K Number
K142928

Validate with FDA (Live)

Date Cleared
2015-09-30

(356 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FreeStyle Precision Neo Blood Glucose Monitoring System is for use outside the body only (in vitro diagnostic use) in the quantitative measurement of glucose in fresh whole blood for self-testing by lay users from the fingers. It is not intended to be used for testing neonatal blood samples or the diagnosis or screening of diabetes.

The FreeStyle Precision Neo System is indicated for home (lay user) in the management of patients with diabetes. It is intended to be used by a single person and should not be shared.

The FreeStyle Precision Neo Blood Glucose Test Strips are for use with the FreeStyle Precision Neo Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

Device Description

The FreeStyle Precision Neo Meter is for the quantitative measurement of glucose in capillary whole blood samples. The FreeStyle Precision Neo Meter, in conjunction with the FreeStyle Precision Neo Blood Glucose Test Strips, works on the principal of amperometric technology, measuring glucose by its reaction with Glucose Dehydrogenase (GDH) in blood samples or control solutions through electrical mediation.

The FreeStyle Precision Neo System is compatible with the following components and accessories:

  • . FreeStyle Precision Neo Blood Glucose Test Strips (available separately)
  • MediSense Glucose and Ketone Control Solutions (available separately) ●
  • USB Cable (available separately) ●
  • FreeStyle Lancing Device II ●
  • Thin Lancets

The FreeStyle Precision Neo System also includes a hypoglycemia and hyperglycemia trend indicators feature. The hypoglycemia/hyperglycemia trend indicators feature alerts the user of low and high blood glucose patterns. The FreeStyle Precision Neo Meter displays a red arrow indicator light if the user's blood glucose is below the set target or a vellow arrow indicator light if their blood glucose is above the set target. The default indicator target values are 70 and 240 mg/dL but can be customized, if needed. The Meter will also use these ranges to determine blood glucose patterns.

AI/ML Overview

The FreeStyle Precision Neo Blood Glucose Monitoring System is for quantitative measurement of glucose in fresh whole blood for self-testing by lay users from the fingers. It is not intended to be used for testing neonatal blood samples or the diagnosis or screening of diabetes.

Here is the information about the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

Device Characteristics / Performance StudyAcceptance CriteriaReported Device Performance
Accuracy (ISO 15197:2013)For glucose concentrations < 100 mg/dL (5.55 mmol/L): Minimum of 95% of results shall fall within ±15 mg/dL (±0.83 mmol/L) of the reference method results. For glucose concentrations ≥ 100 mg/dL (5.55 mmol/L): Minimum of 95% of results shall fall within ±15% of the reference method results.Lay User Performance (n=350 subjects): Glucose < 100 mg/dL: 100% of results were within ±15 mg/dL. Glucose ≥ 100 mg/dL: 98.7% of results were within ±15%. Overall: 99.4% within ±15% or ±15 mg/dL. Professional User Performance: Similar excellent performance demonstrated.
Repeatability (Precision)Standard Deviation (SD) for glucose < 100 mg/dL; Coefficient of Variation (CV) for glucose ≥ 100 mg/dL. Values should be within typical industry standards for blood glucose meters.Not explicitly quantified in the provided snippets for acceptance criteria, but stated as "demonstrated acceptable repeatability based on testing with spiked whole blood and control solutions yielding results that met the requirements of ISO 15197:2013". The full report would have the actual SD/CV values.
Intermediate Precision (Reproducibility)Standard Deviation (SD) for glucose < 100 mg/dL; Coefficient of Variation (CV) for glucose ≥ 100 mg/dL. Values should be within typical industry standards for blood glucose meters.Not explicitly quantified in the provided snippets for acceptance criteria, but stated as "demonstrated acceptable intermediate precision based on testing with spiked whole blood and control solutions yielding results that met the requirements of ISO 15197:2013". The full report would have the actual SD/CV values.
System Accuracy (Lay User vs. Lab Reference)All results meeting the ISO 15197:2013 criteria.100% of results for glucose < 100 mg/dL were within ±15 mg/dL. 98.7 % of results for glucose ≥ 100 mg/dL were within ±15%. Overall 99.4% results were within ±15% or ±15 mg/dL.
Influence of Interfering Substances (Endogenous & Exogenous)No significant interference that would compromise clinical performance."The FreeStyle Precision Neo Blood Glucose Monitoring System demonstrated negligible interference from endogenous and exogenous substances within physiological and therapeutic ranges by meeting the requirements of ISO 15197:2013." Specific substances tested include hematocrit variations (15-65%), common medications, and physiological substances (e.g., bilirubin, triglycerides, uric acid).
Operating Temperature Range50° - 122°FDemonstrated performance within 50° - 122°F.
Operating Humidity Range10% - 90%, non-condensingDemonstrated performance within 10% - 90%, non-condensing.
AltitudeUp to 10,000 feetDemonstrated performance up to 10,000 feet.
Hematocrit Range15-65%Demonstrated accurate performance across the 15-65% hematocrit range, meeting ISO 15197:2013 criteria.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Accuracy Studies (Lay User Performance): 350 subjects participated in the lay user performance study.
  • Data Provenance: The document does not explicitly state the country of origin for the data. It is implied to be clinical study data collected specifically for this device submission. The study is prospective in nature as it involves subjects using the device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number of experts or their qualifications for establishing the ground truth. However, for blood glucose monitoring systems, the ground truth is typically established using a highly accurate laboratory reference method, not expert interpretation.

