K Number
K182463

Validate with FDA (Live)

Date Cleared
2019-11-06

(422 days)

Product Code
Regulation Number
876.5130
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This product is used for temporary urethral catheterization or indwelling catheterization and bladder irrigation.

Device Description

The Silicone Foley Catheter uses the pressure of the bladder to excrete urine. The balloon of the catheter can be inflated and fixed the catheter in place after being injected with a sterile liquid. Foley catheter (2-way) can be used for bladder drainage, and Foley catheter (3-way) also can be used for bladder irrigation. Female type Foley Catheter is equipped with the advancer which can be used instead of tweezers during the catheterization. The Silicone Foley Catheter is for single use, is intended for short-term (less than 14 days) use, and is provided with sterile.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device, the Silicone Foley Catheter (K182463). This type of submission aims to demonstrate that a new device is substantially equivalent to a legally marketed predicate device, rather than proving its safety and effectiveness through extensive clinical trials as an FDA approval would. Therefore, the information typically requested in your prompt (acceptance criteria, specific study designs, sample sizes, expert ground truth, MRMC studies, standalone performance with effect sizes, ground truth types for training and test sets) is generally not present in a 510(k) summary for this type of device.

Instead, the submission focuses on non-clinical testing to demonstrate that the new device performs comparably to the predicate and meets relevant standards.

Here's an analysis based on the provided document, addressing your points where information is available or noting its absence:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly provide a table of acceptance criteria with corresponding device performance metrics in the format you requested for a diagnostic AI device. Instead, it lists various non-clinical tests and states that "Results demonstrate that the device is safe and effective in meeting user requirements in accordance with its intended use."

The non-clinical testing includes:

  • Bench testing of functional performance: This would involve testing physical characteristics and function against established standards (e.g., ASTM F623-99(2013), ISO 2859-1:1999, ISO 8536-4:2010, USP). The acceptance criteria would be defined within these standards and company specifications, but these specific criteria and the detailed performance results are not provided in this summary.
  • Biocompatibility testing: Compliance with ISO 10993 series (e.g., ISO 10993-3, -5, -6, -9, -10, -11) would set the acceptance criteria for material safety. The results are summarized as meeting requirements.
  • Sterilization validation: Compliance with ISO 11135, ISO 11138, ISO 10993-7, ISO 11737 series.
  • Shelf life validation: This would involve stability testing over time to ensure the device remains safe and effective (e.g., ASTM F623-99(2013)).
  • Simulation Shipment validation: Compliance with ASTM F88, F1929, F1886.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided. For a device like a Foley catheter, "test sets" in the context of diagnostic AI performance studies (as implied by your prompt) are not typically used. The testing would involve physical samples of the catheter for bench, biocompatibility, and sterilization tests. The number of samples for these tests is standard for regulatory submissions but not detailed here.

The document mentions "Clinical Evaluation Report, CER-FC-001(A2)" and "Silicone Foley Catheter Post Market Clinical Follow-Up Study, PMCF-2018-01(A1)". While these suggest some clinical data, the summary does not detail sample sizes, data provenance, or whether this constitutes a "test set" in the diagnostic AI sense.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable and therefore not provided. For a physical device like a Foley catheter, the concept of "ground truth" established by experts for a test set (as in image interpretation for AI) does not directly apply. Performance is determined by objective physical and biological tests against established standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable and therefore not provided. Adjudication methods are typically relevant for human interpretation tasks, especially in diagnostic AI studies.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No such study was performed or is mentioned. MRMC studies are specific to evaluating AI in diagnostic image interpretation or similar tasks where human interaction with AI is a factor. This device is a physical medical device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a physical catheter, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

As mentioned, the concept of "ground truth" as you've defined it for diagnostic AI does not directly apply. For a physical device, ground truth would be established through:

  • Functional specifications: The device must meet predefined physical dimensions, strength, flow rates, balloon inflation characteristics, etc., as per engineering designs and adopted standards.
  • Biocompatibility standards: The materials used must not cause adverse biological reactions, verified through in vitro and in vivo tests following ISO 10993.
  • Sterility assurance: The device must be sterile, verified through sterilization validation and bioburden testing.
  • Clinical outcomes (implied): The mentioned Clinical Evaluation Report and Post Market Clinical Follow-Up Study would evaluate patient outcomes related to device use, but the details are not provided here.

8. The sample size for the training set

This information is not applicable. The device is a physical product, not an AI algorithm trained on data.

