Aesculap T-Space PEEK and XP Spinal System
K151056 · Aesculap Implant Systems, LLC · MAX · Aug 13, 2015 · Orthopedic
Device Facts
| Record ID | K151056 |
| Device Name | Aesculap T-Space PEEK and XP Spinal System |
| Applicant | Aesculap Implant Systems, LLC |
| Product Code | MAX · Orthopedic |
| Decision Date | Aug 13, 2015 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3080 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The T-Space Spinal Implant System are indicated for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 Spondylolisthesis at involved levels. These patients may have had previous non-fusion spinal surgery at the involved spinal level(s). The AIS T-Space Spinal Implant System is intended for use with supplemental spinal fixation systems that have been cleared for use in the lumbrosacral spine (i.e., posterior pedicle screw and rod systems, and anterior screw and rod systems). The AIS T-Space Spinal Implant System implants can be used individually or in pairs. The AIS T-Space Spinal Implant System is also intended for use with autogenous bone graft. Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the Aesculap device.
Device Story
Intervertebral body fusion device; implanted into vertebral body space to improve spinal stability and support fusion. Available in coated and uncoated versions; various shapes/sizes to match patient anatomy. Used in conjunction with supplemental spinal fixation systems (pedicle screw/rod or anterior plate/screw systems) and autogenous bone graft. Implanted by surgeons in clinical settings. Provides mechanical support to facilitate fusion; benefits patients by stabilizing degenerated spinal segments.
Clinical Evidence
Bench testing only. Performed static torsion, axial compression, shear resistance, and subsidence per ASTM F2077 and F2267. Coated implants underwent additional testing per FDA guidance for metallic plasma sprayed coatings, including microstructure (ASTM F1854), tensile strength (ASTM F1147), shear strength (ASTM F1044), shear fatigue (ASTM F1160), and abrasion resistance (ASTM F1978). Results demonstrate performance meeting or exceeding predicates.
Technological Characteristics
Intervertebral fusion implant. Materials: PEEK-Optima (ASTM F2026) for uncoated; titanium vacuum plasma spray coating (Plasmapore, ISO 5832-3) for coated version. Includes tantalum radiographic markers (ASTM F-560). Mechanical principle: load-bearing spacer for spinal fusion. No software or energy source.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1). DDD defined as discogenic back pain with radiographic confirmation. Includes patients with up to Grade 1 Spondylolisthesis or retrolisthesis and those with prior non-fusion surgery at the target level. Requires 6 months of failed non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
- Spinal Elements Lucent System (K122967)
- AIS PEEK VBR and Intervertebral Body Fusion Systems (K060762 / K071983)
- CeSpace XP Intervertebral Body Fusion Device (K123909)
- ProSpace XP Intervertebral Body Fusion Device (K132421)
Related Devices
- K173022 — SeaSpine Ventura NanoMetalene System · SeaSpine Orthopedics Corporation · Oct 26, 2017
- K192132 — SeaSpine Beachside System · SeaSpine Orthopedics Corporation · Dec 3, 2019
- K163707 — NuVasive CoRoent Ti-C System · Nu Vasive, Incorporated · May 1, 2017
- K182406 — Altus Spine Interbody Fusion System · Altus Partners, LLC · Feb 22, 2019
- K163481 — HALF DOME Posterior Lumbar Interbody System · Astura Medical · May 3, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of three human profiles facing to the right, arranged in a way that they appear to be interconnected or overlapping. The profiles are rendered in a dark color, creating a silhouette effect. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profile graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 13, 2015
Aesculap Implant Systems, LLC Ms. Lisa M. Boyle Regulatory Affairs Manager 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K151056
Trade/Device Name: Aesculap T-Space PEEK and XP Spinal System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: July 16, 2015 Received: July 17, 2015
Dear Ms. Boyle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 – Ms. Lisa M. Boyle
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151056
Device Name Aesculap T-Space PEEK and XP Spinal System
Indications for Use (Describe)
When used as an Intervertebral Body Fusion System:
The T-Space Spinal Implant System are indicated for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 Spondylolisthesis at involved levels. These patients may have had previous non-fusion spinal surgery at the involved spinal level(s). The AIS T-Space Spinal Implant System is intended for use with supplemental spinal fixation systems that have been cleared for use in the lumbrosacral spine (i.e., posterior pedicle screw and rod systems, and anterior screw and rod systems). The AIS T-Space Spinal Implant System implants can be used individually or in pairs. The AIS T-Space Spinal Implant System is also intended for use with autogenous bone graft.
Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the Aesculap device.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### B. 510(k) SUMMARY (as required by 21 CFR 807.92)
#### Aesculap® Implant Systems(AIS) - T-Space Spinal Implant System
August 10, 2015
| COMPANY: | Aesculap®Implant Systems (AIS), LLC.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 3005673311 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT: | Lisa M. Boyle, Regulatory Affairs Manager<br>610-984-9274 (phone)<br>610-791-6882 (fax)<br>lisa.boyle@aesculap.com |
| TRADE NAME: | Aesculap T-Space PEEK and XP Spinal System |
| COMMON NAME: | Intervertebral Fusion Device |
| CLASSIFICATION NAME: | Orthosis, Spinal Intervertebral Fusion |
| REGULATION NUMBER: | 21 CFR 888.3080 |
| PRODUCT CODE: | MAX |
| REVIEW PANEL: | Orthopedics |
### INDICATIONS FOR USE
When used as an Intervertebral Body Fusion System:
The T-Space Spinal Implant System are indicated for spinal fusion procedures at one or two contiguous levels (L2-S1) in skeletally mature patients with degenerative disc disease (DDD). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. DDD patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at involved levels. These patients may have had previous non-fusion spinal surgery at the involved spinal level(s). The AIS T-Space Spinal Implant System is intended for use with supplemental spinal fixation systems that have been cleared for use in the lumbrosacral spine (i.e., posterior pedicle screw and rod systems, anterior plate systems, and anterior screw and rod systems). The AIS T-Space Spinal Implants can be used individually or in pairs. The AIS T-Space Spinal Implant System is also intended for use with autogenous bone graft.
Patients must have undergone a regimen of at least six (6) months of non-operative treatment prior to being treated with the Aesculap device.
#### DEVICE DESCRIPTION
The T-Space Spinal Implant System is an intervertebral body fusion device that is implanted into the vertebral body space to improve stability of the spine while supporting fusion. The implants are offered in a coated and uncoated version and a variety of shapes and sizes to meet the requirements of the individual patient anatomy. The uncoated PEEK version is manufactured from PEEK - Optima (per ASTM F2026). The coated version will be the same in design as the uncoated PEEK implant with a titanium layer and a vacuum plasma spray coating (Plasmapore® - per ISO 5832-3). The device will have tantalum radiographic markers per ASTM F-560.
{5}------------------------------------------------
## TECHNOLOGICAL CHARACTERISTICS(compared to Predicate(s))
The components of the T-Space Spinal Implant System are offered in the same similar shapes and sizes as the predicate devices. The material used for the Aesculap® Implant Systems device is the same as that used to manufacture the predicate devices.
# PERFORMANCE DATA
As recommended by the FDA Guidance for Spinal System 510(k)'s, non-clinical testing was performed to demonstrate that the AIS T-Space implants (coated and uncoated) are substantially equivalent to other predicate devices. The following testing was performed:
- . Static torsion per ASTM F2077
- Static and dynamic axial compression per ASTM F2077 ●
- Shear resistance evaluation per ASTM F2267 .
- . Subsidence per ASTM F2267
In addition to FDA's Spine Guidance. Aesculap has also completed non-clinical testing recommended in the "Guidance for Industry on the Testing of Metallic Plasma Sprayed Coatings on Orthopedic Implants to Support Reconsideration of Postmarket Surveillance Requirements." The following tests were performed:
- Microstructure of the coating per ASTM F1854
- Static Tensile Strength per ASTM F1147
- Static Shear Strength per ASTM F1044
- Shear Fatigue Test per ASTM F1160
- Abrasion Resistance per ASTM F1978
The results of these tests showed that the T-Space Spinal Implant System meets or exceeds the performance of the predicate devices, and the device is therefore found to be substantially equivalent.
# SUBSTANTIAL EQUIVALENCE
AIS believes that the T-Space Spinal Implant System is substantially equivalent to the design of the Spinal Elements Lucent System (K122967) and AIS PEEK VBR and Intervertebral Body Fusion Systems (K071983). The Plasmapore® coating has been used and cleared in a number of legally marketed product lines manufactured by Aesculap (hip, knee, and spinal implants) for many years. The most recent Spinal Implants to be cleared with the Plasmapore® coating is the CeSpace XP Intervertebral Body Fusion Device (K123909) and ProSpace XP Intervertebral Body Fusion Device (K132421).
# PRIMARY PREDICATES:
- Spinal Elements Lucent System (K122967) ●
### ADDITIONAL PREDICATES:
- AIS PEEK VBR and Intervertebral Body Fusion Systems (K060762 / K071983). ●
- CeSpace XP Intervertebral Body Fusion Device (K123909) and.
- ProSpace XP Intervertebral Body Fusion Device (K132421). .