(252 days)
For the transfer of solutions from one container to another.
The B. Braun Universal Spike is a sterile, single use, disposable fluid transfer device consisting of a closure-piercing component capable of piercing and penetrating the closure of a fluid container, an air-inlet component which allows the dispensing from both rigid and pressure collapsible dispensing containers, and a distal female luer lock adapter which provides access for the attachment of a syringe or fluid transfer set. The closure-piercing component and luer lock adapter are covered by protective caps which prevent contamination of the device until its point of use.
The provided text describes a 510(k) premarket notification for a medical device called the "Universal Spike." The submission focuses on demonstrating substantial equivalence to a predicate device through performance testing. However, the document does not describe an AI/ML-driven device or study. All performance criteria and testing mentioned relate to the physical and functional characteristics of the Universal Spike (a fluid transfer device), not to any algorithmic performance.
Therefore, the requested information specifically related to AI/ML device acceptance criteria, performance metrics, sample sizes for test/training sets, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance cannot be extracted from this document as it is not relevant to the type of device being described.
Summary based on the provided document (not an AI/ML device):
The document details the submission for the Universal Spike, a sterile, single-use, disposable fluid transfer device. The submission aims to demonstrate substantial equivalence to a predicate device, the B. Braun Dispensing Pin with One-Way Valve (K943181), rather than proving the performance of an AI/ML algorithm.
1. Table of Acceptance Criteria and Reported Device Performance:
The document lists performance tests conducted, implying that successful completion of these tests (meeting referenced standards or specified criteria like "no leakage" or "minimum tensile force") demonstrates the device meets acceptance criteria for substantial equivalence.
| Characteristic / Test | Acceptance Criteria (Implied / Stated) | Reported Device Performance (Implied) |
|---|---|---|
| Chemical Analysis | Meets ISO 8536-4, ISO 10993-18 standards | Successfully met standards |
| Visual Control | Meets ISO 22413, ISO 8536 standards | Successfully met standards |
| Fragmentation Test | Meets ISO 22413 standard | Successfully met standard |
| Penetration Force | Meets ISO 8536-2 standard | Successfully met standard |
| Spike retention / seal ability | Meets ISO 8536-2 standard | Successfully met standard |
| Patency Test | Meets ISO 8536-4 standard | Successfully met standard |
| Pressure Test / Leakage | No leakage detected when subjected to air pressure of 450 mmHg | No leakage detected |
| Pull Test (cap and spike) | Withstands a minimum tensile force of separation between 4.41-44.13 Newtons | Withstood required tensile force |
| Free Flow | Meets ISO 22413 standard | Successfully met standard |
| Particulate Contamination | Meets USP <788> standard | Successfully met standard |
| Luer testing | Meets ISO 594-1 standard | Successfully met standard |
| Biocompatibility (Cytotoxicity, Hypersensitivity, | Meets ISO 10993-5, ISO 10993-10 (x2), ISO 10993-11 (x2), ASTM F756, USP <151> standards | Successfully met standards |
| Intracutaneous Reactivity, Acute Systemic | ||
| Toxicity, Hemocompatibility, Pyrogenicity) | ||
| Overall Conclusion | Demonstrates substantial equivalence to predicate device based on functional and performance testing | Device shown to be substantially equivalent to predicate |
2. Sample size used for the test set and the data provenance: Not applicable. This is not an AI/ML study, but performance testing of a physical medical device. The document does not specify the number of units tested for each criterion, nor does it refer to "data provenance" in the context of diagnostic data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth, in the context of AI/ML, refers to a definitively correct answer for diagnostic tasks. This document describes the physical and functional testing of a device, not a diagnostic algorithm.
4. Adjudication method for the test set: Not applicable. This concept is relevant for expert disagreement resolution in diagnostic studies.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document is not about AI assistance for human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithmic device.
7. The type of ground truth used: For the physical and functional tests, the "ground truth" is defined by the technical standards (e.g., ISO, USP, ASTM) and specified criteria (e.g., "no leakage," "minimum tensile force"). There isn't a "ground truth" as understood in diagnostic AI.
8. The sample size for the training set: Not applicable. No training set is involved for this physical device's evaluation.
