Irrigation Tubing Bipolar Cord Set

K180908 · New Deantronics Taiwan , Ltd. · GEI · May 18, 2018 · General, Plastic Surgery

Device Facts

Record IDK180908
Device NameIrrigation Tubing Bipolar Cord Set
ApplicantNew Deantronics Taiwan , Ltd.
Product CodeGEI · General, Plastic Surgery
Decision DateMay 18, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Irrigation Tubing Bipolar Cord Sets are intended to provide irrigation and energy simultaneously to bipolar forceps specifically designed for irrigation.

Device Story

Device is an electrosurgical accessory consisting of an integrated bipolar cord and irrigation tubing. It connects an RF electrosurgical generator to bipolar forceps and a saline source to the forceps. Co-extrusion manufacturing integrates the cord and tubing into a single unit to improve surgical organization. The cord uses 24 AWG copper wires with PVC insulation; the fluid catheter uses PVC and silicone tubing. Operated by a surgeon, the device facilitates simultaneous coagulation and irrigation. The saline flow is controlled by a surgeon-controlled fluid controller (rotary or gravity feed). The device is single-use and provided sterile. It benefits patients by enabling organized, simultaneous irrigation and coagulation during electrosurgical procedures.

Clinical Evidence

Bench testing only. Testing included electrical safety and performance verification against pre-defined acceptance criteria. Compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2 was demonstrated. No clinical data was required or provided.

Technological Characteristics

Integrated bipolar cord and irrigation tubing. Materials: PVC (cable insulation, tubing, connectors), silicone (tubing), ABS (proximal spike), polypropylene (clamp). Connectivity: Connects to RF generator and saline source. Dimensions: 3.6m electrical cable, 4.4m tubing. Sterilization: Provided sterile. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2. No software or algorithm.

