3M Protemp Cement Temporary Dental Cement

K170330 · 3M Company · EMA · Feb 6, 2017 · Dental

Device Facts

Record IDK170330
Device Name3M Protemp Cement Temporary Dental Cement
Applicant3M Company
Product CodeEMA · Dental
Decision DateFeb 6, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3275
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

Temporary cementation of provisional restorations. Temporary cementation of final restorations.

Device Story

3M Protemp Cement is a radiopaque, non-eugenol, self-curing temporary dental cement; includes option for tack light-curing of excess material for easier cleanup. Supplied as base and catalyst paste in proprietary automix delivery dispenser with self-sealing cap; designed for lower extrusion force and reduced waste compared to traditional systems. Used by dental professionals in clinical settings for temporary fixation of provisional or final restorations. Benefits include eugenol-free formulation for patients with sensitivities and improved handling via automix delivery.

Clinical Evidence

Bench testing only. Evaluated working/setting times, bonding strength, film thickness, compressive strength, extrusion force, radiopacity, dimensional change, and fluoride release. Biocompatibility assessed per ISO 10993 series and ISO 7405. No animal or clinical testing submitted.

Technological Characteristics

Non-eugenol, methacrylate and water-based resin cement. Self-curing with optional tack light-cure. Contains inorganic fillers; releases fluoride. Radiopaque. Automix/multi-dose delivery system. Storage: 2°-27° C. Biocompatibility per ISO 10993-1, -3, -5, -6, -10, -11, -12 and ISO 7405.

Indications for Use

Indicated for temporary cementation of provisional and final dental restorations in patients requiring temporary fixation.