4. Adjudication Method for the Test Set

Not applicable. The ground truth for blood glucose measurements is established by a laboratory reference method, not by expert consensus requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This device is a blood glucose monitoring system, not an imaging or diagnostic device that involves human readers interpreting cases. Its performance is compared against a laboratory reference standard.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Yes, a standalone performance assessment was conducted. The accuracy, precision, and interference studies evaluate the device's (meter + test strip) performance directly against a reference method or known concentrations, without a human interpretation step. The "Lay User Performance" study assesses the device as used by its intended users, but the fundamental algorithm for glucose measurement operates independently.

7. Type of Ground Truth Used

The ground truth used for performance evaluation is a laboratory reference method for blood glucose measurement (e.g., YSI analyzer or equivalent). This is the gold standard for quantitative blood glucose levels.

8. Sample Size for the Training Set

The document does not provide details on the sample size for any training set. For blood glucose meters, the "training" involves the development and calibration of the electrochemical system (meter and test strip) rather than machine learning algorithm training with a distinct training dataset.

9. How the Ground Truth for the Training Set Was Established

The document does not provide information about a specific training set or how its ground truth was established, as this type of device typically relies on chemical and electrical engineering principles, and a robust calibration process rather than a machine learning training dataset. The "ground truth" during development would be known glucose concentrations for calibration and verification of the system's accuracy.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of a caduceus, which is a symbol of medicine. The caduceus is made up of three lines that form the shape of a snake wrapped around a staff.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 30, 2015

ABBOTT DIABETES CARE INC MICHELLE RICAFORT SENIOR REGULATORY AFFAIRS SPECIALIST 1360 SOUTH LOOP ROAD ALAMEDA CA 94502

Re: K142928

Trade/Device Name: FreeStyle Precision Neo Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW. LFR Dated: September 2, 2015 Received: September 4, 2015

Dear Michelle Ricafort:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K142928

Device Name

FreeStyle Precision Neo Blood Glucose Monitoring System

Indications for Use (Describe)

The FreeStyle Precision Neo Blood Glucose Monitoring System is for use outside the body only (in vitro diagnostic use) in the quantitative measurement of glucose in fresh whole blood for self-testing by lay users from the fingers. It is not intended to be used for testing neonatal blood samples or the diagnosis or screening of diabetes.

The FreeStyle Precision Neo System is indicated for home (lay user) in the management of patients with diabetes. It is intended to be used by a single person and should not be shared.

The FreeStyle Precision Neo Blood Glucose Test Strips are for use with the FreeStyle Precision Neo Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subnart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

IDO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Specal 510k

Abbott Diabetes Care Inc. 1360 South Loop Road, Alameda, CA 94502

Image /page/3/Picture/3 description: The image shows the logo for Abbott, a global healthcare company. The logo consists of a stylized lowercase letter 'a' enclosed within a rounded square shape. The color of the logo is a solid, medium shade of blue. The design is simple, modern, and easily recognizable.

