewire Source | | | • Guidewire Straightener | | #### VII. PERFORMANCE DATA The following performance data was provided in support of the substantial equivalence. - Dimensional Analysis ■ - 트 Sensor Accuracy - Guide Wire / Device Compatibility ■ - Flexibility - Tip Deflection - 트 Flexibility - Kink Resistance - " Flexibility - Sensor Functionality - Pouch Peel Strength - 트 Dilator to Hemostasis Valve Separation Force - L929 MEM Elution - ISO - L929 Neutral Red Uptake Test (4 Concentrations) - ISO - I Rabbit Pyrogen Test (Material Mediated - ISO) - 트 Rabbit Blood Hemolysis Complete ASTM The modified Endophys Pressure Sensing Sheath met all specified criteria and based on the design verification performance, the conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as Page 4 {6}------------------------------------------------ effective, and performs as well as or better than the legally marketed device. #### SUBSTANTIAL EQUIVALENCE VIII. The indications for use for the predicate device is substantially equivalent to the indications for use for the proposed PSS Kit. Furthermore, the PSS Kit has the same intended use, patient population, and anatomical sites as well as similar technological characteristics as the predicate device. The differences in technological characteristics have been analyzed and addressed through testing. Any differences in the technological characteristics between the devices do not raise different questions of safety or effectiveness. Thus, the Endophys Pressure Sensing Sheath Kit is substantially equivalent to the predicate device. #### CONCLUSION IX. The PSS Kit contains a catheter introducer sheath with an integrated fiber optic pressure transducer and dilator for standard interventional procedures. The PSS Kit has the same intended use, patient population, and anatomical sites as well as similar technological characteristics as the predicate device. The differences in technological characteristics have been analyzed and addressed through testing. As such, the PSS Kit is substantially equivalent to the predicate device. #### X. SUMMARY The PSS Kit is substantially equivalent to the predicate device.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%