9. How the ground truth for the training set was established

This information is not applicable for the same reason as point 8.

In summary:

The provided document is a 510(k) summary for a physical medical device. The regulatory pathway for this type of device primarily relies on demonstrating substantial equivalence to a predicate device through non-clinical performance and safety testing, rather than the extensive clinical efficacy studies typical for novel therapies or the rigorous diagnostic performance studies for AI algorithms that your questions implicitly assume.

The "device performance" relies on meeting established international and national standards for physical characteristics, biocompatibility, and sterility, which are evaluated through a battery of non-clinical tests. Clinical data is mentioned (CER, PMCF), suggesting real-world performance observation, but details on methodologies, sample sizes, and specific clinical endpoints are not included in this summary.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 6, 2019

Biosensors International Pte. Ltd. Jane Chen Jiayan Junior Associate, Regulatory Affairs 36 Jalan Tukang 619266 Singapore Singapore

Re: K182463

Trade/Device Name: Silicone Foley Catheter Regulation Number: 21 CFR 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZL Dated: September 18, 2019 Received: September 25, 2019

Dear Jane Chen Jiayan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Glenn B. Bell, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182463

Device Name Silicone Foley Catheter

Indications for Use (Describe)

This product is used for temporary urethral catheterization or indwelling catheterization and bladder irrigation.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Biosensors International. The logo consists of a blue symbol that resembles a stylized letter "B" made up of three curved lines stacked on top of each other. Below the symbol, the words "BIOSENSORS" are written in a bold, sans-serif font, and underneath that, the word "INTERNATIONAL" is written in a smaller, sans-serif font with a trademark symbol next to it.

Section 5: 510(k) Summary

[As required by section 807.92(c)]

Silicone Foley Catheter (K182463)

1. Applicant Information

Company Name:Biosensors International Pte Ltd
Company Address:36 Jalan Tukang, Singapore 619266
Establishment Number:2084493
Contact Person:Jane Chen JiayanManager, QA/RAPhone: +65 6213 5777 (Ext: 735)Fax: +65 6213 5737Email: ja.chen@biosensors.com
Date Prepared:06 Nov 2019

2. Device Name and Classification

Trade Name:Silicone Foley Catheter
Regulation Number:21 CFR 876.5130
Regulation Name:Urological Catheter and Accessories
Regulatory Class:II
Common Name:Catheter, Retention Type, Balloon
Product Code:EZL

3. Predicate Device

Disposable Silicone Foley Catheter (K130908)

4. Device Description

The Silicone Foley Catheter uses the pressure of the bladder to excrete urine. The balloon of the catheter can be inflated and fixed the catheter in place after being injected with a sterile liquid. Foley catheter (2-way) can be used for bladder drainage, and Foley catheter (3-way) also can be used for bladder irrigation.

Female type Foley Catheter is equipped with the advancer which can be used instead of tweezers during the catheterization.

The Silicone Foley Catheter is for single use, is intended for short-term (less than 14 days) use, and is provided with sterile.

Intended population: Pediatric, Male and Female

Biosensors International Pte Ltd 36 Jalan Tukang, Singapore 619 Phone: +65 6213 577 Fax: +65 6213 5737 o. Req No. 198904428 www.biosensors-ccp.c

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for Biosensors International. The logo features a stylized blue symbol resembling three interconnected "B" shapes stacked vertically. Below the symbol, the word "BIOSENSORS" is written in a bold, sans-serif font. Underneath "BIOSENSORS", the word "INTERNATIONAL" is written in a smaller, sans-serif font, accompanied by the trademark symbol.

5. Indications for Use

This product is used for temporary urethral catheterization or indwelling catheterization and bladder irrigation.

6. Summary of Technological Characteristics of Device in relation to Predicate Device

The technology of Biosensors Silicone Foley Catheter (K182463) is substantially equivalent to the Disposable Silicone Foley Catheter manufactured by Guangdong Baihe Medical Technology Co.,Ltd (K130908), with similar indications for use and design construction. The safety and effectiveness of Biosensors Silicone Foley Catheter have been assessed and discussed further in section 7 below.