Regulatory Classification

Identification

Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features the department's name in a circular arrangement around a stylized emblem. The emblem consists of three abstract human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 6, 2017 3M ESPE Dental Products c/o Mark Job Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, MN 55313 Re: K170330 Trade/Device Name: 3M™ Protemp™ Cement Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: Class II Product Code: EMA Dated: January 31, 2017 Received: February 2, 2017 Dear Mark Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Susan Runno DDS, MA Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the 3M logo. The logo is in red and features the number "3" followed by the letter "M". The font is bold and sans-serif. The logo is simple and recognizable. #### Indications for Use Statement এ. DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration #### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. 510(k) Number (if known) Device Name 3MTM ProtempTM Cement Temporary Dental Cement Indications for Use (Describe) - · Temporary cementation of provisional restorations - · Temporary cementation of final restorations Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) Page 1 of 1 {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the 3M logo. The logo is red and features the number "3" followed by the letter "M". The font is bold and sans-serif. The logo is simple and recognizable. 3. 510(k) Summary > 3M ESPE Dental Products 2510 Conway Avenue St. Paul, MN 55144-1000 Image /page/3/Picture/4 description: The image shows the 3M company logo. The logo is red and features the number "3" followed by the letter "M". The logo is simple and recognizable, and it is often used to represent the 3M company in various contexts. # 510(k) Summary This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92. | 510(k) Submitter | 3M ESPE Dental Products<br>2510 Conway Avenue<br>St. Paul, MN 55144-1000 USA | |------------------|------------------------------------------------------------------------------| |------------------|------------------------------------------------------------------------------| | Contact person | Idalia M. Stark, MS | |----------------|-------------------------------| | | Regulatory Affairs Specialist | | | Phone: (651) 733-0168 | | | Fax: (651) 737-9665 | | | imstark@mmm.com | | Submission Date: | January 31, 2017 | |------------------------|------------------------------------------------| | Trade Name: | 3M™ Protemp™ Cement<br>Temporary Dental Cement | | Classification Name: | Dental Cement | | Regulation Number: | 21 CFR 872.3275 | | Product Code: | EMA | | Classification Panel: | Dental | | Device Classification: | Class II | Predicate Device: Temp-Bond™ NE (K003658) Reference Device: RelyX™ Luting Plus Automix Resin Modified Glass Ionomer Cement (K11185) {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the 3M logo. The logo is in red and features the number '3' followed by the letter 'M'. The font is bold and sans-serif. The logo is simple and recognizable. ### Description of Device: 3M™ Protemp™ Cement Temporary Dental Cement is a radiopaque, temporary cement used for temporary or provisional fixation of dental restorations. It is a selfcuring cement, with an option for tack light curing of any excess cement for ease of clean up. As a non-eugenol cement, Protemp™ Cement Temporary Dental Cement is suitable for use on patients with sensitivity to eugenol. Protemp™ Cement Temporary Dental Cement consists of a base and catalyst paste packaged in a proprietary automix delivery dispenser with a self-sealing cap. The automix delivery dispenser offers convenience over traditional hand mixed cements. Compared to leading automix systems, the proprietary delivery dispenser is designed to generate less waste and also has a lower extrusion force for ease of use. It is available in two shades. Opaque and Translucent. | | Protemp™ Cement<br>Temporary Dental Cement | Temp-Bond™ NE<br>K003658 | |---------------------|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | Temporary Dental Cement | Temporary Dental Cement | | Indications for Use | Temporary cementation of provisional restorations. Temporary cementation of final restorations. | Temporary cement used for the cementation of temporary restorations.1 Temp-Bond is a temporary cement for trial cementing restorations or cementing temporary crowns and bridges.2 | ## Intended Use/Indications for Use Comparison: 1 Indications from FDA 510(k) clearance letter enclosure. 2 Indications from Kerr website: https://www.kerr-restoratives/temp-bond-temporary-dental-cement {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the 3M logo. The logo is in red and features the number "3" followed by the letter "M". The font is bold and sans-serif. The logo is simple and recognizable. # Technological Properties Comparison: | PROPERTIES | Protemp™ Cement<br>Temporary Dental<br>Cement | Temp-Bond™ NE<br>K003658 | |------------------------------------------------------------------|-------------------------------------------------------------------------|--------------------------------------------------| | Formulation | Non-eugenol<br>Methacrylate and<br>water based resin | Non-eugenol<br>Zinc oxide | | Intended Use | Temporary dental<br>cement | Temporary dental cement | | Recommended Cure<br>mode | Self-curing<br>with option for tack<br>light curing of excess<br>cement | Self-curing | | Inorganic Fillers | Yes | Yes | | Releases Fluoride | Yes* | No | | Radiopaque | Yes | Yes | | Manufacturer<br>Recommended<br>Working Time<br>(minutes:seconds) | 1:30 | ≥ 1:30 | | Manufacturer<br>Recommended<br>Setting Time<br>(minutes:seconds) | 5:00 after seating in<br>the mouth | ≤ 7:00 after mixing | | Available shades | Opaque/Translucent | Universal | | Dispensing System | Automix/Multi-dose | Automix/Multi-dose Unidose® Tube-Tube/Multi-dose | | Storage conditions | 2°-27° C/<br>35°-80° F | Ambient temperature | * - Fluoride release profile supplied in 510(k) per FDA Guidance Dental Cements – Premarket Notification (August 18, 1998) {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the 3M logo. The logo is in red and features the number 3 followed by the letter M. The logo is simple and recognizable. ## Performance Testing: Bench testing consistent with the clinical application and intended use was used to evaluate the performance of Protemp™ Cement Temporary Dental Cement compared to the predicate device Temp-Bond™ NE (K003658). Properties evaluated include: working and setting times, bonding strength, optimum film thickness, compressive strength, extrusion force, radiopacity, dimensional change and fluoride release profile. Based on the testing results, 3M ESPE Dental Products concludes that Protemp™ Cement Temporary Dental Cement is substantially equivalent to the predicate device Temp-Bond™ NE. ### Animal and Clinical Testing No animal or clinical testing was submitted in this 510(k). ### Biocompatibility Testing 3M's biocompatibility assessment, developed for this new product using standard risk assessment techniques of ISO14971:2007 and consideration of FDA & internationally recognized guidelines including: - ISO 10993-1;2009 ● - ISO 10993-3:2014 ● - ISO 10993-5:2009 ● - ISO 10993-6:2007 ● - ISO 10993-10:2010 ● - ISO 10993-11:2006 - ISO 10993-12:2012 . - ISO 7405:2008/Amd 1 2013 The assessment concluded that the product is biocompatible for its intended use. ### Conclusion Protemp™ Cement Temporary Dental Cement is substantially equivalent to the predicate device Temp-Bond™ NE in terms of intended use, indications for use, dispensing system, technological characteristics, and physical and mechanical properties. The information presented in this 510(k) submission demonstrate that {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the 3M logo. The logo is red and features the number 3 followed by the letter M. The logo is simple and recognizable. Protemp™ Cement Temporary Dental Cement is as safe, as effective and is demonstrated to be substantially equivalent to the predicate device.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%