9. How the ground truth for the training set was established: Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, representing the department's focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 2, 2015
B. Braun Medical, Inc. Ms. Tracy Maddock Senior Regulatory Affairs Specialist 901 Marcon Boulevard Allentown, Pennsylvania 18109
Re: K150787
Trade/Device Name: Universal Spike Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: LHI Dated: October 28, 2015 Received: October 29, 2015
Dear Ms. Maddock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Tracy Maddock.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
Device Name Universal Spike
Indications for Use (Describe)
For the transfer of solutions from one container to another.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED,
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
{3}------------------------------------------------
D Bruun Medical Inc. STOtk) Premarket Notification l'inversal Spike
| 5. | 510(k) Summary | K150787 |
|---|---|---|
| SUBMITTER: | B. Braun Medical Inc.901 Marcon BoulevardAllentown, PA 18109-9341610-266-0500Contact: Tracy Maddock, RACSr. Regulatory Affairs SpecialistPhone: (610) 596-2545Fax: (610) 266-4962E-mail: Tracy.Maddock@bbraun.com | |
| DATE: | December 2, 2015 | |
| DEVICE NAME: | Universal Spike | |
| COMMON ORUSUAL NAME: | IV Fluid Transfer Set | |
| DEVICECLASSIFICATION: | Class II per 21 CFR 880.5440Intravascular Administration Set, product code LHIClassification Panel: General Hospital | |
| PREDICATE DEVICES: | B. Braun Dispensing Pin with One-Way Valve, K943181 |
DESCRIPTION:
The B. Braun Universal Spike is a sterile, single use, disposable fluid transfer device consisting of a closure-piercing component capable of piercing and penetrating the closure of a fluid container, an air-inlet component which allows the dispensing from both rigid and pressure collapsible dispensing containers, and a distal female luer lock adapter which provides access for the attachment of a syringe or fluid transfer set. The closure-piercing component and luer lock adapter are covered by protective caps which prevent contamination of the device until its point of use.
INTENDED USE INDICATIONS FOR USE:
For the transfer of solutions from one container to another.
{4}------------------------------------------------
A Brunin Medical Inc. 519(k) Premarket Notification l'niversal Spike
SUBSTANTIAL EQUIVALENCE:
The Universal Spike is substantially equivalent to the B. Braun Dispensing Pin with One-Way Valve (K943181).
Comparison of Technological Characteristics with the Predicate Device
The proposed device has the same intended use and the same or similar technological characteristics as the predicate device. The proposed spike is similar in design, having the same type of components such as a piercing component with protective cap, an air inlet and a distal luer lock adapter with cap as the predicate device. These components are also constructed of similar materials. Both devices are provided as sterile packaged devices sterilized by ethylene oxide for single use.
A table summarizing the comparison between the Universal Spike and the predicate device is provided below.
| Characteristic | Predicate Device - K943181Dispensing Pin with One-WayValve | ProposedUniversal Spike |
|---|---|---|
| Intended Use | Intended for the transfer of solutionsfrom one container to another | For the transfer of solutions from onecontainer to another |
| Components | Vented or non-vented spike with orwithout air filter and distal femaleluer lock hub (can be attached tosyringe or IV fluid transfer set) | Vented spike with air inlet filter anddistal female luer lock connector (canbe attached to syringe or fluid transferset) |
| Materials ofconstruction | Spike Guard – LDPE | Spike Cap - LDPE |
| Piercing Device - ABS | Spike Body - ABS | |
| Air InletHousing - PolypropyleneFilter - Pallflex | Air InletHousing - PolypropyleneFilter -3 micron membrane | |
| Female Luer Lock - Polycarbonate | Female Luer Lock - ABS | |
| Female Luer Lock Cap - HDPE | Female Luer Lock cap –Polypropylene | |
| Back Check ValveHousing Top and Bottom -Polycarbonate | N/A | |
| Punched Disk - Silicone | ||
| Configurations | • Normal sized spike or lowervolume, mini spike• Air-inlet vents or non-vented | • Normal sized spike• Air inlet filter |
| SterilizationMethod | Ethylene oxide | Ethylene oxide |
{5}------------------------------------------------
B. Brunn Medical Inc. 510fk) Premarket Nortfication I 'niversal Spike
Performance Testing
The proposed Universal Spike was subjected to functional and performance testing to demonstrate that the device performs as intended.
The following testing was performed on the subject device to the performance criteria listed or included in the referenced standard:
- . Chemical Analysis (ISO 8536-4, ISO 10993-18)
- Visual Control (ISO 22413, ISO 8536) .
- . Fragmentation Test (ISO 22413)
- . Penetration Force (ISO 8536-2)
- Spike retention / seal ability (ISO 8536-2) �
- . Patency Test (ISO 8536-4)
- Pressure Test / Leakage (no leakage shall be detected when device is subjected to air . pressure of 450 mmHg)
- Pull Test (cap and spike must withstand a minimum tensile force of separation between 4,41- 44.13 Newtons
- Free Flow (ISO 22413) .
- Particulate Contamination (USP <788>) �
- Luer testing (ISO 594-1)
- Biocompatibility (ISO 10993-1)
- Cytotoxicity (MEM Elution, ISO 10993-5) 0
- 0 Delayed-type hypersensitivity (Guinea Pig Maximization, ISO 10993-10)
- Intracutaneous Reactivity (ISO 10993-10) 0
- Acute Systemic Toxicity (ISO 10993-11) 0
- Hemocompatibility (Hemolysis, ASTM F756, ISO 10993-4) 0
- Material-Mediated Pyrogenicity (Rabbit Pyrogen, USP <151>, ISO 10993-11) 0
No clinical testing was performed as this device does not require clinical studies to demonstrate substantial equivalence with the predicate device.
CONCLUSION:
Results of functional and performance testing conducted on the proposed device demonstrate that the Universal Spike is substantially equivalent to the predicate device.
§ 880.5440 Intravascular administration set.
(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.