Subject Device – K182463Predicate Device - K130908
Common NameCatheter, Retention Type, BalloonSame
Product CodeEZLSame
Regulation Number876.5130Same
ClassIISame
PopulationPediatric, male and femaleSame
Indications for UseThis product is used for temporary urethral catheterization or indwelling catheterization and bladder irrigation.Similar
French Size6 - 30 FrSimilar
Length$410 \pm 10mm$310mm and 400mm
LumenTwo-way and Three-waySame
BalloonYesSame
Balloon Size5ml, 5-15ml, 20-30ml, 30ml1.5ml, 3ml, 5ml, 10ml, 15ml, 20ml, 30ml
Single UseYesSame
SterileYesSame
Main Shaft MaterialSiliconeSame
Validity Period2 YearsSame
Indwelling TimeNo more than 14 daysNo more than 30 days
Use with AccessoryYesNo

Comparison between Subject Device and Predicate Device:

Biosensors International Pte Ltd 36 Jalan Tukang, Singapore 619266 Phone: +65 6213 5777 Fax: +65 6213 5737 Co. Reg No. 198904428G www.biosensors-ccp.con

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains the logo for Biosensors International. The logo features a stylized blue symbol resembling three interconnected arches stacked vertically. Below the symbol, the word "BIOSENSORS" is written in bold, uppercase letters. Underneath "BIOSENSORS", the word "INTERNATIONAL" is written in a smaller font size, followed by the trademark symbol.

7. Description of Non-Clinical Testing

Testina includes:

  • . Bench testing of the functional performance
  • l Biocompatibility testing
  • I Sterilization validation
  • Shelf life validation
  • . Simulation Shipment validation.

Standards applied to Non-clinical Testing includes:

  • l Functional performance testing: ASTM F623-99(2013), ISO 2859-1:1999, ISO 8536-4:2010, USP
  • . Biocompatibility testing: ISO 10993-3: 2014 Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity ISO 10993-5:2009 Part 5: Tests for in vitro cytotoxicity

ISO 10993-5:2009 Part 5: Tests for in vitro cytotoxicity
ISO 10993-9:2019 Part 9: Tests for potential degradation

  • ISO 10993-6: 2016 Part 6: Tests for local effects after implantation
  • ISO 10993-10: 2010 Part 10: Tests for irritation and skin sensitization
  • ISO 10993-11: 2017 Part 11: Tests for systemic toxicity
  • . Sterilization validation: ISO 11135: 2014, ISO11138-1:2017, ISO 11138-2:2017, ISO 10993-7: 2008/AC:2009, ISO 11737-1:2018, ISO 11737-2: 2009
  • Shelf life and simulation shipment validation: ASTM F623-99(2013), ASTMF88/F88M-15, ASTMF1929-15, ASTM F1886/F1886M-16, ISO 11135-2014, EN ISO11607-1:2017, EN ISO11607-2:2017

Results demonstrate that the device is safe and effective in meeting user requirements in accordance with its intended use.

8. Description of Clinical data

Clinical Data includes:

  • Clinical Evaluation Report, CER-FC-001(A2) ■
  • . Silicone Foley Catheter Post Market Clinical Follow-Up Study, PMCF-2018-01(A1)
  • . Validation protocol and report for simulation model test of Silicone Foley Catheter -Advancer, VPFC-2018-009(01) and TRFC-2018-009 (01)
  • Clinical Data - Evaluation Form of Silicone Foley Catheter

Results of Clinical evaluation, simulation model test with advancer and Clinical Data demonstrate the safety and the effectiveness of Silicone Foley Catheter and using with the accessory - Advancer.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Biosensors International. The logo consists of a stylized blue symbol resembling three stacked, rounded rectangles, positioned above the company name. The text "BIOSENSORS" is written in a bold, sans-serif font, and below it, "INTERNATIONAL" is written in a smaller, sans-serif font with a trademark symbol.

9. Conclusion

The Silicone Foley Catheter has the similar intended use, technological characteristics and performance characteristics as its predicate device. Performance test results demonstrate that the subject device meets its intended use. It is for these reasons that the subject device can be found substantially equivalent with its predicate device.

Biosensors International Pte Ltd 36 Jalan Tukang, Singapore 619266
Phone: +65 6213 5777
Fax: +65 6213 5777
Co. Reg No. 1989044280G
www.biosensors-cop.com

§ 876.5130 Urological catheter and accessories.

(a)
Identification. A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).(b)
Classification. (1) Class II (performance standards).(2) Class I for the ureteral stylet (guidewire), stylet for gastrourological catheter, ureteral catheter adapter, ureteral catheter connector, and ureteral catheter holder. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.