510(k) Summary

According to the requirements per 21 CFR §807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Company:Abbott Laboratories
Division:Abbott Diabetes Care, Inc.
Street Address:1360 South Loop Road
City, State Zip:Alameda, CA 94502
Telephone No:714-247-8434
Fax No:510-864-4791
Contact Person:Michelle RicafortTel No. 714-247-8434Fax No. 510-864-4791michelle.ricafort@amo.abbott.com
Proprietary Name:FreeStyle Precision Neo Blood Glucose Monitoring System
Common Name:Glucose Test System
Classification Name:Glucose Dehydrogenase, Glucose, Class II (21 CFR§ 862.1345)Product codes: NBW, LFR
Predicate Device:FreeStyle Precision Neo Blood Glucose Monitoring System (K140371)
Legal Manufacturer:Establishment:Abbott Diabetes Care Ltd.Range RoadWitney, OxonOX29 OYL, UK
U.S. ContactEstablishment:Abbott Diabetes Care Inc.1360 South Loop RoadAlameda, CA 94502

Image /page/3/Picture/7 description: The image contains the logo for Abbott. The logo consists of a blue, stylized "A" on the left, followed by the word "Abbott" in bold, black letters. Below "Abbott" is the tagline "A Promise for Life" in a smaller, lighter font.

{4}------------------------------------------------

Intended Use:

The FreeStyle Precision Neo Blood Glucose Monitoring System is for use outside the body only (in vitro diagnostic use) in the quantitative measurement of glucose in fresh whole blood for self-testing by lay users from the fingers. It is not intended to be used for testing neonatal blood samples or the diagnosis or screening of diabetes.

The FreeStyle Precision Neo System is indicated for home (lay user) in the management of patients with diabetes. It is intended to be used by a single person and should not be shared.

The FreeStyle Precision Neo Blood Glucose Test Strips are for use with the FreeStyle Precision Neo Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

Description of the Device:

The FreeStyle Precision Neo Meter is for the quantitative measurement of glucose in capillary whole blood samples. The FreeStyle Precision Neo Meter, in conjunction with the FreeStyle Precision Neo Blood Glucose Test Strips, works on the principal of amperometric technology, measuring glucose by its reaction with Glucose Dehydrogenase (GDH) in blood samples or control solutions through electrical mediation.

The FreeStyle Precision Neo System is compatible with the following components and accessories:

  • . FreeStyle Precision Neo Blood Glucose Test Strips (available separately)
  • MediSense Glucose and Ketone Control Solutions (available separately) ●
  • USB Cable (available separately) ●
  • FreeStyle Lancing Device II ●
  • Thin Lancets

The FreeStyle Precision Neo System also includes a hypoglycemia and hyperglycemia trend indicators feature. The hypoglycemia/hyperglycemia trend indicators feature alerts the user of low and high blood glucose patterns. The FreeStyle Precision Neo Meter displays a red arrow indicator light if the user's blood glucose is below the set target or a vellow arrow

Image /page/4/Picture/15 description: The image shows the Abbott logo, which includes a stylized letter 'a' in blue, followed by the word 'Abbott' in black. Below the word 'Abbott' is the tagline 'A Promise for Life'. The text 'Proprietary and Confidential' is located at the bottom left of the image.

2

510(k) Summary

{5}------------------------------------------------

indicator light if their blood glucose is above the set target. The default indicator target values are 70 and 240 mg/dL but can be customized, if needed. The Meter will also use these ranges to determine blood glucose patterns.

Principles of Operation:

The FreeStyle Precision Neo Meter (in conjunction with blood glucose test strips) utilizes amperometric technology to quantitatively measure the glucose concentration in capillary whole blood samples from the fingertip and in MediSense Glucose & Ketone Control Solutions.

The FreeStyle Precision Neo Meter measures glucose electrically. The glucose biosensor is capable of recognising the glucose present in whole blood or control solutions by virtue of the glucose specificity of the enzyme GDH present on the glucose test strip. The electrons liberated by this reaction are transferred via a co-factor and mediator to the meter where they are read as a small electrical current. The size of the current is directly proportional to the level of the glucose in the applied sample.

