ScopeFlow Pure(TM)
K181418 · Partnership Medical Limited · OCX · Jul 11, 2019 · Gastroenterology, Urology
Device Facts
| Record ID | K181418 |
| Device Name | ScopeFlow Pure(TM) |
| Applicant | Partnership Medical Limited |
| Product Code | OCX · Gastroenterology, Urology |
| Decision Date | Jul 11, 2019 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.1500 |
| Device Class | Class 2 |
Intended Use
The ScopeFlow Pure™ is intended to be used in in conjunction with a flexible endoscope to insufflate air/CO2 and supply sterile water during endoscopic procedures. The ScopeFlow Pure™ is compatible with the different sterile water bottles available. It is intended to be used for 24 hours and then disposed of.
Device Story
ScopeFlow Pure™ is a single-use endoscopic accessory system; connects to external sterile water source and flexible endoscope (Olympus, Pentax, Fujinon models). Device utilizes dual-parallel PVC tubing to convey sterile water for lens cleaning/visual field improvement and to insufflate air/CO2. Hybrid configurations integrate with EndoStream™ units to supply irrigation from a single sterile water source. Operated by clinical staff during endoscopic procedures; provides continuous irrigation/insufflation support. Benefits include improved visualization and procedural efficiency. Device is disposed of after 24 hours of use.
Clinical Evidence
Bench testing only. Performance evaluation included sterilization validation per ISO 11135, biocompatibility testing (cytotoxicity, intracutaneous irritation, sensitization) per ISO 10993, and functional integrity testing.
Technological Characteristics
Materials: Medical-grade plastic housing, medical-grade TPE/silicone seals, PVC tubing. Principle: Dual-parallel tube irrigation/insufflation. Connectivity: Mechanical interface to endoscope and sterile water bottle. Sterilization: Ethylene Oxide. Form factor: 15 configurations based on endoscope compatibility (Olympus, Pentax, Fujinon).
Indications for Use
Indicated for use in conjunction with flexible endoscopes to insufflate air/CO2 and supply sterile water during endoscopic procedures. Compatible with various sterile water bottles; intended for 24-hour use before disposal.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
- EndoSmart Bottle, Rinse and Insufflation System (K140753)
- Endo SmartCap (K971125)
Related Devices
- K140753 — ENDO SMART BOTTLE, RINSE AND INSUFFLATION SYSTEM · Medivators · Jul 15, 2014
- K173650 — AquaShield System CO2- Fujifilm · United States Endoscopy Group, Inc. · Jan 5, 2018
- K102855 — UNIVERSAL IRRIGATION SOLUTION HYBRID · Byrne Medical, Inc. · Nov 19, 2010
- K221310 — Endoscopic Clean Connecting Tubes, Endoscopic CO2 Source Tubing · Changzhou Endoclean Medical Device Co., Ltd. · Oct 7, 2022
- K153362 — AquaShield system CO2, AquaShield system-PENTAX · United States Endoscopy Group, Inc. · Dec 16, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 11, 2019
Partnership Medical Limited Silbiano Gonzales Consultant QRC Consulting, LLC 10422 Huebner Road, Apt# 508 San Antonio, TX 78240
Re: K181418
Trade/Device Name: ScopeFlow Pure™ Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OCX Dated: June 6, 2019 Received: June 10, 2019
Dear Silbiano Gonzales:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen, Ph.D. Acting Assistant Division Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181418
Device Name ScopeFlow Pure™
Use (Describe)
The ScopeFlow Pure™ is designed to be used in conjunction with a flexible endoscope to insufflate air/CO2 and supply The Scoperiow Purem is designed to be used in voopeFlow Pure™ is compatible with the different sterile water bottles available. It is intended to be used for 24 hours and then disposed of
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the I he burden time for this collection of information to coes, gather and maintain the data needed and completer as any other aspect time to review instruction. Search existing ductions regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### l. SUBMITTER
| Applicant Name:<br>Address: | Partnership Medical Limited<br>Hartley House, Galveston Grove<br>Fenton, Stoke on Trent, ST4 3PE |
|-----------------------------|--------------------------------------------------------------------------------------------------|
| Contact Person: | Rob Hartley, Director |
| US Contact / Correspondent: | QRC Consulting, LLC.<br>Silbiano Gonzales, Consultant |
| Address: | 5623 Hamilton Wolfe Rd.,<br>Apt# 921 San Antonio, TX<br>78240 |
| Phone:<br>Email: | (210) 287-9558<br>silbianog@sbcglobal.net |
#### DEVICE II.
| Trade Name: | |
|-----------------|--|
| Catalog Number: | |
ScopeFlow Pure™
| CATALOG (PART) NUMBERS | | | | |
|------------------------|-----------|------------|-------------------|-------------------|
| SFP145 | SFP145CO2 | SFP145CO2S | SFP1451<br>HYBRID | SFP1452<br>HYBRID |
| SFP160 | SFP160CO2 | SFP160CO2S | SFP1601<br>HYBRID | SFP1602<br>HYBRID |
| SFP165 | SFP165CO2 | SFP165CO2S | SFP1651<br>HYBRID | SFP1652<br>HYBRID |
Common or Usual Name: Regulation Name: Regulation Number: Device Class: Product Code:
Endoscopic Irrigation/Suction System Endoscope and accessories 21 CFR 876.1500 ll ОСХ
The Gastroenterology/Urology devices panel has classified Endoscopic Irrigation/Suction System as Class II under 21 CFR §870.1500. OCX is the product code that has been assigned for these types of devices.
