Polychloroprene Powder Free Sterile Surgical Gloves, White, Tested for Use with Chemotherapy Drugs
K182176 · Pt. Medisafe Technologies · KGO · Mar 28, 2019 · General, Plastic Surgery
Device Facts
| Record ID | K182176 |
| Device Name | Polychloroprene Powder Free Sterile Surgical Gloves, White, Tested for Use with Chemotherapy Drugs |
| Applicant | Pt. Medisafe Technologies |
| Product Code | KGO · General, Plastic Surgery |
| Decision Date | Mar 28, 2019 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4460 |
| Device Class | Class 1 |
Intended Use
This surgeon's glove is a device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. This glove is also tested for use with Chemotherapy Drugs and its permeation time is listed as below.
Device Story
Polychloroprene powder-free sterile surgical glove; synthetic rubber construction. Used by operating room personnel to provide a barrier against contamination during surgical procedures. Tested for permeation resistance against various chemotherapy drugs (e.g., Carmustine, Cisplatin, Cyclophosphamide, Thiotepa). Provides protection to the wearer when handling hazardous chemotherapy agents. Output is a physical barrier; no electronic or software components.
Clinical Evidence
Bench testing only. Permeation testing performed for 14 chemotherapy drugs to determine breakthrough detection times. Results provided for Carmustine (37.5 min) and Thiotepa (58.3 min), with others exceeding 240 minutes.
Technological Characteristics
Material: Polychloroprene (synthetic rubber). Form factor: Sterile surgical glove. Powder-free. Tested for chemotherapy drug permeation resistance.
Indications for Use
Indicated for use by operating room personnel to protect surgical wounds from contamination. Tested for use with specific chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Related Devices
- K173258 — Polychloroprene Powder Free Sterile Surgical Gloves, White, Tested for Use with Chemotherapy · Pt. Medisafe Technologies · Apr 20, 2018
- K201774 — Polychloroprene Powder Free Sterile Surgical Gloves, Low Dermatitis Potential and Tested for Use with Chemotherapy Drugs · Pt. Medisafe Technologies · Sep 16, 2020
- K172420 — Sterile Powder Free Nitrile Surgical Glove, White Colored. Tested for Use with Chemotherapy Drugs, with Non-Pyrogenic Labeling Claim · Wear Safe (Malaysia) Sdn. Bhd. · Sep 29, 2017
- K142190 — Polychloroprene Powder-free Surgical Glove (White) Tested for Use with Chemotherapy Drugs; Polychloroprene Powder-free Surgical Glove (Green) Tested for Use with Chemotherapy Drugs · Pt. Medisafe Technologies · Jul 14, 2015
- K171898 — Sterile Polyisoprene Powder Free Surgical Gloves, Tested for Use With Chemotherapy Drugs · Better Care Plastic Technology Co., Ltd. · Aug 18, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue.
March 28, 2019
PT. Medisafe Technologies Punitha Samy Group Regulatory Affairs Manager JI Batang Kuis. Gg Tambak Rejo Pasar IX, Desa Buntu Bedimbar Tanjung Morawa, Medan, 20362 Id
Re: K182176
Trade/Device Name: Polychloroprene Powder Free Sterile Surgical Gloves, White, Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I Product Code: KGO, LZC Dated: February 22, 2019 Received: March 1, 2019
Dear Punitha Samy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Clarence W. Murray lii III -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182176
#### Device Name
Polychloroprene Powder Free Sterile Surgical Gloves, White, Tested for Use with Chemotherapy Drugs
#### Indications for Use (Describe)
This surgeon's glove is a device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. This glove is also tested for use with Chemotherapy Drugs and its permeation time is listed as below.
| Test Chemotherapy Drug and Concentration | Breakthrough Detection Time (Minutes) |
|--------------------------------------------------|---------------------------------------|
| Carmustine (BCNU) 3.3 mg/ml (3,300 ppm) | 37.5 (47.8, 38.3, 37.5) |
| Cisplatin 1.0 mg/ml (1,000ppm) | >240 min |
| Cyclophosphamide (Cytoxan) 20 mg/ml (20,000 ppm) | >240 min |
| Dacarabzine (DTIC) 10.0 mg/ml (10,000 ppm) | >240 min |
| Doxorubicin Hydrochloride 2.0 mg/ml (2,000 ppm) | >240 min |
| Etoposide (Toposar) 20.0 mg/ml (20,000 ppm) | >240 min |
| Fluorouracil 50.0 mg/ml (50,000 ppm) | >240 min |
| Ifosfamide 50.0 mg/ml (50,000 ppm) | >240 min |
| Methotrexate 25 mg/ml (25, 000 ppm) | >240 min |
| Mechlorethamine HCI 1.0 mg/ml (1,000 ppm) | >240 min |
| Melphalan 5 mg/ml (5,000 ppm) | >240 min |
| Paclitaxel (Taxol) 6.0 mg/ml (6,000 ppm) | >240 min |
| Thiotepa 10.0 mg/ml (10,000 ppm) | 58.3 (69.8, 68.6, 58.3) |
| Vincristine Sulfate 1.0 mg/ml (1,000 ppm) | >240 min |
Please note that the following drugs have low permeation times: Carmustine (3.3 mg/ml): 37.5 minutes ThioTEPA (10.0 mg/ml): 58.3 minutes
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."