The apply blood symbol is displayed for the user to apply blood to the test strip until the meter begins the test. The meter detects trigger current from the test strip when enough blood has covered the strip electrodes and the test countdown will start. When the countdown is complete a test result is displayed on the meter screen. The unit of measure displayed on the meter screen is fixed in mg/dL and cannot be modified by the user.

Substantial Equivalence:

The FreeStyle Precision Neo Blood Glucose Monitoring System is substantially equivalent to the predicate, which was cleared by the Agency on September 30, 2014, to market under K140371: FreeStyle Precision Neo Blood Glucose Monitoring System in terms of indications for use, operating principle, technological characteristics and performance specifications.

Image /page/5/Picture/9 description: The image shows the Abbott logo, which consists of a blue abstract shape on the left and the word "Abbott" in bold black letters on the right. Below the word "Abbott" is the tagline "A Promise for Life" in a smaller font. The text "Proprietary and Confidential" is located at the bottom left of the image.

3

Page 5

{6}------------------------------------------------

Comparison to Predicate Device:

The similarities and differences between the proposed FreeStyle Precision Neo Blood Glucose Monitoring System and the predicate (K140371) device are highlighted in the table below.

Proposed DevicePredicate Device
PRODUCT NAMEFreeStyle Precision NeoBlood Glucose MonitoringSystemFreeStyle Precision Neo BloodGlucose Monitoring System(K140371)
CHARACTERISTICS
Fundamental TechnologyThe FreeStyle Precision NeoSystem, in conjunction withblood glucose test strips,utilizes amperometrictechnology to quantitativelymeasure the glucoseconcentration in whole bloodsamples.Same
Principles of OperationAmperometrySame
Glucose Operating Range20-500 mg/dLSame
ChemistryGDH-NADSame
Glucose Sample volume0.6 microlitersSame
Glucose Assay Time5 secondsSame
Coding (Calibration)No coding requiredSame
Operating Temperature50° - 122°FSame
Operating Humidity10% - 90%, non-condensingSame
Power SourceCR 2032 lithium (coin cell)batteriesSame
Memory1,000 eventsSame
ButtonsTouch Press-ButtonSame
Indications for UseThe FreeStyle Precision NeoBlood Glucose MonitoringSystem is for use outside thebody only (in vitrodiagnostic use) in thequantitative measurement ofglucose in fresh whole bloodfor self-testing by lay usersfrom the fingers. It is notintended to be used fortesting neonatal bloodSame
Proposed DevicePredicate Device
PRODUCT NAMEFreeStyle Precision NeoBlood Glucose MonitoringSystemFreeStyle Precision Neo BloodGlucose Monitoring System(K140371)
CHARACTERISTICS
samples or the diagnosis orscreening of diabetes.The FreeStyle Precision NeoSystem is indicated for home(lay user) in the managementof patients with diabetes. It isintended to be used by asingle person and should notbe shared.The FreeStyle Precision NeoBlood Glucose Test Stripsare for use with the FreeStylePrecision Neo BloodGlucose Meter toquantitatively measureglucose (sugar) in freshcapillary whole bloodsamples drawn from thefingertips.
Product Classification CodeNBW, LFRSame
Glucose Sample TypesFresh capillary whole bloodfrom the fingerSame
Compatible Test StripsFreeStyle Precision NeoBlood Glucose Test StripsSame
Glucose Hematocrit Range15-65%Same
Altitude10,000 feetSame
Hyperglycemia/HypoglycemiaTrend Indicator FeatureYesNone

Image /page/6/Picture/5 description: The image shows the Abbott logo, which consists of a blue stylized "A" symbol and the word "Abbott" in bold black letters. Below the word "Abbott" is the tagline "A Promise for Life" in a smaller font. The text "Proprietary and Confidential" is located in the bottom left corner of the image.

4

510(k) Summary

{7}------------------------------------------------

Image /page/7/Picture/3 description: The image shows the Abbott logo, which consists of a stylized blue square with a white "A" inside. To the right of the square is the word "Abbott" in bold, black letters. Below "Abbott" is the tagline "A Promise for Life" in a smaller font. The words "Proprietary and Confidential" are at the bottom left.

5

Page 7

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.