PART 876 -- GASTROENTEROLOGY-UROLOGY-UROLOGY DEVICES
{4}------------------------------------------------
Subpart B--Diagnostic Devices
Sec. 876.1500 Endoscope and accessories
(a) Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and mav include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
(b) Classification. (1) Class II (performance standards).
(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b) (2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in 876.9.
Product Code: OCX
Device: Endoscopic irrigation/suction system
Definition: To supply sterile water, other solutions and/or suction to endoscopes during endoscopic procedures.
#### III. DEVICE DESCRIPTION
The ScopeFlow Pure™ is available in fifteen (15) configurations:
| CATALOG (PART) NUMBERS | | | | |
|------------------------|-----------|------------|-------------------|-------------------|
| SFP145 | SFP145CO2 | SFP145CO2S | SFP1451<br>HYBRID | SFP1452<br>HYBRID |
| SFP160 | SFP160CO2 | SFP160CO2S | SFP1601<br>HYBRID | SFP1602<br>HYBRID |
| SFP165 | SFP165CO2 | SFP165CO2S | SFP1651<br>HYBRID | SFP165<br>HYBRID |
{5}------------------------------------------------
The ScopeFlow Pure™ is used to convey sterile water from an external water source to a flexible endoscope to aid its function.
The ScopeFlow Pure™ consists of:
1) PVC tubing is used provide the method of transferring the sterile water from the external source to the flexible endoscope. This sterile water is used to clean the lens of the flexible endoscope during operation or to assist in improving the flexible endoscopes visual field.
2) An endoscope connector at the distal end of the device. For the ScopeFlow Pure™ the three different connectors are designed to fit Olympus, Pentax and Fujinon endoscopes. The connectors are manufactured from medical grade plastic, which is used as an outer housing and medical grade TPE or silicone which are used as seals to prevent water and air from escaping during use.
The ScopeFlow Pure™ functions by using a two parallel tubes configuration to supply air to the water bottle and provide water to the flexible endoscope.
The HYBRID configuration consists of the ScopeFlow Pure™ and an EndoStream™ units being used together. The HYBRID models function in the same way but instead of using a separate sterile bottle for irrigation of the lens and irrigation of the GI Tract, one bottle of sterile water is used as the feed for both, the Hybrid has only one screwcap.
{6}------------------------------------------------
The ScopeFlow Pure™ is available in fifteen (15) configurations, according to endoscope manufacturer:
| PFE/PML Product Code | Endoscope type |
|----------------------|----------------------------------------------------------------------------------------------------|
| SFP145 | ScopeFlow Pure™ for Olympus 140 & 240 series endoscopes and higher |
| SFP145CO2 | ScopeFlow Pure™ for Olympus 140 & 240 series endoscopes and higher with CO2 |
| SFP145CO2S | ScopeFlow Pure™ for Olympus 140 & 240 series endoscopes and higher with CO2 (short) |
| SFP1451 HYBRID | ScopeFlow Pure™ for Olympus 140 & 240 series endoscopes and higher with CO2 and EndoStream™ PFE130 |
| SFP1452 HYBRID | ScopeFlow Pure™ for Olympus 140 & 240 series endoscopes and higher with CO2 and EndoStream™ PFE230 |
| SFP160 | ScopeFlow Pure™ for Pentax endoscopes |
| SFP160CO2 | ScopeFlow Pure™ for Pentax endoscopes with CO2 |
| SFP160CO2S | ScopeFlow Pure™ for Pentax endoscopes with CO2 (short) |
| SFP1601 HYBRID | ScopeFlow Pure™ for Pentax endoscopes with CO2 (short) and EndoStream™ PFE130 |
| SFP1602 HYBRID | ScopeFlow Pure™ for Pentax endoscopes with CO2 (short) and EndoStream™ PFE230 |
| PFE/PML Product<br>Code | Endoscope type |
|-------------------------|-------------------------------------------------------------------------|
| SFP165 | ScopeFlow Pure™ for Fujinon endoscopes G5<br>and newer |
| SFP165CO2 | ScopeFlow Pure™ for Fujinon endoscopes G5<br>and newer with CO2 |
| SFP165CO2S | ScopeFlow Pure™ for Fujinon endoscopes G5<br>and newer with CO2 (short) |
| SFP1651 HYBRID | ScopeFlow Pure™ for Fujinon endoscopes<br>with CO2 & EndoStream™ PFE230 |
| SFP1652 HYBRID | ScopeFlow Pure™ for Fujinon endoscopes<br>with CO2 & EndoStream™ PFE130 |
{7}------------------------------------------------
#### IV. INDICATIONS FOR USE
The ScopeFlow Pure™ is intended to be used in in conjunction with a flexible endoscope to insufflate air/CO2 and supply sterile water during endoscopic procedures. The ScopeFlow Pure™ is compatible with the different sterile water bottles available. It is intended to be used for 24 hours and then disposed of.