Indications for Use

Indicated for use with bipolar forceps specifically designed for irrigation to provide simultaneous irrigation and electrosurgical energy.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. May 18, 2018 New Deantronics Taiwan Ltd. % Mr. Craig Coombs Coombs Medical Device Consulting, Inc 1193 Sherman St Alameda, California 94501 Re: K180908 Trade/Device Name: Irrigation Tubing Bipolar Cord Sets Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: April 4, 2018 Received: April 6, 2018 Dear Mr. Coombs: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); {1}------------------------------------------------ # Page 2 - Mr. Craig Coombs K180908 and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). > Sincerely, Jennif Steven For Binita S. Ashar, M.D., M.B.A., F.A.C.S Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K180908 Device Name Irrigation Tubing Bipolar Cord Sets Indications for Use (Describe) The Irrigation Tubing Bipolar Cord Sets are intended to provide irrigation and energy simultaneously to bipolar forceps specifically designed for irrigation. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X | Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for New Deantronics. The word "NEW" is in blue, and the word "DEANTRONICS" is in blue as well. There is a blue circle with a red dot in the middle between the two words. # Section 5: 510(k) Summary ## A. Device Information: | Category | Comments | |------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Sponsor: | New Deantronics Taiwan Ltd.<br>12F., No.51, Sec. 4, Chong Yang Rd.,<br>Tu Cheng Dist,<br>New Taipei City 23675, Taiwan R.O.C.<br>Tel: (886) 2-2268-1726<br>Fax: (886) 2-2268-3800 | | Correspondent Contact<br>Information: | Craig Coombs<br>Coombs Medical Device Consulting<br>1193 Sherman Street<br>Alameda, CA 94501<br>Tel: 510-337-0140 | | Device Common Name: | Electrosurgical accessory | | Device Classification Number: | 21 CFR 878.4400 | | Device Classification &<br>Product Code: | Class 2,<br>GEI | | Device Proprietary Name: | Irrigation Tubing Bipolar Cord Sets | ### Predicate Device Information: | Predicate Device: | Codman Integrated Bipolar Cord and<br>Tubing Set | |----------------------------------------------------|------------------------------------------------------------------| | Predicate Device Manufacturer: | Codman & Shurtleff, Inc. | | Predicate Device Common Name: | Electrosurgical accessory<br>Irrigation Tubing Bipolar Cord Sets | | Predicate Device Premarket Notification # | K163106 | | Predicate Device Classification: | 21 CFR 878.4400 | | Predicate Device Classification &<br>Product Code: | Class 2,<br>GEI | ### B. Date Summary Prepared 4 April 2018 # C. Description of Device The Irrigation Tubing Bipolar Cord Se is designed to connect to both an electrical generator and bipolar forceps via an electrically conducting cord and to a saline source for irrigation via a catheter to provide simultaneous coagulation capabilities and irrigation capabilities. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for New Deantronics. The word "NEW" is in blue, outlined in white, and is located on the left side of the logo. To the right of the word "NEW" is a blue circle with a red dot in the center. The word "DEANTRONICS" is in blue and is located to the right of the circle. A co-extrusion manufacturing process is applied to integrate a bipolar cord with irrigation tubing into one unit. This integrated Irrigation Tubing Bipolar Cord set are helpful to keep the bipolar electrosurgery organized. The electrically conducting cord is formed with a pair of copper wires each coaxially surrounded by an outer insulation sheath. The insulated bipolar cord (24 AWG) provides safety protection to patients and surgical staff. The fluid catheter is formed of PVC and silicone tubing connected by connectors and is defined as fluid passage pathway. # D. Indications for Use The Irrigation Tubing Bipolar Cord Sets are intended to provide irrigation and energy simultaneously to bipolar forceps specifically designed for irrigation. ## E. Comparison to Predicate Device As described below, the application New Deantronics Bipolar Cord Tubing Set is substantially equivalent in intended use, technology, design and physician use to the predicate Codman Integrated Bipolar Cord Tubing and Set (K163106). | Feature | Application Device:<br>Irrigation Tubing Bipolar<br>Cord Sets | Predicate Device:<br>Codman Integrated<br>Bipolar Cord Tubing<br>and Set (K163106) | Pertinence of Feature<br>to Consideration of<br>Substantial<br>Equivalence. | |-------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | The Irrigation Tubing<br>Bipolar Cord Sets are<br>intended to provide<br>irrigation and energy<br>simultaneously to bipolar<br>forceps specifically<br>designed for irrigation. | The CODMAN<br>Integrated Bipolar Cord<br>and Tubing Sets are<br>intended to provide<br>irrigation and energy<br>simultaneously to<br>bipolar forceps<br>specifically designed for<br>irrigation. | The Indications for Use<br>of the two devices are<br>identical, except for<br>name of the devices | | Product Code | GEI | GEI | Identical | | Technology | | | | | Mechanism of<br>Operation | The integrated Bipolar<br>Cord Set and Irrigation<br>tubing is designed to<br>connect an RF electrical<br>generator to bipolar<br>coagulating forceps via<br>an electrically conducting<br>cord. It is also intended<br>to connect a saline<br>source, via the irrigation<br>tubing. The saline flow<br>through the tubing is<br>controlled by a surgeon-<br>controlled fluid | Same | Identical | | Feature | Application Device:<br>Irrigation Tubing Bipolar<br>Cord Sets | Predicate Device:<br>Codman Integrated<br>Bipolar Cord Tubing<br>and Set (K163106) | Pertinence of Feature<br>to Consideration of<br>Substantial<br>Equivalence. | | | controller. This allows<br>for simultaneous,<br>controlled, coagulation<br>and irrigation<br>capabilities. | | | | Compatible w/<br>Monopolar or Bipolar<br>RF administration? | Bipolar only | Bipolar only | Identical | | Integral Electrical<br>Cable and Fluid<br>Tubing? | Yes | Yes | Identical | | Fluid feed<br>compatibility | Rotary or Gravity | Rotary or Gravity | Identical | | Design - Electrical<br>Cable | | | | | Cable length | 3.6 meters | 3.4 meters | Functionally Identical | | Wire Gauge | 24 AWG | 24 AWG | Identical | | Materials for cable<br>insulation | PVC | PVC | Identical | | Generator plug<br>design | Electrodes Consolidated<br>into a single plug with<br>banana plug connectors | Two separate plugs with<br>banana plug connectors | Visually different but<br>electrically identical. | | Plug Overmold | PVC | PVC | Identical | | Design - Saline<br>Tubing | | | | | Tubing length | 4.4 meters | 4.8 meters | Functionally identical | | Materials for<br>irrigation tubing | PVC and silicone | PVC and silicone | Identical | | Tubing Inner<br>Diameter | PVC: 3.0 mm<br>Silicone (Gravity Feed):<br>3.0 mm<br>Silicone (Rotary Pump):<br>0.8 mm | PVC: 3.0 mm<br>Silicone (Gravity Feed):<br>3.0 mm<br>Silicone (Rotary Pump):<br>0.8mm | Identical | | Tubing Flow rate<br>(Rotary type) | Min: 1.15cc/min (+/-30%)<br>Max: 16 cc/min (+/-30%) | Min: 1.15cc/min (+/-30%)<br>Max: 16 cc/min (+/-30%) | Identical | | Tubing Flow rate<br>(Gravity type) | Min: < 6 cc/min<br>Max: > 21 cc/min | Min: < 6 cc/min<br>Max: > 21 cc/min | Identical | | Proximal Spike<br>adapter for Saline<br>source | ABS | ABS | Identical | | Feature | Application Device:<br>Irrigation Tubing Bipolar<br>Cord Sets | Predicate Device:<br>Codman Integrated<br>Bipolar Cord Tubing<br>and Set (K163106) | Pertinence of Feature<br>to Consideration of<br>Substantial<br>Equivalence. | | Tubing Sections<br>Adaptor | PVC | PVC | Identical | | Tubing Clamp | Polypropylene | Celanese and Polyacetal | These materials do not<br>come in contact with the<br>patient or the saline<br>stream. Both clamps<br>have identical<br>functionality. No new<br>questions of safety or<br>efficacy are raised by<br>the material differences. | | Other Attributes | | | | | Single Use or<br>Reusable? | Single Use | Single Use | Identical | | Provided Sterilized? | Yes | Yes | Identical | | Performance/ Safety<br>Testing in accordance<br>with: | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2 | Identical | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for "NEW DEANTRONICS". The word "NEW" is in blue, and the word "DEANTRONICS" is in blue as well. There is a blue circle with a red dot in the middle between the two words. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for "NEW DEANTRONICS". The word "NEW" is in blue, and the word "DEANTRONICS" is in a larger, bolded blue font. There is a blue circle with a red dot in the middle between the two words. New Deantronics concludes that the devices are substantially equivalent. ### F. Summary of Supporting Data Electrical safety testing and bench testing has demonstrated that the performance of New Deantronics Irrigation Tubing Bipolar Cord Sets meet the requirements of its pre-defined acceptance criteria and intended uses. The results of the non-clinical testing demonstrate that the irrigation tubing bipolar cord sets are as safe and effective as the predicate devices. New Deantronics concludes that the New Deantronics Irrigation Tubing Bipolar Cord Sets are substantially equivalent to the predicate Codman Integrated Bipolar Cord Tubing and Set (K163106).
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%