#### V. PREDICATE DEVICES
The ScopeFlow Pure™ is designed to be used with commercially available endoscopes and intended to be used in conjunction with a flexible endoscope to insufflate air and supply sterile water during endoscopic procedures. The ScopeFlow Pure™ is compatible with different commercially available sterile water bottles.
The ScopeFlow Pure™ legally marketed predicates, to which Partnership Medical Limited is claiming equivalence are:
| Predicate Device for the ScopeFlow Pure™ | | |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Company | Predicate Device Name | FDA 510(k) Number |
| Medivators | EndoSmart Bottle, Rinse and<br>Insufflation System [Model: 100145U /<br>100150U / 100160U / 100165CO2U /<br>100145CO2U / 100150CO2U /<br>100145CO2EXTU / 100150CO2EXTU /<br>100165CO2EXTU] | K140753 |
| Endo<br>SmartCap<br>Company | The Endo SmartCap [Model: 100145 /<br>100150] | K971125 |
{8}------------------------------------------------
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS VI. WITH THE PREDICATE DEVICES
The ScopeFlow Pure™ is substantially equivalent to both the Medivators (Endo Smart Bottle Rinse and Insufflation System, K140753) and the Endo SmartCap Company (Endo Smart Cap, K971125).
| Characteristic<br>Compared | Our Device | Predicate 1 | Predicate 2 |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Partnership<br>Medical, Ltd | Endo<br>SmartCap™<br>Company | Medivators,<br>Inc. |
| Device | ScopeFlow<br>Pure™ | The Endo<br>SmartCap™ | Endo Smart<br>Bottle, Rinse<br>and Insufflation<br>System |
| 510(k) Number | - | K971125 | K140753 |
| Indication | The ScopeFlow<br>Pure™ is intended<br>to be used in<br>conjunction with a<br>flexible endoscope<br>to insufflate<br>air/CO2 and supply<br>sterile water during<br>endoscopic<br>procedures.<br>The ScopeFlow<br>Pure™ is<br>compatible with the<br>different sterile<br>water bottles<br>available. It is<br>intended to be<br>used for 24<br>hours and then<br>disposed of. | ENDO<br>SMARTCAP™<br>is intended to<br>supply sterile<br>water to series<br>10, 100 and<br>130 Olympus<br>endoscopes<br>when connected<br>to a commercially<br>available sterile<br>water bottle. | EndoSmart<br>Bottle is<br>intended to be<br>used with an<br>air or CO2<br>source and/or<br>a pump along<br>with a sterile<br>water source to<br>supply air or<br>CO2 and<br>sterile water to<br>an endoscope<br>during<br>endoscopic<br>procedures. |
{9}------------------------------------------------
| Characteristic<br>Compared | Our Device | Predicate 1 | Predicate 2 |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|----------------|----------------|
| Product<br>Code/Class | OCX | OCX | OCX |
| Principle of<br>Operation | used in<br>conjunction with<br>a flexible<br>endoscope to<br>insufflate air<br>and supply<br>sterile water<br>during<br>endoscopic<br>procedures. | Similar | Similar |
| Difference | No difference | No difference | No difference |
| Materials of<br>Construction | Per section 16 | Identical | Identical |
| Sterilization | Ethylene<br>Oxide | Ethylene Oxide | Ethylene Oxide |
## VII. PERFORMANCE DATA
Performance testing was completed (per respective standards) for sterilization validation (ISO 11135), package and shelf-life testing and biocompatibility.
The ScopeFlow Pure™ product was evaluated for cytotoxicity, intracutaneous irritation and sensitization per biocompatibility requirements according to ISO 10993.
The ScopeFlow Pure™ products were also tested to demonstrate functionality and performance integrity.
#### VIII. CONCLUSIONS
Partnership Medical Limited considers the ScopeFlow Pure™ to be substantially equivalent to the legally marketed predicate device listed above, and safe and effective for the intended use. The conclusion is based on similarities in indications for use, materials, performance testing, technological characteristics, principle of operation and design features. Any differences do not raise any new issues of safety or effectiveness.